Skip to main content
Erschienen in: Radiological Physics and Technology 4/2020

Open Access 05.11.2020

Systolic modified Look–Locker inversion recovery myocardial T1 mapping improves the accuracy of T1 and extracellular volume fraction measurements of patients with high heart rate or atrial fibrillation

verfasst von: Hirohiko Shinbo, Satoshi Tomioka, Toshihiko Ino, Keiko Koyama

Erschienen in: Radiological Physics and Technology | Ausgabe 4/2020

Abstract

Image data for T1 mapping are generally acquired during mid-diastole period. However, T1 mapping tends to fail for patients with high heart rate or atrial fibrillation because of short or irregular R-R interval. Focusing on the evidence that the timing of systole is more stable than that of diastole from the R wave, we compared systolic T1 mapping with conventional diastolic T1 mapping for all participants (n = 58) by visual scoring of T1 calculation error. The systolic scores were significantly better than the diastolic scores (p < 0.05). This advantage of the systolic scores was confirmed selectively for patients with atrial fibrillation (p < 0.05, n = 19). The successful number of nonrigid image registration alignment for extracellular volume fraction (ECV) analysis also increased significantly for systolic images compared with diastolic images (p < 0.05). Thus, systolic T1 mapping improves the accuracy of T1 values and ECV analysis.
Hinweise
The original version of this article was revised due to a retrospective Open Access Cancellation.
A correction to this article is available online at https://​doi.​org/​10.​1007/​s12194-020-00596-x.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

1 Introduction

Late gadolinium enhancement (LGE) is good at detecting focal myocardial fibrosis but less effective in detecting diffuse myocardial fibrosis. Fifty-nine percent of dilated cardiomyopathy is not contrasted with delayed enhancement [1]. However, T1 mapping is clinically useful for quantitative evaluation of diffuse myocardial fibrosis [25]. In addition, myocardial characteristics, such as edema and focal and diffuse fibrosis, can be quantitatively evaluated with extracellular volume fraction (ECV) analysis, and the ECV can be calculated from both precontrast and postcontrast T1 maps [68]. The modified Look–Locker inversion recovery (MOLLI) sequence for measuring T1 values is widely useful because of its high precision [9, 10]. In the MOLLI sequence, several images are acquired, while T1 is changed in the process of recovering the signal after longitudinal magnetization by the inversion recovery pulse. With the MOLLI sequence, we can calculate T1 values for each pixel from the signal intensity of the acquired data. The MOLLI sequence can obtain stable T1 values through time-based data acquisition [11]. T1 mapping is generally acquired during the mid-diastole phase, when left ventricular volume change is minimal, as the optimal cardiac phase for patients with low heart rate. However, irregularities among patients with high heart rate and atrial fibrillation are commonly observed on cardiac magnetic resonance (CMR) imaging, and these patients have irregular and short R-R intervals. Therefore, diastolic T1 mapping cannot accommodate irregular and short R-R intervals and cannot be acquired in the optimal cardiac phase. This would increase calculation errors resulting from movements, and thus, the correct T1 values could not be obtained in the past. In some pilot studies, systolic T1 mapping is performed when the myocardium is the thickest, and few calculation errors resulting from movements are equally or more appropriate than diastolic T1 mapping in the quantitative evaluation of the myocardium because partial volume effects are reduced, and motion artifacts are minimized during tachycardia [1215]. Therefore, we visually evaluated calculation errors resulting from movements (motion artifacts during tachycardia) using the T1 mapping confidence map (error map) in the magnetic resonance system. Traditionally, during diastole, the alignment of both precontrast and postcontrast auto-analysis non-rigid image registration for ECV is mostly unsuccessful. Non-rigid image registration alignment is standard for every workstation's ECV analysis software, and the success of non-rigid image registration alignment is highly dependent on the parallel shift amount, rotation shift amount, and scaling deformation amount. If alignment fails, the left ventricular myocardium in three slices of the left ventricular base, middle, and apex with both precontrast and postcontrast images needs to be manually reextracted from the endocardial and epicardial boundaries. This re-extraction reduces the accuracy of the ECV analysis. On the contrary, successfully by auto-analysis, non-rigid image registration alignment improves the accuracy of ECV analysis because the subendocardial, mid-wall, and epicardial segment of the left ventricular myocardium matched precontrast and postcontrast images. In this study, we compared the conventional diastole and systole in T1 mapping and ECV.

2 Materials and methods

2.1 Study population

We recruited 20 healthy volunteers (10 men and 10 women; mean age, 45 ± 12 years) and 38 patients (23 men and 15 women; mean age, 61 ± 12 years) without contraindications to CMR imaging. Twenty healthy volunteers underwent non-contrast examinations. Thirty-eight patients were administered contrast agents, and T1 maps were obtained from both precontrast and postcontrast images; further, the ECV analysis was also performed. None of the healthy volunteers had any evidence of or risk factors for cardiovascular disease. Patients were recruited consecutively and had diagnoses of atrial fibrillation (n = 19), dilated cardiomyopathy (n = 4), ischemic cardiomyopathy (n = 2), hypertrophic cardiomyopathy (n = 2), old myocardial infarction (n = 4), cardiac sarcoidosis (n = 3), and other diseases (n = 4). The study protocol was approved by our institutional review board. The study was conducted in accordance with the Declaration of Helsinki. Informed consent was obtained from all participants before the CMR imaging. A detailed explanation of the contrast agents was provided to each participant.

2.2 Determination of the MOLLI sequence sampling scheme

The MOLLI sequence is acquired with both precontrast and postcontrast T1 maps by time-based data acquisition. The time-based MOLLI sequence has fixed acquisition and pauses (recovery time) (Fig. 1). Precontrast T1 mapping is set to acquisition time (5 s) + pause time (3 s) + acquisition time (3 s). Postcontrast T1 mapping is set to acquisition time (4 s) + pause time (1 s) + acquisition time (3 s) + pause time (1 s) + acquisition time (2 s) [16]. Because the R-R interval is short at high heart rates, a fixed recovery pause time (precontrast pause time of 3 s, postcontrast pause time of 1 s) can fully recover longitudinal magnetization before the next inversion pulse is applied. We adopted a time-based MOLLI sequence that is independent of heartbeat, from low to high heartbeats.

2.3 Trigger delay in diastole and systole

Acquisition timing was visually determined in short-axis cine images; when the left ventricular area was at its maximum, the delay time for diastolic data acquisition was calculated. Subsequently, when the left ventricular area was minimal, the delay time for systolic data acquisition was calculated [17].

2.4 CMR image acquisition

CMR imaging was performed with a 1.5-T magnetic resonance system (Ingenia, release 5.3.1.3, and dS Torso receiver coil; Philips Healthcare, Amsterdam, the Netherlands). With electrocardiographic gating, data for all sequences were acquired while participants held their breath. Cine images, based on electrocardiography-gated steady-state free precession imaging, were acquired in the short-axis view, and in four-, two-, and three-chamber views, and were reconstructed for 24 phases/heartbeat. The cine sequence parameters were as follows: echo time (TE), 1.16 ms; repetition time (TR), 2.3 ms; flip angle (FA), 60°; slice thickness, 10 mm; field of view, 380 × 380 mm2; in-plane resolution, 1.70 × 1.41 mm2; acquisition matrix, 224 × 236; and sensitivity encoding (SENSE) phase direction factor of 2.5. LGE sequence parameters with three-dimensional inversion recovery T1-turbo field echo were as follows: TE, 2.3 ms; TR, 4.7 ms; FA, 15°; slice thickness, 10 mm; field of view, 380 × 380 mm2; in-plane resolution, 1.48 × 1.65 mm2; acquisition matrix, 256 × 231; SENSE phase direction factor of 2.5; SENSE slice direction factor of 1.5; linear profile order; and radial turbo direction. LGE images were acquired 10–15 min after an injection of 0.1 mmol/kg gadobutrol (Gadovist, Bayer, Germany). T1 mapping sequence parameters were as follows: TE, 0.88 ms; TR, 1.97 ms; FA, 35°; slice thickness, 10 mm; field of view, 300 × 300 mm2; in-plane resolution, 2.0 × 2.0 mm2; acquisition matrix, 152 × 150; SENSE phase direction factor of 2.0; and linear profile order. T1 mapping images were acquired in basal, midventricular, and apical short-axis planes during diastole and systole before and 15 min after an injection of 0.1 mmol/kg gadobutrol.

2.5 Image analysis

2.5.1 Left ventricular functions

All image datasets were transferred to a workstation (Intelli Space Portal, version 7.0.2.20700; Philips Healthcare). Left ventricular ejection fraction (LVEF), end-diastolic volume (EDV), and end-systolic volumes (ESV) were evaluated. The LVEF was calculated as follows:
LVEF (%) = (EDV–ESV)/EDV × 100.

2.5.2 T1 mapping

T1 mapping images obtained from all participants were transferred to a workstation (Ziostation2, T1 map, version 2.4.2.3; Ziosoft, Inc., Tokyo, Japan). To measure T1 values, the T1 mapping confidence map was used to manually set a region of interest (ROI) of 5 mm2 at the middle of the septum in the mid-wall of the left ventricle (Fig. 2). The confidence map is displayed as an error when the standard deviation deviates significantly for each TI pixel collected in the T1 recovery curve. Therefore, the confidence map will show the calculation errors resulting from movements as blackout pixels. The ROI did not include the pixel of calculation errors resulting from movements; it was set in a reliable range [18].

2.5.3 Visual evaluation

Visual evaluation was conducted on the precontrast T1 mapping confidence map for all participants. Figure 3 shows the visual evaluation method. The blackout pixels represent the calculation errors of the T1 value. Scoring was performed based on the number of blackout pixels in the heart area. If the number of pixels indicating calculation errors was more than 50, the score was set to 1. If it was between 20 and 50, the score was 2. If it was less than 20 pixels, the score was 3. An experienced reader with more than 5 years of cardiac magnetic resonance imaging radiology technician performed a visual evaluation of the confidence map calculation error. Visual evaluation was calculated separately for all participants, for patients with atrial fibrillation alone, and for each heart rate.

2.5.4 ECV

In 38 patients, the ECV was calculated from the left ventricular myocardial T1 values, and the left ventricular cavity blood T1 values acquired pre- and post-contrast. ECV was calibrated with the blood hematocrit level at a workstation for an offline analysis (Ziostation2, T1 map, version 2.4.2.3) [19]. The blood hematocrit level is the result of a blood sample taken the day before CMR. The ECV was calculated as
$$ {\text{ECV }} = \, ({1} - {\text{hematocrit}})\left( {\frac{1}{{\text{T1 myocardium post}}} - \frac{1}{{\text{T1 myocardium pre}}}} \right)/\left( {\frac{1}{{\text{T1 cavity blood post}}} - \frac{1}{{\text{T1 cavity blood pre}}}} \right) $$
Using the Bull's-Eye plot of ECV, we also calculated the average value of five segments of the left ventricular myocardial septum collected during diastole and systole (Fig. 4). The left ventricular septum was chosen because the myocardium is thicker than the anterior wall, lateral wall, and posterior wall. These facilitate segments of endocardial, mid-wall, and epicardial. In addition, the ECV was analyzed on an offline analysis workstation, auto-analysis non-rigid image registration alignment with both precontrast and postcontrast images was performed (Fig. 5), and the number of successful cases by auto-analysis was counted. If the alignment failed, the left ventricular myocardium in three slices of the left ventricular base, middle, and apex with both precontrast and postcontrast images was manually reextracted from the endocardial and epicardial boundaries.

2.6 Statistical analyses

Data are expressed as the means ± standard deviations. The normality of the diastolic and systolic T1 data in healthy volunteers and patients was assessed using the Kolmogorov–Smirnov test. Because of the normality of the data, the Pearson correlation coefficient (r) was used for the analysis, and the data were compared using paired t-test and one-way analysis of variance. The normality of the visual evaluation data in all participants and that of ECV data was assessed using the Kolmogorov–Smirnov test. Visual evaluation data and ECV data did not strictly follow a Gaussian distribution. Therefore, the statistical tests used were non-parametric, and the Kruskal–Wallis test was used. Statistical significance was defined as a p-value < 0.05. All statistical analyses were performed using R software, version 3.4.1 [20].

3 Results

3.1 Clinical characteristics of study participants

Table 1 shows the demographic data, heart rate, and heart rate fluctuation during scanning and left ventricular functional indices of healthy volunteers and patients.
Table 1
Characteristics and left-ventricle functional indices of healthy volunteers and patients
 
Healthy volunteers
Patients
 
(n = 20)
(n = 38)
Sex (man), n (%)
10 (50%)
23 (61%)
Age (years)
45.2 ± 11.9
61.4 ± 12.3
Height (cm)
163.9 ± 7.6
161.9 ± 7.6
Weight (kg)
61.8 ± 7.0
58.8 ± 10.7
BMI (kg/m2)
21.8 ± 3.6
23.58 ± 4.1
Heart rate (bpm)
65.4 ± 4.4
68.9 ± 11.2
HR fluctuation (bpm)
8.7 ± 3.7
16.4 ± 7.4
LVEF (%)
61.29 ± 7.58
48.63 ± 18.28
LVEDV (mL)
128.7 ± 23.5
87.2 ± 16.5
LVESV (mL)
58.1 ± 12.8
48.1 ± 21.4
Data are expressed as means ± standard deviations
BMI body mass index, HR heart rate, LV left ventricle, EF ejection fraction, EDV end-diastolic volume, ESV end-systolic volume

3.2 Comparison between diastolic and systolic T1 values

Left ventricular myocardial T1 values during diastole and systole were strongly and positively correlated in healthy volunteers (r = 0.94, p < 0.05) (Fig. 6). Table 2 lists left ventricular myocardial T1 values in diastole and systole of the healthy volunteers and patients. Left ventricular myocardial T1 systolic values were significantly lower than T1 diastolic values in both healthy volunteers (diastole 998.7 ± 17.1 ms vs systole 988.7 ± 24.9 ms; p < 0.05) and patients (diastole 1014.8 ± 45.9 ms vs systole 1006.9 ± 45.4 ms; p < 0.05).
Table 2
Result of the LV–T1 values between diastole and systole (ms)
 
Diastole
Systole
p-value
Patients (n = 38)
1014.8 ± 45.9
1006.9 ± 45.4
p < 0.05
Healthy volunteers (n = 20)
998.7 ± 17.1
988.7 ± 24.9
p < 0.05
Total (n = 58)
1008.7 ± 36.0
1000.0 ± 37.8
p < 0.05
Data expressed as mean ± standard deviation

3.3 Visual evaluation

According to the visual evaluation of the diastolic and systolic T1 mapping confidence map images, in all participants, the systolic score was significantly higher than the diastolic score (diastole 2.03 ± 0.51 vs systole 2.25 ± 0.57; p < 0.05) (Table 3). In patients with atrial fibrillation, the systolic score was also significantly higher than the diastolic score (diastole 1.63 ± 0.53 vs systole 2.18 ± 0.59; p < 0.05) (Table 3). T1 mapping visual evaluation for each heart rate also revealed that when the heart rates were 40–60 bpm, the diastolic score was higher than the systolic score, whereas at heart rates of 61–120 bpm, the systolic score was higher than the diastolic score (Fig. 7).
Table 3
Visual evaluation of all participants and AF patients only
 
Diastole
Systole
p-value
All participants (n = 58)
2.03 ± 0.51
2.25 ± 0.57
p < 0.05
AF patients only (n = 19)
1.63 ± 0.53
2.18 ± 0.59
p < 0.05
Data expressed as mean ± standard deviation

3.4 ECV

ECV showed significantly lower systolic scores and T1 values. As a supplement, we included cases in which auto-analysis failed and manual analysis was performed for reasons of low success during diastole (diastole 0.34 ± 0.05 vs systole 0.33 ± 0.05; p < 0.05) (Fig. 8). Diastolic and systolic ECV were strongly and positively correlated in patients (r = 0.98, p < 0.05) (Fig. 9). Auto-analysis non-rigid image registration alignment for ECV analysis in low–high heart rate and atrial fibrillation cases was significantly more successful during systole, and systolic ECV showed improved accuracy (heart rate 40–70 bpm, diastole 7 vs systole 12; p < 0.05) (heart rate 71–120 bpm, diastole 2 vs systole 22; p < 0.05) (Fig. 10).

4 Discussion

In this study, diastolic and systolic T1 values were significantly different but strongly correlated in the normal myocardium of 20 healthy volunteers. In addition, systolic T1 values were lower than diastolic values [14]. The reason for this is that during diastole, the left ventricular myocardial thickness is lower and is affected by the partial volume effect of adjacent tissues, and includes blood within a set ROI, which may have resulted in slightly higher diastolic T1 values. In all participants and patients with atrial fibrillation, systolic T1 mapping had better visual evaluation with significant difference. A low heart rate of 40–60 bpm showed good visual evaluation even in diastolic T1 mapping because of the optimal phase at mid-diastole [15]. At a high heart rate, the motion-free time in diastole is shortened more than in systole. In patients with atrial fibrillation, although the R-R interval varies in each cardiac cycle, the variation in timing of systolic acquisition is smaller than that of diastolic acquisition [11, 15]. Therefore, systole acquisition resulted in more evaluable images [21, 22]. As with T1 mapping, ECV was significantly lower during systole than during diastole, and the systolic values showed a strong positive correlation with the diastolic values. These findings concur with those of some studies [1417]. Auto-analysis precontrast and postcontrast non-rigid image registration alignment for ECV analysis in both the heart rate 40–70 bpm group, and heart rate 71–120 bpm and atrial fibrillation groups was significantly more successful during systole. The success of non-rigid image registration alignment is highly dependent on the parallel shift amount, rotation shift amount, and scaling deformation amount. In systole, the left ventricular myocardium maintains its thickness and the partial volume effect is reduced, and calculation errors resulting from movements are reduced. Accordingly, successfully by auto-analysis, non-rigid image registration alignment improves the accuracy of ECV analysis because the subendocardial, mid-wall, and epicardial segment of the left ventricular myocardium matched precontrast and postcontrast images. A limitation of this study was the small study population. Previous consensus statements suggested that 20 is the sample size required to obtain a power of normal values for T1 mapping [23]. Therefore, the control data obtained from the 20 healthy volunteers in this study are probably reliable. Moreover, during diastole and systole, the acquisition section for T1 mapping may appear different because of the myocardium strain. In the future, further studies on not only the short-axis section but also the two- and four-chamber sections will be necessary.

5 Conclusion

Systolic MOLLI myocardial T1 mapping improves the accuracy of T1 values and ECV analysis by reducing calculation errors resulting from movements. It is applicable to patients with low–high heart rates and atrial fibrillation.

Acknowledgements

This manuscript was partly supported by Akiyoshi Ohtsuka Fellowship of the Japanese Society of Radiological Technology for improvement in English expression of a draft version of the manuscript.

Compliance with ethical standards

Conflict of Interest

The authors declare that they have no conflict of interest.

Research involving human participants

All procedures performed in studies involving human participants were in accordance with the ethical standards of the Institutional Review Board and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This article does not contain any studies performed with animals.
Informed consent was obtained from all individual participants included in the study.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat McCrohon JA, Moon JCC, Prasad SK, McKenna WJ, Lorenz CH, Coats AJS, Pennell DJ. Differentiation of heart failure related to dilated cardiomyopathy and coronary artery disease using gadolinium-enhanced cardiovascular magnetic resonance. Circulation. 2003;108:54–9. CrossRef McCrohon JA, Moon JCC, Prasad SK, McKenna WJ, Lorenz CH, Coats AJS, Pennell DJ. Differentiation of heart failure related to dilated cardiomyopathy and coronary artery disease using gadolinium-enhanced cardiovascular magnetic resonance. Circulation. 2003;108:54–9. CrossRef
2.
Zurück zum Zitat Bulluck H, Maestrini V, Rosmini S, Abdel-Gadir A, Treibel TA, Castelletti S, Bucciarelli-Ducci C, Manisty C, Moon JC. Myocardial T1 mapping. Circ J. 2015;79:487–94. CrossRef Bulluck H, Maestrini V, Rosmini S, Abdel-Gadir A, Treibel TA, Castelletti S, Bucciarelli-Ducci C, Manisty C, Moon JC. Myocardial T1 mapping. Circ J. 2015;79:487–94. CrossRef
3.
Zurück zum Zitat Bulluck H, Hammond-Haley M, Fontana M, Knight DS, Sirker A, Herrey AS, Manisty C, Kellman P, Moon JC, Hausenloy DJ. Quantification of both the area-at-risk and acute myocardial infarct size in ST-segment elevation myocardial infarction using T1-mapping. J Cardiovasc Magn Reson. 2017;19:57. CrossRef Bulluck H, Hammond-Haley M, Fontana M, Knight DS, Sirker A, Herrey AS, Manisty C, Kellman P, Moon JC, Hausenloy DJ. Quantification of both the area-at-risk and acute myocardial infarct size in ST-segment elevation myocardial infarction using T1-mapping. J Cardiovasc Magn Reson. 2017;19:57. CrossRef
4.
Zurück zum Zitat Taylor AJ, Salerno M, Dharmakumar R, Jerosch-Herold M. T1 mapping: basic techniques and clinical applications. JACC Cardiovasc Imaging. 2016;9:67–81. CrossRef Taylor AJ, Salerno M, Dharmakumar R, Jerosch-Herold M. T1 mapping: basic techniques and clinical applications. JACC Cardiovasc Imaging. 2016;9:67–81. CrossRef
5.
Zurück zum Zitat Ferreira VM, Piechnik SK, Dall’Armellina E, Karamitsos TD, Francis JM, Ntusi N, Holloway C, Choudhury RP, Kardos A, Robson MD, Friedrich MG. Neubauer S Native T1-mapping detects the location, extent and patterns of acute myocarditis without the need for gadolinium contrast agents. J Cardiovasc Magn Reson. 2014;16:36. CrossRef Ferreira VM, Piechnik SK, Dall’Armellina E, Karamitsos TD, Francis JM, Ntusi N, Holloway C, Choudhury RP, Kardos A, Robson MD, Friedrich MG. Neubauer S Native T1-mapping detects the location, extent and patterns of acute myocarditis without the need for gadolinium contrast agents. J Cardiovasc Magn Reson. 2014;16:36. CrossRef
6.
Zurück zum Zitat Puntmann VO, Peker E, Chandrashekhar Y, Nagel E. T1 mapping in characterizing myocardial disease: a comprehensive review. Circ Res. 2016;119:277–99. CrossRef Puntmann VO, Peker E, Chandrashekhar Y, Nagel E. T1 mapping in characterizing myocardial disease: a comprehensive review. Circ Res. 2016;119:277–99. CrossRef
7.
Zurück zum Zitat Pagano JJ, Chow K, Khan A, Michelakis E, Paterson I, Oudit GY, Thompson RB. Reduced right ventricular native myocardial T1 in An-derson-fabry disease: comparison to pulmonary hypertension and healthy controls. PLoS ONE. 2016;11:e0157565. CrossRef Pagano JJ, Chow K, Khan A, Michelakis E, Paterson I, Oudit GY, Thompson RB. Reduced right ventricular native myocardial T1 in An-derson-fabry disease: comparison to pulmonary hypertension and healthy controls. PLoS ONE. 2016;11:e0157565. CrossRef
8.
Zurück zum Zitat Germain P, El Ghannudi S, Jeung MY, Ohlmann P, Epailly E, Roy C, Gangi A. Native T1 mapping of the heart–a pictorial review. Clin Med In-sights Cardiol. 2014;8:1–11. Germain P, El Ghannudi S, Jeung MY, Ohlmann P, Epailly E, Roy C, Gangi A. Native T1 mapping of the heart–a pictorial review. Clin Med In-sights Cardiol. 2014;8:1–11.
9.
Zurück zum Zitat Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, et al. Myocardial T1 mapping with MRI: comparison of Look-Locker and MOLLI sequences. J Magn Reson Imaging. 2011;34:1367–73. CrossRef Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, et al. Myocardial T1 mapping with MRI: comparison of Look-Locker and MOLLI sequences. J Magn Reson Imaging. 2011;34:1367–73. CrossRef
10.
Zurück zum Zitat Raman FS, Kawel-Boehm N, Gai N, Freed M, Han J, Liu CY, et al. Modified Look-Locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners. J Cardi-ovasc Magn Reson. 2013;15:64. CrossRef Raman FS, Kawel-Boehm N, Gai N, Freed M, Han J, Liu CY, et al. Modified Look-Locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners. J Cardi-ovasc Magn Reson. 2013;15:64. CrossRef
11.
Zurück zum Zitat Wang Y, Vidan E, Bergman GW. Cardiac motion of coronary arteries: variability in the rest period and implications for coronary MR angiography. Radiology. 1999;213:751–8. CrossRef Wang Y, Vidan E, Bergman GW. Cardiac motion of coronary arteries: variability in the rest period and implications for coronary MR angiography. Radiology. 1999;213:751–8. CrossRef
12.
Zurück zum Zitat Arai H, Kawakubo M, Sanui K, et al. Accurate and robust systolic myocardial T1 mapping using saturation recovery with individualized delay time: comparison with diastolic T1 mapping. Radiol Phys Technol. 2018;11:415–22. CrossRef Arai H, Kawakubo M, Sanui K, et al. Accurate and robust systolic myocardial T1 mapping using saturation recovery with individualized delay time: comparison with diastolic T1 mapping. Radiol Phys Technol. 2018;11:415–22. CrossRef
13.
Zurück zum Zitat Tessa C, Diciotti S, Landini N, Lilli A, Del Meglio J, Salvatori L, Giannelli M, Greiser A, Vignali C, Casolo G. Myocardial T1 and T2 mapping in diastolic and systolic phase. Int J Cardiovasc Imaging. 2015;31:1001–10. CrossRef Tessa C, Diciotti S, Landini N, Lilli A, Del Meglio J, Salvatori L, Giannelli M, Greiser A, Vignali C, Casolo G. Myocardial T1 and T2 mapping in diastolic and systolic phase. Int J Cardiovasc Imaging. 2015;31:1001–10. CrossRef
14.
Zurück zum Zitat Reiter U, Reiter G, Dorr K, Greiser A, Mader-thaner R, Fuchsjager M. Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology. 2014;271:365–72. CrossRef Reiter U, Reiter G, Dorr K, Greiser A, Mader-thaner R, Fuchsjager M. Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology. 2014;271:365–72. CrossRef
15.
Zurück zum Zitat Ferreira VM, Wijesurendra RS, Liu A, Greiser A, Casadei B, Robson MD, Neubauer S, Piechnik SK. Systolic ShMOLLI myocardial T1-mapping for improved robustness to partial-volume effects and applications in tachyarrhythmias. J Cardiovasc Magn Reson. 2015;17:77. CrossRef Ferreira VM, Wijesurendra RS, Liu A, Greiser A, Casadei B, Robson MD, Neubauer S, Piechnik SK. Systolic ShMOLLI myocardial T1-mapping for improved robustness to partial-volume effects and applications in tachyarrhythmias. J Cardiovasc Magn Reson. 2015;17:77. CrossRef
16.
Zurück zum Zitat Kellman Peter. Michael S Hansen T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson. 2014;16(1):2. CrossRef Kellman Peter. Michael S Hansen T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson. 2014;16(1):2. CrossRef
17.
Zurück zum Zitat Kellman Peter. Andrew E Arai and Hui Xue T1 and extracellular volume mapping in the heart: estimation of error maps and the influence of noise on precision. J Cardiovasc Magn Reson. 2013;15:56. CrossRef Kellman Peter. Andrew E Arai and Hui Xue T1 and extracellular volume mapping in the heart: estimation of error maps and the influence of noise on precision. J Cardiovasc Magn Reson. 2013;15:56. CrossRef
18.
Zurück zum Zitat Messroghli DR, Radjenovic A, Kozerke S, Hig-gins DM, Sivananthan MU, Ridgway JP. Modified look-locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52:141–6. CrossRef Messroghli DR, Radjenovic A, Kozerke S, Hig-gins DM, Sivananthan MU, Ridgway JP. Modified look-locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52:141–6. CrossRef
19.
Zurück zum Zitat Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH. Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012;33(10):1268–78. CrossRef Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH. Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012;33(10):1268–78. CrossRef
20.
Zurück zum Zitat Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transpl. 2013;48:452–8. CrossRef Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transpl. 2013;48:452–8. CrossRef
21.
Zurück zum Zitat Zhao L, Li S, Ma X, Greiser A, Zhang T, An J, Bai R, Dong J, Fan Z. Systolic MOLLI T1 mapping with heart-rate-dependent pulse sequence sampling scheme is feasible in patients with atrial fibrillation. J Cardiovasc Magn Reson. 2016;18:13. CrossRef Zhao L, Li S, Ma X, Greiser A, Zhang T, An J, Bai R, Dong J, Fan Z. Systolic MOLLI T1 mapping with heart-rate-dependent pulse sequence sampling scheme is feasible in patients with atrial fibrillation. J Cardiovasc Magn Reson. 2016;18:13. CrossRef
22.
Zurück zum Zitat Chrysostomakis SI, Karalis IK, Simantirakis EN, et al. Angiotensin II type 1 receptor inhibition is associated with reduced tachyarrhythmia-induced ventricular interstitial fibrosis in a goat atrial fibrillation model. Cardiovasc Drugs Ther. 2007;21:357–65. CrossRef Chrysostomakis SI, Karalis IK, Simantirakis EN, et al. Angiotensin II type 1 receptor inhibition is associated with reduced tachyarrhythmia-induced ventricular interstitial fibrosis in a goat atrial fibrillation model. Cardiovasc Drugs Ther. 2007;21:357–65. CrossRef
23.
Zurück zum Zitat Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, Gatehouse PD, Arai AE, Friedrich MG, Neubauer S, Schulz-Menger J, Schelbert EB. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92. CrossRef Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, Gatehouse PD, Arai AE, Friedrich MG, Neubauer S, Schulz-Menger J, Schelbert EB. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92. CrossRef
Metadaten
Titel
Systolic modified Look–Locker inversion recovery myocardial T1 mapping improves the accuracy of T1 and extracellular volume fraction measurements of patients with high heart rate or atrial fibrillation
verfasst von
Hirohiko Shinbo
Satoshi Tomioka
Toshihiko Ino
Keiko Koyama
Publikationsdatum
05.11.2020
Verlag
Springer Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 4/2020
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-020-00594-z

Weitere Artikel der Ausgabe 4/2020

Radiological Physics and Technology 4/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.