01.04.2011 | Interventional Neuroradiology

The effect of stents on intra-aneurysmal hemodynamics: in vitro evaluation of a pulsatile sidewall aneurysm using laser Doppler anemometry

verfasst von: Franziska Dorn, Franz Niedermeyer, Andrea Balasso, Dieter Liepsch, Thomas Liebig

Erschienen in: Neuroradiology | Ausgabe 4/2011

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Hemodynamic modification by means of flow diversion is increasingly used for treatment of intracranial aneurysms. Despite of promising results, there is still a paucity of methods to reliably predict long-term success of this technique. Laser Doppler anemometry (LDA) can be used to quantify the influence of stents on intra-aneurysmal flow in vitro.

Methods

All experiments were performed with a pulsatile model of a sidewall aneurysm. A physiologic flow was created with a circulatory experimental setup, and a transparent non-Newtonian glycerol–water solution was used to substitute human blood. Flow velocity was measured with a one-component LDA system, recording flow components parallel and perpendicular to the parent vessel. Three different stents (Solitaire, Silk, Phenox flow diverter) were deployed over the aneurysm neck, respectively.

Results

Flow reduction was 67.59% (inflow zone), 9.65% (dome) and 37.94% (outflow zone) by the Solitaire stent. The Silk stent reduced the flow by 58.15% (inflow zone), 89.06% (dome) and 90.06% (outflow zone). The Phenox flow diverter reduced the flow by 96.76% (inflow zone), 90% (dome) and 90.91% (outflow zone) when positioned with narrow stent struts but increased the velocity of up to seven times compared to the unstented model when placed with loose strut packing in the proximal part of the aneurysm.

Conclusion

LDA is a feasible method to quantify intra-aneurysmal flow and flow reduction efficacy of stents in vitro. Flow reduction was negligible with a standard self-expanding stent. For dedicated flow diverters, it depended both on stent design and on appropriate positioning.
Literatur
1.
Zurück zum Zitat Ahlhelm F, Roth C, Kaufmann R, Schulte-Altedorneburg G, Romeike BF, Reith W (2007) Treatment of wide-necked intracranial a-neurysms with a novel self-expanding two-zonal endovascular stent device. Neuroradiology 49(12):1023–1028PubMedCrossRef Ahlhelm F, Roth C, Kaufmann R, Schulte-Altedorneburg G, Romeike BF, Reith W (2007) Treatment of wide-necked intracranial a-neurysms with a novel self-expanding two-zonal endovascular stent device. Neuroradiology 49(12):1023–1028PubMedCrossRef
2.
Zurück zum Zitat Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA (2007) Cloft HJ.A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38(8):2346–2352PubMedCrossRef Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA (2007) Cloft HJ.A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38(8):2346–2352PubMedCrossRef
3.
Zurück zum Zitat Vanninen R, Manninen H, Ronkainen A (2003) Broad-based intracranial aneurysms: thrombosis induced by stent placement. AJNR Am J Neuroradiol 24(2):263–266PubMed Vanninen R, Manninen H, Ronkainen A (2003) Broad-based intracranial aneurysms: thrombosis induced by stent placement. AJNR Am J Neuroradiol 24(2):263–266PubMed
4.
Zurück zum Zitat Pumar JM, Lete I, Pardo MI, Vázquez-Herrero F, Blanco M (2008) LEO stent monotherapy for the endovascular reconstruction of fusiform aneurysms of the middle cerebral artery. AJNR Am J Neuroradiol 29(9):1775–1776PubMedCrossRef Pumar JM, Lete I, Pardo MI, Vázquez-Herrero F, Blanco M (2008) LEO stent monotherapy for the endovascular reconstruction of fusiform aneurysms of the middle cerebral artery. AJNR Am J Neuroradiol 29(9):1775–1776PubMedCrossRef
5.
Zurück zum Zitat Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, Berez AL, Tran Q, Nelson PK, Fiorella D (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64(4):632–642PubMedCrossRef Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, Berez AL, Tran Q, Nelson PK, Fiorella D (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64(4):632–642PubMedCrossRef
6.
Zurück zum Zitat Liepsch D, Moravec S, Baumgart R (1992) Some flow visualization and laser-Doppler-velocity measurements in a true-to-scale elastic model of a human aortic arch—a new model technique. Biorheology 29(5–6):563–580PubMed Liepsch D, Moravec S, Baumgart R (1992) Some flow visualization and laser-Doppler-velocity measurements in a true-to-scale elastic model of a human aortic arch—a new model technique. Biorheology 29(5–6):563–580PubMed
7.
Zurück zum Zitat Gobin YP, Counord JL, Flaud P, Duffaux J (1994) In vitro study of haemodynamics in a giant saccular aneurysm model; influence of flow dynamics in the parent vessel and effects of coil embolisation. Neuroradiology 36(7):530–536PubMedCrossRef Gobin YP, Counord JL, Flaud P, Duffaux J (1994) In vitro study of haemodynamics in a giant saccular aneurysm model; influence of flow dynamics in the parent vessel and effects of coil embolisation. Neuroradiology 36(7):530–536PubMedCrossRef
8.
Zurück zum Zitat Graves VB, Strother CM, Partington CR, Rappe A (1992) Flow dynamics of lateral carotid artery aneurysms and their effects on coils and balloons: an experimental study in dogs. AJNR Am J Neuroradiol 13(1):189–196PubMed Graves VB, Strother CM, Partington CR, Rappe A (1992) Flow dynamics of lateral carotid artery aneurysms and their effects on coils and balloons: an experimental study in dogs. AJNR Am J Neuroradiol 13(1):189–196PubMed
9.
Zurück zum Zitat Strother CM, Graves VB, Rappe A (1992) Aneurysm hemodynamics: an experimental study. AJNR Am J Neuroradiol 13(4):1089–1095PubMed Strother CM, Graves VB, Rappe A (1992) Aneurysm hemodynamics: an experimental study. AJNR Am J Neuroradiol 13(4):1089–1095PubMed
10.
Zurück zum Zitat Steiger HJ, Perktold K (1997) Computer modeling of intracranial saccular and lateral aneurysms for the study of their hemodynamics. Neurosurgery 41(1):326–327PubMedCrossRef Steiger HJ, Perktold K (1997) Computer modeling of intracranial saccular and lateral aneurysms for the study of their hemodynamics. Neurosurgery 41(1):326–327PubMedCrossRef
11.
Zurück zum Zitat Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB (2009) An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model. Stroke 40(3):952–958PubMedCrossRef Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB (2009) An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model. Stroke 40(3):952–958PubMedCrossRef
12.
Zurück zum Zitat Seong J, Wakhloo AK, Lieber BB (2007) In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng 129(6):863–872PubMedCrossRef Seong J, Wakhloo AK, Lieber BB (2007) In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng 129(6):863–872PubMedCrossRef
13.
Zurück zum Zitat Lieber BB, Livescu V, Hopkins LN, Wakhloo AK (2002) Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow. Ann Biomed Eng 30(6):768–777PubMedCrossRef Lieber BB, Livescu V, Hopkins LN, Wakhloo AK (2002) Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow. Ann Biomed Eng 30(6):768–777PubMedCrossRef
14.
Zurück zum Zitat Ford MD, Nikolov HN, Milner JS, Lownie SP, Demont EM, Kalata W, Loth F, Holdsworth DW, Steinman DA (2008) PIV-measured versus CFD-predicted flow dynamics in anatomically realistic cerebral aneurysm models. J Biomech Eng 130(2):021015PubMedCrossRef Ford MD, Nikolov HN, Milner JS, Lownie SP, Demont EM, Kalata W, Loth F, Holdsworth DW, Steinman DA (2008) PIV-measured versus CFD-predicted flow dynamics in anatomically realistic cerebral aneurysm models. J Biomech Eng 130(2):021015PubMedCrossRef
15.
Zurück zum Zitat Trager AL, Sadasivan C, Seong J, Lieber BB (2009) Correlation between angiographic and particle image velocimetry quantifications of flow diverters in an in vitro model of elastase-induced rabbit aneurysms. J Biomech Eng 131(3):034506PubMedCrossRef Trager AL, Sadasivan C, Seong J, Lieber BB (2009) Correlation between angiographic and particle image velocimetry quantifications of flow diverters in an in vitro model of elastase-induced rabbit aneurysms. J Biomech Eng 131(3):034506PubMedCrossRef
16.
Zurück zum Zitat Augsburger L, Farhat M, Reymond P, Fonck E, Kulcsar Z, Stergiopulos N, Rüfenacht DA (2009) Effect of flow diverter porosity on intraaneurysmal blood flow. Klin Neuroradiol 19(3):204–214PubMedCrossRef Augsburger L, Farhat M, Reymond P, Fonck E, Kulcsar Z, Stergiopulos N, Rüfenacht DA (2009) Effect of flow diverter porosity on intraaneurysmal blood flow. Klin Neuroradiol 19(3):204–214PubMedCrossRef
17.
Zurück zum Zitat Shojima M, Oshima M, Takagi K et al (2004) Magnitude and role of wall shear stress on cerebral aneurysm: computational fluid dynamic study of 20 middle cerebral artery aneurysms. Stroke 35(11):2500–2505PubMedCrossRef Shojima M, Oshima M, Takagi K et al (2004) Magnitude and role of wall shear stress on cerebral aneurysm: computational fluid dynamic study of 20 middle cerebral artery aneurysms. Stroke 35(11):2500–2505PubMedCrossRef
18.
Zurück zum Zitat Steinman DA, Milner JS, Norley CJ, Lownie SP, Holdsworth DW (2003) Image-based computational simulation of flow dynamics in a giant intracranial aneurysm. AJNR Am J Neuroradiol 24(4):559–566PubMed Steinman DA, Milner JS, Norley CJ, Lownie SP, Holdsworth DW (2003) Image-based computational simulation of flow dynamics in a giant intracranial aneurysm. AJNR Am J Neuroradiol 24(4):559–566PubMed
19.
Zurück zum Zitat Bousel L, Rayz V, McCulloch C et al (2008) Aneurysm growth occurs at region of low wall shear stress: patient-specific correlation of hemodynamics and growth in a longitudinal study. Stroke 39(11):2997–3002CrossRef Bousel L, Rayz V, McCulloch C et al (2008) Aneurysm growth occurs at region of low wall shear stress: patient-specific correlation of hemodynamics and growth in a longitudinal study. Stroke 39(11):2997–3002CrossRef
20.
Zurück zum Zitat Castro MA, Putman CM, Cebral JR (2006) Computational fluid dynamics modeling of intracranial aneurysms: effects of parent artery segmentation on intra-aneurysmal hemodynamics. AJNR Am J Neuroradiol 27(8):1703–1709PubMed Castro MA, Putman CM, Cebral JR (2006) Computational fluid dynamics modeling of intracranial aneurysms: effects of parent artery segmentation on intra-aneurysmal hemodynamics. AJNR Am J Neuroradiol 27(8):1703–1709PubMed
21.
Zurück zum Zitat Cebral JR, Castro MA, Appanaboyina S, Putman CM, Millan D, Frangi A (2005) Efficient pipeline for image-based patient-specific analysis of cerebral aneurysm hemodynamics: technique and sensitivity. IEEE Trans Med Imaging 24(4):457–467PubMedCrossRef Cebral JR, Castro MA, Appanaboyina S, Putman CM, Millan D, Frangi A (2005) Efficient pipeline for image-based patient-specific analysis of cerebral aneurysm hemodynamics: technique and sensitivity. IEEE Trans Med Imaging 24(4):457–467PubMedCrossRef
22.
Zurück zum Zitat Cebral JR, Castro MA, Burgess JE, Pergolizzi RS, Sheridan MJ, Putman CM (2005) Characterization of cerebral aneurysms for assessing risk of rupture by using patient-specific computational hemodynamics models. Am J Neuroradiol 26(10):2550–2559PubMed Cebral JR, Castro MA, Burgess JE, Pergolizzi RS, Sheridan MJ, Putman CM (2005) Characterization of cerebral aneurysms for assessing risk of rupture by using patient-specific computational hemodynamics models. Am J Neuroradiol 26(10):2550–2559PubMed
23.
Zurück zum Zitat Mantha A, Karmonik C, Benndorf G, Strother C, Metcalfe R (2006) Hemodynamics in a cerebral artery before and after the formation of an aneurysm. AJNR Am J Neuroradiol 27(5):1113–1118PubMed Mantha A, Karmonik C, Benndorf G, Strother C, Metcalfe R (2006) Hemodynamics in a cerebral artery before and after the formation of an aneurysm. AJNR Am J Neuroradiol 27(5):1113–1118PubMed
24.
Zurück zum Zitat Benndorf G, Ionescu M (2009) Y Alvarado MV, Hipp J, Metcalfe R. Wall shear stress in intracranial self-expanding stents studied using ultra-high-resolution 3D reconstructions. AJNR Am J Neuroradiol 30(3):479–486PubMedCrossRef Benndorf G, Ionescu M (2009) Y Alvarado MV, Hipp J, Metcalfe R. Wall shear stress in intracranial self-expanding stents studied using ultra-high-resolution 3D reconstructions. AJNR Am J Neuroradiol 30(3):479–486PubMedCrossRef
25.
Zurück zum Zitat Larrabide I, Radaelli A, Frangi A (2008) Fast virtual stenting with deformable meshes: application to intracranial aneurysms. Med Image Comput Comput Assist Interv 11(Pt 2):790–797PubMed Larrabide I, Radaelli A, Frangi A (2008) Fast virtual stenting with deformable meshes: application to intracranial aneurysms. Med Image Comput Comput Assist Interv 11(Pt 2):790–797PubMed
26.
Zurück zum Zitat Ohta M, Wetzel SG, Dantan P, Bachelet C, Lovblad KO, Yilmaz H, Flaud P, Rüfenacht DA (2005) Rheological changes after stenting of an cerebral aneurysm: a finite element modeling approach. Cardiovasc Intervent Radiol 28(6):768–772PubMedCrossRef Ohta M, Wetzel SG, Dantan P, Bachelet C, Lovblad KO, Yilmaz H, Flaud P, Rüfenacht DA (2005) Rheological changes after stenting of an cerebral aneurysm: a finite element modeling approach. Cardiovasc Intervent Radiol 28(6):768–772PubMedCrossRef
27.
Zurück zum Zitat Hirabayashi M, Ohta M, Rüfenacht DA, Chopard B (2003) Characterization of flow reduction properties in an aneurysm due to a stent. Phys Rev E Stat Nonlin Soft Matter Phys 68(2 Pt 1):021918PubMedCrossRef Hirabayashi M, Ohta M, Rüfenacht DA, Chopard B (2003) Characterization of flow reduction properties in an aneurysm due to a stent. Phys Rev E Stat Nonlin Soft Matter Phys 68(2 Pt 1):021918PubMedCrossRef
Metadaten
Titel
The effect of stents on intra-aneurysmal hemodynamics: in vitro evaluation of a pulsatile sidewall aneurysm using laser Doppler anemometry
verfasst von
Franziska Dorn
Franz Niedermeyer
Andrea Balasso
Dieter Liepsch
Thomas Liebig
Publikationsdatum
01.04.2011
Verlag
Springer-Verlag
Erschienen in
Neuroradiology / Ausgabe 4/2011
Print ISSN: 0028-3940
Elektronische ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-010-0723-4

Weitere Artikel der Ausgabe 4/2011

Neuroradiology 4/2011 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Was nützt die Kraniektomie bei schwerer tiefer Hirnblutung?

17.05.2024 Hirnblutung Nachrichten

Eine Studie zum Nutzen der druckentlastenden Kraniektomie nach schwerer tiefer supratentorieller Hirnblutung deutet einen Nutzen der Operation an. Für überlebende Patienten ist das dennoch nur eine bedingt gute Nachricht.

Thrombektomie auch bei großen Infarkten von Vorteil

16.05.2024 Ischämischer Schlaganfall Nachrichten

Auch ein sehr ausgedehnter ischämischer Schlaganfall scheint an sich kein Grund zu sein, von einer mechanischen Thrombektomie abzusehen. Dafür spricht die LASTE-Studie, an der Patienten und Patientinnen mit einem ASPECTS von maximal 5 beteiligt waren.

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.