Abstract

Objective

To investigate the use of retrievable inferior vena cava (IVC) filters in the current practice and analyze the causes of filter left in situ despite initial plans for retrieval.

Methods

A systematic search of all English-language studies of retrievable IVC filters was conducted, including clinical trials and observational studies published between January 1984 and March 2016.

Results

A total of 103 studies were identified, including 10 filter models in 20,319 patients. Placement indications were reported for 11,128 (54.8%) patients, including therapeutic (n = 6270; 56.3%) and prophylactic (n = 4858; 43.7%) indications. A total of 13,224 (65.1%) filters were left as permanent devices; 7095 (34.9%) filters were removed. The reasons for filter nonretrieval among the 5308 (40.1%) reported cases were primary permanent indication (21.2%; 1127/5308), death (19.4%; 1031/5308), ongoing pulmonary embolism (PE) protection (19.0%; 1011/5308), failed retrieval (13.7%; 725/5308), loss to follow-up (13.0%; 689/5308), discontinued care (4.4%; 235/5308), physician oversight (4.0%; 213/5308), patient morbidity (2.8%; 149/5308), and patient refusal (2.4%; 128/5308). A total of 7820 patients presented for filter retrieval, and 7095 filters (90.7%) were successfully removed, with a mean indwelling time of 106.6 ± 47.3 days. Breakthrough PE was reported in 2.1% (191/9169) of patients. Filter tilt, recurrent deep vein thrombosis, penetration, IVC thrombosis, migration, and fracture occurred in 7.7% (798/10,348), 7.1% (362/5092), 5.4% (379/7001), 3.9% (345/8788), 1.4% (160/11,679), and 0.5% (50/9509) of patients, respectively.

Conclusions

Approximately two-thirds of retrievable filters were not retrieved even though more than 85% of the filters were initially intended for temporary use. The major reasons for filter left in situ despite initial plans for retrieval were death, need for ongoing PE protection, failed retrieval, loss to follow-up, discontinued care, and physician oversight.
Literatur
1.
Zurück zum Zitat Grassi C, Swan T, Cardella J, Meranze S, Oglevie S, Omary R, Roberts A, Sacks D, Silverstein M, Towbin R. Quality improvement guidelines for percutaneous permanent inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol. 2001;12:137–41.CrossRefPubMed Grassi C, Swan T, Cardella J, Meranze S, Oglevie S, Omary R, Roberts A, Sacks D, Silverstein M, Towbin R. Quality improvement guidelines for percutaneous permanent inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol. 2001;12:137–41.CrossRefPubMed
2.
Zurück zum Zitat Angel LF, Tapson V, Galgon RE, Restrepo MI, Kaufman J. Systematic review of the use of retrievable inferior vena cava filters. J Vasc Interv Radiol. 2011;22:1522–30.CrossRefPubMed Angel LF, Tapson V, Galgon RE, Restrepo MI, Kaufman J. Systematic review of the use of retrievable inferior vena cava filters. J Vasc Interv Radiol. 2011;22:1522–30.CrossRefPubMed
3.
Zurück zum Zitat Malgor RD, Labropoulos N. A systematic review of symptomatic duodenal perforation by inferior vena cava filters. J Vasc Surg. 2012;55:856–61.CrossRefPubMed Malgor RD, Labropoulos N. A systematic review of symptomatic duodenal perforation by inferior vena cava filters. J Vasc Surg. 2012;55:856–61.CrossRefPubMed
4.
Zurück zum Zitat Caplin DM, Nikolic B, Kalva SP, Ganguli S, Saad WE, Zuckerman DA, Society of Interventional Radiology Standards of Practice C. Quality improvement guidelines for the performance of inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol. 2011;22:1499–506.PubMed Caplin DM, Nikolic B, Kalva SP, Ganguli S, Saad WE, Zuckerman DA, Society of Interventional Radiology Standards of Practice C. Quality improvement guidelines for the performance of inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol. 2011;22:1499–506.PubMed
5.
Zurück zum Zitat Molvar C. Inferior vena cava filtration in the management of venous thromboembolism: filtering the data. Semin Interv Radiol. 2012;29:204–17.CrossRef Molvar C. Inferior vena cava filtration in the management of venous thromboembolism: filtering the data. Semin Interv Radiol. 2012;29:204–17.CrossRef
6.
Zurück zum Zitat Yunus TE, Tariq N, Callahan RE, Niemeyer DJ, Brown OW, Zelenock GB, Shanley CJ. Changes in inferior vena cava filter placement over the past decade at a large community-based academic health center. J Vasc Surg. 2008;47:157–65.CrossRefPubMed Yunus TE, Tariq N, Callahan RE, Niemeyer DJ, Brown OW, Zelenock GB, Shanley CJ. Changes in inferior vena cava filter placement over the past decade at a large community-based academic health center. J Vasc Surg. 2008;47:157–65.CrossRefPubMed
7.
Zurück zum Zitat Moher D, Liberati A, Tetzlaff J, Altman DG, Prisma G. Preferred reporting items for systematic reviews and meta-analyses: the prisma statement. PLoS Med. 2009;6:e1000097.CrossRefPubMedPubMedCentral Moher D, Liberati A, Tetzlaff J, Altman DG, Prisma G. Preferred reporting items for systematic reviews and meta-analyses: the prisma statement. PLoS Med. 2009;6:e1000097.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Morales JP, Li X, Irony TZ, Ibrahim NG, Moynahan M, Cavanaugh KJ. Decision analysis of retrievable inferior vena cava filters in patients without pulmonary embolism. J Vasc Surg Venous Lymphat Disord. 2013;1:376–84.CrossRefPubMed Morales JP, Li X, Irony TZ, Ibrahim NG, Moynahan M, Cavanaugh KJ. Decision analysis of retrievable inferior vena cava filters in patients without pulmonary embolism. J Vasc Surg Venous Lymphat Disord. 2013;1:376–84.CrossRefPubMed
9.
Zurück zum Zitat Tsui B, An T, Moon E, King R, Wang W. Retrospective review of 516 implantations of option inferior vena cava filters at a single health care system. J Vasc Interv Radiol. 2016;27:345–53.CrossRefPubMed Tsui B, An T, Moon E, King R, Wang W. Retrospective review of 516 implantations of option inferior vena cava filters at a single health care system. J Vasc Interv Radiol. 2016;27:345–53.CrossRefPubMed
10.
Zurück zum Zitat Zhu X, Tam MD, Bartholomew J, Newman JS, Sands MJ, Wang W. Retrievability and device-related complications of the G2 filter: a retrospective study of 139 filter retrievals. J Vasc Interv Radiol. 2011;22:806–12.CrossRefPubMed Zhu X, Tam MD, Bartholomew J, Newman JS, Sands MJ, Wang W. Retrievability and device-related complications of the G2 filter: a retrospective study of 139 filter retrievals. J Vasc Interv Radiol. 2011;22:806–12.CrossRefPubMed
11.
Zurück zum Zitat Ray CE Jr, Mitchell E, Zipser S, Kao EY, Brown CF, Moneta GL. Outcomes with retrievable inferior vena cava filters: a multicenter study. J Vasc Interv Radiol. 2006;17:1595–604.CrossRefPubMed Ray CE Jr, Mitchell E, Zipser S, Kao EY, Brown CF, Moneta GL. Outcomes with retrievable inferior vena cava filters: a multicenter study. J Vasc Interv Radiol. 2006;17:1595–604.CrossRefPubMed
12.
Zurück zum Zitat Charles HW, Black M, Kovacs S, Gohari A, Arampulikan J, McCann JW, Clark TW, Bashar M, Steiger D. G2 inferior vena cava filter: retrievability and safety. J Vasc Interv Radiol. 2009;20:1046–51.CrossRefPubMed Charles HW, Black M, Kovacs S, Gohari A, Arampulikan J, McCann JW, Clark TW, Bashar M, Steiger D. G2 inferior vena cava filter: retrievability and safety. J Vasc Interv Radiol. 2009;20:1046–51.CrossRefPubMed
13.
Zurück zum Zitat Decousus H, Leizorovicz A, Parent F, Page Y, Tardy B, Girard P, Laporte S, Faivre R, Charbonnier B, Barral FG, Huet Y, Simonneau G. A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis. N Engl J Med. 1998;338:409–15.CrossRefPubMed Decousus H, Leizorovicz A, Parent F, Page Y, Tardy B, Girard P, Laporte S, Faivre R, Charbonnier B, Barral FG, Huet Y, Simonneau G. A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis. N Engl J Med. 1998;338:409–15.CrossRefPubMed
14.
Zurück zum Zitat Jia Z, Wu A, Tam M, Spain J, McKinney JM, Wang W. Caval penetration by inferior vena cava filters: a systematic literature review of clinical significance and management. Circulation. 2015;132:944–52.CrossRefPubMed Jia Z, Wu A, Tam M, Spain J, McKinney JM, Wang W. Caval penetration by inferior vena cava filters: a systematic literature review of clinical significance and management. Circulation. 2015;132:944–52.CrossRefPubMed
15.
Zurück zum Zitat Darcy MD, Cardella JF, Hunter DW, Smith TP, Castaneda-Zuniga WR, Lund G, Amplatz K. Experience with the amplatz retrievable vena caval filter. Work in progress. Radiology. 1986;161:611–4.CrossRefPubMed Darcy MD, Cardella JF, Hunter DW, Smith TP, Castaneda-Zuniga WR, Lund G, Amplatz K. Experience with the amplatz retrievable vena caval filter. Work in progress. Radiology. 1986;161:611–4.CrossRefPubMed
16.
Zurück zum Zitat Lee MJ, Valenti D, de Gregorio MA, Minocha J, Rimon U, Pellerin O. The cirse retrievable ivc filter registry: retrieval success rates in practice. Cardiovasc Interv Radiol. 2015;38:1502–7.CrossRef Lee MJ, Valenti D, de Gregorio MA, Minocha J, Rimon U, Pellerin O. The cirse retrievable ivc filter registry: retrieval success rates in practice. Cardiovasc Interv Radiol. 2015;38:1502–7.CrossRef
17.
Zurück zum Zitat Hammond CJ, Bakshi DR, Currie RJ, Patel JV, Kinsella D, McWilliams RG, Watkinson A, Nicholson AA. Audit of the use of IVC filters in the UK: experience from three centres over 12 years. Clin Radiol. 2009;64:502–10.CrossRefPubMed Hammond CJ, Bakshi DR, Currie RJ, Patel JV, Kinsella D, McWilliams RG, Watkinson A, Nicholson AA. Audit of the use of IVC filters in the UK: experience from three centres over 12 years. Clin Radiol. 2009;64:502–10.CrossRefPubMed
18.
Zurück zum Zitat Peterson EA, Yenson PR, Liu D, Lee AY. Predictors of attempted inferior vena cava filters retrieval in a tertiary care centre. Thromb Res. 2014;134:300–4.CrossRefPubMed Peterson EA, Yenson PR, Liu D, Lee AY. Predictors of attempted inferior vena cava filters retrieval in a tertiary care centre. Thromb Res. 2014;134:300–4.CrossRefPubMed
19.
Zurück zum Zitat Mansour A, Ismael Y, Abdel-Razeq H. Inferior vena cava filters in patients with advanced-stage cancer. Hematol Oncol Stem Cell Ther. 2014;7:136–41.CrossRefPubMed Mansour A, Ismael Y, Abdel-Razeq H. Inferior vena cava filters in patients with advanced-stage cancer. Hematol Oncol Stem Cell Ther. 2014;7:136–41.CrossRefPubMed
20.
Zurück zum Zitat Mahrer A, Zippel D, Garniek A, Golan G, Bensaid P, Simon D, Rimon U. Retrievable vena cava filters in major trauma patients: prevalence of thrombus within the filter. Cardiovasc Interv Radiol. 2008;31:785–9.CrossRef Mahrer A, Zippel D, Garniek A, Golan G, Bensaid P, Simon D, Rimon U. Retrievable vena cava filters in major trauma patients: prevalence of thrombus within the filter. Cardiovasc Interv Radiol. 2008;31:785–9.CrossRef
21.
Zurück zum Zitat Durack JC, Westphalen AC, Kekulawela S, Bhanu SB, Avrin DE, Gordon RL, Kerlan RK. Perforation of the IVC: rule rather than exception after longer indwelling times for the gunther tulip and celect retrievable filters. Cardiovasc Interv Radiol. 2012;35:299–308.CrossRef Durack JC, Westphalen AC, Kekulawela S, Bhanu SB, Avrin DE, Gordon RL, Kerlan RK. Perforation of the IVC: rule rather than exception after longer indwelling times for the gunther tulip and celect retrievable filters. Cardiovasc Interv Radiol. 2012;35:299–308.CrossRef
22.
Zurück zum Zitat Wang W, Zhou D, Obuchowski N, Spain J, An T. Fracture and migration of celect inferior vena cava filters: a retrospective review of 741 consecutive implantations. J Vasc Interv Radiol. 2013;24:1719–22.CrossRefPubMed Wang W, Zhou D, Obuchowski N, Spain J, An T. Fracture and migration of celect inferior vena cava filters: a retrospective review of 741 consecutive implantations. J Vasc Interv Radiol. 2013;24:1719–22.CrossRefPubMed
23.
Zurück zum Zitat Hull JE, Robertson SW. Bard recovery filter: evaluation and management of vena cava limb perforation, fracture, and migration. J Vasc Interv Radiol. 2009;20:52–60.CrossRefPubMed Hull JE, Robertson SW. Bard recovery filter: evaluation and management of vena cava limb perforation, fracture, and migration. J Vasc Interv Radiol. 2009;20:52–60.CrossRefPubMed
24.
Zurück zum Zitat Sutphin PD, Reis SP, McKune A, Ravanzo M, Kalva SP, Pillai AK. Improving inferior vena cava filter retrieval rates with the define, measure, analyze, improve, control methodology. J Vasc Interv Radiol. 2015;26:491–8.CrossRefPubMed Sutphin PD, Reis SP, McKune A, Ravanzo M, Kalva SP, Pillai AK. Improving inferior vena cava filter retrieval rates with the define, measure, analyze, improve, control methodology. J Vasc Interv Radiol. 2015;26:491–8.CrossRefPubMed
25.
Zurück zum Zitat Stoneburner Jr CGR, McKinney B, Wang W, Sella D, Paz-Fumagalli R, Frey G, Starke L, Gregory D, McKinney J. Implementing a system for inferior vena cava filter follow-up and retrieval: a quality improvement approach. In: The International symposium on endovascular therapy (ISET); 2016, p. 6–10. Stoneburner Jr CGR, McKinney B, Wang W, Sella D, Paz-Fumagalli R, Frey G, Starke L, Gregory D, McKinney J. Implementing a system for inferior vena cava filter follow-up and retrieval: a quality improvement approach. In: The International symposium on endovascular therapy (ISET); 2016, p. 6–10.
26.
Zurück zum Zitat Minocha J, Idakoji I, Riaz A, Karp J, Gupta R, Chrisman HB, Salem R, Ryu RK, Lewandowski RJ. Improving inferior vena cava filter retrieval rates: impact of a dedicated inferior vena cava filter clinic. J Vasc Interv Radiol. 2010;21:1847–51.CrossRefPubMed Minocha J, Idakoji I, Riaz A, Karp J, Gupta R, Chrisman HB, Salem R, Ryu RK, Lewandowski RJ. Improving inferior vena cava filter retrieval rates: impact of a dedicated inferior vena cava filter clinic. J Vasc Interv Radiol. 2010;21:1847–51.CrossRefPubMed
27.
Zurück zum Zitat Zhou D, Spain J, Moon E, McLennan G, Sands MJ, Wang W. Retrospective review of 120 celect inferior vena cava filter retrievals: experience at a single institution. J Vasc Interv Radiol. 2012;23:1557–63.CrossRefPubMed Zhou D, Spain J, Moon E, McLennan G, Sands MJ, Wang W. Retrospective review of 120 celect inferior vena cava filter retrievals: experience at a single institution. J Vasc Interv Radiol. 2012;23:1557–63.CrossRefPubMed
28.
Zurück zum Zitat Scher D, Venbrux A, Okapal K, Gabriel G, Dufour R, Chun A, Sarin S, Akman A. Retrieval of trapease and optease inferior vena cava filters with extended dwell times. J Vasc Interv Radiol. 2015;26:1519–25.CrossRefPubMed Scher D, Venbrux A, Okapal K, Gabriel G, Dufour R, Chun A, Sarin S, Akman A. Retrieval of trapease and optease inferior vena cava filters with extended dwell times. J Vasc Interv Radiol. 2015;26:1519–25.CrossRefPubMed
29.
Zurück zum Zitat Kuo WT, Odegaard JI, Rosenberg JK, Hofmann LV. Excimer laser-assisted removal of embedded inferior vena cava filters: a single-center prospective study. Circ Cardiovasc Interv. 2013;6:560–6.CrossRefPubMed Kuo WT, Odegaard JI, Rosenberg JK, Hofmann LV. Excimer laser-assisted removal of embedded inferior vena cava filters: a single-center prospective study. Circ Cardiovasc Interv. 2013;6:560–6.CrossRefPubMed
30.
Zurück zum Zitat Kuo WT, Cupp JS, Louie JD, Kothary N, Hofmann LV, Sze DY, Hovsepian DM. Complex retrieval of embedded ivc filters: alternative techniques and histologic tissue analysis. Cardiovasc Interv Radiol. 2012;35:588–97.CrossRef Kuo WT, Cupp JS, Louie JD, Kothary N, Hofmann LV, Sze DY, Hovsepian DM. Complex retrieval of embedded ivc filters: alternative techniques and histologic tissue analysis. Cardiovasc Interv Radiol. 2012;35:588–97.CrossRef
31.
Zurück zum Zitat Teo TKB, Angle JF, Shipp JI, Bluett MK, Gilliland CA, Turba UC, Matsumoto AH. Incidence and management of inferior vena cava filter thrombus detected at time of filter retrieval. J Vasc Interv Radiol. 2011;22:1514–20.CrossRefPubMed Teo TKB, Angle JF, Shipp JI, Bluett MK, Gilliland CA, Turba UC, Matsumoto AH. Incidence and management of inferior vena cava filter thrombus detected at time of filter retrieval. J Vasc Interv Radiol. 2011;22:1514–20.CrossRefPubMed
Metadaten
Titel
Utility of Retrievable Inferior Vena Cava Filters: A Systematic Literature Review and Analysis of the Reasons for Nonretrieval of Filters with Temporary Indications
verfasst von
Zhongzhi Jia
Tyler A. Fuller
J. Mark McKinney
Ricardo Paz-Fumagalli
Gregory T. Frey
David M. Sella
Thuong Van Ha
Weiping Wang
Publikationsdatum
22.01.2018
Verlag
Springer US
Erschienen in
CardioVascular and Interventional Radiology / Ausgabe 5/2018
Print ISSN: 0174-1551
Elektronische ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-018-1880-9

Weitere Artikel der Ausgabe 5/2018

CardioVascular and Interventional Radiology 5/2018 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.