A computer-simulated liver phantom (virtual liver phantom) for multidetector computed tomography evaluation

verfasst von: Yoshinori Funama, Kazuo Awai, Osamu Miyazaki, Yoshiharu Nakayama, Da Liu, Taiga Goto, Yasuyuki Yamashita, Shinichi Hori

Erschienen in: European Radiology | Ausgabe 4/2006

Einloggen, um Zugang zu erhalten

Abstract

Objective

The purpose of study was to develop a computer-simulated liver phantom for hepatic CT studies. A computer-simulated liver phantom was mathematically constructed on a computer workstation.

Materials and methods

The computer-simulated phantom was calibrated using real CT images acquired by an actual four-detector CT. We added an inhomogeneous texture to the simulated liver by referring to CT images of chronically damaged human livers. The mean CT number of the simulated liver was 60 HU and we added numerous 5-to 10-mm structures with 60±10 HU/mm. To mimic liver tumors we added nodules measuring 8, 10, and 12 mm in diameter with CT numbers of 60±10, 60±15, and 60±20 HU. Five radiologists visually evaluated similarity of the texture of the computer-simulated liver phantom and a real human liver to confirm the appropriateness of the virtual liver images using a five-point scale.

Results

The total score was 44 in two radiologists, and 42, 41, and 39 in one radiologist each. They evaluated that the textures of virtual liver were comparable to those of human liver.

Conclusions

Our computer-simulated liver phantom is a promising tool for the evaluation of the image quality and diagnostic performance of hepatic CT imaging.
Literatur
1.
Zurück zum Zitat Brix G, Lechel U, Veit R et al (2004) Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study. Eur Radiol 14:1275–1284 Brix G, Lechel U, Veit R et al (2004) Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study. Eur Radiol 14:1275–1284
2.
Zurück zum Zitat Begemann PG, van Stevendaal U, Manzke R et al (2005) Evaluation of spatial and temporal resolution for ECG-gated 16-row multidetector CT using a dynamic cardiac phantom. Eur Radiol 15:1015–1026 Begemann PG, van Stevendaal U, Manzke R et al (2005) Evaluation of spatial and temporal resolution for ECG-gated 16-row multidetector CT using a dynamic cardiac phantom. Eur Radiol 15:1015–1026
3.
Zurück zum Zitat Gurung J, Khan MF, Maataoui A et al (2005) Multislice CT of the pelvis: dose reduction with regard to image quality using 16-row CT. Eur Radiol 15:1898–1905 Gurung J, Khan MF, Maataoui A et al (2005) Multislice CT of the pelvis: dose reduction with regard to image quality using 16-row CT. Eur Radiol 15:1898–1905
4.
Zurück zum Zitat Hyodoh H, Katagiri Y, Sakai T, Hyodoh K, Akiba H, Hareyama M (2005) Creation of individual ideally shaped stents using multi-slice CT: in vitro results from the semi-automatic virtual stent (SAVS) designer. Eur Radiol 15:1623–1628 Hyodoh H, Katagiri Y, Sakai T, Hyodoh K, Akiba H, Hareyama M (2005) Creation of individual ideally shaped stents using multi-slice CT: in vitro results from the semi-automatic virtual stent (SAVS) designer. Eur Radiol 15:1623–1628
5.
Zurück zum Zitat Gupta AK, Nelson RC, Johnson GA et al (2003) Optimization of eight-element multi-detector row helical CT technology for evaluation of the abdomen. Radiology 227:739–745 Gupta AK, Nelson RC, Johnson GA et al (2003) Optimization of eight-element multi-detector row helical CT technology for evaluation of the abdomen. Radiology 227:739–745
6.
Zurück zum Zitat Verdun FR, Denys A, Valley JF et al (2002) Detection of low-contrast objects: experimental comparison of single-and multi-detector row CT with a phantom. Radiology 223:426–431 Verdun FR, Denys A, Valley JF et al (2002) Detection of low-contrast objects: experimental comparison of single-and multi-detector row CT with a phantom. Radiology 223:426–431
7.
Zurück zum Zitat Suess C, Kalender WA, Coman JM (1999) New low-contrast resolution phantoms for computed tomography. Med Phys 26:296–302 Suess C, Kalender WA, Coman JM (1999) New low-contrast resolution phantoms for computed tomography. Med Phys 26:296–302
8.
Zurück zum Zitat Muramatsu Y, Tsuda Y, Nakamura Y et al (2003) The development and use of a chest phantom for optimizing scanning techniques on a variety of low-dose helical computed tomography devices. J Comput Assist Tomogr 27:364–374 Muramatsu Y, Tsuda Y, Nakamura Y et al (2003) The development and use of a chest phantom for optimizing scanning techniques on a variety of low-dose helical computed tomography devices. J Comput Assist Tomogr 27:364–374
9.
Zurück zum Zitat Kalender WA, Wolf H, Suess C et al (1999) Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 9:323–328 Kalender WA, Wolf H, Suess C et al (1999) Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 9:323–328
10.
Zurück zum Zitat Frush DP, Slack CC, Hollingsworth CL et al (2002) Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients. AJR Am J Roentgenol 179:1107–1113 Frush DP, Slack CC, Hollingsworth CL et al (2002) Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients. AJR Am J Roentgenol 179:1107–1113
11.
Zurück zum Zitat Hu H (1999) Multi-slice helical CT: scan and reconstruction. Med Phys 26:5–18 Hu H (1999) Multi-slice helical CT: scan and reconstruction. Med Phys 26:5–18
12.
Zurück zum Zitat Sabiston DC (1997) Textbook of urgery, 15 edn. Saunders, Philadelphia Sabiston DC (1997) Textbook of urgery, 15 edn. Saunders, Philadelphia
13.
Zurück zum Zitat Gies M, Kalender WA, Wolf H et al (1999) Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys 26:2235–2247 Gies M, Kalender WA, Wolf H et al (1999) Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys 26:2235–2247
14.
Zurück zum Zitat Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ (2005) Multi-detector row CT systems and image-reconstruction techniques. Radiology 235:756–773 Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ (2005) Multi-detector row CT systems and image-reconstruction techniques. Radiology 235:756–773
15.
Zurück zum Zitat Obuchowski NA (2003) Receiver operating characteristic curves and their use in radiology. Radiology 229:3–8 Obuchowski NA (2003) Receiver operating characteristic curves and their use in radiology. Radiology 229:3–8
16.
Zurück zum Zitat Partan G, Mayrhofer R, Urban M et al. W (2003) Diagnostic performance of liquid crystal and cathode-ray-tube monitors in brain computed tomography. Eur Radiol 13:2397–2401 Partan G, Mayrhofer R, Urban M et al. W (2003) Diagnostic performance of liquid crystal and cathode-ray-tube monitors in brain computed tomography. Eur Radiol 13:2397–2401
17.
Zurück zum Zitat Mayo JR, Kim KI, MacDonald SL et al (2004) Reduced radiation dose helical chest CT: effect on reader evaluation of structures and lung findings. Radiology 232:749–756 Mayo JR, Kim KI, MacDonald SL et al (2004) Reduced radiation dose helical chest CT: effect on reader evaluation of structures and lung findings. Radiology 232:749–756
18.
Zurück zum Zitat Kalra MK, Maher MM, Toth TL et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628 Kalra MK, Maher MM, Toth TL et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628
19.
Zurück zum Zitat Hamberg LM, Rhea JT, Hunter GJ et al (2003) Multi-detector row CT: radiation dose characteristics. Radiology 226:762–772 Hamberg LM, Rhea JT, Hunter GJ et al (2003) Multi-detector row CT: radiation dose characteristics. Radiology 226:762–772
20.
Zurück zum Zitat Nickoloff EL, Dutta AK, Lu ZF (2003) Influence of phantom diameter, kVp and scan mode upon computed tomography dose index. Med Phys 30:395–402 Nickoloff EL, Dutta AK, Lu ZF (2003) Influence of phantom diameter, kVp and scan mode upon computed tomography dose index. Med Phys 30:395–402
21.
Zurück zum Zitat Sigal-Cinqualbre AB, Hennequin R, Abada HT et al (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231:169–174 Sigal-Cinqualbre AB, Hennequin R, Abada HT et al (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231:169–174
22.
Zurück zum Zitat Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435 Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435
23.
Zurück zum Zitat Chabat F, Yang GZ, Hansell DM (2003) Obstructive lung diseases: texture classification for differentiation at CT. Radiology 228:871–877 Chabat F, Yang GZ, Hansell DM (2003) Obstructive lung diseases: texture classification for differentiation at CT. Radiology 228:871–877
24.
Zurück zum Zitat Byng JW, Boyd NF, Fishell E, Jong RA, Yaffe MJ (1996) Automated analysis of mammographic densities. Phys Med Biol 41:909–923 Byng JW, Boyd NF, Fishell E, Jong RA, Yaffe MJ (1996) Automated analysis of mammographic densities. Phys Med Biol 41:909–923
25.
Zurück zum Zitat Rikxoort EM (2004) Texture analysis. Internal report. Nijmegen Institute for Cognition and Information Rikxoort EM (2004) Texture analysis. Internal report. Nijmegen Institute for Cognition and Information
26.
Zurück zum Zitat Caldwell CB, Stappelton SJ, Holdsworth DW et al (1990) Characterization of mammographic parenchymal pattern by fractal dimension. Phys Med Biol 35:235–247 Caldwell CB, Stappelton SJ, Holdsworth DW et al (1990) Characterization of mammographic parenchymal pattern by fractal dimension. Phys Med Biol 35:235–247
Metadaten
Titel
A computer-simulated liver phantom (virtual liver phantom) for multidetector computed tomography evaluation
verfasst von
Yoshinori Funama
Kazuo Awai
Osamu Miyazaki
Yoshiharu Nakayama
Da Liu
Taiga Goto
Yasuyuki Yamashita
Shinichi Hori
Publikationsdatum
01.04.2006
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 4/2006
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-0012-5

Weitere Artikel der Ausgabe 4/2006

European Radiology 4/2006 Zur Ausgabe

Book Reviews

Recent literature

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.