A phase-cycled temperature-sensitive fast spin echo sequence with conductivity bias correction for monitoring of mild RF hyperthermia with PRFS

verfasst von: Mingming Wu, Hendrik T. Mulder, Yuval Zur, Silke Lechner-Greite, Marion I. Menzel, Margarethus M. Paulides, Gerard C. van Rhoon, Axel Haase

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 3/2019

Einloggen, um Zugang zu erhalten

Abstract

Objective

Mild hyperthermia (HT) treatments are generally monitored by phase-referenced proton resonance frequency shift calculations. A novel phase and thus temperature-sensitive fast spin echo (TFSE) sequence is introduced and compared to the double echo gradient echo (DEGRE) sequence.

Theory and methods

For a proton resonance frequency shift (PRFS)-sensitive TFSE sequence, a phase cycling method is applied to separate even from odd echoes. This method compensates for conductivity change-induced bias in temperature mapping as does the DEGRE sequence. Both sequences were alternately applied during a phantom heating experiment using the clinical setup for deep radio frequency HT (RF-HT). The B0 drift-corrected temperature values in a region of interest around temperature probes are compared to the temperature probe data and further evaluated in Bland–Altman plots. The stability of both methods was also tested within the thighs of three volunteers at a constant temperature using the subcutaneous fat layer for B0-drift correction.

Results

During the phantom heating experiment, on average TFSE temperature maps achieved double temperature-to-noise ratio (TNR) efficiency in comparison with DEGRE temperature maps. In-vivo images of the thighs exhibit stable temperature readings of ± 1 °C over 25 min of scanning in three volunteers for both methods. On average, the TNR efficiency improved by around 25% for in vivo data.

Conclusion

A novel TFSE method has been adapted to monitor temperature during mild HT.
Literatur
1.
Zurück zum Zitat Kampinga HH (2006) Cell biological effects of hyperthermia alone or combined with radiation or drugs: a short introduction to newcomers in the field. Int J Hyperth 22(3):pp. 191–196CrossRef Kampinga HH (2006) Cell biological effects of hyperthermia alone or combined with radiation or drugs: a short introduction to newcomers in the field. Int J Hyperth 22(3):pp. 191–196CrossRef
2.
Zurück zum Zitat Wust P, Gellermann J, Harder C, Tilly W, Rau B, Dinges S, Schlag P, Budach V, Felix R (1998) Rationale for using invasive thermometry for regional hyperthermia of pelvic tumors. Int J Radiat Oncol Biol Phys 41(5):pp. 1129–1137CrossRef Wust P, Gellermann J, Harder C, Tilly W, Rau B, Dinges S, Schlag P, Budach V, Felix R (1998) Rationale for using invasive thermometry for regional hyperthermia of pelvic tumors. Int J Radiat Oncol Biol Phys 41(5):pp. 1129–1137CrossRef
3.
Zurück zum Zitat van der Zee J, Peer-Valstar JN, Rietveld PJ, de Graaf-Strukowska L, van Rhoon GC (1998) Practical limitations of interstitial thermometry during deep hyperthermia. Int J Radiat Oncol Biol Phys 40(5):pp. 1205–1212 van der Zee J, Peer-Valstar JN, Rietveld PJ, de Graaf-Strukowska L, van Rhoon GC (1998) Practical limitations of interstitial thermometry during deep hyperthermia. Int J Radiat Oncol Biol Phys 40(5):pp. 1205–1212
4.
Zurück zum Zitat Gellermann J, Wlodarczyk W, Ganter H, Nadobny J, Fähling H, Seebass M, Felix R, Wust P (2005) A practical approach to thermography in a hyperthermia/magnetic resonance hybrid system: validation in a heterogeneous phantom. Int J Radiat Oncol Biol Phys 61(1):pp. 267–277CrossRef Gellermann J, Wlodarczyk W, Ganter H, Nadobny J, Fähling H, Seebass M, Felix R, Wust P (2005) A practical approach to thermography in a hyperthermia/magnetic resonance hybrid system: validation in a heterogeneous phantom. Int J Radiat Oncol Biol Phys 61(1):pp. 267–277CrossRef
5.
Zurück zum Zitat McDannold N (2005) Quantitative MRI-based temperature mapping based on the proton resonant frequency shift: review of validation studies. Int J Hyperth 21(6):pp. 533–546CrossRef McDannold N (2005) Quantitative MRI-based temperature mapping based on the proton resonant frequency shift: review of validation studies. Int J Hyperth 21(6):pp. 533–546CrossRef
6.
Zurück zum Zitat Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, Suzuki Y (1995) A precise and fast temperature mapping using water proton chemical shift. Magn Reson Med 34(6):pp. 814–823CrossRef Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, Suzuki Y (1995) A precise and fast temperature mapping using water proton chemical shift. Magn Reson Med 34(6):pp. 814–823CrossRef
7.
Zurück zum Zitat Gellermann J, Wlodarczyk W, Hildebrandt B, Ganter H, Nicolau A, Rau B, Tilly W, Fähling H, Nadobny J, Felix R, Wust P (2005) Noninvasive magnetic resonance thermography of recurrent rectal carcinoma in a 1.5 tesla hybrid system. Cancer Res 65(13):pp. 5872–5880CrossRef Gellermann J, Wlodarczyk W, Hildebrandt B, Ganter H, Nicolau A, Rau B, Tilly W, Fähling H, Nadobny J, Felix R, Wust P (2005) Noninvasive magnetic resonance thermography of recurrent rectal carcinoma in a 1.5 tesla hybrid system. Cancer Res 65(13):pp. 5872–5880CrossRef
8.
Zurück zum Zitat Gellermann J, Hildebrandt B, Issels R, Ganter H, Wlodarczyk W, Budach V, Felix R, Tunn PU, Reichardt P, Wust P (2006) Noninvasive magnetic resonance thermography of soft tissue sarcomas during regional hyperthermia. Cancer 107(6):pp. 1373–1382CrossRef Gellermann J, Hildebrandt B, Issels R, Ganter H, Wlodarczyk W, Budach V, Felix R, Tunn PU, Reichardt P, Wust P (2006) Noninvasive magnetic resonance thermography of soft tissue sarcomas during regional hyperthermia. Cancer 107(6):pp. 1373–1382CrossRef
9.
Zurück zum Zitat Peters RD, Henkelman RM (2000) Proton-resonance frequency shift MR thermometry is affected by changes in the electrical conductivity of tissue. Magn Reson Med 43(1):pp. 62–71 Peters RD, Henkelman RM (2000) Proton-resonance frequency shift MR thermometry is affected by changes in the electrical conductivity of tissue. Magn Reson Med 43(1):pp. 62–71
12.
Zurück zum Zitat Edelstein WA, Glover GH, Hardy CJ, Redington RW (1986) The intrinsic signal-to-noise ratio in NMR imaging. Magn Reson Med 3(4):pp. 604–618CrossRef Edelstein WA, Glover GH, Hardy CJ, Redington RW (1986) The intrinsic signal-to-noise ratio in NMR imaging. Magn Reson Med 3(4):pp. 604–618CrossRef
13.
Zurück zum Zitat Ronnau J, Haimov S, Gogineni S (1994) The effect of signal-to-noise ratio on phase measurements with polarimetric radars. Remote Sens Rev 9(1–2):pp. 27–37 Ronnau J, Haimov S, Gogineni S (1994) The effect of signal-to-noise ratio on phase measurements with polarimetric radars. Remote Sens Rev 9(1–2):pp. 27–37
14.
Zurück zum Zitat Winter L, Oberacker E, Paul K, Ji Y, Oezerdem C, Ghadjar P, Thieme A, Budach V, Wust P, Niendorf T (2016) Magnetic resonance thermometry: methodology, pitfalls and practical solutions. Int J Hyperth 32(1):63–75CrossRef Winter L, Oberacker E, Paul K, Ji Y, Oezerdem C, Ghadjar P, Thieme A, Budach V, Wust P, Niendorf T (2016) Magnetic resonance thermometry: methodology, pitfalls and practical solutions. Int J Hyperth 32(1):63–75CrossRef
15.
Zurück zum Zitat Zur Y (2015) A new time shifted fast spin echo thermometry sequence. In: Proceedings of the 23rd scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto, p 4054 Zur Y (2015) A new time shifted fast spin echo thermometry sequence. In: Proceedings of the 23rd scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto, p 4054
16.
Zurück zum Zitat Vogel MW, Pattynama PM, Lethimonnier FL, Le Roux P (2003) Use of fast spin echo for phase shift magnetic resonance thermometry. J Magn Reson Imaging 18(4):507–512CrossRefPubMed Vogel MW, Pattynama PM, Lethimonnier FL, Le Roux P (2003) Use of fast spin echo for phase shift magnetic resonance thermometry. J Magn Reson Imaging 18(4):507–512CrossRefPubMed
17.
Zurück zum Zitat Splice Schick F (1997) Sub-second diffusion-sensitive MR imaging using a modified fast spin-echo acquisition mode. Magn Reson Med 38(4):pp. 638–644 Splice Schick F (1997) Sub-second diffusion-sensitive MR imaging using a modified fast spin-echo acquisition mode. Magn Reson Med 38(4):pp. 638–644
18.
Zurück zum Zitat Paysen H, Paul K, Pham M, Winter L, Niendorf T (2017) Toward hybrid MR thermometry in aqueous and adipose tissue using simultaneous dual contrast weighting with double echo rare imaging. In: Proceedings of the 25th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, p 1181 Paysen H, Paul K, Pham M, Winter L, Niendorf T (2017) Toward hybrid MR thermometry in aqueous and adipose tissue using simultaneous dual contrast weighting with double echo rare imaging. In: Proceedings of the 25th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, p 1181
21.
Zurück zum Zitat Carr HY, Purcell EM (1954) Effects of diffusion on free precession in nuclear magnetic resonance experiments. Phys Rev 94(3):630–638CrossRef Carr HY, Purcell EM (1954) Effects of diffusion on free precession in nuclear magnetic resonance experiments. Phys Rev 94(3):630–638CrossRef
22.
Zurück zum Zitat Meiboom S, Gill D (1958) Modified spin-echo method for measuring nuclear relaxation times. Rev Sci Instrum 29(8):688–691CrossRef Meiboom S, Gill D (1958) Modified spin-echo method for measuring nuclear relaxation times. Rev Sci Instrum 29(8):688–691CrossRef
23.
Zurück zum Zitat Norris DG, Bornert P (1993) Coherence and interference in ultrafast RARE experiments. J Magn Reson A 105(2):123–127CrossRef Norris DG, Bornert P (1993) Coherence and interference in ultrafast RARE experiments. J Magn Reson A 105(2):123–127CrossRef
24.
Zurück zum Zitat Zur Y, Chen W (2014) A technique to eliminate artifacts in 3D fast spin echo imaging. In: Proceedings of the 22nd scientific meeting, International Society for Magnetic Resonance in Medicine, Milan, p 1648 Zur Y, Chen W (2014) A technique to eliminate artifacts in 3D fast spin echo imaging. In: Proceedings of the 22nd scientific meeting, International Society for Magnetic Resonance in Medicine, Milan, p 1648
26.
Zurück zum Zitat Zur Y (2016) Improved thermometry based on a fast spin echo sequence. In: Proceedings of the 24th scientific meeting, International Society for Magnetic Resonance in Medicine, Singapore, p 3598 Zur Y (2016) Improved thermometry based on a fast spin echo sequence. In: Proceedings of the 24th scientific meeting, International Society for Magnetic Resonance in Medicine, Singapore, p 3598
27.
Zurück zum Zitat El-Sharkawy AM, Schär M, Bottomley PA, Atalar E (2006) Monitoring and correcting spatio-temporal variations of the MR scanner’s static magnetic field. Magn Reson Mater Phy 19(5):223–236CrossRef El-Sharkawy AM, Schär M, Bottomley PA, Atalar E (2006) Monitoring and correcting spatio-temporal variations of the MR scanner’s static magnetic field. Magn Reson Mater Phy 19(5):223–236CrossRef
28.
Zurück zum Zitat Bing C, Staruch RM, Tillander M, Köhler MO, Mougenot C, Ylihautala M, Laetsch TW, Chopra R (2016) Drift correction for accurate PRF-shift MR thermometry during mild hyperthermia treatments with MR-HIFU. Int J Hyperth 32(6):pp. 673–687CrossRef Bing C, Staruch RM, Tillander M, Köhler MO, Mougenot C, Ylihautala M, Laetsch TW, Chopra R (2016) Drift correction for accurate PRF-shift MR thermometry during mild hyperthermia treatments with MR-HIFU. Int J Hyperth 32(6):pp. 673–687CrossRef
29.
Zurück zum Zitat Hofstetter LW, Yeo DT, Dixon WT, Kempf JG, Davis CE, Foo TK (2012) Fat-referenced MR thermometry in the breast and prostate using IDEAL. J Magn Reson Imaging 36(3):722–732CrossRefPubMed Hofstetter LW, Yeo DT, Dixon WT, Kempf JG, Davis CE, Foo TK (2012) Fat-referenced MR thermometry in the breast and prostate using IDEAL. J Magn Reson Imaging 36(3):722–732CrossRefPubMed
30.
Zurück zum Zitat Bowman RR (1976) A probe for measuring temperature in radio-frequency-heated material. IEEE Trans Microw Theory Tech 24(1):43–45CrossRef Bowman RR (1976) A probe for measuring temperature in radio-frequency-heated material. IEEE Trans Microw Theory Tech 24(1):43–45CrossRef
31.
Zurück zum Zitat Busse RF, Brau AC, Vu A, Michelich CR, Bayram E, Kijowski R, Reeder SB, Rowley HA (2008) Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo. Magn Reson Med 60(3):640–649CrossRefPubMedPubMedCentral Busse RF, Brau AC, Vu A, Michelich CR, Bayram E, Kijowski R, Reeder SB, Rowley HA (2008) Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo. Magn Reson Med 60(3):640–649CrossRefPubMedPubMedCentral
32.
Zurück zum Zitat Craciunescu OI, Stauffer PR, Soher BJ, Wyatt CR, Arabe O, Maccarini R, Das SK, Cheng KS, Wong TZ, Jones EL, Dewhirst MW, Vujaskovic Z, MacFall JR (2009) Accuracy of real time noninvasive temperature measurements using magnetic resonance thermal imaging in patients treated for high grade extremity soft tissue sarcomas. Med Phys 36(11):4848–4858CrossRefPubMedPubMedCentral Craciunescu OI, Stauffer PR, Soher BJ, Wyatt CR, Arabe O, Maccarini R, Das SK, Cheng KS, Wong TZ, Jones EL, Dewhirst MW, Vujaskovic Z, MacFall JR (2009) Accuracy of real time noninvasive temperature measurements using magnetic resonance thermal imaging in patients treated for high grade extremity soft tissue sarcomas. Med Phys 36(11):4848–4858CrossRefPubMedPubMedCentral
33.
Zurück zum Zitat Simonis FFJ, Petersen ET, Lagendijk JJW, van den Berg CAT (2016) Feasibility of measuring thermoregulation during RF heating of the human calf muscle using MR based methods. Magn Reson Med 75:1743–1751CrossRefPubMed Simonis FFJ, Petersen ET, Lagendijk JJW, van den Berg CAT (2016) Feasibility of measuring thermoregulation during RF heating of the human calf muscle using MR based methods. Magn Reson Med 75:1743–1751CrossRefPubMed
34.
Zurück zum Zitat Simonis FFJ, Raajimakers AJE, Lagendijk JJW, van den Berg CAT (2017) Validating subject-specific RF and thermal simulations in the calf muscle using MR-based temperature measurements. Magn Reson Med 77:1691–1700CrossRefPubMed Simonis FFJ, Raajimakers AJE, Lagendijk JJW, van den Berg CAT (2017) Validating subject-specific RF and thermal simulations in the calf muscle using MR-based temperature measurements. Magn Reson Med 77:1691–1700CrossRefPubMed
35.
Zurück zum Zitat Dadakova T, Gellermann J, Voigt O, Korvink JG, Pavlina JM, Hennig J, Bock M (2015) Fast PRF-based MR thermometry using double-echo EPI: in vivo comparison in a clinical hyperthermia setting. Magn Reson Mater Phy 28:305–314CrossRef Dadakova T, Gellermann J, Voigt O, Korvink JG, Pavlina JM, Hennig J, Bock M (2015) Fast PRF-based MR thermometry using double-echo EPI: in vivo comparison in a clinical hyperthermia setting. Magn Reson Mater Phy 28:305–314CrossRef
37.
Zurück zum Zitat Yuan J, Mei CS, Panych LP, McDannold NJ, Madore B (2012) Towards fast and accurate temperature mapping with proton resonance frequency-based MR thermometry. Quant Imaging Med Surg 2(1):21PubMedPubMedCentral Yuan J, Mei CS, Panych LP, McDannold NJ, Madore B (2012) Towards fast and accurate temperature mapping with proton resonance frequency-based MR thermometry. Quant Imaging Med Surg 2(1):21PubMedPubMedCentral
38.
Zurück zum Zitat Gellermann J, Faehling H, Mielec M, Cho CH, Budach V, Wust P (2008) Image artifacts during MRT hybrid hyperthermia—causes and elimination. Int J Hyperthermia 24(4):327–335CrossRefPubMed Gellermann J, Faehling H, Mielec M, Cho CH, Budach V, Wust P (2008) Image artifacts during MRT hybrid hyperthermia—causes and elimination. Int J Hyperthermia 24(4):327–335CrossRefPubMed
Metadaten
Titel
A phase-cycled temperature-sensitive fast spin echo sequence with conductivity bias correction for monitoring of mild RF hyperthermia with PRFS
verfasst von
Mingming Wu
Hendrik T. Mulder
Yuval Zur
Silke Lechner-Greite
Marion I. Menzel
Margarethus M. Paulides
Gerard C. van Rhoon
Axel Haase
Publikationsdatum
04.12.2018
Verlag
Springer International Publishing
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 3/2019
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-018-0725-5

Weitere Artikel der Ausgabe 3/2019

Magnetic Resonance Materials in Physics, Biology and Medicine 3/2019 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.