Precision of T1-relaxation time measurements in the hepatic portal vein: influence of measurement technique and sequence parameters

verfasst von: Svein Are Sirirud Vatnehol, Per Kristian Hol, Atle Bjørnerud, Mahmood Amiry-Moghaddam, Camilla Haglerød, Tryggve Holck Storås

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 3/2019

Einloggen, um Zugang zu erhalten

Abstract

Objective

To investigate the effects of a range of parameter settings on T1 measurement stability in the portal vein using the T1-mapping sequences Look-Locker (LL) and Modified Look-Locker inversion recovery (MOLLI).

Materials and methods

Ten different versions of LL and MOLLI sequences were tested and compared to a reference sequence provided by the MR manufacturer. Ten healthy volunteers were imaged multiple times on two separate scan days at 3T. The mean T1 values and coefficient of variation (CoV) were calculated for each of the ten sequences and compared to the reference sequence.

Results

Six of the tested sequences had T1 values close to the reference sequence; among those, three sequences achieved lower CoV than the reference sequence. Lowest CoV was achieved using a non-triggered LL sequence with 5 beat readout and a 45o flip angle (mean T1 1733 ms ± 89 ms, CoV 1.3% ± 0.58%).

Conclusion

T1-measurements in the hepatic portal vein can be performed with high precision using either MOLLI or LL sequences provided that LL sampling duration is sufficiently long and flip angle sufficiently high. The advantage of constant timing outweighed the advantage of ECG-triggering.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Ganong WF (1999) Review of medical physiology, 19th edn. Appleton & Lange, Stamford Ganong WF (1999) Review of medical physiology, 19th edn. Appleton & Lange, Stamford
2.
Zurück zum Zitat Shen Y, Goerner FL, Snyder C, Morelli JN, Hao D, Hu D, Li X, Runge VM (2015) T1 relaxivities of gadolinium-based magnetic resonance contrast agents in human whole blood at 1.5, 3, and 7 T. Invest Radiol 50(5):330–338CrossRefPubMed Shen Y, Goerner FL, Snyder C, Morelli JN, Hao D, Hu D, Li X, Runge VM (2015) T1 relaxivities of gadolinium-based magnetic resonance contrast agents in human whole blood at 1.5, 3, and 7 T. Invest Radiol 50(5):330–338CrossRefPubMed
3.
Zurück zum Zitat Liu P, Chalak LF, Krishnamurthy LC, Mir I, Peng SL, Huang H, Lu H (2016) T1 and T2 values of human neonatal blood at 3 Tesla: dependence on hematocrit, oxygenation, and temperature. Magn Reson Med 75(4):1730–1735CrossRefPubMed Liu P, Chalak LF, Krishnamurthy LC, Mir I, Peng SL, Huang H, Lu H (2016) T1 and T2 values of human neonatal blood at 3 Tesla: dependence on hematocrit, oxygenation, and temperature. Magn Reson Med 75(4):1730–1735CrossRefPubMed
4.
Zurück zum Zitat Zhang X, Petersen ET, Ghariq E, De Vis JB, Webb AG, Teeuwisse WM, Hendrikse J, van Osch MJ (2013) In vivo blood T(1) measurements at 1.5 T, 3 T, and 7 T. Magn Reson Med 70(4):1082–1086CrossRefPubMed Zhang X, Petersen ET, Ghariq E, De Vis JB, Webb AG, Teeuwisse WM, Hendrikse J, van Osch MJ (2013) In vivo blood T(1) measurements at 1.5 T, 3 T, and 7 T. Magn Reson Med 70(4):1082–1086CrossRefPubMed
5.
Zurück zum Zitat Piechnik SK, Ferreira VM, Lewandowski AJ, Ntusi NA, Banerjee R, Holloway C, Hofman MB, Sado DM, Maestrini V, White SK, Lazdam M, Karamitsos T, Moon JC, Neubauer S, Leeson P, Robson MD (2013) Normal variation of magnetic resonance T1 relaxation times in the human population at 1.5 T using ShMOLLI. J Cardiovasc Magn Reson 15(13):11 Piechnik SK, Ferreira VM, Lewandowski AJ, Ntusi NA, Banerjee R, Holloway C, Hofman MB, Sado DM, Maestrini V, White SK, Lazdam M, Karamitsos T, Moon JC, Neubauer S, Leeson P, Robson MD (2013) Normal variation of magnetic resonance T1 relaxation times in the human population at 1.5 T using ShMOLLI. J Cardiovasc Magn Reson 15(13):11
6.
Zurück zum Zitat Varela M, Hajnal JV, Petersen ET, Golay X, Merchant N, Larkman DJ (2011) A method for rapid in vivo measurement of blood T1. NMR Biomed 24(1):80–88CrossRefPubMed Varela M, Hajnal JV, Petersen ET, Golay X, Merchant N, Larkman DJ (2011) A method for rapid in vivo measurement of blood T1. NMR Biomed 24(1):80–88CrossRefPubMed
7.
Zurück zum Zitat Treibel TA, Nasis A, Fontana M, Maestrini V, Castelletti S, Bhuva AN, Rosmini S, Abdel-Gadir A, Bulluck H, Kellman P, Piechnik SK, Robson MD, Moon J (2015) An instantaneous ECV with no blood sampling: using native blood T1 for hematocrit is as good as standard ECV. J Cardiovasc Magn Reson 17(1):Q129CrossRefPubMedCentral Treibel TA, Nasis A, Fontana M, Maestrini V, Castelletti S, Bhuva AN, Rosmini S, Abdel-Gadir A, Bulluck H, Kellman P, Piechnik SK, Robson MD, Moon J (2015) An instantaneous ECV with no blood sampling: using native blood T1 for hematocrit is as good as standard ECV. J Cardiovasc Magn Reson 17(1):Q129CrossRefPubMedCentral
8.
Zurück zum Zitat Look DC, Locker DR (1970) Time saving in measurement of NMR end EPR relaxation times. Rev Sci Instrum 41:250–251CrossRef Look DC, Locker DR (1970) Time saving in measurement of NMR end EPR relaxation times. Rev Sci Instrum 41:250–251CrossRef
9.
Zurück zum Zitat Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP (2004) Modified Look-Locker inversion recovery (MOLLI) for high-resolution T 1 mapping of the heart. Magn Reson Med 52(1):141–146CrossRefPubMed Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP (2004) Modified Look-Locker inversion recovery (MOLLI) for high-resolution T 1 mapping of the heart. Magn Reson Med 52(1):141–146CrossRefPubMed
10.
Zurück zum Zitat Qin Q, Strouse JJ, van Zijl PC (2011) Fast measurement of blood T 1 in the human jugular vein at 3 Tesla. Magn Reson Med 65(5):1297–1304CrossRefPubMed Qin Q, Strouse JJ, van Zijl PC (2011) Fast measurement of blood T 1 in the human jugular vein at 3 Tesla. Magn Reson Med 65(5):1297–1304CrossRefPubMed
11.
Zurück zum Zitat Li W, Liu P, Lu H, Strouse JJ, van Zijl PC, Qin Q (2016) Fast measurement of blood T 1 in the human carotid artery at 3T: accuracy, precision, and reproducibility. Magn Reson Med 77(6):7 Li W, Liu P, Lu H, Strouse JJ, van Zijl PC, Qin Q (2016) Fast measurement of blood T 1 in the human carotid artery at 3T: accuracy, precision, and reproducibility. Magn Reson Med 77(6):7
12.
Zurück zum Zitat Vassiliou V, Heng EL, Nyktari E, Greiser A, Firmin D, Pennell DJ, Gatehouse P, Prasad SK (2015) Effect of temperature and heart rate variability on Phantom T1 maps. J Cardiovasc Magn Reson 17(1):W24CrossRefPubMedCentral Vassiliou V, Heng EL, Nyktari E, Greiser A, Firmin D, Pennell DJ, Gatehouse P, Prasad SK (2015) Effect of temperature and heart rate variability on Phantom T1 maps. J Cardiovasc Magn Reson 17(1):W24CrossRefPubMedCentral
13.
Zurück zum Zitat Scheffler K, Lehnhardt S (2003) Principles and applications of balanced SSFP techniques. Eur Radiol 13(11):2409–2418CrossRefPubMed Scheffler K, Lehnhardt S (2003) Principles and applications of balanced SSFP techniques. Eur Radiol 13(11):2409–2418CrossRefPubMed
14.
Zurück zum Zitat Deichmann RH, Haase A (1992) Quantification of T1 Values by SNAPSHOT-FLASH NMR Imaging. J Magn Reson 96(3):608–612 Deichmann RH, Haase A (1992) Quantification of T1 Values by SNAPSHOT-FLASH NMR Imaging. J Magn Reson 96(3):608–612
15.
Zurück zum Zitat Taylor AJ, Salerno M, Dharmakumar R, Jerosch-Herold M (2016) T1 mapping: basic techniques and clinical applications. JACC Cardiovasc Imaging 9(1):67–81CrossRefPubMed Taylor AJ, Salerno M, Dharmakumar R, Jerosch-Herold M (2016) T1 mapping: basic techniques and clinical applications. JACC Cardiovasc Imaging 9(1):67–81CrossRefPubMed
17.
Zurück zum Zitat Kalavagunta C, Michaeli S, Metzger GJ (2014) In-vitro Gd-DTPA relaxometry studies in oxygenated venous human blood and aqueous solution at 3 and 7T. Contrast Media Mol Imaging 9(2):169–176CrossRefPubMedPubMedCentral Kalavagunta C, Michaeli S, Metzger GJ (2014) In-vitro Gd-DTPA relaxometry studies in oxygenated venous human blood and aqueous solution at 3 and 7T. Contrast Media Mol Imaging 9(2):169–176CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Shimada K, Nagasaka T, Shidahara M, Machida Y, Tamura H (2012) In vivo measurement of longitudinal relaxation time of human blood by inversion-recovery fast gradient-echo MR imaging at 3T. Magn Reson Med Sci 11(4):265–271CrossRefPubMed Shimada K, Nagasaka T, Shidahara M, Machida Y, Tamura H (2012) In vivo measurement of longitudinal relaxation time of human blood by inversion-recovery fast gradient-echo MR imaging at 3T. Magn Reson Med Sci 11(4):265–271CrossRefPubMed
19.
Zurück zum Zitat Wu W-C, Jain V, Li C, Giannetta M, Hurt H, Wehrli FW, Wang DJJ (2010) In vivo venous blood T(1) measurement using inversion recovery true-FISP in children and adults. Magn Reson Med 64(4):1140–1147CrossRefPubMedPubMedCentral Wu W-C, Jain V, Li C, Giannetta M, Hurt H, Wehrli FW, Wang DJJ (2010) In vivo venous blood T(1) measurement using inversion recovery true-FISP in children and adults. Magn Reson Med 64(4):1140–1147CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Zhang X, Petersen ET, Ghariq E, Vis JB, Webb AG, Teeuwisse WM, Hendrikse J, Osch MJP (2013) In vivo blood T1 measurements at 1.5 T, 3 T, and 7 T. Magn Reson Med 70(4):1082–1086CrossRefPubMed Zhang X, Petersen ET, Ghariq E, Vis JB, Webb AG, Teeuwisse WM, Hendrikse J, Osch MJP (2013) In vivo blood T1 measurements at 1.5 T, 3 T, and 7 T. Magn Reson Med 70(4):1082–1086CrossRefPubMed
21.
Zurück zum Zitat McRobbie DW (2007) MRI from picture to proton, 2nd edn. Cambridge University Press, Cambridge McRobbie DW (2007) MRI from picture to proton, 2nd edn. Cambridge University Press, Cambridge
22.
Zurück zum Zitat Gallix BP, Taourel P, Dauzat M, Bruel JM, Lafortune M (1997) Flow pulsatility in the portal venous system: a study of Doppler sonography in healthy adults. Am J Roentgenol 169(1):141–144CrossRef Gallix BP, Taourel P, Dauzat M, Bruel JM, Lafortune M (1997) Flow pulsatility in the portal venous system: a study of Doppler sonography in healthy adults. Am J Roentgenol 169(1):141–144CrossRef
Metadaten
Titel
Precision of T1-relaxation time measurements in the hepatic portal vein: influence of measurement technique and sequence parameters
verfasst von
Svein Are Sirirud Vatnehol
Per Kristian Hol
Atle Bjørnerud
Mahmood Amiry-Moghaddam
Camilla Haglerød
Tryggve Holck Storås
Publikationsdatum
02.01.2019
Verlag
Springer International Publishing
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 3/2019
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-018-00731-1

Weitere Artikel der Ausgabe 3/2019

Magnetic Resonance Materials in Physics, Biology and Medicine 3/2019 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.