Abstract

Purpose

To describe the methodology of constructing a three-dimensional (3D) printed model of the glenohumeral joint, to serve as an interventional phantom for fluoroscopy-guided shoulder arthrography training.

Materials and methods

The osseous structures, intra-articular space and skin surface of the shoulder were digitally extracted as separate 3D meshes from a normal CT arthrogram of the shoulder, using commercially available software. The osseous structures were 3D-printed in gypsum, a fluoroscopically radiopaque mineral, using binder jet technology. The joint capsule was 3D printed with rubber-like TangoPlus material, using PolyJet technology. The capsule was secured to the humeral head and glenoid to create a sealed intra-articular space. A polyamide mold of the skin was printed using selective laser sintering. The joint was stabilized inside the mold, and the surrounding soft tissues were cast in silicone of varying densities. Fluoroscopically-guided shoulder arthrography was performed using anterior, posterior, and rotator interval approaches. CT arthrographic imaging of the phantom was also performed.

Results

A life-size phantom of the glenohumeral joint was constructed. The radiopaque osseous structures replicated in-vivo osseous corticomedullary differentiation, with dense cortical bone and less dense medullary cancellous bone. The glenoid labrum was successfully integrated into the printed capsule, and visualized on CT arthrography. The phantom was repeatedly used to perform shoulder arthrography using all three conventional approaches, and simulated the in vivo challenges of needle guidance.

Conclusions

3D printing of a complex capsule, such as the glenohumeral joint, is possible with this technique. Such a model can serve as a valuable training tool.
Literatur
1.
Zurück zum Zitat Lomasney LM, Choi H, Jayanthi N. Magnetic resonance arthrography of the upper extremity. Radiol Clin N Am. 2013;51(2):227–37.CrossRefPubMed Lomasney LM, Choi H, Jayanthi N. Magnetic resonance arthrography of the upper extremity. Radiol Clin N Am. 2013;51(2):227–37.CrossRefPubMed
2.
Zurück zum Zitat Malfair D. Therapeutic and diagnostic joint injections. Radiol Clin N Am. 2008;46(3):439–53.CrossRefPubMed Malfair D. Therapeutic and diagnostic joint injections. Radiol Clin N Am. 2008;46(3):439–53.CrossRefPubMed
3.
Zurück zum Zitat Magee T. 3-T MRI of the shoulder: is MR arthrography necessary? AJR Am J Roentgenol. 2009;192(1):86–92.CrossRefPubMed Magee T. 3-T MRI of the shoulder: is MR arthrography necessary? AJR Am J Roentgenol. 2009;192(1):86–92.CrossRefPubMed
4.
Zurück zum Zitat Probyn LJ, White LM, Salonen DC, Tomlinson G, Boynton EL. Recurrent symptoms after shoulder instability repair: direct MR arthrographic assessment--correlation with second-look surgical evaluation. Radiology. 2007;245(3):814–23.CrossRefPubMed Probyn LJ, White LM, Salonen DC, Tomlinson G, Boynton EL. Recurrent symptoms after shoulder instability repair: direct MR arthrographic assessment--correlation with second-look surgical evaluation. Radiology. 2007;245(3):814–23.CrossRefPubMed
5.
Zurück zum Zitat Jacobson JA, Lin J, Jamadar DA, Hayes CW. Aids to successful shoulder arthrography performed with a fluoroscopically guided anterior approach. Radiographics. 2003;379;23(2):373–8; discussion 379. Jacobson JA, Lin J, Jamadar DA, Hayes CW. Aids to successful shoulder arthrography performed with a fluoroscopically guided anterior approach. Radiographics. 2003;379;23(2):373–8; discussion 379.
6.
Zurück zum Zitat Farmer KD, Hughes PM. MR arthrography of the shoulder: fluoroscopically guided technique using a posterior approach. AJR Am J Roentgenol. 2002;178(2):433–4.CrossRefPubMed Farmer KD, Hughes PM. MR arthrography of the shoulder: fluoroscopically guided technique using a posterior approach. AJR Am J Roentgenol. 2002;178(2):433–4.CrossRefPubMed
7.
Zurück zum Zitat Depelteau H, Bureau NJ, Cardinal E, Aubin B, Brassard P. Arthrography of the shoulder: a simple fluoroscopically guided approach for targeting the rotator cuff interval. AJR Am J Roentgenol. 2004;182(2):329–32.CrossRefPubMed Depelteau H, Bureau NJ, Cardinal E, Aubin B, Brassard P. Arthrography of the shoulder: a simple fluoroscopically guided approach for targeting the rotator cuff interval. AJR Am J Roentgenol. 2004;182(2):329–32.CrossRefPubMed
8.
Zurück zum Zitat Friedman T, Michalski M, Goodman TR, Brown JE. 3D printing from diagnostic images: a radiologist's primer with an emphasis on musculoskeletal imaging-putting the 3D printing of pathology into the hands of every physician. Skeletal Radiol. 2016;45(3):307–21.CrossRefPubMed Friedman T, Michalski M, Goodman TR, Brown JE. 3D printing from diagnostic images: a radiologist's primer with an emphasis on musculoskeletal imaging-putting the 3D printing of pathology into the hands of every physician. Skeletal Radiol. 2016;45(3):307–21.CrossRefPubMed
10.
Zurück zum Zitat Numminen K, Sipila O, Makisalo H. Preoperative hepatic 3D models: virtual liver resection using three-dimensional imaging technique. Eur J Radiol. 2005;56(2):179–84.CrossRefPubMed Numminen K, Sipila O, Makisalo H. Preoperative hepatic 3D models: virtual liver resection using three-dimensional imaging technique. Eur J Radiol. 2005;56(2):179–84.CrossRefPubMed
11.
Zurück zum Zitat Esses SJ, Berman P, Bloom AI, Sosna J. Clinical applications of physical 3D models derived from MDCT data and created by rapid prototyping. AJR Am J Roentgenol. 2011;196(6):W683–8.CrossRefPubMed Esses SJ, Berman P, Bloom AI, Sosna J. Clinical applications of physical 3D models derived from MDCT data and created by rapid prototyping. AJR Am J Roentgenol. 2011;196(6):W683–8.CrossRefPubMed
12.
Zurück zum Zitat Vukicevic M, Mosadegh B, Min JK, Little SH. Cardiac 3D printing and its future directions. J Am Coll Cardiol Imaging. 2017;10(2):171–84.CrossRef Vukicevic M, Mosadegh B, Min JK, Little SH. Cardiac 3D printing and its future directions. J Am Coll Cardiol Imaging. 2017;10(2):171–84.CrossRef
13.
Zurück zum Zitat Javan R, Herrin D, Tangestanipoor A. Understanding spatially complex segmental and branch anatomy using 3D printing: liver, lung, prostate, coronary arteries, and circle of Willis. Acad Radiol. 2016;23(9):1183–9.CrossRefPubMed Javan R, Herrin D, Tangestanipoor A. Understanding spatially complex segmental and branch anatomy using 3D printing: liver, lung, prostate, coronary arteries, and circle of Willis. Acad Radiol. 2016;23(9):1183–9.CrossRefPubMed
14.
Zurück zum Zitat Tai BL, Rooney D, Stephenson F, Liao P, Sagher O, Shih AJ, et al. Development of a 3D-printed external ventricular drain placement simulator: technical note. J Neurosurg. 2015;123(4):1070–6.CrossRefPubMed Tai BL, Rooney D, Stephenson F, Liao P, Sagher O, Shih AJ, et al. Development of a 3D-printed external ventricular drain placement simulator: technical note. J Neurosurg. 2015;123(4):1070–6.CrossRefPubMed
15.
Zurück zum Zitat O'Reilly MK, Reese S, Herlihy T, Geoghegan T, Cantwell CP, Feeney RNM, et al. Fabrication and assessment of 3D printed anatomical models of the lower limb for anatomical teaching and femoral vessel access training in medicine. Anat Sci Educ. 2016;9(1):71–9.CrossRefPubMed O'Reilly MK, Reese S, Herlihy T, Geoghegan T, Cantwell CP, Feeney RNM, et al. Fabrication and assessment of 3D printed anatomical models of the lower limb for anatomical teaching and femoral vessel access training in medicine. Anat Sci Educ. 2016;9(1):71–9.CrossRefPubMed
16.
Zurück zum Zitat Dias TR, Alves Junior JDC, Abdala N. Learning curve of radiology residents during training in fluoroscopy-guided facet joint injections. Radiol Bras. 2017;50(3):162–9.CrossRefPubMedPubMedCentral Dias TR, Alves Junior JDC, Abdala N. Learning curve of radiology residents during training in fluoroscopy-guided facet joint injections. Radiol Bras. 2017;50(3):162–9.CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Ali S, Qandeel M, Ramakrishna R, Yang CW. Virtual simulation in enhancing procedural training for fluoroscopy-guided lumbar puncture: a pilot study. Acad Radiol. 2018;25(2):235–9.CrossRefPubMed Ali S, Qandeel M, Ramakrishna R, Yang CW. Virtual simulation in enhancing procedural training for fluoroscopy-guided lumbar puncture: a pilot study. Acad Radiol. 2018;25(2):235–9.CrossRefPubMed
18.
Zurück zum Zitat Javan R, Bansal M, Tangestanipoor A. A prototype hybrid gypsum-based 3-dimensional printed training model for computed tomography-guided spinal pain management. J Comput Assist Tomogr. 2016;40(4):626–31.CrossRefPubMed Javan R, Bansal M, Tangestanipoor A. A prototype hybrid gypsum-based 3-dimensional printed training model for computed tomography-guided spinal pain management. J Comput Assist Tomogr. 2016;40(4):626–31.CrossRefPubMed
19.
Zurück zum Zitat Pham DL, Xu C, Prince JL. Current methods in medical image segmentation. Annu Rev Biomed Eng. 2000;2:315–37.CrossRefPubMed Pham DL, Xu C, Prince JL. Current methods in medical image segmentation. Annu Rev Biomed Eng. 2000;2:315–37.CrossRefPubMed
20.
Zurück zum Zitat Mitsouras D, Liacouras P, Imanzadeh A, Giannopoulos AA, Cai T, Kumamaru KK, et al. Medical 3D printing for the radiologist. Radiographics. 2015;35(7):1965–88.CrossRefPubMedPubMedCentral Mitsouras D, Liacouras P, Imanzadeh A, Giannopoulos AA, Cai T, Kumamaru KK, et al. Medical 3D printing for the radiologist. Radiographics. 2015;35(7):1965–88.CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Faulkner AR, Bourgeois AC, Bradley YC, Pasciak AS. A robust and inexpensive phantom for fluoroscopically guided lumbar puncture training. Simul Healthc. 2015;10(1):54–8.CrossRefPubMed Faulkner AR, Bourgeois AC, Bradley YC, Pasciak AS. A robust and inexpensive phantom for fluoroscopically guided lumbar puncture training. Simul Healthc. 2015;10(1):54–8.CrossRefPubMed
23.
Zurück zum Zitat Dimmick S, Jones M, Challen J, Iedema J, Wattuhewa U, Coucher J. CT-guided procedures: evaluation of a phantom system to teach accurate needle placement. Clin Radiol. 2007;62(2):166–71.CrossRefPubMed Dimmick S, Jones M, Challen J, Iedema J, Wattuhewa U, Coucher J. CT-guided procedures: evaluation of a phantom system to teach accurate needle placement. Clin Radiol. 2007;62(2):166–71.CrossRefPubMed
24.
Zurück zum Zitat Boddu SR, Corey A, Peterson R, Saindane AM, Hudgins PA, Chen Z, et al. Fluoroscopic-guided lumbar puncture: fluoroscopic time and implications of body mass index—a baseline study. AJNR Am J Neuroradiol. 2014;35(8):1475–80.CrossRefPubMed Boddu SR, Corey A, Peterson R, Saindane AM, Hudgins PA, Chen Z, et al. Fluoroscopic-guided lumbar puncture: fluoroscopic time and implications of body mass index—a baseline study. AJNR Am J Neuroradiol. 2014;35(8):1475–80.CrossRefPubMed
25.
26.
Zurück zum Zitat Chen RK, Shih AJ. Multi-modality gellan gum-based tissue-mimicking phantom with targeted mechanical, electrical, and thermal properties. Phys Med Biol. 2013;58(16):5511–25.CrossRefPubMed Chen RK, Shih AJ. Multi-modality gellan gum-based tissue-mimicking phantom with targeted mechanical, electrical, and thermal properties. Phys Med Biol. 2013;58(16):5511–25.CrossRefPubMed
27.
Zurück zum Zitat Ryan JR, Chen T, Nakaji P, Frakes DH, Gonzalez LF. Ventriculostomy simulation using patient-specific ventricular anatomy, 3D printing, and hydrogel casting. World Neurosurg. 2015;84(5):1333–9.CrossRefPubMed Ryan JR, Chen T, Nakaji P, Frakes DH, Gonzalez LF. Ventriculostomy simulation using patient-specific ventricular anatomy, 3D printing, and hydrogel casting. World Neurosurg. 2015;84(5):1333–9.CrossRefPubMed
28.
Zurück zum Zitat Morais P, Tavares JMRS, Queiros S, Veloso F, D'hooge J, Vilaca JL. Development of a patient-specific atrial phantom model for planning and training of inter-atrial interventions. Med Phys. 2017;44(11):5638–49.CrossRefPubMed Morais P, Tavares JMRS, Queiros S, Veloso F, D'hooge J, Vilaca JL. Development of a patient-specific atrial phantom model for planning and training of inter-atrial interventions. Med Phys. 2017;44(11):5638–49.CrossRefPubMed
29.
Zurück zum Zitat Javan R, Cho AL. An assembled prototype multimaterial three-dimensional-printed model of the neck for computed tomography- and ultrasound-guided interventional procedures. J Comput Assist Tomogr. 2017;41(6):941–8.CrossRefPubMed Javan R, Cho AL. An assembled prototype multimaterial three-dimensional-printed model of the neck for computed tomography- and ultrasound-guided interventional procedures. J Comput Assist Tomogr. 2017;41(6):941–8.CrossRefPubMed
Metadaten
Titel
A prototype assembled 3D-printed phantom of the glenohumeral joint for fluoroscopic-guided shoulder arthrography
verfasst von
Ramin Javan
Amy L. Ellenbogen
Nicholas Greek
Shawn Haji-Momenian
Publikationsdatum
09.06.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 5/2019
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-018-2979-4

Weitere Artikel der Ausgabe 5/2019

Skeletal Radiology 5/2019 Zur Ausgabe

Test Yourself: Question

Painful right elbow mass

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.