In their article in
Critical Care, Sponholz et al. [
1] compared two devices: the molecular adsorbent recirculating system (MARS) and single pass albumin dialysis (SPAD). As Sponholz et al. claim, SPAD is not an effective method for achieving a significant decrease in the concentration of bile acids, creatinine or urea, because a single SPAD procedure induces a 7.62 % (median) reduction in the level of bile acids and an increase in the concentrations of creatinine and urea by respectively 5.04 % (median) and 4.69 % (median), compared with pre-dialysis levels [
1]. In contrast, our as yet unpublished study demonstrates that SPAD can effectively reduce not only the level of bile acids but also the concentration of ammonia. Based on our study, the application of the SPAD method with a dialysate flow rate of 1000 ml/h for a total of around 10 h reduces the level of bile acids by 21.9 % (median) and ammonia by 16.25 % (median).
A question arises as to the possible causes of such discrepancies. Since continuous venovenous haemodialysis was performed in both studies, the divergence cannot be attributed to the technique. In our view, this divergence is associated with several factors. One of these factors is the difference in dialysate flow rate. An albumin dialysate flow rate of 700 ml/h is highly insufficient. Taking into consideration the results obtained by Sponholz et al., an increase in the albumin dialysate flow rate to at least 1000 ml/h appears to be a necessary precondition to ensure effective elimination of bile acids. Also, the methodology adopted for the study is questionable. In our opinion, comparing the effectiveness of urea and creatinine elimination at radically different dialysate flow rates (SPAD 700 ml/h vs MARS 2000 ml/h) is unjustified and may lead to erroneous conclusions, which in fact occurred [
1].
Differences in efficacy between the MARS and SPAD procedures reported in the article in question may also be related to inappropriately selected haemofilters in the extracorporeal blood purification systems under comparison.
Furthermore, Sponholz et al. [
1] pointed out metabolic derangements and electrolyte disturbances, particularly in SPAD using regional citrate anticoagulation. In our view, increasing the dialysate flow rate would significantly contribute to preventing the irregularities noted in the study. On the other hand, we believe that the risk of citrate overdose during liver dialysis, but also during continuous venovenous haemodialysis, is very low [
2‐
4]. In our medical centre, we have been successfully performing continuous venovenous haemodialysis procedures with regional citrate anticoagulation in patients with severe liver dysfunction, sometimes for up to several weeks. We have never noted any symptoms of overdose apart from end-stage liver failure.
Authors’ response
Christoph Sponholz3, Steffen Mitzner4, 5, Michael Bauer3, 6 and Andreas Kortgen3, 6
(3)Department of Anaesthesiology and Critical Care Medicine, Jena University Hospital, Jena, Germany
(4)Division of Nephrology, Department of Medicine, Rostock University Medical Centre, Rostock, Germany
(5)Fraunhofer Institute for Cell Therapy and Immunology, Extracorporeal Immunomodulation Project Group, Rostock, Germany
(6)Center for Sepsis Control and Care, Integrated Treatment and Research Center, Jena University Hospital, Jena, Germany
Corresponding author
Andreas Kortgen
Department of Anaesthesiology and Critical Care Medicine, Jena University Hospital, Erlanger Allee 101
07747 Jena, Germany.
Phone: +49 3641 9 – 32 31 84
eMail: Andreas.kortgen@med.uni-jena.de
We appreciate the interest of Piechota and Piechota in our study. Indeed, SPAD is not a standardised method either with respect to the albumin concentration in the dialysate solution or to dialysate flow rates. The rationale for using dialysis flow rates of 700 ml/h in our study was based on previously published data on the effective detoxification capacity of a 4 % albumin-containing dialysis solution in vitro [
5] and our own experience in critically ill patients [
6]. Moreover, the flow rate of 700 ml/h allows 7 h of treatment duration similar to MARS using one 5000 ml dialysis bag, resulting in significant changes in total bilirubin levels. However, neither bile acids nor creatinine or urea levels could be reduced in this setting. As discussed extensively, enhancing the dialysis flow rates and treatment times in SPAD could potentially lead to higher detoxification capacities accompanied by higher treatment costs, particularly for higher amounts of albumin. With respect to treatment times the same holds true for MARS, when using a new treatment kit.
Furthermore, Piechota and Piechota raised the question of metabolic and electrolyte disturbances during SPAD administration and citrate anticoagulation in our study. Indeed, we believe that these disturbances were related to the dialysis flow rate, resulting in relative citrate overdosing, and discussed this issue extensively in our publication [
1]. Increasing the dialysis flow rate could reduce the risk of citrate overdosing and accumulation. We demonstrated that regional citrate administration can be used effectively and safely in the critical care setting even in patients with severe liver dysfunction [
7,
8].
We agree with Piechota and Piechota that the dialysis haemofilter impacts on efficacy of SPAD. Reviewing the literature, successful SPAD procedures were published running with different haemofilters, including polysulfone filters as used in the current study (AV1000s polysulfone ultraflux haemofilter; Fresenius Medical Care, Bad Homburg, Germany). Thus, the haemofilter is unlikely responsible for our findings.
To summarise, the SPAD procedure is published to run at different albumin concentrations, dialysis modes and flow rates and with different haemofilters. Currently little is known about the efficacy of SPAD running in different modalities and no review is available summarising all of these issues. Our study thus represents one piece in the puzzle for the discussion of evaluating and standardising the SPAD procedure.
Moreover, our study underlines that the often-performed clinical routine of only assessing the course of bilirubin levels to evaluate efficacy of extracorporeal liver support is insufficient.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (
http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (