Bronchodilator delivery with metered-dose inhaler during mechanical ventilation

verfasst von: Dimitris Georgopoulos, Eleni Mouloudi, Eumorfia Kondili, Maria Klimathianaki

Erschienen in: Critical Care | Ausgabe 4/2000

Einloggen, um Zugang zu erhalten

Abstract

The delivery of bronchodilators with metered-dose inhaler (MDI) in mechanically ventilated patients has attracted considerable interest in recent years. This is because the use of the MDI has several advantages over the nebulizer, such as reduced cost, ease of administration, less personnel time, reliability of dosing and a lower risk of contamination. A spacer device is fundamental in order to demonstrate the efficacy of the bronchodilatory therapy delivered by MDI. Provided that the technique of administration is appropriate, MDIs are as effective as nebulizers, despite a significantly lower dose of bronchodilator given by the MDI.
Literatur
1.
Zurück zum Zitat Boucher BA, Kuhl DA, Coffey BC, Fabian TC: Drug use in a trauma intensive care unit. Am J Hosp Pharm 1990, 47: 805-810.PubMed Boucher BA, Kuhl DA, Coffey BC, Fabian TC: Drug use in a trauma intensive care unit. Am J Hosp Pharm 1990, 47: 805-810.PubMed
2.
Zurück zum Zitat Dhand R, Tobin MJ: Inhaled bronchodilator therapy in mechanically ventilated patients. Am J Respir Crit Care Med 1997, 156: 3-10.CrossRefPubMed Dhand R, Tobin MJ: Inhaled bronchodilator therapy in mechanically ventilated patients. Am J Respir Crit Care Med 1997, 156: 3-10.CrossRefPubMed
3.
Zurück zum Zitat Dhand R, Tobin MJ: Bronhodilator delivery with metered-dose inhalers in mechanically-ventilated patients. Eur Respir J 1996, 9: 585-595. 10.1183/09031936.96.09030585CrossRefPubMed Dhand R, Tobin MJ: Bronhodilator delivery with metered-dose inhalers in mechanically-ventilated patients. Eur Respir J 1996, 9: 585-595. 10.1183/09031936.96.09030585CrossRefPubMed
4.
Zurück zum Zitat Wright PE, Carmichael LC, Bernard GR: Effect of bronchodilators on lung mechanics in the acute respiratory distress syndrome (ARDS). Chest 1994, 106: 1517-1523.CrossRefPubMed Wright PE, Carmichael LC, Bernard GR: Effect of bronchodilators on lung mechanics in the acute respiratory distress syndrome (ARDS). Chest 1994, 106: 1517-1523.CrossRefPubMed
5.
Zurück zum Zitat Georgopoulos D, Brochard L: Ventilatory strategies in acute exacerbations of chronic obstructive pulmonary disease. Eur Respir Monit 1998, 8: 12-44. Georgopoulos D, Brochard L: Ventilatory strategies in acute exacerbations of chronic obstructive pulmonary disease. Eur Respir Monit 1998, 8: 12-44.
6.
Zurück zum Zitat Georgopoulos D, Burchardi H: Ventilatory strategies in adult patients with status asthmaticus. Eur Respir Monit 1998, 8: 45-83. Georgopoulos D, Burchardi H: Ventilatory strategies in adult patients with status asthmaticus. Eur Respir Monit 1998, 8: 45-83.
7.
Zurück zum Zitat Summer W, Elston R, Tharpe L, Nelson S, Haponic EF: Aerosol bronchodilator delivery methods: relative impact on pulmonary function and cost of respiratory care. Arch Intern Mer 1989, 149: 618-623. 10.1001/archinte.149.3.618CrossRef Summer W, Elston R, Tharpe L, Nelson S, Haponic EF: Aerosol bronchodilator delivery methods: relative impact on pulmonary function and cost of respiratory care. Arch Intern Mer 1989, 149: 618-623. 10.1001/archinte.149.3.618CrossRef
8.
Zurück zum Zitat Bowton DL, Goldsmoth WM, Haponic EF: Substitution of metered-dose inhalers for hand nebulizers: success and cost-saving in a large, acute care hospital. Chest 1998, 101: 305-308.CrossRef Bowton DL, Goldsmoth WM, Haponic EF: Substitution of metered-dose inhalers for hand nebulizers: success and cost-saving in a large, acute care hospital. Chest 1998, 101: 305-308.CrossRef
9.
Zurück zum Zitat Alvine GF, Rodgers P, Fitzsimmons KM, Ahrens RC: Disposable jet nebulizers; how reliable are they? Chest 1992, 101: 316-319.CrossRefPubMed Alvine GF, Rodgers P, Fitzsimmons KM, Ahrens RC: Disposable jet nebulizers; how reliable are they? Chest 1992, 101: 316-319.CrossRefPubMed
10.
Zurück zum Zitat Hamill RJ, Houston ED, Georghiu PR, Wright CE, Koza MA, Cadle RM, Goeptert PA, Lewis DA, Zenon GJ, Clarridge JE: An outbreak of Burkholderia (formerly Pseudomonas ) cepacia respiratory track colonisation and infection associated with nebulized albuterol therapy. Ann Intern Med 1995, 122: 762-766.CrossRefPubMed Hamill RJ, Houston ED, Georghiu PR, Wright CE, Koza MA, Cadle RM, Goeptert PA, Lewis DA, Zenon GJ, Clarridge JE: An outbreak of Burkholderia (formerly Pseudomonas ) cepacia respiratory track colonisation and infection associated with nebulized albuterol therapy. Ann Intern Med 1995, 122: 762-766.CrossRefPubMed
11.
Zurück zum Zitat Beaty CD, Ritz Rh, Benson Ms: Continuous in-line nebulizers complicate pressure support ventilation. Chest 1989, 96: 1360-1363.CrossRefPubMed Beaty CD, Ritz Rh, Benson Ms: Continuous in-line nebulizers complicate pressure support ventilation. Chest 1989, 96: 1360-1363.CrossRefPubMed
12.
Zurück zum Zitat Hess D: Aerosol bronchodilator delivery during mechanical ventilation nebulizer or inhaler? Chest 1991, 100: 1103-1104.CrossRefPubMed Hess D: Aerosol bronchodilator delivery during mechanical ventilation nebulizer or inhaler? Chest 1991, 100: 1103-1104.CrossRefPubMed
13.
Zurück zum Zitat Manthous CA, Hall JB, Schmidt GA, Wood LD: Metered-dose inhaler versus nebulized albuterol in mechanically ventilated patients. Am Rev Respir Dis 1993, 148: 1567-1570.CrossRefPubMed Manthous CA, Hall JB, Schmidt GA, Wood LD: Metered-dose inhaler versus nebulized albuterol in mechanically ventilated patients. Am Rev Respir Dis 1993, 148: 1567-1570.CrossRefPubMed
14.
Zurück zum Zitat Fink JB, Dhand R, Duarte AG, Jenne JW, Tobin MJ: Deposition of aerosol from metered-dose inhaler during mechanical ventilation: an in vitro model. Am J Respir Crit Care Med 1996, 154: 382-387.CrossRefPubMed Fink JB, Dhand R, Duarte AG, Jenne JW, Tobin MJ: Deposition of aerosol from metered-dose inhaler during mechanical ventilation: an in vitro model. Am J Respir Crit Care Med 1996, 154: 382-387.CrossRefPubMed
15.
Zurück zum Zitat Crogan SJ, Bishop MJ: Delivery efficiency of metered-dose aerosols given via endotracheal tube. Anesthesiology 1989, 70: 1008-1010.CrossRefPubMed Crogan SJ, Bishop MJ: Delivery efficiency of metered-dose aerosols given via endotracheal tube. Anesthesiology 1989, 70: 1008-1010.CrossRefPubMed
16.
Zurück zum Zitat Fuller HD, Dolovich MB, Postmituck G, Wong Pack W, Newhouse MT: Pressurized aerosol versus jet aerosol delivery to mechanically-ventilated patients: comparison of dose to the lungs. Am Rev Respir Dis 1990, 141: 440-444.CrossRefPubMed Fuller HD, Dolovich MB, Postmituck G, Wong Pack W, Newhouse MT: Pressurized aerosol versus jet aerosol delivery to mechanically-ventilated patients: comparison of dose to the lungs. Am Rev Respir Dis 1990, 141: 440-444.CrossRefPubMed
17.
Zurück zum Zitat Ebert J, Adams AB, Green-Eide B: An evaluation of MDI spacers and adapters: their effect on the respirable volume of medication. Respir Care 1992, 37: 862-868.PubMed Ebert J, Adams AB, Green-Eide B: An evaluation of MDI spacers and adapters: their effect on the respirable volume of medication. Respir Care 1992, 37: 862-868.PubMed
18.
Zurück zum Zitat Fuller HD, Dolovich MB, Chambers C, Newhouse MT: Aerosol delivery during mechanical ventilation: a predictive in vitro model. J Aerosol Med 1992, 5: 251-259.CrossRef Fuller HD, Dolovich MB, Chambers C, Newhouse MT: Aerosol delivery during mechanical ventilation: a predictive in vitro model. J Aerosol Med 1992, 5: 251-259.CrossRef
19.
Zurück zum Zitat Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ: Reconciling in vitro measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency enhancing factors. Am J Respir Crit Care Med 1999, 159: 63-68.CrossRefPubMed Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ: Reconciling in vitro measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency enhancing factors. Am J Respir Crit Care Med 1999, 159: 63-68.CrossRefPubMed
20.
Zurück zum Zitat Diot P, Morra L, Smaldone GC: Albuterol delivery in a model of mechanical ventilation: comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med 1995, 152: 1391-1394.CrossRefPubMed Diot P, Morra L, Smaldone GC: Albuterol delivery in a model of mechanical ventilation: comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med 1995, 152: 1391-1394.CrossRefPubMed
21.
Zurück zum Zitat Rau JL, Harwood RJ, Groff JL: Evaluation of a reservoir devise for metered dose bronchodilator delivery to intubated adults: an in vitro study. Chest 1992, 102: 924-930.CrossRefPubMed Rau JL, Harwood RJ, Groff JL: Evaluation of a reservoir devise for metered dose bronchodilator delivery to intubated adults: an in vitro study. Chest 1992, 102: 924-930.CrossRefPubMed
22.
Zurück zum Zitat Fuller HD, Dolovich MB, Turpie FH, Newhouse MT: Efficiency of bronchodilator aerosol delivery to the lungs from the metered dose inhaler in mechanically ventilated patients: a study comparing four different actuator devices. Chest 1994, 105: 214-218.CrossRefPubMed Fuller HD, Dolovich MB, Turpie FH, Newhouse MT: Efficiency of bronchodilator aerosol delivery to the lungs from the metered dose inhaler in mechanically ventilated patients: a study comparing four different actuator devices. Chest 1994, 105: 214-218.CrossRefPubMed
23.
Zurück zum Zitat Fok TF, al-Essa M, Monkman S, Dolovich M, Girard L, Coates G, Kirpalani H: Delivery of metered dose inhaler aerosols to paralyzed and nonparalyzed rabbits. Crit Care Med 1997, 25: 140-144. 10.1097/00003246-199701000-00026CrossRefPubMed Fok TF, al-Essa M, Monkman S, Dolovich M, Girard L, Coates G, Kirpalani H: Delivery of metered dose inhaler aerosols to paralyzed and nonparalyzed rabbits. Crit Care Med 1997, 25: 140-144. 10.1097/00003246-199701000-00026CrossRefPubMed
24.
Zurück zum Zitat Marik P, Hogan J, Krikorian J: A comparison of bronchodilator therapy delivered by nebulization and metered-dose inhaler in mechanically ventilated patients. Chest 1999, 115: 1653-1657. 10.1378/chest.115.6.1653CrossRefPubMed Marik P, Hogan J, Krikorian J: A comparison of bronchodilator therapy delivered by nebulization and metered-dose inhaler in mechanically ventilated patients. Chest 1999, 115: 1653-1657. 10.1378/chest.115.6.1653CrossRefPubMed
25.
Zurück zum Zitat Duarte AG, Dhand R, Ralston R, Fink J, Fahey P, Tobin MJ, Jenne JW: Serum albuterol levels in mechanically ventilated patients and healthy subjects after metered-dose inhaler administration. Am J Respir Crit Care Med 1996, 154: 1658-1663.CrossRefPubMed Duarte AG, Dhand R, Ralston R, Fink J, Fahey P, Tobin MJ, Jenne JW: Serum albuterol levels in mechanically ventilated patients and healthy subjects after metered-dose inhaler administration. Am J Respir Crit Care Med 1996, 154: 1658-1663.CrossRefPubMed
26.
Zurück zum Zitat Gottfried SB, Rossi A, Higgs BD, Zocchi L, Grassino A, Milic-Emili J: Respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1849-1858.PubMed Gottfried SB, Rossi A, Higgs BD, Zocchi L, Grassino A, Milic-Emili J: Respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1849-1858.PubMed
27.
Zurück zum Zitat Bates JHT, Rossi A, Milic-Emili J: Analysis of the behaviour of the respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1840-1848.PubMed Bates JHT, Rossi A, Milic-Emili J: Analysis of the behaviour of the respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1840-1848.PubMed
28.
Zurück zum Zitat Tuxen DV, Lane S: The effects of ventilatory pattern on hyperinflation, airway pressures and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987, 136: 872-879.CrossRefPubMed Tuxen DV, Lane S: The effects of ventilatory pattern on hyperinflation, airway pressures and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987, 136: 872-879.CrossRefPubMed
29.
Zurück zum Zitat Georgopoulos D, Giannouli E, Patakas D: Effects of extrinsic positive end-expiratory pressure on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1993, 19: 197-203.CrossRefPubMed Georgopoulos D, Giannouli E, Patakas D: Effects of extrinsic positive end-expiratory pressure on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1993, 19: 197-203.CrossRefPubMed
30.
Zurück zum Zitat Dawson SV, Elliott EA: Wave-speed limitation on expiratory flow-a unifying concept. Am Rev Respir Dis 1997, 140: 1-3. Dawson SV, Elliott EA: Wave-speed limitation on expiratory flow-a unifying concept. Am Rev Respir Dis 1997, 140: 1-3.
31.
Zurück zum Zitat Gay PC, Patel HG, Nelson SB, Gilles B, Hubmayr RD: Metered dose inhalers for bronchodilator delivery in intubated, mechanically ventilated patients. Chest 1991, 99: 66-71.CrossRefPubMed Gay PC, Patel HG, Nelson SB, Gilles B, Hubmayr RD: Metered dose inhalers for bronchodilator delivery in intubated, mechanically ventilated patients. Chest 1991, 99: 66-71.CrossRefPubMed
32.
Zurück zum Zitat Georgopoulos D, Mitrouska I, Markopoulou K, Patakas D, Anthonisen NR: Effects of breathing pattern on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1995, 21: 880-886.CrossRefPubMed Georgopoulos D, Mitrouska I, Markopoulou K, Patakas D, Anthonisen NR: Effects of breathing pattern on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1995, 21: 880-886.CrossRefPubMed
33.
Zurück zum Zitat Tobin MJ: Monitoring respiratory mechanics in spontaneously breathing patients. In: Principles and Practice of Intensive Care Monitoring. Edited by Tobin MJ. New York: McGraw-Hill, Incorporated, Health Professions Division, 1998, 617-654. Tobin MJ: Monitoring respiratory mechanics in spontaneously breathing patients. In: Principles and Practice of Intensive Care Monitoring. Edited by Tobin MJ. New York: McGraw-Hill, Incorporated, Health Professions Division, 1998, 617-654.
34.
Zurück zum Zitat Mancebo J, Amaro P, Lorino H, Lemaire F, Harf A, Brochard L: Effects of albuterol inhalation on the work of breathing during weaning from mechanical ventilation. Am Rev Respir Dis 1991, 144: 95-100.CrossRefPubMed Mancebo J, Amaro P, Lorino H, Lemaire F, Harf A, Brochard L: Effects of albuterol inhalation on the work of breathing during weaning from mechanical ventilation. Am Rev Respir Dis 1991, 144: 95-100.CrossRefPubMed
35.
Zurück zum Zitat Mead J, Lindgren I, Gaensler AE: The mechanical properties of the lungs in emphysema. J Clin Invest 1956, 34: 1005-1016.CrossRef Mead J, Lindgren I, Gaensler AE: The mechanical properties of the lungs in emphysema. J Clin Invest 1956, 34: 1005-1016.CrossRef
36.
Zurück zum Zitat Fernandez A, Lazaro A, Garcia A, Aragon C, Cerda E: Bronchodilators in patients with obstructive pulmonary disease on mechanical ventilation: utilisation of metered-dose inhalers. Am Rev Respir Dis 1990, 141: 164-168.CrossRefPubMed Fernandez A, Lazaro A, Garcia A, Aragon C, Cerda E: Bronchodilators in patients with obstructive pulmonary disease on mechanical ventilation: utilisation of metered-dose inhalers. Am Rev Respir Dis 1990, 141: 164-168.CrossRefPubMed
37.
Zurück zum Zitat Dhand R, Jubran A, Tobic MT: Bronchodilator delivery by metered-dose inhaler in ventilator support patients. Am J Respir Crit Care Med 1995, 151: 1827-1833.CrossRefPubMed Dhand R, Jubran A, Tobic MT: Bronchodilator delivery by metered-dose inhaler in ventilator support patients. Am J Respir Crit Care Med 1995, 151: 1827-1833.CrossRefPubMed
38.
Zurück zum Zitat Dhand R, Duarte AG, Jubran A, Jenne JW, Fink JB, Fahey PJ, Tobin MJ: Dose-response to bronchodilator delivered by metered dose inhaler in ventilator-support patients. Am J Respir Crit Care Med 1996, 154: 388-393.CrossRefPubMed Dhand R, Duarte AG, Jubran A, Jenne JW, Fink JB, Fahey PJ, Tobin MJ: Dose-response to bronchodilator delivered by metered dose inhaler in ventilator-support patients. Am J Respir Crit Care Med 1996, 154: 388-393.CrossRefPubMed
39.
Zurück zum Zitat Mouloudi E, Katsanoulas K, Anastasaki M, Askitopoulou E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of end-inspiratory pause. Eur Respir J 1998, 12: 165-169. 10.1183/09031936.98.12010165CrossRefPubMed Mouloudi E, Katsanoulas K, Anastasaki M, Askitopoulou E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of end-inspiratory pause. Eur Respir J 1998, 12: 165-169. 10.1183/09031936.98.12010165CrossRefPubMed
40.
Zurück zum Zitat Mouloudi E, Katsanoulas K, Anastasaki M, Hoing S, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of tidal volume. Intensive Care Med 1999, 25: 1215-1221. 10.1007/s001340051049CrossRefPubMed Mouloudi E, Katsanoulas K, Anastasaki M, Hoing S, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of tidal volume. Intensive Care Med 1999, 25: 1215-1221. 10.1007/s001340051049CrossRefPubMed
41.
Zurück zum Zitat Mouloudi E, Prinianakis G, Kondili E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of flow pattern. Eur Respir J 2000, in press. Mouloudi E, Prinianakis G, Kondili E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of flow pattern. Eur Respir J 2000, in press.
42.
Zurück zum Zitat Guerin G, Chevre A, Dessirier P, Poncet T, Becquemin MH, Dequin PF, Le Guellec C, Jacques D: Inhaled fenoterol-ipratropium bromide in mechanically ventilated patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1999, 159: 1036-1042.CrossRefPubMed Guerin G, Chevre A, Dessirier P, Poncet T, Becquemin MH, Dequin PF, Le Guellec C, Jacques D: Inhaled fenoterol-ipratropium bromide in mechanically ventilated patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1999, 159: 1036-1042.CrossRefPubMed
43.
Zurück zum Zitat Mouloudi E, Katsanoulas K, Hoing S, Prinianakis G, Zeaki M, Georgopoulos D: Duration of effect of bronchodilator delivered by metered dose inhaler in mechanically ventilated COPD patients. Intensive Care Med 1999, 25(Suppl 1): S129. Mouloudi E, Katsanoulas K, Hoing S, Prinianakis G, Zeaki M, Georgopoulos D: Duration of effect of bronchodilator delivered by metered dose inhaler in mechanically ventilated COPD patients. Intensive Care Med 1999, 25(Suppl 1): S129.
Metadaten
Titel
Bronchodilator delivery with metered-dose inhaler during mechanical ventilation
verfasst von
Dimitris Georgopoulos
Eleni Mouloudi
Eumorfia Kondili
Maria Klimathianaki
Publikationsdatum
01.08.2000
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 4/2000
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc698

Weitere Artikel der Ausgabe 4/2000

Critical Care 4/2000 Zur Ausgabe

Tipps für den Umgang mit Behandlungsfehlern

01.05.2024 DGIM 2024 Kongressbericht

Es ist nur eine Frage der Zeit, bis es zu einem Zwischenfall kommt und ein Behandlungsfehler passiert. Doch wenn Ärztinnen und Ärzte gut vorbereitet sind, schaffen es alle Beteiligten den Umständen entsprechend gut durch diese Krise. 

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.