Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 4/2020

Open Access 27.08.2019 | Original Research

Cardiac output estimation by multi-beat analysis of the radial arterial blood pressure waveform versus intermittent pulmonary artery thermodilution: a method comparison study in patients treated in the intensive care unit after off-pump coronary artery bypass surgery

verfasst von: Gillis Greiwe, Valerie Peters, Alexander Hapfelmeier, Stefano Romagnoli, Mathias Kubik, Bernd Saugel

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 4/2020

Abstract

Cardiac output (CO) is a key hemodynamic variable that can be minimally invasively estimated by pulse wave analysis. Multi-beat analysis is a novel pulse wave analysis method. In this prospective observational clinical method comparison study, we compared CO estimations by multi-beat analysis with CO measured by intermittent pulmonary artery thermodilution (PATD) in adult patients treated in the intensive care unit (ICU) after off-pump coronary artery bypass surgery (OPCAB). We included patients after planned admission to the ICU after elective OPCAB who were monitored with a radial arterial catheter and a pulmonary artery catheter. At seven time points, we determined CO using intermittent PATD (PATD-CO; reference method) and simultaneously recorded the radial arterial blood pressure waveform that we later used to estimate CO using multi-beat analysis (MBA-CO; test method) with the Argos monitor (Retia Medical; Valhalla, NY, USA). Blood pressure waveforms impaired by inappropriate damping properties or artifacts were excluded. We compared PATD-CO and MBA-CO using Bland–Altman analysis accounting for repeated measurements, the percentage error, and the concordance rate derived from four-quadrant plot analysis (15% exclusion zone). We analyzed 167 CO values of 31 patients. Mean PATD-CO was 5.30 ± 1.22 L/min and mean MBA-CO was 5.55 ± 1.82 L/min. The mean of the differences between PATD-CO and MBA-CO was 0.08 ± 1.10 L/min (95% limits of agreement: − 2.13 L/min to + 2.29 L/min). The percentage error was 40.7%. The four-quadrant plot-derived concordance rate was 88%. CO estimation by multi-beat analysis of the radial arterial blood pressure waveform (Argos monitor) shows reasonable agreement compared with CO measured by intermittent PATD in adult patients treated in the ICU after OPCAB.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

1 Introduction

Cardiac output (CO)—as a main determinant of oxygen delivery—is a key hemodynamic variable to guide therapy in critically ill patients with circulatory shock [1]. Invasive indicator dilution methods such as pulmonary artery thermodilution (PATD) [2] and transpulmonary thermodilution [3] remain the clinical reference methods to assess CO and novel minimally-invasive hemodynamic monitoring tools should be validated against these invasive reference methods [46]. Invasive uncalibrated pulse wave analysis requiring only an arterial catheter is increasingly used as a minimally invasive method for stroke volume—and thus CO—estimation in surgical and critically ill patients [57]. The term pulse wave analysis comprises several markedly different approaches and algorithms to estimate stroke volume by analyzing the arterial blood pressure waveform [8, 9]. As a main advantage, pulse wave analysis enables CO to be estimated continuously. It is therefore recommended for real-time CO estimation during functional tests of fluid responsiveness such as the passive leg raising [10] or fluid challenge test [11].
Multi-beat analysis is a novel method to estimate stroke volume by mathematically analyzing the radial blood pressure waveform over time scales greater than a single cardiac cycle without external calibration. This enables beat-to-beat waveform variations to be considered and effects of pulse wave reflections to be minimized; eventually, by analyzing the waveform over several cardiac cycles, the multi-beat analysis method aims to identify the pure exponential pressure decay of the arterial blood pressure waveform without confounding wave reflections in the periphery of the arterial vessel tree and estimates stroke volume using a proprietary formula and biometric data [12, 13]. Based on experimental [13] and retrospective [12, 14, 15] pilot studies Retia Medical (Valhalla, NY, USA) developed a commercially available bedside CO monitor using multi-beat analysis—the Argos monitor—that was recently cleared by the US Food and Drug Administration (FDA). In this clinical method comparison study, we compared CO estimations by multi-beat analysis of the radial arterial blood pressure waveform (Argos monitor) to CO measured by intermittent PATD in adult patients treated in the intensive care unit (ICU) after off-pump coronary artery bypass surgery (OPCAB).

2 Methods

2.1 Study design and protocol

This was a CO method comparison study performed in the Center of Anesthesiology and Intensive Care Medicine at the University Medical Center Hamburg-Eppendorf (Hamburg, Germany) between August 2017 and February 2018. The ethics committee approved this study (Ethikkommission der Ärztekammer Hamburg, Hamburg, Germany) and participants gave written informed consent prior to enrollment. Adult patients (age ≥ 18 years) who were scheduled for elective OPCAB with planned postoperative ICU admission and in whom hemodynamic monitoring with a radial artery catheter and a pulmonary artery catheter (PAC) was planned for clinical reasons not related to the study were eligible for inclusion in the study. Patients could not be included if they could not give or refused to give informed consent or had severe heart valve insufficiency or arrhythmias. Patients were excluded if the clinical situation did not allow for conducting study measurements (e.g. agitated patient, transfer to the OR for redo surgery, removal of the arterial line or the PAC).
After the surgical procedure, patients were transferred to the ICU and standard monitoring was established. Correct positioning of the PAC was verified radiographically and by continuous pulmonary artery pressure assessment. After admission to the ICU, we measured CO using PATD (PATD-CO; reference method) at seven time points (with approximately 30 min in between sets) and simultaneously recorded the radial blood pressure waveform for later offline estimation of CO using multi-beat analysis (MBA-CO; test method; Argos monitor, Retia Medical, Valhalla, NY, USA). No study related intervention was performed. Patients were treated according to routine care in the operating room and in the ICU.

2.2 Study measurements

At each of the seven time points, we performed and averaged five single PATD measurements, each performed by injecting a 10 mL bolus of ice-cold saline solution. We used a 7 French thermodilution pulmonary artery catheter (PAC) (Arrow; Teleflex Medical Europe Ltd; Co Westmeath, Ireland). Single PATD boluses were randomly injected over the respiratory cycle. For the time span of each set of PATD measurements we extracted the blood pressure signal sampled at 100 Hz from the ICU monitoring network via a data grabbing software (Dräger Data Grabber; Dräger; Lübeck, Germany). Waveform segments starting 20 s before the first and stopping 20 s after the last PATD measurement were included in the analysis. These blood pressure waveform segments were fed into the Argos monitor for off-line analysis via a 3.5 mm stereo cable, using an MP150 digital-to-analog converter (BIOPAC System; Goleta, CA, USA). Required patient characteristics (gender, age, height, body weight) were entered into the Argos monitor. We used the Argos CO monitor in the 20-s mode, i.e., the monitor estimates and displays CO every 5 s using multi-beat analysis of the arterial blood pressure waveform over 20 s. MBA-CO values estimated during the time span of the corresponding PATD-CO measurements were averaged.
Before the final analysis, three researchers independently visually inspected the recorded arterial blood pressure waveforms for inappropriate damping properties (underdamping/resonance artifacts or overdamping) and artifacts (flushing of the arterial line or motion artifacts). Waveforms impaired by inappropriate damping properties (missed during routine clinical care despite regularly performed fast-flush tests [16]) or artifacts were excluded from further analysis.

2.3 Statistical analysis

We calculated the mean ± standard deviation (SD) to describe continuous patient data and absolute and relative frequencies to describe qualitative patient characteristics. We calculated the mean CO values ± SD for PATD-CO and MBA-CO. The correlation between PATD-CO and MBA-CO was computed within subjects taking repeated observations into account and removing the between subject variability (as required for method comparison studies) [17]. To assess the agreement between PATD-CO and MBA-CO we performed Bland–Altman analysis accounting for multiple observations per patient [18] and calculated the mean and SD of the differences, the 95% limits of agreement (= mean difference ± 1.96 × SD of the difference), and the percentage error [19]. To assess the trending ability of MBA-CO compared with PATD-CO we calculated the concordance rate based on four-quadrant plot analysis showing relative CO changes using a central exclusion zone of 15% [20, 21]. For statistical analyses we used R version 3.5.0 (The R Foundation for Statistical Computing, Vienna, Austria).

3 Results

A total of 284 paired PATD-CO and MBA-CO measurements in 42 patients were recorded. Two patients had to be excluded because of malposition of the PAC and one patient was excluded because of excess motion artifacts of the blood pressure waveform. 106 segments were excluded because of inappropriate damping properties or artifacts of the blood pressure waveform. We included 167 paired CO values of 31 patients in the final analysis.
Basic patient characteristics are shown in Table 1. The relation and within-subject correlation of PATD-CO and MBA-CO is shown in Fig. 1. Mean PATD-CO and MBA-CO were 5.30 ± 1.22 L/min and 5.55 ± 1.82 L/min, respectively. Bland–Altman analysis revealed a mean of the differences between PATD-CO and MBA-CO of 0.08 ± 1.10 L/min (95% limits of agreement: − 2.13 L/min to + 2.29 L/min) (Fig. 2). The percentage error was 40.7%. Four-quadrant plot analysis is shown in Fig. 3. The concordance between changes in PATD-CO and MBA-CO was 88%.
Table 1
Baseline patient characteristics
Number of included patients (n)
31
Age (years)
71 ± 12
Sex, female n (%)
6 (19)
Height (m)
1.74 ± 0.08
Weight (kg)
87 ± 19
BMI (kg/m2)
28.8 ± 5.8
Data are displayed as absolute and relative frequencies or mean and standard deviation

4 Discussion

In this method comparison study, CO estimations by multi-beat analysis of the radial arterial blood pressure waveform (Argos monitor) showed reasonable agreement and trending ability compared with CO measured by intermittent PATD in adult patients treated in the ICU after OPCAB.
This is the first prospective method comparison study evaluating the CO measurement performance of the novel multi-beat analysis algorithm now included in the recently FDA-cleared Argos monitor in comparison with intermittent PATD—the clinical reference method—in patients treated in the ICU after cardiac surgery. Previous studies investigated former versions of the multi-beat analysis algorithm in comparison with experimental gold standard CO measurements (aortic flow probe) in animals [13] and retrospectively in comparison with thermodilution CO measurements in 15 [12] and 169 [15] patients from a freely accessible ICU database. These previous studies showed promising results with regard to the CO measurement performance of the multi-beat analysis method. However, in these previous studies, the agreement and trending ability of the multi-beat analysis method with the reference method were assessed using calibrated multi-beat analysis-derived CO estimates, i.e., the CO estimates were calibrated to the reference CO values and multi-beat analysis was used as a calibrated pulse wave analysis method [8, 9]. In contrast, in our study, we used multi-beat analysis as a—so-called—”uncalibrated” pulse wave analysis method only using arterial blood pressure waveform data and biometric information to estimate CO [8, 9]. Therefore, our findings are not indiscriminately comparable to results of these previous studies.
In 2010, Peyton and Chong [22] published an extensive meta-analysis on minimally-invasive methods for CO estimation in surgical and ICU patients. For pulse wave analysis methods, this meta-analysis yielded an overall pooled mean of the differences of 0.00 L/min (± 95% confidence interval, ± 0.09 L/min), a pooled standard deviation of the differences between paired measurements of 1.22 L/min, and a pooled percentage error of 41.3% (± 95% confidence interval, ± 2.7%). Of note is that this meta-analysis included both uncalibrated and calibrated pulse wave analysis methods. Compared with these pooled results our study investigating uncalibrated multi-beat analysis revealed a slightly lower standard deviation of the mean of the differences and percentage error.
Four-quadrant plot analysis and the concordance rate are used to describe the ability of a test method to adequately and timely detect changes in a hemodynamic variable measured with a reference method. In our study, the concordance rate between changes in MBA-CO and PATD-CO (88%) reflects good trending ability of multi-beat analysis. The concordance rate is markedly higher than the ones observed in recent studies comparing the latest version of an established pulse wave analysis algorithm (FloTrac; Edwards Lifesciences, Irvine; CA, USA) with pulmonary artery thermodilution in patients having cardiac surgery using cardiopulmonary bypass [23, 24].
In general, irrespective of the specific method used, a key prerequisite for pulse wave analysis to provide accurate and reliable estimations of hemodynamic variables is an optimal signal quality of the arterial blood pressure waveform. Artifacts in the arterial blood pressure waveform can be caused by movement or by inappropriate damping properties (underdamping/resonance artifacts or overdamping) of the monitoring system consisting of the arterial catheter, a fluid-filled tubing system, and a pressure transducer [16, 2527]. In accordance with our study, it has been demonstrated that underdamping/resonance artifacts occur in about one-third of cardiovascular patients in the operating room or ICU [26, 27]. Underdamping/resonance artifacts can result in a marked overestimation of systolic arterial blood pressure and incorrect estimation of pulse wave analysis-derived hemodynamic variables such as stroke volume, CO, and pulse pressure variation [26, 27]. Therefore, during routine clinical care, we repeatedly perform “fast-flush tests” [16, 25] to evaluate the dynamic response of the arterial blood pressure monitoring system. To be able to objectively describe the actual measurement performance of a pulse wave analysis method in comparison with a reference method in a method comparison study it is essential to only include impeccable arterial blood pressure waveforms not disturbed by artifacts or inappropriate damping properties of the arterial catheter/tubing system. We, therefore, carefully checked the quality of the recorded arterial blood pressure waveforms and excluded waveform segments that were likely to be altered by inappropriate damping properties (underdamping/resonance or overdamping) and other artifacts. On the one hand, excluding artifactual waveforms is a key prerequisite to draw meaningful conclusions from method comparison studies on pulse wave analysis algorithms. On the other hand, excluding waveforms might increase the risk for selection bias and limit the generalizability of our findings to routine clinical conditions.
Another limitation of our study is that we did not use the multi-beat analysis monitor at the bedside, but recorded arterial waveform segments during routine care and later performed off-line analysis to estimate MBA-CO. In addition, because this was a single center study including patients after OPCAB surgery our results might not be generalizable to other settings or patient collectives. Further, our study setting was not designed to evaluate the trending ability of multi-beat analysis because the patients were in a relatively stable hemodynamic state and no CO-changing interventions were performed between study measurements. This resulted in a limited number of eligible observations for tending analysis.

5 Conclusion

CO estimations by multi-beat analysis of the radial arterial blood pressure waveform (Argos monitor) show reasonable agreement compared with CO measured by intermittent PATD in adult patients treated in the ICU after off-pump coronary artery bypass surgery (OPCAB). For further assessment of the trending ability additional studies are needed.

Acknowledgements

We thank Ricardo Toledo-Crow, PhD (Advanced Science Research Center, City University of New York; New York, NY, USA) for his support with off-line analysis, i.e., feeding arterial blood pressure waveforms into the Argos CO monitor (Retia Medical) for CO estimation.

Compliance with ethical standards

Conflict of interest

GG has received travel expenses from Vygon (Ecouen, France). SR has received honoraria for giving lectures from Baxter (Rome, Italy), BBraun (Melsungen, Germany), Orion Pharma (Espoo, Finland), Medigas (Assago, Italy), and Vygon. SR has received honoraria for consulting from Vygon and Medtronic (Minneapolis, MN, USA). SR has received institutional restricted research grants from Pall International (Milano, Italy) and Baxter (Rome, Italy). BS has received institutional restricted research grants from Retia Medical (Valhalla, NY, USA). BS collaborates with Pulsion Medical Systems (Feldkirchen, Germany) as a member of the medical advisory board and has received institutional restricted research grants, honoraria for giving lectures, and refunds of travel expenses from Pulsion Medical Systems. BS has received research support and honoraria for giving lectures from Edwards Lifesciences (Irvine, CA, USA). BS has received institutional restricted research grants, honoraria for giving lectures, and refunds of travel expenses from CNSystems Medizintechnik (Graz, Austria). BS has received institutional restricted research grants, honoraria for consulting, and refunds of travel expenses from Tensys Medical (San Diego, CA, USA). BS has received honoraria for giving lectures from Philips Medizin Systeme Böblingen (Böblingen, Germany). For all other authors there is no conflict of interest to declare.

Research involving human participants and/or animals

This method comparison study was reviewed and approved by the ethics committee (Ethikkomission der Ärztekammer Hamburg, Hamburg, Germany). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent was obtained from all individual participants included in the study.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.
The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
13.
Zurück zum Zitat Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng. 2006;53(3):459–67.CrossRef Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng. 2006;53(3):459–67.CrossRef
14.
15.
16.
Zurück zum Zitat Gardner RM. Direct blood pressure measurement—dynamic response requirements. Anesthesiology. 1981;54(3):227–36.CrossRef Gardner RM. Direct blood pressure measurement—dynamic response requirements. Anesthesiology. 1981;54(3):227–36.CrossRef
17.
Zurück zum Zitat Bland JM, Altman DG. Calculating correlation coefficients with repeated observations: part 1—correlation within subjects. BMJ. 1995;310(6977):446.CrossRef Bland JM, Altman DG. Calculating correlation coefficients with repeated observations: part 1—correlation within subjects. BMJ. 1995;310(6977):446.CrossRef
19.
Zurück zum Zitat Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15(2):85–91.CrossRef Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15(2):85–91.CrossRef
25.
Zurück zum Zitat Ortega R, Connor C, Kotova F, Deng W, Lacerra C. Use of pressure transducers. N Eng J Med. 2017;376(14):e26.CrossRef Ortega R, Connor C, Kotova F, Deng W, Lacerra C. Use of pressure transducers. N Eng J Med. 2017;376(14):e26.CrossRef
26.
Zurück zum Zitat Romagnoli S, Ricci Z, Quattrone D, Tofani L, Tujjar O, Villa G, Romano SM, De Gaudio AR. Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study. Crit Care. 2014;18(6):644.CrossRef Romagnoli S, Ricci Z, Quattrone D, Tofani L, Tujjar O, Villa G, Romano SM, De Gaudio AR. Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study. Crit Care. 2014;18(6):644.CrossRef
27.
Zurück zum Zitat Romagnoli S, Romano SM, Bevilacqua S, Lazzeri C, Gensini GF, Pratesi C, Quattrone D, Dini D, De Gaudio AR. Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems. J Crit Care. 2011;26(4):415–22.CrossRef Romagnoli S, Romano SM, Bevilacqua S, Lazzeri C, Gensini GF, Pratesi C, Quattrone D, Dini D, De Gaudio AR. Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems. J Crit Care. 2011;26(4):415–22.CrossRef
Metadaten
Titel
Cardiac output estimation by multi-beat analysis of the radial arterial blood pressure waveform versus intermittent pulmonary artery thermodilution: a method comparison study in patients treated in the intensive care unit after off-pump coronary artery bypass surgery
verfasst von
Gillis Greiwe
Valerie Peters
Alexander Hapfelmeier
Stefano Romagnoli
Mathias Kubik
Bernd Saugel
Publikationsdatum
27.08.2019
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 4/2020
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00374-0

Weitere Artikel der Ausgabe 4/2020

Journal of Clinical Monitoring and Computing 4/2020 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.