Comparison between automated breast volume scanner (ABVS) versus hand-held ultrasound as a second look procedure after magnetic resonance imaging

verfasst von: Rossano Girometti, Martina Zanotel, Viviana Londero, Massimo Bazzocchi, Chiara Zuiani

Erschienen in: European Radiology | Ausgabe 9/2017

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the agreement between automated breast volume scanner (ABVS) and conventional ultrasound (US) as a second-look (SL) tool for assessing additional findings found on MRI.

Methods

Over a 7-month period, we prospectively assigned to SL-US and SL-ABVS all patients undergoing 1.5 T breast MRI in whom additional findings were found. Five experienced breast radiologists independently interpreted SL-US and SL-ABVS in blinded sessions to evaluate the detection rate of MRI findings and assign them to BI-RADS categories. We calculated the agreement between the two methods in assessing MRI findings as significant (BI-RADS 3-5) versus not significant (BI-RADS 1-2), as well as their cancer detection rate.

Results

In a population of 131 patients, SL-ABVS and SL-US showed a comparable detection rate of MRI findings (69.3 vs. 71.5%) (p > 0.05; McNemar test), with an almost perfect agreement in assessing them as significant or not (k = 0.94). This translated into a comparably high cancer detection rate (83.8% for SL-ABVS vs. 87.0% for SL-US). Only 1/31 cancers was missed by SL-ABVS.

Conclusions

SL-ABVS and SL-US are nearly equivalent in assessing the significance of MRI findings, leading to a comparable cancer detection rate. SL-ABVS has the potential to replace SL-US in the SL scenario.

Key Points

SL-ABVS shows almost perfect agreement with SL-US in assessing MRI findings.
SL-ABVS shows a comparably high cancer detection rate with respect to SL-US.
SL-ABVS has the potential to replace SL-US in evaluating additional MRI findings.
Literatur
1.
Zurück zum Zitat Sardanelli F, Boetes C, Borisch B et al (2010) Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 46:1296–1316CrossRefPubMed Sardanelli F, Boetes C, Borisch B et al (2010) Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 46:1296–1316CrossRefPubMed
2.
Zurück zum Zitat Lee CH, Dershaw DD, Kopans D et al (2010) Breast cancer screening with imaging: recommendations from the Society of Breast Imaging and the ACR on the use of mammography, breast MRI, breast ultrasound, and other technologies for the detection of clinically occult breast cancer. J Am Coll Radiol 7:18–27CrossRefPubMed Lee CH, Dershaw DD, Kopans D et al (2010) Breast cancer screening with imaging: recommendations from the Society of Breast Imaging and the ACR on the use of mammography, breast MRI, breast ultrasound, and other technologies for the detection of clinically occult breast cancer. J Am Coll Radiol 7:18–27CrossRefPubMed
3.
Zurück zum Zitat Boetes C, Mus RD, Holland R et al (1995) Breast tumors: comparative accuracy of MR imaging relative to mammography and ultrasound for demonstrating extent. Radiology 197:743–747CrossRefPubMed Boetes C, Mus RD, Holland R et al (1995) Breast tumors: comparative accuracy of MR imaging relative to mammography and ultrasound for demonstrating extent. Radiology 197:743–747CrossRefPubMed
4.
Zurück zum Zitat Fischer U, Kopka L, Grabbe E (1999) Breast carcinoma: effect of preoperative contrast-enhanced MR imaging on the therapeutic approach. Radiology 213:881–888CrossRefPubMed Fischer U, Kopka L, Grabbe E (1999) Breast carcinoma: effect of preoperative contrast-enhanced MR imaging on the therapeutic approach. Radiology 213:881–888CrossRefPubMed
5.
Zurück zum Zitat Krämer S, Schulz-Wendtland R, Hagedorn K, Bautz W, Lang N (1998) Magnetic resonance imaging and its role in the diagnosis of multicentric breast cancer. Anticancer Res 18:2163–2164PubMed Krämer S, Schulz-Wendtland R, Hagedorn K, Bautz W, Lang N (1998) Magnetic resonance imaging and its role in the diagnosis of multicentric breast cancer. Anticancer Res 18:2163–2164PubMed
6.
Zurück zum Zitat Hlawatsch A, Teifke A, Schmidt M, Thelen M (2002) Preoperative assessment of breast cancer: sonography versus MR imaging. AJR Am J Roentgenol 179:1493–1501CrossRefPubMed Hlawatsch A, Teifke A, Schmidt M, Thelen M (2002) Preoperative assessment of breast cancer: sonography versus MR imaging. AJR Am J Roentgenol 179:1493–1501CrossRefPubMed
7.
Zurück zum Zitat Berg WA, Gutierrez L, NessAiver MS et al (2004) Diagnostic accuracy of mammography, clinical examination, US, and MR imaging in preoperative assessment of breast cancer. Radiology 233:830–849CrossRefPubMed Berg WA, Gutierrez L, NessAiver MS et al (2004) Diagnostic accuracy of mammography, clinical examination, US, and MR imaging in preoperative assessment of breast cancer. Radiology 233:830–849CrossRefPubMed
8.
Zurück zum Zitat Hata T, Takahashi H, Watanabe K et al (2004) Magnetic resonance imaging for preoperative evaluation of breast cancer: a comparative study with mammography and ultrasonography. J Am Coll Surg 198:190–197CrossRefPubMed Hata T, Takahashi H, Watanabe K et al (2004) Magnetic resonance imaging for preoperative evaluation of breast cancer: a comparative study with mammography and ultrasonography. J Am Coll Surg 198:190–197CrossRefPubMed
9.
Zurück zum Zitat Van Goethem M, Schelfout K, Dijckmans L et al (2004) MR mammography in the pre-operative staging of breast cancer in patients with dense breast tissue: comparison with mammography and ultrasound. Eur Radiol 14:809–816CrossRefPubMed Van Goethem M, Schelfout K, Dijckmans L et al (2004) MR mammography in the pre-operative staging of breast cancer in patients with dense breast tissue: comparison with mammography and ultrasound. Eur Radiol 14:809–816CrossRefPubMed
10.
Zurück zum Zitat Schnall MD, Blume J, Bluemke DA et al (2005) MRI detection of distinct incidental cancer in women with primary breast cancer studied in IBMC 6883. J Surg Oncol 92:32–38CrossRefPubMed Schnall MD, Blume J, Bluemke DA et al (2005) MRI detection of distinct incidental cancer in women with primary breast cancer studied in IBMC 6883. J Surg Oncol 92:32–38CrossRefPubMed
11.
Zurück zum Zitat Houssami N, Ciatto S, Macaskill P et al (2008) Accuracy and surgical impact of magnetic resonance imaging in breast cancer staging: systematic review and meta-analysis in detection of multifocal and multicentric cancer. J Clin Oncol 26:3248–3258CrossRefPubMed Houssami N, Ciatto S, Macaskill P et al (2008) Accuracy and surgical impact of magnetic resonance imaging in breast cancer staging: systematic review and meta-analysis in detection of multifocal and multicentric cancer. J Clin Oncol 26:3248–3258CrossRefPubMed
12.
Zurück zum Zitat Beran L, Liang W, Nims T, Paquelet J, Sickle-Santanello B (2005) Correlation of targeted ultrasound with magnetic resonance imaging abnormalities of the breast. Am J Surg 190:592–594CrossRefPubMed Beran L, Liang W, Nims T, Paquelet J, Sickle-Santanello B (2005) Correlation of targeted ultrasound with magnetic resonance imaging abnormalities of the breast. Am J Surg 190:592–594CrossRefPubMed
13.
Zurück zum Zitat Aracava MM, Chojniak R, Souza JA, Bitencourt AG, Marques EF (2014) Identification of occult breast lesions detected by magnetic resonance imaging with targeted ultrasound: a prospective study. Eur J Radiol 83:516–519CrossRefPubMed Aracava MM, Chojniak R, Souza JA, Bitencourt AG, Marques EF (2014) Identification of occult breast lesions detected by magnetic resonance imaging with targeted ultrasound: a prospective study. Eur J Radiol 83:516–519CrossRefPubMed
14.
Zurück zum Zitat Leung JW (2011) Utility of second-look ultrasound in the evaluation of MRI-detected breast lesions. Semin Roentgenol 46:260–274CrossRefPubMed Leung JW (2011) Utility of second-look ultrasound in the evaluation of MRI-detected breast lesions. Semin Roentgenol 46:260–274CrossRefPubMed
16.
Zurück zum Zitat Kuhl C (2007) The current status of breast MR ImagingPart I. Choice of technique, image interpretation, diagnostic accuracy, and transfer to clinical practice. Radiology 244:356–378CrossRefPubMed Kuhl C (2007) The current status of breast MR ImagingPart I. Choice of technique, image interpretation, diagnostic accuracy, and transfer to clinical practice. Radiology 244:356–378CrossRefPubMed
17.
Zurück zum Zitat Spick C, Baltzer PA (2014) Diagnostic utility of second-look US for breast lesions identified at MR imaging: systematic review and meta-analysis. Radiology 273:401–409CrossRefPubMed Spick C, Baltzer PA (2014) Diagnostic utility of second-look US for breast lesions identified at MR imaging: systematic review and meta-analysis. Radiology 273:401–409CrossRefPubMed
18.
Zurück zum Zitat Orel SG, Rosen M, Mies C, Schnall MD (2006) MR imaging-guided 9-gauge vacuum-assisted core-needle breast biopsy: initial experience. Radiology 238:54–61CrossRefPubMed Orel SG, Rosen M, Mies C, Schnall MD (2006) MR imaging-guided 9-gauge vacuum-assisted core-needle breast biopsy: initial experience. Radiology 238:54–61CrossRefPubMed
19.
Zurück zum Zitat Brennan SB, Sung JS, Dershaw DD, Liberman L, Morris EA (2011) Cancellation of MR imaging-guided breast biopsy due to lesion nonvisualization: frequency and follow-up. Radiology 261:92–99CrossRefPubMed Brennan SB, Sung JS, Dershaw DD, Liberman L, Morris EA (2011) Cancellation of MR imaging-guided breast biopsy due to lesion nonvisualization: frequency and follow-up. Radiology 261:92–99CrossRefPubMed
20.
Zurück zum Zitat Meeuwis C, Mann RM, Mus RD et al (2011) MRI-guided breast biopsy at 3T using a dedicated large core biopsy set: feasibility and initial results. Eur J Radiol 79:257–261CrossRefPubMed Meeuwis C, Mann RM, Mus RD et al (2011) MRI-guided breast biopsy at 3T using a dedicated large core biopsy set: feasibility and initial results. Eur J Radiol 79:257–261CrossRefPubMed
21.
Zurück zum Zitat Berg WA, Blume JD, Cormack JB, Mendelson EB (2006) Operator dependence of physician-performed whole-breast US: lesion detection and characterization. Radiology 241:355–365CrossRefPubMed Berg WA, Blume JD, Cormack JB, Mendelson EB (2006) Operator dependence of physician-performed whole-breast US: lesion detection and characterization. Radiology 241:355–365CrossRefPubMed
23.
Zurück zum Zitat Meng Z, Chen C, Zhu Y et al (2015) Diagnostic performance of the automated breast volume scanner: a systematic review of inter-rater reliability/agreement and meta-analysis of diagnostic accuracy for differentiating benign and malignant breast lesions. Eur Radiol 25:3638–3647CrossRefPubMed Meng Z, Chen C, Zhu Y et al (2015) Diagnostic performance of the automated breast volume scanner: a systematic review of inter-rater reliability/agreement and meta-analysis of diagnostic accuracy for differentiating benign and malignant breast lesions. Eur Radiol 25:3638–3647CrossRefPubMed
24.
Zurück zum Zitat Chae EY, Shin HJ, Kim HJ et al (2013) Diagnostic performance of automated breast ultrasound as a replacement for a hand-held second-look ultrasound for breast lesions detected initially on magnetic resonance imaging. Ultrasound Med Biol 39:2246–2254CrossRefPubMed Chae EY, Shin HJ, Kim HJ et al (2013) Diagnostic performance of automated breast ultrasound as a replacement for a hand-held second-look ultrasound for breast lesions detected initially on magnetic resonance imaging. Ultrasound Med Biol 39:2246–2254CrossRefPubMed
25.
Zurück zum Zitat Halshtok-Neiman O, Shalmon A, Rundstein A, Servadio Y, Gotleib M, Sklair-Levy M (2015) Use of automated breast volumetric sonography as a second-look tool for findings in breast magnetic resonance imaging. Isr Med Assoc J 17:410–413PubMed Halshtok-Neiman O, Shalmon A, Rundstein A, Servadio Y, Gotleib M, Sklair-Levy M (2015) Use of automated breast volumetric sonography as a second-look tool for findings in breast magnetic resonance imaging. Isr Med Assoc J 17:410–413PubMed
26.
Zurück zum Zitat Kim Y, Kang BJ, Kim SH, Lee EJ (2016) Prospective study comparing two second-look ultrasound techniques: handheld ultrasound and an automated breast volume scanner. J Ultrasound Med 35:2103–2112CrossRefPubMed Kim Y, Kang BJ, Kim SH, Lee EJ (2016) Prospective study comparing two second-look ultrasound techniques: handheld ultrasound and an automated breast volume scanner. J Ultrasound Med 35:2103–2112CrossRefPubMed
27.
Zurück zum Zitat D'Orsi CJ, Sickles EA, Mendelson EB et al (2013) ACR BI-RADS® atlas, breast imaging reporting and data system. American College of Radiology, Reston D'Orsi CJ, Sickles EA, Mendelson EB et al (2013) ACR BI-RADS® atlas, breast imaging reporting and data system. American College of Radiology, Reston
28.
Zurück zum Zitat Seigel DG, Podgor MJ, Remaley NA (1992) Acceptable values of kappa for comparison of two groups. Am J Epidemiol 135:571–578CrossRefPubMed Seigel DG, Podgor MJ, Remaley NA (1992) Acceptable values of kappa for comparison of two groups. Am J Epidemiol 135:571–578CrossRefPubMed
29.
Zurück zum Zitat Xu C, Wei S, Xie Y, Guan X, Yang B (2016) Three-dimensional assessment of automated breast volume scanner compared with handheld ultrasound in pre-operative breast invasive ductal carcinomas: a pilot study of 51 cases. Ultrasound Med Biol 42:2089–2096CrossRefPubMed Xu C, Wei S, Xie Y, Guan X, Yang B (2016) Three-dimensional assessment of automated breast volume scanner compared with handheld ultrasound in pre-operative breast invasive ductal carcinomas: a pilot study of 51 cases. Ultrasound Med Biol 42:2089–2096CrossRefPubMed
30.
Zurück zum Zitat An YY, Kim SH, Kang BJ (2015) The image quality and lesion characterization of breast using automated whole-breast ultrasound: a comparison with handheld ultrasound. Eur J Radiol 84:1232–1235CrossRefPubMed An YY, Kim SH, Kang BJ (2015) The image quality and lesion characterization of breast using automated whole-breast ultrasound: a comparison with handheld ultrasound. Eur J Radiol 84:1232–1235CrossRefPubMed
31.
Zurück zum Zitat Chang JM, Cha JH, Park JS, Kim SJ, Moon WK (2015) Automated breast ultrasound system (ABUS): reproducibility of mass localization, size measurement, and characterization on serial examinations. Acta Radiol 56:1163–1170CrossRefPubMed Chang JM, Cha JH, Park JS, Kim SJ, Moon WK (2015) Automated breast ultrasound system (ABUS): reproducibility of mass localization, size measurement, and characterization on serial examinations. Acta Radiol 56:1163–1170CrossRefPubMed
32.
Zurück zum Zitat Li N, Jiang YX, Zhu QL et al (2013) Accuracy of an automated breast volume ultrasound system for assessment of the pre-operative extent of pure ductal carcinoma in situ: comparison with a conventional handheld ultrasound examination. Ultrasound Med Biol 39:2255–2263CrossRefPubMed Li N, Jiang YX, Zhu QL et al (2013) Accuracy of an automated breast volume ultrasound system for assessment of the pre-operative extent of pure ductal carcinoma in situ: comparison with a conventional handheld ultrasound examination. Ultrasound Med Biol 39:2255–2263CrossRefPubMed
33.
Zurück zum Zitat Kelly KM, Dean J, Comulada WS, Lee SJ (2010) Breast cancer detection using automated whole breast ultrasound and mammography in radiographically dense breasts. Eur Radiol 20:734–742CrossRefPubMed Kelly KM, Dean J, Comulada WS, Lee SJ (2010) Breast cancer detection using automated whole breast ultrasound and mammography in radiographically dense breasts. Eur Radiol 20:734–742CrossRefPubMed
34.
Zurück zum Zitat Kelly KM, Dean J, Lee SJ, Comulada WS (2010) Breast cancer detection: radiologists' performance using mammography with and without automated whole-breast ultrasound. Eur Radiol 20:2557–2564CrossRefPubMedPubMedCentral Kelly KM, Dean J, Lee SJ, Comulada WS (2010) Breast cancer detection: radiologists' performance using mammography with and without automated whole-breast ultrasound. Eur Radiol 20:2557–2564CrossRefPubMedPubMedCentral
35.
Zurück zum Zitat Wilczek B, Wilczek HE, Rasouliyan L, Leifland K (2016) Adding 3D automated breast ultrasound to mammography screening in women with heterogeneously and extremely dense breasts: report from a hospital-based, high-volume, single-center breast cancer screening program. Eur J Radiol 85:1554–1563CrossRefPubMed Wilczek B, Wilczek HE, Rasouliyan L, Leifland K (2016) Adding 3D automated breast ultrasound to mammography screening in women with heterogeneously and extremely dense breasts: report from a hospital-based, high-volume, single-center breast cancer screening program. Eur J Radiol 85:1554–1563CrossRefPubMed
36.
Zurück zum Zitat Brem RF, Tabár L, Duffy SW et al (2015) Assessing improvement in detection of breast cancer with three-dimensional automated breast US in women with dense breast tissue: the SomoInsight Study. Radiology 274:663–673CrossRefPubMed Brem RF, Tabár L, Duffy SW et al (2015) Assessing improvement in detection of breast cancer with three-dimensional automated breast US in women with dense breast tissue: the SomoInsight Study. Radiology 274:663–673CrossRefPubMed
37.
Zurück zum Zitat Carbonaro LA, Tannaphai P, Trimboli RM, Verardi N, Fedeli MP, Sardanelli F (2012) Contrast enhanced breast MRI: spatial displacement from prone to supine patient’s position. Preliminary results. Eur J Radiol 81:e771–e774CrossRefPubMed Carbonaro LA, Tannaphai P, Trimboli RM, Verardi N, Fedeli MP, Sardanelli F (2012) Contrast enhanced breast MRI: spatial displacement from prone to supine patient’s position. Preliminary results. Eur J Radiol 81:e771–e774CrossRefPubMed
38.
Zurück zum Zitat Trop I, Labelle M, David J, Mayrand MH, Lalonde L (2010) Second-look targeted studies after breast magnetic resonance imaging: practical tips to improve lesion identification. Curr Probl Diagn Radiol 39:200–211CrossRefPubMed Trop I, Labelle M, David J, Mayrand MH, Lalonde L (2010) Second-look targeted studies after breast magnetic resonance imaging: practical tips to improve lesion identification. Curr Probl Diagn Radiol 39:200–211CrossRefPubMed
39.
Zurück zum Zitat Kim SH, Kang BJ, Choi JJ et al (2013) Radiologists’ performance for detecting lesions and the interobserver variability of automated whole breast ultrasound. Korean J Radiol 14:154–163CrossRefPubMedPubMedCentral Kim SH, Kang BJ, Choi JJ et al (2013) Radiologists’ performance for detecting lesions and the interobserver variability of automated whole breast ultrasound. Korean J Radiol 14:154–163CrossRefPubMedPubMedCentral
40.
Zurück zum Zitat Wang ZL, Xu JH, Li JL, Huang Y, Tang J (2012) Comparison of automated breast volume scanning to hand-held ultrasound and mammography. Radiol Med 117:1287–1293CrossRefPubMed Wang ZL, Xu JH, Li JL, Huang Y, Tang J (2012) Comparison of automated breast volume scanning to hand-held ultrasound and mammography. Radiol Med 117:1287–1293CrossRefPubMed
41.
Zurück zum Zitat Golatta M, Baggs C, Schweitzer-Martin M et al (2015) Evaluation of an automated breast 3D-ultrasound system by comparing it with hand-held ultrasound (HHUS) and mammography. Arch Gynecol Obstet 291:889–895CrossRefPubMed Golatta M, Baggs C, Schweitzer-Martin M et al (2015) Evaluation of an automated breast 3D-ultrasound system by comparing it with hand-held ultrasound (HHUS) and mammography. Arch Gynecol Obstet 291:889–895CrossRefPubMed
42.
Zurück zum Zitat Brem RF, Lenihan MJ, Lieberman J, Torrente J (2015) Screening breast ultrasound: past, present, and future. AJR Am J Roentgenol 204:234–240CrossRefPubMed Brem RF, Lenihan MJ, Lieberman J, Torrente J (2015) Screening breast ultrasound: past, present, and future. AJR Am J Roentgenol 204:234–240CrossRefPubMed
Metadaten
Titel
Comparison between automated breast volume scanner (ABVS) versus hand-held ultrasound as a second look procedure after magnetic resonance imaging
verfasst von
Rossano Girometti
Martina Zanotel
Viviana Londero
Massimo Bazzocchi
Chiara Zuiani
Publikationsdatum
24.01.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 9/2017
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4749-4

Weitere Artikel der Ausgabe 9/2017

European Radiology 9/2017 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.