Erschienen in: Journal of Clinical Monitoring and Computing 2/2020

18.05.2019 | Original Research

Comparison of cardiovascular parameter estimation methods using swine data

verfasst von: Tatsuya Arai, Kichang Lee, Richard J. Cohen

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 2/2020

Einloggen, um Zugang zu erhalten

Abstract

In this study, new and existing methods of estimating stroke volume, cardiac output and total peripheral resistance from analysis of the arterial blood pressure waveform were tested over a wide range of conditions. These pulse contour analysis methods (PCMs) were applied to data obtained in six swine during infusion of volume, phenylephrine, dobutamine, isoproterenol, esmolol and nitroglycerine as well as during progressive hemorrhage. Performance of PCMs was compared using true end-ejection pressures as well as estimated end-ejection pressures. There was considerable overlap in the accuracies of the PCMs when using true end-ejection measures. However, for perhaps the most clinically relevant condition, where radial artery pressure is the input, only Wesseling’s Corrected Impedance method and the Kouchoukos Correction method achieved statistically superior results. We introduced a method of estimating end-ejection by determining when the systolic pressure dropped to a value equal to the sum of the end-diastolic pressure plus a fraction of the pulse pressure. The most accurate estimation of end-ejection was obtained when that fraction was set to 60% for the central arterial pressure and to 50% for the femoral and radial arterial pressures. When the estimated end-ejection measures were used for the PCMs that depend on end-ejection measures and when radial artery pressure was used as the input, only Wesseling’s Corrected Impedance method and the modified Herd’s method achieved statistically superior results. This study provides a systematic comparison of multiple PCMs’ ability to estimate stroke volume, cardiac output, and total peripheral resistance and introduces a new method of estimating end-systole.
Literatur
1.
Zurück zum Zitat Barcroft H, Edholm OG, McMichael J, Sharpey-Schafer EP. Posthaemorrhagic fainting. Study by cardiac output and forearm flow. Lancet. 1944;1:489–91.CrossRef Barcroft H, Edholm OG, McMichael J, Sharpey-Schafer EP. Posthaemorrhagic fainting. Study by cardiac output and forearm flow. Lancet. 1944;1:489–91.CrossRef
2.
Zurück zum Zitat Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ. A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol. 1971;27:392–6.CrossRef Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ. A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol. 1971;27:392–6.CrossRef
3.
Zurück zum Zitat Ganz W, Swan HJ. Measurement of blood flow by thermodilution. Am J Cardiol. 1972;29:241–6.CrossRef Ganz W, Swan HJ. Measurement of blood flow by thermodilution. Am J Cardiol. 1972;29:241–6.CrossRef
4.
Zurück zum Zitat Manecke GR Jr, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR. An unusual case of pulmonary artery catheter malfunction. Anesth Analg. 2002;95:302–4.PubMed Manecke GR Jr, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR. An unusual case of pulmonary artery catheter malfunction. Anesth Analg. 2002;95:302–4.PubMed
5.
Zurück zum Zitat Vender JS, Gilbert HC. Monitoring the anesthetized patient. 3rd ed. Philadelphia: Lippincott-Raven Publishers; 1997. Vender JS, Gilbert HC. Monitoring the anesthetized patient. 3rd ed. Philadelphia: Lippincott-Raven Publishers; 1997.
6.
Zurück zum Zitat Monnet X, Teboul JL. Transpulmonary thermodilution: advantages and limits. Crit Care. 2017;21:147.CrossRef Monnet X, Teboul JL. Transpulmonary thermodilution: advantages and limits. Crit Care. 2017;21:147.CrossRef
7.
Zurück zum Zitat Huter L, Schwarzkopf KR, Schubert NP, Schreiber T. The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model. J Cardiotheorac Vasc Anesth. 2007;21:659–63.CrossRef Huter L, Schwarzkopf KR, Schubert NP, Schreiber T. The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model. J Cardiotheorac Vasc Anesth. 2007;21:659–63.CrossRef
8.
Zurück zum Zitat Staier K, Wilhelm M, Wiesenack C, Thoma M, Keyl C. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol. 2012;29:431–7.CrossRef Staier K, Wilhelm M, Wiesenack C, Thoma M, Keyl C. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol. 2012;29:431–7.CrossRef
9.
Zurück zum Zitat Mihm FG, Gettinger A, Hanson CW 3rd, Gilbert HC, Stover EP, Vender JS, Beerle B, Haddow G. A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med. 1998;26:1346–50.CrossRef Mihm FG, Gettinger A, Hanson CW 3rd, Gilbert HC, Stover EP, Vender JS, Beerle B, Haddow G. A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med. 1998;26:1346–50.CrossRef
10.
Zurück zum Zitat Botero M, Kirby D, Lobato EB, Staples ED, Gravenstein N. Measurement of cardiac output before and after cardiopulmonary bypass: comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing. J Cardiothorac Vasc Anesth. 2004;18:563–72.CrossRef Botero M, Kirby D, Lobato EB, Staples ED, Gravenstein N. Measurement of cardiac output before and after cardiopulmonary bypass: comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing. J Cardiothorac Vasc Anesth. 2004;18:563–72.CrossRef
11.
Zurück zum Zitat Fares WH, Blanchard SK, Stouffer GA, Chang PP, Rosamond WD, Ford HJ, Aris RM. Thermodilution and Fick cardiac outputs differ: impact on pulmonary hypertension evaluation. Can Respir J. 2012;19:261–6.CrossRef Fares WH, Blanchard SK, Stouffer GA, Chang PP, Rosamond WD, Ford HJ, Aris RM. Thermodilution and Fick cardiac outputs differ: impact on pulmonary hypertension evaluation. Can Respir J. 2012;19:261–6.CrossRef
12.
Zurück zum Zitat Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weil J, Singer D. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Crit Care. 2014;18:603.CrossRef Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weil J, Singer D. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Crit Care. 2014;18:603.CrossRef
13.
Zurück zum Zitat Schubert S, Schmitz T, Weiss M, Nagdyman N, Huebler M, Alexi-Meskishvili V, Berger F, Stiller B. Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetry. J Clin Monit Comput. 2008;22:299–307.CrossRef Schubert S, Schmitz T, Weiss M, Nagdyman N, Huebler M, Alexi-Meskishvili V, Berger F, Stiller B. Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetry. J Clin Monit Comput. 2008;22:299–307.CrossRef
14.
Zurück zum Zitat Scolletta S, Franchi F, Romagnoli S, Carla R, Donati A, Fabbri LP, Forfori F, Alonso-Inigo JM, Laviola S, Mangani V, Maj G, Martinelli G, Mirabella L, Morelli A, Persona P, Payen D, PulseCOval Group. Comparison between doppler-echocardiography and uncalibrated pulse contour method for cardiac output measurement: a multicenter observational study. Crit Care Med. 2016;44:1370–9.CrossRef Scolletta S, Franchi F, Romagnoli S, Carla R, Donati A, Fabbri LP, Forfori F, Alonso-Inigo JM, Laviola S, Mangani V, Maj G, Martinelli G, Mirabella L, Morelli A, Persona P, Payen D, PulseCOval Group. Comparison between doppler-echocardiography and uncalibrated pulse contour method for cardiac output measurement: a multicenter observational study. Crit Care Med. 2016;44:1370–9.CrossRef
15.
Zurück zum Zitat Patterson RP. Fundamentals of impedance cardiography. IEEE Eng Med Biol Mag. 1989;8:35–8.CrossRef Patterson RP. Fundamentals of impedance cardiography. IEEE Eng Med Biol Mag. 1989;8:35–8.CrossRef
16.
Zurück zum Zitat Erlanger J, Hooker DR. An experimental study of blood-pressure and of pulse-pressure in man. Johns Hopkins Hosp Rep. 1904;12:145–378. Erlanger J, Hooker DR. An experimental study of blood-pressure and of pulse-pressure in man. Johns Hopkins Hosp Rep. 1904;12:145–378.
17.
Zurück zum Zitat Herd JA, Leclair NR, Simon W. Arterial pressure pulse contours during hemorrhage in anesthetized dogs. J Appl Physiol. 1966;21:1864–8.CrossRef Herd JA, Leclair NR, Simon W. Arterial pressure pulse contours during hemorrhage in anesthetized dogs. J Appl Physiol. 1966;21:1864–8.CrossRef
18.
Zurück zum Zitat Jones WB, Hefner LL, Bancroft WH Jr, Klip W. Velocity of blood flow and stroke volume obtained from the pressure pulse. J Clin Investig. 1959;38:2087–90.CrossRef Jones WB, Hefner LL, Bancroft WH Jr, Klip W. Velocity of blood flow and stroke volume obtained from the pressure pulse. J Clin Investig. 1959;38:2087–90.CrossRef
19.
Zurück zum Zitat Kouchoukos NT, Sheppard LC, McDonald DA. Estimation of stroke volume in the dog by a pulse contour method. Circ Res. 1970;26:611–23.CrossRef Kouchoukos NT, Sheppard LC, McDonald DA. Estimation of stroke volume in the dog by a pulse contour method. Circ Res. 1970;26:611–23.CrossRef
20.
Zurück zum Zitat Liljestrand G, Zander E. Vergleichende Bestimmung des Minutenvolumens des Herzens beim Menschen mittels der Stickoxydulmethode und durch Blutdruckmessung. Zeitschrift fur die gesamte experimentelle Medizin. 1928;59:105–22.CrossRef Liljestrand G, Zander E. Vergleichende Bestimmung des Minutenvolumens des Herzens beim Menschen mittels der Stickoxydulmethode und durch Blutdruckmessung. Zeitschrift fur die gesamte experimentelle Medizin. 1928;59:105–22.CrossRef
21.
Zurück zum Zitat Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng. 2006;53:459–67.CrossRef Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng. 2006;53:459–67.CrossRef
22.
Zurück zum Zitat Parlikar T, Heldt T, Ranade GV, Verghese G. Model-based estimation of cardiac output and total peripheral resistance. Comput Cardiol. 2007;34:379–82. Parlikar T, Heldt T, Ranade GV, Verghese G. Model-based estimation of cardiac output and total peripheral resistance. Comput Cardiol. 2007;34:379–82.
23.
Zurück zum Zitat Sun JX, Reisner AT, Saeed M, Heldt T, Mark RG. The cardiac output from blood pressure algorithms trial. Crit Care Med. 2009;37:72–80.CrossRef Sun JX, Reisner AT, Saeed M, Heldt T, Mark RG. The cardiac output from blood pressure algorithms trial. Crit Care Med. 2009;37:72–80.CrossRef
24.
Zurück zum Zitat Verdouw PD, Beaune J, Roelandt J, Hugenholtz PG. Stroke volume from central aortic pressure? A critical assessment of the various formulae as to their clinical value. Basic Res Cardiol. 1975;70:377–89.CrossRef Verdouw PD, Beaune J, Roelandt J, Hugenholtz PG. Stroke volume from central aortic pressure? A critical assessment of the various formulae as to their clinical value. Basic Res Cardiol. 1975;70:377–89.CrossRef
25.
Zurück zum Zitat Wesseling KH, De Werr B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Its model basis and experimental verification. Adv Cardiovasc Phys. 1983;5:16–52. Wesseling KH, De Werr B, Weber JAP, Smith NT. A simple device for the continuous measurement of cardiac output. Its model basis and experimental verification. Adv Cardiovasc Phys. 1983;5:16–52.
26.
Zurück zum Zitat Arai T, Lee K, Marini RP, Cohen RJ. Estimation of changes in instantaneous aortic blood flow by the analysis of arterial blood pressure. J Appl Physiol. 2012;112:1832–8.CrossRef Arai T, Lee K, Marini RP, Cohen RJ. Estimation of changes in instantaneous aortic blood flow by the analysis of arterial blood pressure. J Appl Physiol. 2012;112:1832–8.CrossRef
27.
Zurück zum Zitat Parlikar T. Modeling and monitoring of cardiovascular dynamics for patients in critical care. Cambridge: Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology; 2007. Parlikar T. Modeling and monitoring of cardiovascular dynamics for patients in critical care. Cambridge: Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology; 2007.
28.
Zurück zum Zitat Bourgeois MJ, Gilbert BK, Von Bernuth G, Wood EH. Continuous determination of beat to beat stroke volume from aortic pulse pressures in the dog. Circ Res. 1976;39:15–24.CrossRef Bourgeois MJ, Gilbert BK, Von Bernuth G, Wood EH. Continuous determination of beat to beat stroke volume from aortic pulse pressures in the dog. Circ Res. 1976;39:15–24.CrossRef
Metadaten
Titel
Comparison of cardiovascular parameter estimation methods using swine data
verfasst von
Tatsuya Arai
Kichang Lee
Richard J. Cohen
Publikationsdatum
18.05.2019
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 2/2020
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00322-y

Weitere Artikel der Ausgabe 2/2020

Journal of Clinical Monitoring and Computing 2/2020 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.