06.01.2022 | Correction
Correction to: Lag times to steady state drug delivery by continuous intravenous infusion: direct comparison of peristaltic and syringe pump performance identifies contributions from infusion system dead volume and pump startup characteristics
Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 4/2022
Einloggen, um Zugang zu erhaltenExcerpt
In the original publication of the article, the heading ‘dark teal’ was inadvertently included in Table 1. The original article has been corrected and the corrected Table 1 is provided here.
Fraction of intended rate
|
Mean (minutes)
|
Statistics (Bonferroni corrected)
|
Statistics (uncorrected)
|
|||
---|---|---|---|---|---|---|
Peristaltic pump (n = 5)
|
Syringe pump (n = 5)
|
98.3% CI
|
P value
|
95% CI
|
P value
|
|
Adult model
|
||||||
5%
|
2.73
|
5.06
|
[− 4.89, 0.23]
|
0.070
|
[− 4.16, − 0.50]
|
0.023
|
50%
|
4.10
|
6.96
|
[− 5.84, 0.13]
|
0.059
|
[− 4.97, − 0.74]
|
0.020
|
95%
|
5.69
|
10.37
|
[− 10.11, 0.75]
|
0.083
|
[− 8.54, − 0.82]
|
0.028
|
Intermediate modela
|
||||||
25%
|
17.52
|
15.84
|
[0.87, 2.50]
|
0.001
|
[1.07, 2.30]
|
< 0.001
|
50%
|
19.13
|
17.41
|
[0.52, 2.92]
|
0.011
|
[0.83, 2.61]
|
0.004
|
75%
|
23.40
|
19.78
|
[0.69, 6.54]
|
0.021
|
[1.44, 5.79]
|
0.007
|
Fraction of intended rate
|
Mean (minutes)
|
Statistics (Bonferroni corrected)
|
Statistics (uncorrected)
|
|||
---|---|---|---|---|---|---|
Peristaltic pump (n = 8)
|
Syringe pump (n = 8)
|
98.3% CI
|
P value
|
95% CI
|
P value
|
|
Pediatric model
|
||||||
5%
|
0.83
|
5.92
|
[− 10.23, 0.04]
|
0.053
|
[− 9.13, − 1.05]
|
0.018
|
50%
|
19.11
|
14.91
|
[2.57, 5.84]
|
< 0.001
|
[2.90, 5.50]
|
< 0.001
|
95%
|
31.49
|
17.99
|
[6.64, 20.35]
|
< 0.001
|
[8.19, 18.80]
|
< 0.001
|