Desflurane usage during anesthesia with and without N2O using FLOW-i Automatic Gas Control with three different wash-in speeds

verfasst von: Robrecht De Medts, Rik Carette, Andre M. De Wolf, Jan F. A. Hendrickx

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 4/2018

Einloggen, um Zugang zu erhalten

Abstract

AGC® (Automatic Gas Control) is the FLOW-i’s automated low flow tool (Maquet, Solna, Sweden) that target controls the inspired O2 (FIO2) and end-expired desflurane concentration (FAdes) while (by design) exponentially decreasing fresh gas flow (FGF) during wash-in to a maintenance default FGF of 300 mL min−1. It also offers a choice of wash-in speeds for the inhaled agents. We examined AGC performance and hypothesized that the use of lower wash-in speeds and N2O both reduce desflurane usage (Vdes). After obtaining IRB approval and patient consent, 78 ASA I-II patients undergoing abdominal surgery were randomly assigned to 1 of 6 groups (n = 13 each), depending on carrier gas (O2/air or O2/N2O) and wash-in speed (AGC speed 2, 4, or 6) of desflurane, resulting in groups air/2, air/4, air/6, N2O/2, N2O/4, and N2O/6. The target for FIO2 was set at 35%, while the FAdes target was selected so that the AGC displayed 1.3 MAC (corrected for the additive affect of N2O if used). AGC was activated upon starting mechanical ventilation. Varvel’s criteria were used to describe performance of achieving the targets. Patient demographics, end-expired N2O concentration, MAC, FGF, and Vdes were compared using ANOVA. Data are presented as mean ± standard deviation, except for Varvel’s criteria (median ± quartiles). Patient demographics did not differ among the groups. Median performance error was −2–0% for FIO2 and −3–1% for FAdes; median absolute performance error was 1–2% for FIO2 and 0–3% for FAdes. MAC increased faster in N2O groups, but total MAC decreased 0.1–0.25 MAC below that in the O2/air groups after 60 min. The effect of wash-in speed on Vdes faded over time. N2O decreased Vdes by 62%. AGC performance for O2 and desflurane targeting is excellent. After 1 h, the wash-in speeds tested are unlikely to affect desflurane usage. N2O usage decreases Vdes proportionally with its reduction in FAtdes.
Literatur
2.
3.
Zurück zum Zitat De Cooman S, Lecain A, Sosnowski M, De Wolf AM, Hendrickx JFA. Desflurane consumption with the Zeus during automated closed circuit versus low flow anesthesia. Acta Anaesthesiol Belg. 2009;60:35–7. doi:10.1186/1471-2253-8-4.CrossRefPubMed De Cooman S, Lecain A, Sosnowski M, De Wolf AM, Hendrickx JFA. Desflurane consumption with the Zeus during automated closed circuit versus low flow anesthesia. Acta Anaesthesiol Belg. 2009;60:35–7. doi:10.​1186/​1471-2253-8-4.CrossRefPubMed
4.
Zurück zum Zitat Nickalls R, Mapleson W. Age-related iso-MAC charts for isoflurane, sevoflurane and desflurane in man. BJA. 2003;91:170.CrossRef Nickalls R, Mapleson W. Age-related iso-MAC charts for isoflurane, sevoflurane and desflurane in man. BJA. 2003;91:170.CrossRef
5.
Zurück zum Zitat Varvel JR, Donoho DL, Shafer SL. Measuring the predictive performance of computer-controlled infusion pumps. J Pharmacokinet Biopharm. 1992;20:63–94.CrossRef Varvel JR, Donoho DL, Shafer SL. Measuring the predictive performance of computer-controlled infusion pumps. J Pharmacokinet Biopharm. 1992;20:63–94.CrossRef
8.
Zurück zum Zitat Hannivoort LN, Vereecke HE, Proost JH, Heyse BE, Eleveld DJ, Bouillon TW, Struys MM, Luginbühl M. Probability to tolerate laryngoscopy and noxious stimulation response index as general indicators of the anaesthetic potency of sevoflurane, propofol, and remifentanil. Br J Anaesth. 2016;116:624–31.CrossRef Hannivoort LN, Vereecke HE, Proost JH, Heyse BE, Eleveld DJ, Bouillon TW, Struys MM, Luginbühl M. Probability to tolerate laryngoscopy and noxious stimulation response index as general indicators of the anaesthetic potency of sevoflurane, propofol, and remifentanil. Br J Anaesth. 2016;116:624–31.CrossRef
Metadaten
Titel
Desflurane usage during anesthesia with and without N2O using FLOW-i Automatic Gas Control with three different wash-in speeds
verfasst von
Robrecht De Medts
Rik Carette
Andre M. De Wolf
Jan F. A. Hendrickx
Publikationsdatum
09.06.2017
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 4/2018
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-017-0031-1

Weitere Artikel der Ausgabe 4/2018

Journal of Clinical Monitoring and Computing 4/2018 Zur Ausgabe

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Beutel versus Maschine: Beste Beatmungstechnik bei Herzstillstand gesucht

02.05.2024 Kardiopulmonale Reanimation Nachrichten

Stehen die Chancen auf eine Rückkehr der Spontanzirkulation nach Herz-Kreislauf-Stillstand bei manueller oder maschineller Beatmung besser? Und unterscheidet sich das neurologische Outcome nach der Reanimation? Das belgische Herzstillstand-Register liefert die Daten für einen direkten Vergleich zwischen Beutel und Beatmungsgerät.

Tipps für den Umgang mit Behandlungsfehlern

01.05.2024 DGIM 2024 Kongressbericht

Es ist nur eine Frage der Zeit, bis es zu einem Zwischenfall kommt und ein Behandlungsfehler passiert. Doch wenn Ärztinnen und Ärzte gut vorbereitet sind, schaffen es alle Beteiligten den Umständen entsprechend gut durch diese Krise. 

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.