Development and evaluation of a small and mobile Magneto Alert Sensor (MALSE) to support safety requirements for magnetic resonance imaging

verfasst von: Conrad Martin, Tobias Frauenrath, Celal Özerdem, Wolfgang Renz, Thoralf Niendorf

Erschienen in: European Radiology | Ausgabe 10/2011

Einloggen, um Zugang zu erhalten

Abstract

Objective

The purpose of this study is to (i) design a small and mobile Magnetic field ALert SEnsor (MALSE), (ii) to carefully evaluate its sensors to their consistency of activation/deactivation and sensitivity to magnetic fields, and (iii) to demonstrate the applicability of MALSE in 1.5 T, 3.0 T and 7.0 T MR fringe field environments.

Methods

MALSE comprises a set of reed sensors, which activate in response to their exposure to a magnetic field. The activation/deactivation of reed sensors was examined by moving them in/out of the fringe field generated by 7TMR.

Results

The consistency with which individual reed sensors would activate at the same field strength was found to be 100% for the setup used. All of the reed switches investigated required a substantial drop in ambient magnetic field strength before they deactivated.

Conclusions

MALSE is a simple concept for alerting MRI staff to a ferromagnetic object being brought into fringe magnetic fields which exceeds MALSEs activation magnetic field. MALSE can easily be attached to ferromagnetic objects within the vicinity of a scanner, thus creating a barrier for hazardous situations induced by ferromagnetic parts which should not enter the vicinity of an MR-system to occur.
Literatur
1.
Zurück zum Zitat Legge A (2009) A review of the top 10 health technology hazards and how to minimise their risks. Nurs Times 105:17–19PubMed Legge A (2009) A review of the top 10 health technology hazards and how to minimise their risks. Nurs Times 105:17–19PubMed
2.
Zurück zum Zitat ECRI (2008) Top 10 health technology hazards. Health Devices 37:343–350 ECRI (2008) Top 10 health technology hazards. Health Devices 37:343–350
3.
Zurück zum Zitat Robitaille PM, Abduljalil AM, Kangarlu A et al (1998) Human magnetic resonance imaging at 8 T. NMR Biomed 11:263–265PubMedCrossRef Robitaille PM, Abduljalil AM, Kangarlu A et al (1998) Human magnetic resonance imaging at 8 T. NMR Biomed 11:263–265PubMedCrossRef
4.
Zurück zum Zitat Vaughan T, DelaBarre L, Snyder C et al (2006) 9.4 T human MRI: preliminary results. Magn Reson Med 56:1274–1282PubMedCrossRef Vaughan T, DelaBarre L, Snyder C et al (2006) 9.4 T human MRI: preliminary results. Magn Reson Med 56:1274–1282PubMedCrossRef
5.
Zurück zum Zitat Barth M, Meyer H, Kannengiesser SA et al (2010) T2-weighted 3D fMRI using S2-SSFP at 7 tesla. Magn Reson Med 63:1015–1020PubMedCrossRef Barth M, Meyer H, Kannengiesser SA et al (2010) T2-weighted 3D fMRI using S2-SSFP at 7 tesla. Magn Reson Med 63:1015–1020PubMedCrossRef
6.
Zurück zum Zitat Tallantyre EC, Morgan PS, Dixon JE et al (2009) A comparison of 3 T and 7 T in the detection of small parenchymal veins within MS lesions. Invest Radiol 44:491–494PubMedCrossRef Tallantyre EC, Morgan PS, Dixon JE et al (2009) A comparison of 3 T and 7 T in the detection of small parenchymal veins within MS lesions. Invest Radiol 44:491–494PubMedCrossRef
7.
Zurück zum Zitat Vaughan JT, Snyder CJ, DelaBarre LJ et al (2009) Whole-body imaging at 7 T: preliminary results. Magn Reson Med 61:244–248PubMedCrossRef Vaughan JT, Snyder CJ, DelaBarre LJ et al (2009) Whole-body imaging at 7 T: preliminary results. Magn Reson Med 61:244–248PubMedCrossRef
8.
Zurück zum Zitat Umutlu L, Maderwald S, Kraff O et al (2010) Dynamic contrast-enhanced breast MRI at 7 Tesla utilizing a single-loop coil: a feasibility trial. Acad Radiol 17:1050–1056PubMedCrossRef Umutlu L, Maderwald S, Kraff O et al (2010) Dynamic contrast-enhanced breast MRI at 7 Tesla utilizing a single-loop coil: a feasibility trial. Acad Radiol 17:1050–1056PubMedCrossRef
9.
Zurück zum Zitat Umutlu L, Orzada S, Kinner S et al (2011) Renal imaging at 7 Tesla: preliminary results. Eur Radiol 21:841–849PubMedCrossRef Umutlu L, Orzada S, Kinner S et al (2011) Renal imaging at 7 Tesla: preliminary results. Eur Radiol 21:841–849PubMedCrossRef
10.
Zurück zum Zitat van Elderen SG, Versluis MJ, Westenberg JJ et al (2010) Right coronary MR angiography at 7 T: a direct quantitative and qualitative comparison with 3 T in young healthy volunteers. Radiology 257:254–259PubMedCrossRef van Elderen SG, Versluis MJ, Westenberg JJ et al (2010) Right coronary MR angiography at 7 T: a direct quantitative and qualitative comparison with 3 T in young healthy volunteers. Radiology 257:254–259PubMedCrossRef
11.
Zurück zum Zitat Niendorf T, Sodickson DK, Krombach GA et al (2010) Toward cardiovascular MRI at 7 T: clinical needs, technical solutions and research promises. Eur Radiol 20:2806–2816PubMedCrossRef Niendorf T, Sodickson DK, Krombach GA et al (2010) Toward cardiovascular MRI at 7 T: clinical needs, technical solutions and research promises. Eur Radiol 20:2806–2816PubMedCrossRef
12.
Zurück zum Zitat von Knobelsdorff-Brenkenhoff F, Frauenrath T, Prothmann M et al (2010) Cardiac chamber quantification using magnetic resonance imaging at 7 Tesla-a pilot study. Eur Radiol 20:2844–2852CrossRef von Knobelsdorff-Brenkenhoff F, Frauenrath T, Prothmann M et al (2010) Cardiac chamber quantification using magnetic resonance imaging at 7 Tesla-a pilot study. Eur Radiol 20:2844–2852CrossRef
13.
Zurück zum Zitat Frauenrath T, Hezel F, Renz W et al (2010) Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla. J Cardiovasc Magn Reson 12:67PubMedCrossRef Frauenrath T, Hezel F, Renz W et al (2010) Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla. J Cardiovasc Magn Reson 12:67PubMedCrossRef
18.
Zurück zum Zitat McNeil DG. M.R.I.’s Strong Magnets Cited in Accidents. New York Times 2005 19.08.2005 McNeil DG. M.R.I.’s Strong Magnets Cited in Accidents. New York Times 2005 19.08.2005
19.
Zurück zum Zitat Administration FaD. MAUDE data base reports of adverse events involving medical devices. 2010 Administration FaD. MAUDE data base reports of adverse events involving medical devices. 2010
20.
Zurück zum Zitat Chen D. Boy, 6, dies of skull injury during MRI; oxygen tank becomes fatal missile in hospital. New York Times 2001 31.06.2010;B1, B5 Chen D. Boy, 6, dies of skull injury during MRI; oxygen tank becomes fatal missile in hospital. New York Times 2001 31.06.2010;B1, B5
21.
Zurück zum Zitat Colletti PM (2004) Size “H” oxygen cylinder: accidental MR projectile at 1.5 Tesla. J Magn Reson Imaging 19:141–143PubMedCrossRef Colletti PM (2004) Size “H” oxygen cylinder: accidental MR projectile at 1.5 Tesla. J Magn Reson Imaging 19:141–143PubMedCrossRef
22.
Zurück zum Zitat Chaljub G, Kramer LA, Johnson RF 3rd et al (2001) Projectile cylinder accidents resulting from the presence of ferromagnetic nitrous oxide or oxygen tanks in the MR suite. AJR Am J Roentgenol 177:27–30PubMed Chaljub G, Kramer LA, Johnson RF 3rd et al (2001) Projectile cylinder accidents resulting from the presence of ferromagnetic nitrous oxide or oxygen tanks in the MR suite. AJR Am J Roentgenol 177:27–30PubMed
23.
Zurück zum Zitat Kanal E, Barkovich AJ, Bell C et al (2007) ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol 188:1447–1474PubMedCrossRef Kanal E, Barkovich AJ, Bell C et al (2007) ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol 188:1447–1474PubMedCrossRef
25.
Zurück zum Zitat Shellock FG, Crues JV (2004) MR procedures: biologic effects, safety, and patient care. Radiology 232:635–652PubMedCrossRef Shellock FG, Crues JV (2004) MR procedures: biologic effects, safety, and patient care. Radiology 232:635–652PubMedCrossRef
26.
Zurück zum Zitat Shellock FG, Woods TO, Crues JV 3rd (2009) MR labeling information for implants and devices: explanation of terminology. Radiology 253:26–30PubMedCrossRef Shellock FG, Woods TO, Crues JV 3rd (2009) MR labeling information for implants and devices: explanation of terminology. Radiology 253:26–30PubMedCrossRef
27.
Zurück zum Zitat Molyneaux DA, Ceisla J, Tsalikis D (2009) Koninklijke Philips Electronics N.V. (BA Eindhoven, NL) assignee. Method and Apparatus for Ferrous Object and/or Magnetic Field Detection for MRI Safety United States Patent Application 20090266887 Molyneaux DA, Ceisla J, Tsalikis D (2009) Koninklijke Philips Electronics N.V. (BA Eindhoven, NL) assignee. Method and Apparatus for Ferrous Object and/or Magnetic Field Detection for MRI Safety United States Patent Application 20090266887
Metadaten
Titel
Development and evaluation of a small and mobile Magneto Alert Sensor (MALSE) to support safety requirements for magnetic resonance imaging
verfasst von
Conrad Martin
Tobias Frauenrath
Celal Özerdem
Wolfgang Renz
Thoralf Niendorf
Publikationsdatum
01.10.2011
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 10/2011
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-011-2153-z

Weitere Artikel der Ausgabe 10/2011

European Radiology 10/2011 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.