Effect of the Y-piece of the ventilation circuit on ventilation requirements in extremely low birth weight infants

verfasst von: M. Wald, Valerie Jeitler, Karin Lawrenz, M. Weninger, Lieselotte Kirchner

Erschienen in: Intensive Care Medicine | Ausgabe 8/2005

Einloggen, um Zugang zu erhalten

Abstract

Objective

Volutrauma caused by high tidal volumes contributes considerably to the development of bronchopulmonary dysplasia. Yet high tidal volumes are required to overcome dead space. In an experimental arrangement we tested whether reduction of dead space might reduce ventilation requirements and thus reduce volutrauma in preterm infants.

Materials and methods

The time required to eliminate CO2 by standardized mechanical ventilation from a preterm infant’s test lung flooded with CO2 was measured. Four different Y-pieces and flow sensor combinations were tested with and without a device for closed suction: Y-piece without flow sensor; integrated flow sensor; small dead-space flow sensor; and a new dead-space free-flow sensor for preterm infants. CO2 concentrations were measured by a capnograph. Mean CO2 elimination times (±SD) were compared.

Results

Mean CO2 elimination time was 37.5 s (±1.18 s) with and 37.4 s (±0.97 s) without closed suction device for the Y-piece without flow sensor, 47.7 s (±0.82 s) and 45.5 s (±1.18 s) for the integrated flow sensor, 42.5 s (±1.27 s) and 41.1 s (±0.99 s) for the small dead-space flow sensor and 38.3 s (±1.16 s) and 36.8 s (±0.79 s) for the dead-space free-flow sensor.

Conclusion

CO2 elimination time with and without closed suction device was nearly identical for the Y-piece without flow sensor and for the dead-space free-flow sensor. With both systems, ventilation requirements were significantly lower than for the integrated flow sensor and for the small dead-space flow sensor (integrated flow sensor vs dead-space free-flow sensor 23.6 and 24.5%, respectively, small dead-space flow sensor vs dead-space free flow sensor 11.7 and 10.9%, respectively); thus, we think that introduction of the innovative dead-space free-flow sensor into clinical practice might reduce incidence and severity of bronchopulmonary dysplasia by reduction of volutrauma.
Literatur
1.
Zurück zum Zitat Dreyfuss D, Soler P, Basset G, Saumon G (1988) High inflation pressure pulmonary edema. Respective effects of high airway pressure, high tidal volume, and positive end-expiratory pressure. Am Rev Respir Dis 137:1159–164PubMed Dreyfuss D, Soler P, Basset G, Saumon G (1988) High inflation pressure pulmonary edema. Respective effects of high airway pressure, high tidal volume, and positive end-expiratory pressure. Am Rev Respir Dis 137:1159–164PubMed
2.
Zurück zum Zitat Gannon CM (1998) Volutrauma, PaCO2 levels, and neurodevelopmental sequelae following assisted ventilation. Clinics in Perinatology 25:159–175PubMed Gannon CM (1998) Volutrauma, PaCO2 levels, and neurodevelopmental sequelae following assisted ventilation. Clinics in Perinatology 25:159–175PubMed
3.
Zurück zum Zitat Dreyfuss D, Soler P, Saumon G (1995) Mechanical ventilation-induced pulmonary edema. Interaction with previous lung alterations. Am J Respir Crit Care Med 151:1568–1575PubMed Dreyfuss D, Soler P, Saumon G (1995) Mechanical ventilation-induced pulmonary edema. Interaction with previous lung alterations. Am J Respir Crit Care Med 151:1568–1575PubMed
4.
Zurück zum Zitat Hernandez LA, Peevy KJ, Moise AA et al. (1989) Chest wall restriction limits high airway pressure induced lung injury in young rabbits. J Appl Physiol 66:2364PubMed Hernandez LA, Peevy KJ, Moise AA et al. (1989) Chest wall restriction limits high airway pressure induced lung injury in young rabbits. J Appl Physiol 66:2364PubMed
5.
Zurück zum Zitat Attar MA, Donn SM (2002) Mechanisms of ventilator-induced lung injury in premature infants. Semin Neonatol 7:353–360PubMed Attar MA, Donn SM (2002) Mechanisms of ventilator-induced lung injury in premature infants. Semin Neonatol 7:353–360PubMed
6.
Zurück zum Zitat Auten RL, Vozzelli M, Clark RH (2001) Volutrauma. What is it, and how do we avoid it? Clin Perinatol 28:505–515PubMed Auten RL, Vozzelli M, Clark RH (2001) Volutrauma. What is it, and how do we avoid it? Clin Perinatol 28:505–515PubMed
7.
Zurück zum Zitat Ricard JD, Dreyfuss D, Saumon G (2003) Ventilator-induced lung injury. Eur Respir J Suppl. 42:2s–9 s Ricard JD, Dreyfuss D, Saumon G (2003) Ventilator-induced lung injury. Eur Respir J Suppl. 42:2s–9 s
8.
Zurück zum Zitat Nolte S (1992) Respiratory death space and ventilation of newborn infants. Klin Padiatr 204:368–372 [in German]PubMed Nolte S (1992) Respiratory death space and ventilation of newborn infants. Klin Padiatr 204:368–372 [in German]PubMed
9.
Zurück zum Zitat Herrera CM, Gerhard T, Claure N, Everett R, Musante G, Thomas C, Bancalari E (2002) Effects of volume-guaranteed synchronized intermittent mandatory ventilation in preterm infants recovering from respiratory failure. Pediatrics110:529–533 Herrera CM, Gerhard T, Claure N, Everett R, Musante G, Thomas C, Bancalari E (2002) Effects of volume-guaranteed synchronized intermittent mandatory ventilation in preterm infants recovering from respiratory failure. Pediatrics110:529–533
10.
Zurück zum Zitat Wald M, Kalous P, Lawrenz K, Jeitler V, Weninger M, Kirchner L (2005) Dead-space washout by split-flow ventilation. A new method to reduce ventilation needs in premature infants. Intensive Care Med 31:674–679PubMed Wald M, Kalous P, Lawrenz K, Jeitler V, Weninger M, Kirchner L (2005) Dead-space washout by split-flow ventilation. A new method to reduce ventilation needs in premature infants. Intensive Care Med 31:674–679PubMed
11.
Zurück zum Zitat Wald M, Jeitler V, Lawrenz K, Weninger M (2004) Dead space free flow sensor reduces the ventilation effort in very low birth weight infants: an experimental study. Critical Care Suppl 1:P13 Wald M, Jeitler V, Lawrenz K, Weninger M (2004) Dead space free flow sensor reduces the ventilation effort in very low birth weight infants: an experimental study. Critical Care Suppl 1:P13
12.
Zurück zum Zitat Wald M (2004) Elimination of ventilator dead space in premature infants by flow partition: a new method. Critical Care Suppl 1:P14 Wald M (2004) Elimination of ventilator dead space in premature infants by flow partition: a new method. Critical Care Suppl 1:P14
13.
Zurück zum Zitat Lugo RA, Keenan J, Salyer JW (2000) Accumulation of CO(2) in reservoir devices during simulated neonatal mechanical ventilation. Pediatr Pulmonol 30:470–475PubMed Lugo RA, Keenan J, Salyer JW (2000) Accumulation of CO(2) in reservoir devices during simulated neonatal mechanical ventilation. Pediatr Pulmonol 30:470–475PubMed
14.
Zurück zum Zitat Claure N, D’Ugard C, Bancalari E (2003) Elimination of ventilator dead space during synchronized ventilation in premature infants. J Pediatr 143:315–320PubMed Claure N, D’Ugard C, Bancalari E (2003) Elimination of ventilator dead space during synchronized ventilation in premature infants. J Pediatr 143:315–320PubMed
15.
Zurück zum Zitat Bernstein G, Mannino FL, Heldt GP, Callahan JD, Bull DH, Sola A, Ariagno RL, Hoffman GL, Frantz ID III, Troche BI, Roberts JL, Dela Cruz TV, Costa E (1996) Randomized multicenter trial comparing sychronized and conventional intermittent mandatory ventilation in neonates. J Pediatr 128:453–463PubMed Bernstein G, Mannino FL, Heldt GP, Callahan JD, Bull DH, Sola A, Ariagno RL, Hoffman GL, Frantz ID III, Troche BI, Roberts JL, Dela Cruz TV, Costa E (1996) Randomized multicenter trial comparing sychronized and conventional intermittent mandatory ventilation in neonates. J Pediatr 128:453–463PubMed
16.
Zurück zum Zitat Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM (2000) Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 162:2109–2112PubMed Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM (2000) Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 162:2109–2112PubMed
17.
Zurück zum Zitat Stresemann E (1969) Washout of anatomical deadspace for alveolar hypoventilation. Respiration 26:425–434 Stresemann E (1969) Washout of anatomical deadspace for alveolar hypoventilation. Respiration 26:425–434
18.
Zurück zum Zitat Kalous P, Kokstein Z (2003) Instrumental dead space in neonatology, and its elimination by continuous tracheal gas insufflation during conventional ventilation. Acta Paediatr 92:518–524PubMed Kalous P, Kokstein Z (2003) Instrumental dead space in neonatology, and its elimination by continuous tracheal gas insufflation during conventional ventilation. Acta Paediatr 92:518–524PubMed
19.
Zurück zum Zitat Kalous P, Kokstein Z, Basek P, Polzer O (2002) Prediction of efficacy of continuous tracheal gas insufflation on CO2 elimination in mechanically ventilated newborns. Prospective observation study. J Matern Fetal Neonat Med 11(S1):298 Kalous P, Kokstein Z, Basek P, Polzer O (2002) Prediction of efficacy of continuous tracheal gas insufflation on CO2 elimination in mechanically ventilated newborns. Prospective observation study. J Matern Fetal Neonat Med 11(S1):298
20.
Zurück zum Zitat Claure N, D’Ugard C, Bancalari E (2003) Elimination of ventilator dead space during synchronized ventilation in premature infants. J Pediatr 143:315–320PubMed Claure N, D’Ugard C, Bancalari E (2003) Elimination of ventilator dead space during synchronized ventilation in premature infants. J Pediatr 143:315–320PubMed
21.
Zurück zum Zitat Dassieu G (2000) Continuous tracheal gas insufflation in preterm infants with hyaline membrane disease. A prospective randomized rrial. Am J Respir Crit Care Med162:826–831 Dassieu G (2000) Continuous tracheal gas insufflation in preterm infants with hyaline membrane disease. A prospective randomized rrial. Am J Respir Crit Care Med162:826–831
22.
Zurück zum Zitat Miro AM (2000) Auto-positive end-expiratory pressure during tracheal gas insufflation: testing a hypothetical model. Crit Care Med 28:3474–3479PubMed Miro AM (2000) Auto-positive end-expiratory pressure during tracheal gas insufflation: testing a hypothetical model. Crit Care Med 28:3474–3479PubMed
23.
Zurück zum Zitat Davies MW, Woodgate PG (2002) Tracheal gas insufflation for the prevention of morbidity and mortality in mechanically ventilated newborn infants. Cochrane Database Syst Rev (2):CD002973 Davies MW, Woodgate PG (2002) Tracheal gas insufflation for the prevention of morbidity and mortality in mechanically ventilated newborn infants. Cochrane Database Syst Rev (2):CD002973
24.
Zurück zum Zitat Kacmarek RM (2001) Complications of tracheal gas insufflation. Respir Care 46:167–176PubMed Kacmarek RM (2001) Complications of tracheal gas insufflation. Respir Care 46:167–176PubMed
25.
Zurück zum Zitat Nahum A (2001) Animal and lung model studies of tracheal gas insufflation. Respir Care 46:149–157PubMed Nahum A (2001) Animal and lung model studies of tracheal gas insufflation. Respir Care 46:149–157PubMed
26.
Zurück zum Zitat Erb T, Frei FJ (2001) The use of cuffed endotracheal tubes in infants and small children. Anaesthesist 50:395–400PubMed Erb T, Frei FJ (2001) The use of cuffed endotracheal tubes in infants and small children. Anaesthesist 50:395–400PubMed
27.
Zurück zum Zitat Nikischin W, Lange M (2003) Correction of compliance and resistance altered by endotracheal tube leaks. Pediatr Crit Care Med 4:344–352PubMed Nikischin W, Lange M (2003) Correction of compliance and resistance altered by endotracheal tube leaks. Pediatr Crit Care Med 4:344–352PubMed
28.
Zurück zum Zitat Main E, Castle R, Stocks J, James I, Hatch D (2001) The influence of endotracheal tube leak on the assessment of respiratory function in ventilated children. Intensive Care Med 27:1788–1797PubMed Main E, Castle R, Stocks J, James I, Hatch D (2001) The influence of endotracheal tube leak on the assessment of respiratory function in ventilated children. Intensive Care Med 27:1788–1797PubMed
29.
Zurück zum Zitat Dassieu G, Brochard L, Agudze E, Patkai J, Janaud JC, Danan C (1998) Continuous tracheal gas insufflation enables a volume reduction strategy in hyaline membrane disease: technical aspects and clinical results. Intensive Care Med 24:1076–1082PubMed Dassieu G, Brochard L, Agudze E, Patkai J, Janaud JC, Danan C (1998) Continuous tracheal gas insufflation enables a volume reduction strategy in hyaline membrane disease: technical aspects and clinical results. Intensive Care Med 24:1076–1082PubMed
30.
Zurück zum Zitat Martinez-Perez M, Bernabe F, Pena R, Fernandez R, Nahum A, Blanch L (2004) Effects of expiratory tracheal gas insufflation in patients with severe head trauma and acute lung injury. Intensive Care Med 30:2021–2027PubMed Martinez-Perez M, Bernabe F, Pena R, Fernandez R, Nahum A, Blanch L (2004) Effects of expiratory tracheal gas insufflation in patients with severe head trauma and acute lung injury. Intensive Care Med 30:2021–2027PubMed
31.
Zurück zum Zitat Cordero L, Sananes M, Ayers LW (2000) Comparison of a closed (Trach Care MAC) with an open endotracheal suction system in small premature infants. J Perinatol 20:151–156PubMed Cordero L, Sananes M, Ayers LW (2000) Comparison of a closed (Trach Care MAC) with an open endotracheal suction system in small premature infants. J Perinatol 20:151–156PubMed
32.
Zurück zum Zitat Mosca FA, Colnaghi M, Lattanzio M, Bray M, Pugliese S, Fumagalli M (1997) Closed versus open endotracheal suctioning in preterm infants: effects on cerebral oxygenation and blood volume. Biol Neonate 72:9–14PubMed Mosca FA, Colnaghi M, Lattanzio M, Bray M, Pugliese S, Fumagalli M (1997) Closed versus open endotracheal suctioning in preterm infants: effects on cerebral oxygenation and blood volume. Biol Neonate 72:9–14PubMed
Metadaten
Titel
Effect of the Y-piece of the ventilation circuit on ventilation requirements in extremely low birth weight infants
verfasst von
M. Wald
Valerie Jeitler
Karin Lawrenz
M. Weninger
Lieselotte Kirchner
Publikationsdatum
01.08.2005
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 8/2005
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-005-2711-4

Weitere Artikel der Ausgabe 8/2005

Intensive Care Medicine 8/2005 Zur Ausgabe

Announcements

August 2005

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.