Emergency CT head and neck imaging: effects of swimmer’s position on dose and image quality

verfasst von: Fabian G. Mueck, Sebastian Roesch, Lucas Geyer, Michael Scherr, Michael Seidenbusch, Robert Stahl, Zsuzsanna Deak, Stefan Wirth

Erschienen in: European Radiology | Ausgabe 5/2014

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To compare the effects of different arm positions on dose exposure and image quality (IQ) in cervical spine CT after trauma in different patient groups.

Methods

Patients in standard (STD = 126) and in swimmer’s position (SWIM = 254) were included. Body mass index (BMI subgroup 1 = underweight to subgroup 4 = obese), anterior–posterior diameter (AP), left–right diameter (LR), area of an ellipse (AoE) and angle between the humeral heads (optimal STD < 3°, optimal SWIM > 10°) were used as grouping criteria. Computed tomography dose index (CTDI) was documented. Two radiologists rated the IQ at three levels (CV1/2, CV4/5, CV7/T1) using a semi-quantitative scale (0 = not diagnostic, 1 = diagnostic with limitations, 2 = diagnostic without limitations). The Mann–Whitney U test correlations of grouping criteria with dose effects and intra-class correlation (ICC) were calculated.

Results

ICC was 0.87. BMI grouping showed the strongest correlation with dose effects: CTDI of optimal STD versus optimal SWIM positioning was 3.17 mGy versus 2.46 mGy (subgroup 1), 5.47 mGy versus 3.97 mGy (subgroup 2), 7.35 mGy versus 5.96 mGy (subgroup 3) and 8.71 mGy versus 8.18 mGy (subgroup 4). Mean IQ at CV7/T1 was 1.65 versus 1.23 (subgroup 1), 1.27 versus 1.46 (subgroup 2), 1.06 versus 1.46 (subgroup 3), 0.79 versus 1.5 (subgroup 4).

Conclusion

Patients with a BMI > 20 kg/m2 benefited from both potential dose reduction and improved image quality at the critical cervicothoracic junction when swimmer’s position was used.

Key Points

• BMI is a useful metric for personalized optimization in CT for the c-spine.
• Using swimmer’s position, patients can benefit from dose reduction.
• In some patients a superior image quality can be achieved with swimmer’s position.
• For swimmer’s positioning an angle of more than 10° is optimal.
Literatur
1.
Zurück zum Zitat Larson DB, Johnson LW, Schnell BM, Salisbury SR, Forman HP (2011) National trends in CT use in the emergency department: 1995–2007. Radiology 258:164–173PubMedCrossRef Larson DB, Johnson LW, Schnell BM, Salisbury SR, Forman HP (2011) National trends in CT use in the emergency department: 1995–2007. Radiology 258:164–173PubMedCrossRef
3.
Zurück zum Zitat Blackmore CC, Ramsey SD, Mann FA, Deyo RA (1999) Cervical spine screening with CT in trauma patients: a cost-effectiveness analysis. Radiology 212:117–125PubMedCrossRef Blackmore CC, Ramsey SD, Mann FA, Deyo RA (1999) Cervical spine screening with CT in trauma patients: a cost-effectiveness analysis. Radiology 212:117–125PubMedCrossRef
4.
Zurück zum Zitat Fleischmann D, Boas FE (2011) Computed tomography–old ideas and new technology. Eur Radiol 21:510–517PubMedCrossRef Fleischmann D, Boas FE (2011) Computed tomography–old ideas and new technology. Eur Radiol 21:510–517PubMedCrossRef
5.
Zurück zum Zitat Geyer LL, Korner M, Hempel R et al (2013) Evaluation of a dedicated MDCT protocol using iterative image reconstruction after cervical spine trauma. Clin Radiol 68:e391–396PubMedCrossRef Geyer LL, Korner M, Hempel R et al (2013) Evaluation of a dedicated MDCT protocol using iterative image reconstruction after cervical spine trauma. Clin Radiol 68:e391–396PubMedCrossRef
6.
Zurück zum Zitat Kropil P, Cohnen M, Andersen K, Heinen W, Stegmann V, Modder U (2010) Image quality in multidetector CT of paranasal sinuses: potential of dose reduction using an adaptive post-processing filter. Rofo 182:973–978PubMedCrossRef Kropil P, Cohnen M, Andersen K, Heinen W, Stegmann V, Modder U (2010) Image quality in multidetector CT of paranasal sinuses: potential of dose reduction using an adaptive post-processing filter. Rofo 182:973–978PubMedCrossRef
7.
Zurück zum Zitat Kropil P, Lanzman RS, Walther C et al (2010) Dose reduction and image quality in MDCT of the upper abdomen: potential of an adaptive post-processing filter. Rofo 182:248–253PubMedCrossRef Kropil P, Lanzman RS, Walther C et al (2010) Dose reduction and image quality in MDCT of the upper abdomen: potential of an adaptive post-processing filter. Rofo 182:248–253PubMedCrossRef
8.
Zurück zum Zitat Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34:4526–4544PubMedCrossRef Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34:4526–4544PubMedCrossRef
9.
Zurück zum Zitat McCollough CH, Bruesewitz MR, Kofler JM Jr (2006) CT dose reduction and dose management tools: overview of available options. Radiographics 26:503–512PubMedCrossRef McCollough CH, Bruesewitz MR, Kofler JM Jr (2006) CT dose reduction and dose management tools: overview of available options. Radiographics 26:503–512PubMedCrossRef
10.
Zurück zum Zitat Baum U, Anders K, Steinbichler G et al (2004) Improvement of image quality of multislice spiral CT scans of the head and neck region using a raw data-based multidimensional adaptive filtering (MAF) technique. Eur Radiol 14:1873–1881PubMedCrossRef Baum U, Anders K, Steinbichler G et al (2004) Improvement of image quality of multislice spiral CT scans of the head and neck region using a raw data-based multidimensional adaptive filtering (MAF) technique. Eur Radiol 14:1873–1881PubMedCrossRef
11.
Zurück zum Zitat Mulkens TH, Marchal P, Daineffe S et al (2007) Comparison of low-dose with standard-dose multidetector CT in cervical spine trauma. AJNR Am J Neuroradiol 28:1444–1450PubMedCrossRef Mulkens TH, Marchal P, Daineffe S et al (2007) Comparison of low-dose with standard-dose multidetector CT in cervical spine trauma. AJNR Am J Neuroradiol 28:1444–1450PubMedCrossRef
12.
Zurück zum Zitat Russell MT, Fink JR, Rebeles F, Kanal K, Ramos M, Anzai Y (2008) Balancing radiation dose and image quality: clinical applications of neck volume CT. AJNR Am J Neuroradiol 29:727–731PubMedCrossRef Russell MT, Fink JR, Rebeles F, Kanal K, Ramos M, Anzai Y (2008) Balancing radiation dose and image quality: clinical applications of neck volume CT. AJNR Am J Neuroradiol 29:727–731PubMedCrossRef
13.
Zurück zum Zitat Brink M, de Lange F, Oostveen LJ et al (2008) Arm raising at exposure-controlled multidetector trauma CT of thoracoabdominal region: higher image quality, lower radiation dose. Radiology 249:661–670PubMedCrossRef Brink M, de Lange F, Oostveen LJ et al (2008) Arm raising at exposure-controlled multidetector trauma CT of thoracoabdominal region: higher image quality, lower radiation dose. Radiology 249:661–670PubMedCrossRef
14.
Zurück zum Zitat Karlo C, Gnannt R, Frauenfelder T et al (2011) Whole-body CT in polytrauma patients: effect of arm positioning on thoracic and abdominal image quality. Emerg Radiol 18:285–293PubMedCrossRef Karlo C, Gnannt R, Frauenfelder T et al (2011) Whole-body CT in polytrauma patients: effect of arm positioning on thoracic and abdominal image quality. Emerg Radiol 18:285–293PubMedCrossRef
15.
Zurück zum Zitat Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK (2007) Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol 188:547–552PubMedCrossRef Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK (2007) Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol 188:547–552PubMedCrossRef
16.
Zurück zum Zitat Loewenhardt B, Buhl M, Gries A et al (2012) Radiation exposure in whole-body computed tomography of multiple trauma patients: bearing devices and patient positioning. Injury 43:67–72PubMedCrossRef Loewenhardt B, Buhl M, Gries A et al (2012) Radiation exposure in whole-body computed tomography of multiple trauma patients: bearing devices and patient positioning. Injury 43:67–72PubMedCrossRef
17.
Zurück zum Zitat Wirth S, Meindl T, Treitl M, Pfeifer KJ, Reiser M (2006) Comparison of different patient positioning strategies to minimize shoulder girdle artifacts in head and neck CT. Eur Radiol 16:1757–1762PubMedCrossRef Wirth S, Meindl T, Treitl M, Pfeifer KJ, Reiser M (2006) Comparison of different patient positioning strategies to minimize shoulder girdle artifacts in head and neck CT. Eur Radiol 16:1757–1762PubMedCrossRef
18.
Zurück zum Zitat Aberle DR, Abtin F, Brown K (2013) Computed tomography screening for lung cancer: has it finally arrived? Implications of the national lung screening trial. J Clin Oncol 31:1002–1008PubMedCentralPubMedCrossRef Aberle DR, Abtin F, Brown K (2013) Computed tomography screening for lung cancer: has it finally arrived? Implications of the national lung screening trial. J Clin Oncol 31:1002–1008PubMedCentralPubMedCrossRef
19.
Zurück zum Zitat Hoang JK, Yoshizumi TT, Nguyen G et al (2012) Variation in tube voltage for adult neck MDCT: effect on radiation dose and image quality. AJR Am J Roentgenol 198:621–627PubMedCrossRef Hoang JK, Yoshizumi TT, Nguyen G et al (2012) Variation in tube voltage for adult neck MDCT: effect on radiation dose and image quality. AJR Am J Roentgenol 198:621–627PubMedCrossRef
20.
Zurück zum Zitat National Lung Screening Trial Research T, Church TR, Black WC et al (2013) Results of initial low-dose computed tomographic screening for lung cancer. N Engl J Med 368:1980–1991CrossRef National Lung Screening Trial Research T, Church TR, Black WC et al (2013) Results of initial low-dose computed tomographic screening for lung cancer. N Engl J Med 368:1980–1991CrossRef
21.
Zurück zum Zitat Saltzherr TP, Fung Kon Jin PH, Beenen LF, Vandertop WP, Goslings JC (2009) Diagnostic imaging of cervical spine injuries following blunt trauma: a review of the literature and practical guideline. Injury 40:795–800PubMedCrossRef Saltzherr TP, Fung Kon Jin PH, Beenen LF, Vandertop WP, Goslings JC (2009) Diagnostic imaging of cervical spine injuries following blunt trauma: a review of the literature and practical guideline. Injury 40:795–800PubMedCrossRef
23.
Zurück zum Zitat Samei E, Badano A, Chakraborty D et al (2005) Assessment of display performance for medical imaging systems: executive summary of AAPM TG18 report. Med Phys 32:1205–1225PubMedCrossRef Samei E, Badano A, Chakraborty D et al (2005) Assessment of display performance for medical imaging systems: executive summary of AAPM TG18 report. Med Phys 32:1205–1225PubMedCrossRef
25.
Zurück zum Zitat Mueck FG, Michael L, Deak Z et al (2013) Upgrade to lterative image reconstruction (lR) in MDCT imaging: a clinical study for detailed parameter optimization beyond vendor recommendations using the adaptive statistical lterative reconstruction environment (ASIR) Part 2: the chest. Rofo. doi:10.1055/s-0033-1335152 PubMed Mueck FG, Michael L, Deak Z et al (2013) Upgrade to lterative image reconstruction (lR) in MDCT imaging: a clinical study for detailed parameter optimization beyond vendor recommendations using the adaptive statistical lterative reconstruction environment (ASIR) Part 2: the chest. Rofo. doi:10.​1055/​s-0033-1335152 PubMed
26.
Zurück zum Zitat Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166PubMedCrossRef Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166PubMedCrossRef
27.
Zurück zum Zitat Hoppe H, Vock P, Bonel HM, Ozdoba C, Gralla J (2006) A novel multiple-trauma CT-scanning protocol using patient repositioning. Emerg Radiol 13:123–128PubMedCrossRef Hoppe H, Vock P, Bonel HM, Ozdoba C, Gralla J (2006) A novel multiple-trauma CT-scanning protocol using patient repositioning. Emerg Radiol 13:123–128PubMedCrossRef
28.
Zurück zum Zitat Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248:254–263PubMedCrossRef Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248:254–263PubMedCrossRef
29.
Zurück zum Zitat Kane AG, Reilly KC, Murphy TF (2004) Swimmer's CT: improved imaging of the lower neck and thoracic inlet. AJNR Am J Neuroradiol 25:859–862PubMed Kane AG, Reilly KC, Murphy TF (2004) Swimmer's CT: improved imaging of the lower neck and thoracic inlet. AJNR Am J Neuroradiol 25:859–862PubMed
30.
Zurück zum Zitat Zarb F, Rainford L, McEntee MF (2010) AP diameter shows the strongest correlation with CTDI and DLP in abdominal and chest CT. Radiat Prot Dosim 140:266–273CrossRef Zarb F, Rainford L, McEntee MF (2010) AP diameter shows the strongest correlation with CTDI and DLP in abdominal and chest CT. Radiat Prot Dosim 140:266–273CrossRef
Metadaten
Titel
Emergency CT head and neck imaging: effects of swimmer’s position on dose and image quality
verfasst von
Fabian G. Mueck
Sebastian Roesch
Lucas Geyer
Michael Scherr
Michael Seidenbusch
Robert Stahl
Zsuzsanna Deak
Stefan Wirth
Publikationsdatum
01.05.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 5/2014
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3105-1

Weitere Artikel der Ausgabe 5/2014

European Radiology 5/2014 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.