Endotracheal tube cuff pressure in intensive care unit: the need for pressure monitoring

verfasst von: Samir Jaber, Moez El Kamel, Gérald Chanques, Mustapha Sebbane, Séverine Cazottes, Pierre-François Perrigault, Jean-Jacques Eledjam

Erschienen in: Intensive Care Medicine | Ausgabe 5/2007

Einloggen, um Zugang zu erhalten

Excerpt

Sir: Endotracheal tube cuff pressure (CP) should be in the range of 15–22 mmHg to avoid complications related to underinflation (aspiration and leakage) and overinflation (tracheal wall ischemia, stenosis and tracheo-esophageal fistulae). Cuff overinflation (CP > 22 mmHg) is frequent (40–80%), both in intensive care unit (ICU) patients [1, 2] and in patients under general anesthesia [3, 4]. In contrast to the case in the operating room the impact of the implementation of a routine cuff pressure monitoring protocol on reducing the incidence of cuff overinflation in the ICU has not been evaluated. We therefore conducted a prospective before-after study in a 16-bed ICU to (a) evaluate the values of CP before the implementation of a cuff pressure monitoring protocol and (b) to determine the protocol's impact on reducing cuff overinflation 1 month and 3 years later. The study was conducted over three 1-month periods (P1, P2, and P3). During the first period (P1) the paramedical team was not aware of the study. At the end of P1 preliminary results were communicated to the staff during a briefing on the cuff pressure monitoring protocol. This program consisted of the nursing staff performing routine CP measurements and recording the CP values on the patient charts. Measurements were made during every team shift or whenever there was a manipulation that could potentially modify the CP. The objective was to obtain the minimal pressure necessary to prevent an audible leak around the tube cuff without exceeding 22 mmHg. CP, diastolic (DAP), and mean (MAP) arterial pressures were measured daily. Over the three periods 293 measurements were made in 103 patients. Between the three periods there were no significant differences in demographic data or in the number of patients and CP measurements (Table 1). Both mean CP and overinflation rate were significantly higher in P1 than in P2 and P3 (Table 1). In P3 the rate of underinflation was higher than in P2. There were significantly more cases with CP > DAP and CP > MAP in P1 than in P2 and P3 (Table 1). In comparison to our first period, Spittle and Beavis [1] performed a regional telephone request in 22 UK ICUs and found that only 13% of units measured CP routinely. There was a lack of knowledge among the nursing staff as well; 67% of senior nurses did not know the recommended range of intracuff pressure [1]. In another study CP was not monitored in 75% of ICUs [2]. A recent report found that 38% of nurses in 11 different ICUs believed that the only aim of the cuff is to prevent extubation by maintaining the tube in the trachea, and 50% did not consider themselves sufficiently trained to manage CP monitoring [5].
Table 1
Main admission characteristics of the patients and main results obtained during the three periods of the study: before protocol (P1), 1 month after protocol (P2), and 3 years after protocol (P3) (CP cuff pressure, BMI body mass index, SAP systolic arterial pressure, DAP diastolic arterial pressure, MAP mean arterial pressure, SAPS II Simplified Acute Physiology Score)
 
P1
P2
P3
p: P2 vs. P1
p: P3 vs. P1
Number of patients
30
41
32
Number of CP measurements
93
101
99
Age (years)
53 ± 11
51 ± 14
54 ± 14
0.09
0.12
Sex ratio (M/F)
1.0
1.2
1.0
0.73
0.98
BMI
22 ± 6
22 ± 5
21 ± 6
0.76
0.36
SAPS II score
43 ± 4
45 ± 4
45 ± 6
0.48
0.56
Cause of admission
     
0.74
0.62
Medical
39 (42%)
40 (40%)
45 (45%)
   
Surgical
54 (58%)
61 (61%)
54 (55%)
   
Tube size
     
0.41
0.99
7.0–7.5
46 (50%)
56 (55%)
49 (50%)
   
8.0–8.5
47 (50%)
45 (45%)
50 (50%)
   
SAP (mmHg)
140 ± 24
127 ± 32
128 ± 24
0.11
0.21
DAP (mmHg)
67 ± 12
67 ± 14
65 ± 13
0.86
0.75
MAP (mmHg)
90 ± 14
89 ± 15
86 ± 16
0.72
0.68
Mean CP (mmHg)
42 ± 22
26 ± 3
21 ± 19
< 0.001
< 0.001
CP
         
≤ 15 mmHg
5 (6%)
8 (8%)
40 (40%)
0.50
< 0.001
15–21.9 mmHg
16 (17%)
69 (68%)
28 (28%)
< 0.001
0.07
> 22 mmHg
72 (77%)
24 (24%)
31 (31%)
< 0.001
< 0.001
CP > DAP
22 (24%)
3 (3%)
6 (6%)
< 0.001
< 0.001
CP > MAP
10 (12%)
2 (2%)
2 (2%)
0.01
0.01
Literatur
1.
Zurück zum Zitat Spittle C, Beavis S (2001) Do you measure tracheal cuff pressure? A survey of clinical practice. Br J Anaesth 87:344–345 Spittle C, Beavis S (2001) Do you measure tracheal cuff pressure? A survey of clinical practice. Br J Anaesth 87:344–345
2.
Zurück zum Zitat Vyas D, Inweregbu K, Pittard A (2002) Measurement of tracheal tube cuff pressure in critical care. Anaesthesia 57:275–277CrossRefPubMed Vyas D, Inweregbu K, Pittard A (2002) Measurement of tracheal tube cuff pressure in critical care. Anaesthesia 57:275–277CrossRefPubMed
3.
Zurück zum Zitat Braz J, Navarro L, Takata I, Junior P (1999) Endotracheal tube cuff pressure: need for precise measurement. Sao Paulo Med J 117:243–247CrossRefPubMed Braz J, Navarro L, Takata I, Junior P (1999) Endotracheal tube cuff pressure: need for precise measurement. Sao Paulo Med J 117:243–247CrossRefPubMed
4.
Zurück zum Zitat Sengupta P, Sessler D, Maglinger P, Wells S, Vogt A, Durrani J, Wadhwa A (2004) A Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol 29:4–8 Sengupta P, Sessler D, Maglinger P, Wells S, Vogt A, Durrani J, Wadhwa A (2004) A Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol 29:4–8
5.
Zurück zum Zitat Mol D, De Villiers Gdu T, Claassen A, Joubert G (2004) Use and care of an endotracheal/tracheostomy tube cuff-are intensive care unit staff adequately informed? S Afr J Surg 42:14–16PubMed Mol D, De Villiers Gdu T, Claassen A, Joubert G (2004) Use and care of an endotracheal/tracheostomy tube cuff-are intensive care unit staff adequately informed? S Afr J Surg 42:14–16PubMed
Metadaten
Titel
Endotracheal tube cuff pressure in intensive care unit: the need for pressure monitoring
verfasst von
Samir Jaber
Moez El Kamel
Gérald Chanques
Mustapha Sebbane
Séverine Cazottes
Pierre-François Perrigault
Jean-Jacques Eledjam
Publikationsdatum
01.05.2007
Verlag
Springer Berlin Heidelberg
Erschienen in
Intensive Care Medicine / Ausgabe 5/2007
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0594-2

Weitere Artikel der Ausgabe 5/2007

Intensive Care Medicine 5/2007 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.