Evaluation of accuracy of capillary refill index with pneumatic fingertip compression

verfasst von: Koichiro Shinozaki, Kota Saeki, Lee S. Jacobson, Julianne M. Falotico, Timmy Li, Hideaki Hirahara, Katsuyuki Horie, Naoki Kobayashi, Steve Weisner, Joshua W. Lampe, Lance B. Becker

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Capillary refill time (CRT) is a method of measuring a patient’s peripheral perfusion status through a visual assessment performed by a clinician. We developed a new method of measuring CRT using standard pulse oximetry sensor, which was designated capillary refill index (CRI). We evaluated the accuracy of CRI in comparison to CRT image analysis. Thirty healthy adult volunteers were recruited for a derivation study and 30 patients in the emergency department (ED) were for validation. Our high fidelity mechanical device compresses and releases the fingertip to measure changes in blood volume using infrared-light (940 nm). CRT was calculated by image analysis software using recorded fingertip videos. CRI and CRT were measured at: room temperature (ROOM TEMP), 15 °C cold water (COLD), and 38 °C warm water (REWARM). Intra-rater reliability, Bland–Altman plots, and correlation coefficients were used to evaluate the accuracy of the novel CRI method. CRI (4.9 [95% CI 4.5–5.3] s) and CRT (4.0 [3.6–4.3]) in the COLD group were higher than the ROOM TEMP and REWARM groups. High intra-rater reliability was observed in both measurements (0.97 [0.95–0.98] and 0.98 [0.97–0.99], respectively). The Bland–Altman plots suggested a systematic bias: CRI was consistently higher than CRT (difference: + 1.01 s). There was a strong correlation between CRI and CRT (r = 0.89, p < 0.001). ED patients had higher CRI (3.91 [5.05–2.75]) and CRT (2.21 [3.19–1.23]) than those of healthy volunteers at room temperature. The same difference and correlation patterns were verified in the ED setting. CRI was as reliable as CRT by image analysis. The values of CRI was approximately 1 s higher than CRT.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Pickard A, Karlen W, Ansermino JM. Capillary refill time: is it still a useful clinical sign? Anesth Analg. 2011;113:120–3.CrossRef Pickard A, Karlen W, Ansermino JM. Capillary refill time: is it still a useful clinical sign? Anesth Analg. 2011;113:120–3.CrossRef
2.
Zurück zum Zitat Lima A, Bakker J. Noninvasive monitoring of peripheral perfusion. Intensive Care Med. 2005;31:1316–26.CrossRef Lima A, Bakker J. Noninvasive monitoring of peripheral perfusion. Intensive Care Med. 2005;31:1316–26.CrossRef
3.
Zurück zum Zitat Lima A, Jansen TC, van Bommel J, et al. The prognostic value of the subjective assessment of peripheral perfusion in critically ill patients. Crit Care Med. 2009;37:934–8.CrossRef Lima A, Jansen TC, van Bommel J, et al. The prognostic value of the subjective assessment of peripheral perfusion in critically ill patients. Crit Care Med. 2009;37:934–8.CrossRef
4.
Zurück zum Zitat van Genderen ME, Paauwe J, de Jonge J, et al. Clinical assessment of peripheral perfusion to predict postoperative complications after major abdominal surgery early: a prospective observational study in adults. Crit Care. 2014;18:R114.CrossRef van Genderen ME, Paauwe J, de Jonge J, et al. Clinical assessment of peripheral perfusion to predict postoperative complications after major abdominal surgery early: a prospective observational study in adults. Crit Care. 2014;18:R114.CrossRef
5.
Zurück zum Zitat Hernandez G, Ospina-Tascon GA, Damiani LP, et al. Effect of a resuscitation strategy targeting peripheral perfusion status vs serum lactate levels on 28-day mortality among patients with septic shock: the ANDROMEDA-SHOCK randomized clinical trial. JAMA. 2019;321:654–64.CrossRef Hernandez G, Ospina-Tascon GA, Damiani LP, et al. Effect of a resuscitation strategy targeting peripheral perfusion status vs serum lactate levels on 28-day mortality among patients with septic shock: the ANDROMEDA-SHOCK randomized clinical trial. JAMA. 2019;321:654–64.CrossRef
6.
Zurück zum Zitat Shinozaki K, Jacobson LS, Saeki K, et al. Does training level affect the accuracy of visual assessment of capillary refill time? Crit Care. 2019;23:157.CrossRef Shinozaki K, Jacobson LS, Saeki K, et al. Does training level affect the accuracy of visual assessment of capillary refill time? Crit Care. 2019;23:157.CrossRef
7.
Zurück zum Zitat Schriger DL, Baraff L. Defining normal capillary refill: variation with age, sex, and temperature. Ann Emerg Med. 1988;17:932–5.CrossRef Schriger DL, Baraff L. Defining normal capillary refill: variation with age, sex, and temperature. Ann Emerg Med. 1988;17:932–5.CrossRef
9.
Zurück zum Zitat Severinghaus JW, Honda Y. History of blood gas analysis. VII. Pulse oximetry. J Clin Monit. 1987;3:135–8.CrossRef Severinghaus JW, Honda Y. History of blood gas analysis. VII. Pulse oximetry. J Clin Monit. 1987;3:135–8.CrossRef
10.
Zurück zum Zitat Morimura N, Takahashi K, Doi T, et al. A pilot study of quantitative capillary refill time to identify high blood lactate levels in critically ill patients. Emerg Med J. 2015;32:444–8.CrossRef Morimura N, Takahashi K, Doi T, et al. A pilot study of quantitative capillary refill time to identify high blood lactate levels in critically ill patients. Emerg Med J. 2015;32:444–8.CrossRef
11.
Zurück zum Zitat Shinozaki K, Capilupi MJ, Saeki K, et al. Low temperature increases capillary blood refill time following mechanical fingertip compression of healthy volunteers: prospective cohort study. J Clin Monit Comput. 2019;33:259–67.CrossRef Shinozaki K, Capilupi MJ, Saeki K, et al. Low temperature increases capillary blood refill time following mechanical fingertip compression of healthy volunteers: prospective cohort study. J Clin Monit Comput. 2019;33:259–67.CrossRef
12.
Zurück zum Zitat Shinozaki K, Capilupi MJ, Saeki K, et al. Blood refill time: clinical bedside monitoring of peripheral blood perfusion using pulse oximetry sensor and mechanical compression. Am J Emerg Med. 2018;36:2310–2.CrossRef Shinozaki K, Capilupi MJ, Saeki K, et al. Blood refill time: clinical bedside monitoring of peripheral blood perfusion using pulse oximetry sensor and mechanical compression. Am J Emerg Med. 2018;36:2310–2.CrossRef
13.
Zurück zum Zitat Lima A, van Genderen ME, Klijn E, et al. Peripheral vasoconstriction influences thenar oxygen saturation as measured by near-infrared spectroscopy. Intensive Care Med. 2012;38:606–11.CrossRef Lima A, van Genderen ME, Klijn E, et al. Peripheral vasoconstriction influences thenar oxygen saturation as measured by near-infrared spectroscopy. Intensive Care Med. 2012;38:606–11.CrossRef
14.
Zurück zum Zitat Blaxter LL, Morris DE, Crowe JA, et al. An automated quasi-continuous capillary refill timing device. Physiol Meas. 2016;37:83–99.CrossRef Blaxter LL, Morris DE, Crowe JA, et al. An automated quasi-continuous capillary refill timing device. Physiol Meas. 2016;37:83–99.CrossRef
15.
Zurück zum Zitat Whitehead AL, Julious SA, Cooper CL, et al. Estimating the sample size for a pilot randomised trial to minimise the overall trial sample size for the external pilot and main trial for a continuous outcome variable. Stat Methods Med Res. 2016;25:1057–73.CrossRef Whitehead AL, Julious SA, Cooper CL, et al. Estimating the sample size for a pilot randomised trial to minimise the overall trial sample size for the external pilot and main trial for a continuous outcome variable. Stat Methods Med Res. 2016;25:1057–73.CrossRef
16.
Zurück zum Zitat Browne RH. On the use of a pilot sample for sample size determination. Stat Med. 1995;14:1933–40.CrossRef Browne RH. On the use of a pilot sample for sample size determination. Stat Med. 1995;14:1933–40.CrossRef
17.
Zurück zum Zitat John RT, Henricson J, Nilsson GE, et al. Reflectance spectroscopy: to shed new light on the capillary refill test. J Biophotonics. 2018;11:e20170043. John RT, Henricson J, Nilsson GE, et al. Reflectance spectroscopy: to shed new light on the capillary refill test. J Biophotonics. 2018;11:e20170043.
18.
Zurück zum Zitat Shavit I, Brant R, Nijssen-Jordan C, et al. A novel imaging technique to measure capillary-refill time: improving diagnostic accuracy for dehydration in young children with gastroenteritis. Pediatrics. 2006;118:2402–8.CrossRef Shavit I, Brant R, Nijssen-Jordan C, et al. A novel imaging technique to measure capillary-refill time: improving diagnostic accuracy for dehydration in young children with gastroenteritis. Pediatrics. 2006;118:2402–8.CrossRef
19.
Zurück zum Zitat Bakker J, Jansen TC. Don't take vitals, take a lactate. Intensive Care Med. 2007;33:1863–5.CrossRef Bakker J, Jansen TC. Don't take vitals, take a lactate. Intensive Care Med. 2007;33:1863–5.CrossRef
20.
Zurück zum Zitat Jansen TC, van Bommel J, Woodward R, et al. Association between blood lactate levels, Sequential Organ Failure Assessment subscores, and 28-day mortality during early and late intensive care unit stay: a retrospective observational study. Crit Care Med. 2009;37:2369–74.CrossRef Jansen TC, van Bommel J, Woodward R, et al. Association between blood lactate levels, Sequential Organ Failure Assessment subscores, and 28-day mortality during early and late intensive care unit stay: a retrospective observational study. Crit Care Med. 2009;37:2369–74.CrossRef
21.
Zurück zum Zitat Kawaguchi R, Nakada TA, Oshima T, et al. Optimal pressing strength and time for capillary refilling time. Crit Care. 2019;23:4.CrossRef Kawaguchi R, Nakada TA, Oshima T, et al. Optimal pressing strength and time for capillary refilling time. Crit Care. 2019;23:4.CrossRef
Metadaten
Titel
Evaluation of accuracy of capillary refill index with pneumatic fingertip compression
verfasst von
Koichiro Shinozaki
Kota Saeki
Lee S. Jacobson
Julianne M. Falotico
Timmy Li
Hideaki Hirahara
Katsuyuki Horie
Naoki Kobayashi
Steve Weisner
Joshua W. Lampe
Lance B. Becker
Publikationsdatum
08.01.2020
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2021
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00454-1

Weitere Artikel der Ausgabe 1/2021

Journal of Clinical Monitoring and Computing 1/2021 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.