Fluid attenuated inversion recovery (FLAIR) MRI at 7.0 Tesla: comparison with 1.5 and 3.0 Tesla

verfasst von: Jaco J. M. Zwanenburg, Jeroen Hendrikse, Fredy Visser, Taro Takahara, Peter R. Luijten

Erschienen in: European Radiology | Ausgabe 4/2010

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Purpose

To assess fluid-attenuated inversion recovery (FLAIR) magnetic resonance imaging (MRI) at three field strengths, regarding signal-to-noise ratio (SNR), contrast and signal homogeneity, in order to determine the potential gain and current challenges of FLAIR at ultra-high field strength (7 T).

Methods

FLAIR images of five healthy volunteers (age 24 ± 4 years, 4 male) were acquired at 1.5 T, 3 T and 7 T. Image homogeneity and visibility of normal brain structures were evaluated. SNR of grey matter (GM), white matter (WM) and cerebrospinal fluid (CSF) were measured in regions not affected by transmit field heterogeneity.

Results

The SNR (mean ± SD) at 7 T (GM 168 ± 15, WM 125 ± 11) increased slightly more than proportionally, compared with at 1.5 T (GM 30 ± 3, WM 22 ± 2) and 3 T (GM 62 ± 7, WM 44 ± 4). Relative contrast between GM and WM at 7 T (1.35 ± 0.07) was slightly less than at 3 T (1.42 ± 0.14) or 1.5 T (1.37 ± 0.07). Several major fibre bundles became visible at 7 T. One incidentally observed white matter lesion was well visible at all field strengths.

Conclusion

Image homogeneity remains challenging and should be improved by future technical developments. FLAIR imaging at 7 T yields a high SNR,with better contrast for WM substructures and the iron-bearing basal ganglia, and has potential for good conspicuity of WM lesions.

Introduction

T2-weighted fluid attenuated inversion recovery (FLAIR) imaging is one of the cornerstones of MR neuroimaging protocols. With the nulling of the cerebrospinal fluid (CSF), FLAIR imaging results in an excellent contrast between lesions in the white matter or grey matter and the surrounding brain tissue [1]. For instance, FLAIR imaging is regarded as one of the most sensitive imaging techniques for the detection of white matter lesions, either age related or in patients with multiple sclerosis [2, 3]. FLAIR imaging is nowadays a robust sequence at 1.5 T and 3 T, the clinically most commonly used field strengths [4]. Recently, ultra-high field strength magnets of 7 T have become more widely available. Studies showed that spectroscopy [5], gradient echo imaging [6] and MR angiography sequences [7, 8] benefit most from the increased signal-to-noise (SNR) at 7 T. The combination of a higher SNR and a shorter T2* is beneficial for BOLD fMRI [9, 10] and susceptibility-weighted imaging [6]. With the wider availability of 7-T magnets interest has increased in the clinical application of ultra-high field MRI, and recently 7-T imaging results of a multiple sclerosis patient were presented, including a FLAIR image [11]. Thus far, however, no studies have compared the current quality of FLAIR imaging at 7 T with FLAIR at the commonly used field strengths 1.5 T and 3 T. For clinical neuroimaging at 7 T, it is crucial to have a FLAIR imaging protocol with comparable quality to those of lower field strengths. Such a protocol enables an adequate evaluation of abnormalities and a better interpretation of the results from the above-mentioned MR sequences that benefit from an ultra-high field strength.
FLAIR imaging at 7 T is challenging, mainly because of practical issues related to the homogeneity of the magnetic field and limitations in the specific absorption rate (SAR). Image heterogeneity is caused by variations in homogeneity of both the receive field and the transmit field (B 1 +). Variations in the transmit field at 7 T results in spatial variation in excitation and refocusing flip angles, which leads to variable contrast and signal within the acquired images. Inadequate nulling of the CSF may be prevented by the use of an adiabatic inversion pulse, which can achieve full inversion in the presence of substantial inhomogeneities in the main magnetic field and the transmit field [12]. However, adiabatic inversion pulses have a much higher SAR than ordinary pulses. The SAR of an adiabatic pulse can easily be five times higher than the SAR of a plain amplitude modulated pulse.
A more fundamental problem may arise from the field strength dependency of the T 1 relaxation times. Because of the prolonged T 1 relaxation times of grey and white matter (GM and WM), with a relatively constant T 1 of cerebrospinal fluid (CSF) [13], the contrast in FLAIR may be less at ultra-high field strength, as the T 1 contrast counteracts the T 2 contrast. The field-strength-dependent effect of the competing T 1 contrast can be illustrated with straightforward calculations of the magnetisation in an inversion recovery sequence with infinite repetition time. Assuming a perfect inversion pulse and a perfect 90° excitation pulse after the inversion delay TI, the transverse magnetisation following excitation can be shown to be
$$M_{xy} = M_0 \left( {1 - 2{\text{e}}^{ - {{{\text{TI}}} \mathord{\left/ {\vphantom {{{\text{TI}}} {T_1 }}} \right. \kern-\nulldelimiterspace} {T_1 }}} } \right){\text{e}}^{ - {t \mathord{\left/ {\vphantom {t {T_2 }}} \right. \kern-\nulldelimiterspace} {T_2 }}} .$$
In this equation, M xy is the transverse magnetisation, M 0 the longitudinal equilibrium magnetisation, and t the time following excitation. A prolonged T 1 at a given TI implies less relaxation and thus more remaining T 1 contrast. This is illustrated in Fig. 1, where both the relaxation of the magnetisation after inversion and the signal after excitation are shown. The field-strength-dependent T 1 values were taken from Rooney et al. [13]. In order to only illustrate the T 1 effects, a single T 2 for GM (90 ms) and WM (60 ms) for all field strengths was used.
Despite these challenges of FLAIR imaging at 7 T, FLAIR imaging may benefit from the increased SNR at ultra-high field strength, similar to other pulse sequences. The present paper assesses the potential of FLAIR imaging at 7 T by concentrating on the image areas that suffer least from variations in the transmit field. According to our experience, adequate nulling of CSF at 7 T can be obtained in the entire brain by using an adiabatic inversion pulse. The subsequent SAR problem can be alleviated by reducing the number of acquired slices, while an increased slice gap still provides images distributed over the entire brain.
The aim of the present study is to assess fluid-attenuated inversion recovery (FLAIR) magnetic resonance imaging (MRI) at three different field strengths (the most commonly used field strength 1.5 T, high field 3 T and ultra-high field 7 T), regarding signal-to-noise ratio (SNR), contrast and signal homogeneity. This comparison is important in order to understand the potential advantages and current challenges of FLAIR imaging for clinical neuroimaging studies at ultra-high field strength.

Material and methods

Five volunteers (age 24 ± 4 years, 4 male, 1 female) without history of neurological disease were imaged on 1.5-T, 3-T and 7-T MRI (Philips Healthcare), with an 8-channel receive head coil and body transmit coil at 1.5 T and 3 T (max. B 1 13.5 μT), and a 16-channel receive coil with volume T/R coil for transmission (max. B 1 20 μT) at 7 T (Nova Medical, Wilmington, MA, USA). Written informed consent was obtained in accordance with the Institutional Review Board of our hospital.
The parameters of the FLAIR sequence are largely determined by the requirement to null CSF, and the limitations imposed by the SAR constraints, as modelled in the system software. A straightforward comparison of the FLAIR sequences at different field strengths was performed by slightly adjusting the inversion delay (TI) to null CSF on all systems, and by reducing the number of slices at the 7-T system to stay within the SAR limits. Imaging parameters can be found in Table 1. The inversion pulse for CSF suppression consisted of an adiabatic hyperbolic secant pulse (at all field strengths), which proved to give sufficient inversion for the whole brain at 7 T, despite the inhomogeneity in the transmit field. No SENSE acceleration was applied. The default image homogeneity correction of the MRI system was applied at all field strengths, to correct for image intensity variation induced by the receive coil sensitivity pattern. The noise level for each image was measured by repeating each data acquisition with the same receiver gain for data sampling, without applying any RF pulses or gradients.
Table 1
Imaging parameters of the T 2-weighted FLAIR sequences at the three different field strengths
 
1.5 T
3 T
7 T
FOV (mm2)
230 × 183
230 × 183
230 × 183
Acquired resolution (mm2)
0.65 × 0.88
0.65 × 0.88
0.65 × 0.88
Slice thickness (mm)
4
4
4
Slice spacing (mm)
0
0
8
Number of interleaves
3
3
1
Turbo spin-echo factor
26
26
26
Refocusing flip angle (°)
120
120
120
TI (ms)
2,800
2,900
3,100
TR (ms)
14,000
14,000
21,222
TE
125
125
101
Echo spacing
9.3
9.3
7.5
Readout BW (Hz/pixel)
221
221
230
Number of slices
24
24
8
Imaging duration (s)
378
378
191
Estimated SAR (W/kg)
0.3
0.8
1.9
FOV field of view, TI inversion delay, TR repetition time, TE echo time, BW bandwidth, SAR specific absorption rate
A qualitative comparison of the images was performed, regarding heterogeneity in signal intensity and visibility of brain structure at the three field strengths. To assess the SNR, regions of interest (ROIs) were defined at one location in the brain, where signal drop due to transmit field heterogeneity was minimal. The obtained SNR and contrast from these ROIs can be regarded as an indication of the potential of 7-T FLAIR imaging, when the technical limitations of obtaining a homogeneous flip angle are overcome. The used ROIs are illustrated in Fig. 2. The standard deviation of the combined ROIs in the noise image was used as noise metric for the SNR calculation. Relative contrast between GM and WM was defined as SNRGM/SNRWM.
SNR of grey and white matter, and relative contrast between GM and WM were compared among the field strengths using paired t tests. For the relative contrast, single-sided P values are reported, as we expect from theory that the relative contrast will decrease towards higher field strength (yielding a difference in a single direction). A P value < 0.05 was considered to be significant.

Results

Figure 3 shows representative FLAIR images of a single subject at all three field strengths. The 7-T images show high SNR and high contrast between different structures in most of the cerebrum. Marked variation in image intensity was observed, with signal loss in the frontal and parietal operculum of the brain, which increased towards the more caudal slices. This variation is caused by the well-known spatial variation in the transmit field. Despite this variation in the transmit field, the signal from CSF was adequately suppressed in all slices.
The SNR values (mean ± SD) at 7 T (GM 168 ± 15, WM 125 ± 11) increased slightly more than proportionally, compared with at 1.5 T (GM 30 ± 3, WM 22 ± 2) and 3 T (GM 62 ± 7, WM 44 ± 4), as illustrated in Fig. 4. The relative contrast at 7 T (1.35 ± 0.07) was slightly less than that at 3 T (1.42 ± 0.14) or 1.5 T (1.37 ± 0.07), but these differences were not significant (= 0.12 and P = 0.13, respectively). The signal from the CSF was well suppressed at all field strengths. The measured SNR of CSF was 4.3 ± 0.7 for 1.5 T, 5.6 ± 0.9 for 3 T, and 10 ± 3 for 7 T.
Several interesting observations were made. First, at 7 T several white matter structures became much more apparent, which were consistent with the major fibre tracts like the pyramidal tracts and the optic radiation (Fig. 5). Second, there was an increase in contrast in the area of the basal ganglia at high field strength, which is clear from Figs. 3 and 5. This contrast originated mainly from decreased signal in the globus pallidus at 3 T and especially at 7 T. Increased hypointensity was also observed for the nucleus ruber at 3 T and 7 T. Within the thalamus, subnuclear structures became visible at 7 T (Fig. 5). Third, one of the volunteers had a small incidental white matter lesion, which was plainly visible at all three field strengths. At 7 T, the location of this lesion with respect to the corona radiata could clearly be identified (Fig. 6).

Discussion

The SNR of FLAIR imaging showed a clear, approximately linear, increase from 1.5 T to 3 T and 7 T. A trend was found towards a lower relative grey–white matter contrast with FLAIR imaging at 7 T compared with 1.5 T and 3 T, while the contrast between the basal ganglia increased. Furthermore, white matter structures consistent with major fibre bundles became more apparently visible at 7 T. We showed the possibility of an adequate CSF suppression with FLAIR imaging at 7 T, despite the inhomogeneity of the transmit field. However, severe signal heterogeneities were still present in the region of the temporal lobe and in the cerebellum, which needs to be addressed by future technical developments for adequate FLAIR imaging of the full brain.
As expected we showed a clear SNR increase in FLAIR imaging at 7 T compared with 1.5 T and 3 T. These results are in line with previous studies comparing the signal characteristics of FLAIR at 3 T and 1.5 T, although no formal SNR comparisons were made in these studies [1416]. Bachman et al. reported SNR and contrast-to-noise values for 1.5 T and 3 T [17]. If one corrects for the difference in voxel size between 1.5 T and 3 T in that study, it appears that there was also an approximately linear increase in SNR with field strength.
The FLAIR example images suggest that the increased SNR will result in better delineation of white matter lesions, with less noise, from a lower towards a higher field strength. However, a larger 7-T study with patients is needed to investigate the visibility of white matter lesions in different areas in the brain, such as the white matter bundles. Recent studies in MS patients showed that white matter lesions in MS patients are often located in the proximity of veins [11]. For these studies susceptibility-weighted imaging [18] at 7 T might be combined with FLAIR imaging to obtain a high sensitivity for veins and white matter lesions, respectively.
Interestingly, FLAIR imaging at 7 T resulted in a different contrast for several substructures of the brain. Anatomical substructures in the white matter can clearly be distinguished, such as the pyramidal and cortical spinal tracts. Furthermore, we found a clear hypointensity of the globus pallidus and the nucleus ruber at 3 T, but most obviously at 7 T. Also structures in the thalamus became visible at 7 T. The basal ganglia have a relatively high iron content, and hence a relatively short T 2, which may explain the observed hypointensity of these regions.
No significant difference was found in the present study with regard to the relative contrast to noise at different field strengths although a trend was found towards a lower relative contrast at 7 T. As has been illustrated in Fig. 1, the contrast between GM and WM in FLAIR was expected to be less at ultra-high field strength, as T 1 contrast counteracts the T 2 contrast. Still, this effect was relatively small for the GM and WM ROI analysed in the present study, and not significant within the small number of volunteers. The effect of the competing T 1 contrast may be stronger in other regions of the brain, as the contrast varies throughout the brain as a result of differences in T 1 and T 2 values. Especially for the occipital lobe the grey–white matter contrast is known to be different from that of other regions of the brain. The results may also be different from the straightforward simulations, as the simulations did not take into account the use of multiple refocusing pulses with concomitant magnetisation transfer effects, and the variation in T 2 values across field strengths.
It is worth noting that the visibility of white matter substructures at 7 T is similar to that described by Hajnal et al. [1], in the first publication of the FLAIR technique on a 1.0-T MRI system at a long TE (240 ms). By imaging an infant, Hajnal et al. showed that the T 2-weighted FLAIR signal is sensitive to the degree of myelinisation, as myelinated white matter regions appeared dark and unmyelinated white matter appeared bright. They suggested that the observed contrast within white matter may be explained partly by a small susceptibility difference between myelinated fibres and surrounding tissue. Such a susceptibility difference would induce magnetic field gradients around fibre bundles that are not parallel to the main field, leading to diffusion-induced signal loss in a spin echo. Fibres that are parallel to the main field do not produce susceptibility-induced gradients and will hence have higher signal in a spin echo sequence. The degree of such diffusion-induced signal loss would depend obviously on the strength of the susceptibility-induced gradients and the TE. Because the susceptibility-induced gradients increase with increasing field strength, a small susceptibility difference between myelinated fibres and surrounding tissue would perfectly explain the fact that we obtained a similar contrast as Hajnal, at a much shorter echo time.
Li et al. [6] also observed white matter substructures at 7 T, using T 2*-weighted images, and suggested differences in myelin content, iron concentration and/or fibre direction with respect to the main field as potential contrast mechanisms. It should be noted, however, that the contrast mechanism for T 2-weighted FLAIR is different from the commonly used susceptibility-weighted imaging at high field strength, as direct susceptibility effects are reversed in a spin echo. Indirectly, susceptibility effects may play a role as spin diffusion through susceptibility-induced field gradients may yield irreversible signal loss.
A limitation of the present study is that the current SAR limitations did not allow for full brain coverage at 7 T in a reasonable scan time. Reducing the number of slices resulted in a wider temporal spacing of the adiabatic inversion pulses (longer TR) and thus in a reduction in the SAR. Full brain coverage could be obtained at the cost of two extra interleaves, leading to a longer imaging time of approximately 10 min. Moreover, even more slices would be required if one would exploit the gain in SNR to obtain a higher resolution. Therefore, further developments are certainly required to obtain an adequate high resolution T 2-weighted FLAIR protocol at 7 T. A possible solution for full brain coverage within a reasonable imaging time is the implementation of B 1 shimming, which may allow a less SAR demanding inversion pulse to be used. The use of low refocusing flip angles may also greatly reduce SAR, while T 2 contrast can be preserved by increasing the echo train length.
In the present study a slightly shorter echo time was used at 7 T to compensate for a shorter T 2 at this field strength. Owing to the somewhat higher (nominal) B 1 available at 7 T, the shorter echo time was possible without significant change in the receiver bandwidth, which makes the SNR directly comparable among the field strengths. However, the TR was longer at 7 T, which may have led to a difference in the amount of signal loss from magnetisation transfer. Furthermore there was a difference in the receive coils used, with a 16-channel receive coil used at 7T and 8-channel receive coils at 1.5 T and 3 T, which also may have contributed to the observed increase in SNR at 7 T. As the inhomogeneity in both transmit and receive fields requires new solutions with multiple coils at high field strength [19], one may regard the gain in SNR observed in the studied region as an indication of the potential of FLAIR at 7 T.
A limitation that FLAIR imaging has in common with other imaging techniques at 7 T is the sensitivity to variations in transmit field and a relative low RF power in the temporal lobes and the frontal and temporal operculum. Consequently, there is a major drop in signal in these regions. Future developments and optimisation of multichannel transmit options should provide a solution for a more homogeneous intensity throughout the brain.
In conclusion, we demonstrate the potential of T 2-weighted FLAIR imaging at 7 T with a high SNR and contrast that allow the depiction of anatomical detail such as white matter substructures in addition to the depiction of white matter lesions. Furthermore, the increase in contrast of white matter substructures and deep brain structures show that FLAIR imaging at 7 T may be a valuable addition to MR neuroimaging protocols at this ultra-high field strength.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Hajnal JV, De CB, Lewis PD, Baudouin CJ, Cowan FM, Pennock JM, Young IR, Bydder GM (1992) High signal regions in normal white matter shown by heavily T2-weighted CSF nulled IR sequences. J Comput Assist Tomogr 16:506–513PubMedCrossRef Hajnal JV, De CB, Lewis PD, Baudouin CJ, Cowan FM, Pennock JM, Young IR, Bydder GM (1992) High signal regions in normal white matter shown by heavily T2-weighted CSF nulled IR sequences. J Comput Assist Tomogr 16:506–513PubMedCrossRef
2.
Zurück zum Zitat Barkhof F, Scheltens P (2002) Imaging of white matter lesions. Cerebrovasc Dis 13(Suppl 2):21–30CrossRefPubMed Barkhof F, Scheltens P (2002) Imaging of white matter lesions. Cerebrovasc Dis 13(Suppl 2):21–30CrossRefPubMed
3.
Zurück zum Zitat Moraal B, Roosendaal SD, Pouwels PJ, Vrenken H, van Schijndel RA, Meier DS, Guttmann CR, Geurts JJ, Barkhof F (2008) Multi-contrast, isotropic, single-slab 3D MR imaging in multiple sclerosis. Eur Radiol 18:2311–2320CrossRefPubMed Moraal B, Roosendaal SD, Pouwels PJ, Vrenken H, van Schijndel RA, Meier DS, Guttmann CR, Geurts JJ, Barkhof F (2008) Multi-contrast, isotropic, single-slab 3D MR imaging in multiple sclerosis. Eur Radiol 18:2311–2320CrossRefPubMed
4.
Zurück zum Zitat Chagla GH, Busse RF, Sydnor R, Rowley HA, Turski PA (2008) Three-dimensional fluid attenuated inversion recovery imaging with isotropic resolution and nonselective adiabatic inversion provides improved three-dimensional visualization and cerebrospinal fluid suppression compared to two-dimensional flair at 3 tesla. Invest Radiol 43:547–551CrossRefPubMed Chagla GH, Busse RF, Sydnor R, Rowley HA, Turski PA (2008) Three-dimensional fluid attenuated inversion recovery imaging with isotropic resolution and nonselective adiabatic inversion provides improved three-dimensional visualization and cerebrospinal fluid suppression compared to two-dimensional flair at 3 tesla. Invest Radiol 43:547–551CrossRefPubMed
5.
Zurück zum Zitat Tkac I, Gruetter R (2005) Methodology of H-1 NMR spectroscopy of the human brain at very high magnetic fields. Appl Magn Reson 29:139–157CrossRefPubMed Tkac I, Gruetter R (2005) Methodology of H-1 NMR spectroscopy of the human brain at very high magnetic fields. Appl Magn Reson 29:139–157CrossRefPubMed
6.
Zurück zum Zitat Duyn JH, van GP, Li TQ, de Zwart JA, Koretsky AP, Fukunaga M (2007) High-field MRI of brain cortical substructure based on signal phase. Proc Natl Acad Sci U S A 104:11796–11801CrossRefPubMed Duyn JH, van GP, Li TQ, de Zwart JA, Koretsky AP, Fukunaga M (2007) High-field MRI of brain cortical substructure based on signal phase. Proc Natl Acad Sci U S A 104:11796–11801CrossRefPubMed
7.
Zurück zum Zitat Cho ZH, Kang CK, Han JY, Kim SH, Kim KN, Hong SM, Park CW, Kim YB (2008) Observation of the Lenticulostriate Arteries in the Human Brain In Vivo Using 7.0T MR Angiography. Stroke. doi:10.1161/STROKEAHA.107.508002 Cho ZH, Kang CK, Han JY, Kim SH, Kim KN, Hong SM, Park CW, Kim YB (2008) Observation of the Lenticulostriate Arteries in the Human Brain In Vivo Using 7.0T MR Angiography. Stroke. doi:10.​1161/​STROKEAHA.​107.​508002
8.
Zurück zum Zitat Hendrikse J, Zwanenburg JJ, Visser F, Takahara T, Luijten P (2008) Noninvasive depiction of the lenticulostriate arteries with time-of-flight MR angiography at 7.0 T. Cerebrovasc Dis 26:624–629CrossRefPubMed Hendrikse J, Zwanenburg JJ, Visser F, Takahara T, Luijten P (2008) Noninvasive depiction of the lenticulostriate arteries with time-of-flight MR angiography at 7.0 T. Cerebrovasc Dis 26:624–629CrossRefPubMed
9.
Zurück zum Zitat Gati JS, Menon RS, Ugurbil K, Rutt BK (1997) Experimental determination of the BOLD field strength dependence in vessels and tissue. Magn Reson Med 38:296–302CrossRefPubMed Gati JS, Menon RS, Ugurbil K, Rutt BK (1997) Experimental determination of the BOLD field strength dependence in vessels and tissue. Magn Reson Med 38:296–302CrossRefPubMed
10.
Zurück zum Zitat Yacoub E, Shmuel A, Pfeuffer J, van de Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X (2001) Imaging brain function in humans at 7 Tesla. Magn Reson Med 45:588–594CrossRefPubMed Yacoub E, Shmuel A, Pfeuffer J, van de Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X (2001) Imaging brain function in humans at 7 Tesla. Magn Reson Med 45:588–594CrossRefPubMed
11.
Zurück zum Zitat Tallantyre EC, Brookes MJ, Dixon JE, Morgan PS, Evangelou N, Morris PG (2008) Demonstrating the perivascular distribution of MS lesions in vivo with 7-Tesla MRI. Neurology 70:2076–2078CrossRefPubMed Tallantyre EC, Brookes MJ, Dixon JE, Morgan PS, Evangelou N, Morris PG (2008) Demonstrating the perivascular distribution of MS lesions in vivo with 7-Tesla MRI. Neurology 70:2076–2078CrossRefPubMed
12.
Zurück zum Zitat Garwood M, Delabarre L (2001) The return of the frequency sweep: designing adiabatic pulses for contemporary NMR. J Magn Reson 153:155–177CrossRefPubMed Garwood M, Delabarre L (2001) The return of the frequency sweep: designing adiabatic pulses for contemporary NMR. J Magn Reson 153:155–177CrossRefPubMed
13.
Zurück zum Zitat Rooney WD, Johnson G, Li X, Cohen ER, Kim SG, Ugurbil K, Springer CS Jr (2007) Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn Reson Med 57:308–318CrossRefPubMed Rooney WD, Johnson G, Li X, Cohen ER, Kim SG, Ugurbil K, Springer CS Jr (2007) Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn Reson Med 57:308–318CrossRefPubMed
14.
Zurück zum Zitat Phal PM, Usmanov A, Nesbit GM, Anderson JC, Spencer D, Wang P, Helwig JA, Roberts C, Hamilton BE (2008) Qualitative comparison of 3-T and 1.5-T MRI in the evaluation of epilepsy. AJR Am J Roentgenol 191:890–895CrossRefPubMed Phal PM, Usmanov A, Nesbit GM, Anderson JC, Spencer D, Wang P, Helwig JA, Roberts C, Hamilton BE (2008) Qualitative comparison of 3-T and 1.5-T MRI in the evaluation of epilepsy. AJR Am J Roentgenol 191:890–895CrossRefPubMed
15.
Zurück zum Zitat Runge VM, Case RS, Sonnier HL (2006) Advances in clinical 3-tesla neuroimaging. Invest Radiol 41:63–67CrossRefPubMed Runge VM, Case RS, Sonnier HL (2006) Advances in clinical 3-tesla neuroimaging. Invest Radiol 41:63–67CrossRefPubMed
16.
Zurück zum Zitat Schwindt W, Kugel H, Bachmann R, Kloska S, Allkemper T, Maintz D, Pfleiderer B, Tombach B, Heindel W (2003) Magnetic resonance imaging protocols for examination of the neurocranium at 3 T. Eur Radiol 13:2170–2179CrossRefPubMed Schwindt W, Kugel H, Bachmann R, Kloska S, Allkemper T, Maintz D, Pfleiderer B, Tombach B, Heindel W (2003) Magnetic resonance imaging protocols for examination of the neurocranium at 3 T. Eur Radiol 13:2170–2179CrossRefPubMed
17.
Zurück zum Zitat Bachmann R, Reilmann R, Schwindt W, Kugel H, Heindel W, Kramer S (2006) FLAIR imaging for multiple sclerosis: a comparative MR study at 1.5 and 3.0 Tesla. Eur Radiol 16:915–921CrossRefPubMed Bachmann R, Reilmann R, Schwindt W, Kugel H, Heindel W, Kramer S (2006) FLAIR imaging for multiple sclerosis: a comparative MR study at 1.5 and 3.0 Tesla. Eur Radiol 16:915–921CrossRefPubMed
18.
Zurück zum Zitat Haacke EM, Xu Y, Cheng YC, Reichenbach JR (2004) Susceptibility weighted imaging (SWI). Magn Reson Med 52:612–618CrossRefPubMed Haacke EM, Xu Y, Cheng YC, Reichenbach JR (2004) Susceptibility weighted imaging (SWI). Magn Reson Med 52:612–618CrossRefPubMed
19.
Zurück zum Zitat Wiesinger F, van de Moortele PF, Adriany G, De Zanche N, Ugurbil K, Pruessmann KP (2006) Potential and feasibility of parallel MRI at high field. NMR Biomed 19:368–378CrossRefPubMed Wiesinger F, van de Moortele PF, Adriany G, De Zanche N, Ugurbil K, Pruessmann KP (2006) Potential and feasibility of parallel MRI at high field. NMR Biomed 19:368–378CrossRefPubMed
Metadaten
Titel
Fluid attenuated inversion recovery (FLAIR) MRI at 7.0 Tesla: comparison with 1.5 and 3.0 Tesla
verfasst von
Jaco J. M. Zwanenburg
Jeroen Hendrikse
Fredy Visser
Taro Takahara
Peter R. Luijten
Publikationsdatum
01.04.2010
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 4/2010
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-009-1620-2

Weitere Artikel der Ausgabe 4/2010

European Radiology 4/2010 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.