Frequency domain analysis of cerebral near infrared spectroscopy signals during application of an impedance threshold device in spontaneously ventilating volunteers

verfasst von: Douglas A. Colquhoun, Kimberly Naden, Robert H. Thiele

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 4/2016

Einloggen, um Zugang zu erhalten

Abstract

Currently available near infrared spectroscopy (NIRS) devices are unable to discriminate between arterial and venous blood, a potential source of artifact. The purpose of this study was to test the hypothesis that oscillations in NIR signals at the respiratory and cardiac frequency could be attributed to venous and arterial blood, respectively, and thereby isolated. After written informed consent was obtained, a two-wavelength NIRS device was placed over the left frontal cortex in 20 volunteers. After 5 min of unimpeded spontaneous ventilation, an impedance threshold device (ITD, average resistance—7 cm H2O) was applied and an additional two minutes of data recorded. Tissue saturation (StO2) calculated at the ventilatory and cardiac frequencies was compared to non-pulsatile StO2, before and after application of the ITD using spectral peak and power algorithms. The ITD increased non-pulsatile cerebral saturation by 3.6 %. The ITD had no discernable effect on pulsatile estimates of StO2 at either the ventilatory or cardiac frequencies. StO2 estimated at the NIRS spectral peak from 0.75 to 1.75 Hz was 24 % higher than non-pulsatile StO2 (p = 0.0013). There were no other significant differences between pulsatile and non-pulsatile algorithms in the estimation of StO2. In 64 % of cases, both the low (ventilator) and high (cardiac) frequency estimates of StO2 were either both larger or both smaller than non-pulsatile StO2, suggesting that they were interrogating the same vascular bed. Frequency domain analysis cannot reliably separate NIRS waveforms into arterial and venous components.
Literatur
1.
Zurück zum Zitat Ferrari M, Quaresima V. Near infrared brain and muscle oximetry: from the discovery to current applications. J Near Infrared Spectrosc. 2012;20:1–14.CrossRef Ferrari M, Quaresima V. Near infrared brain and muscle oximetry: from the discovery to current applications. J Near Infrared Spectrosc. 2012;20:1–14.CrossRef
2.
Zurück zum Zitat Ghosh A, Elwell C, Smith M. Review article: cerebral near-infrared spectroscopy in adults: a work in progress. Anesth Analg. 2012;115(6):1373–83.CrossRefPubMed Ghosh A, Elwell C, Smith M. Review article: cerebral near-infrared spectroscopy in adults: a work in progress. Anesth Analg. 2012;115(6):1373–83.CrossRefPubMed
3.
Zurück zum Zitat Benni PB, Chen B, Dykes FD, et al. Validation of the CAS neonatal NIRS system by monitoring vv-ECMO patients: preliminary results. Adv Exp Med Biol. 2005;566:195–201.CrossRefPubMed Benni PB, Chen B, Dykes FD, et al. Validation of the CAS neonatal NIRS system by monitoring vv-ECMO patients: preliminary results. Adv Exp Med Biol. 2005;566:195–201.CrossRefPubMed
4.
Zurück zum Zitat Ito H, Ibaraki M, Kanno I, Fukuda H, Miura S. Changes in the arterial fraction of human cerebral blood volume during hypercapnia and hypocapnia measured by positron emission tomography. J Cereb Blood Flow Metab. 2005;25(7):852–7.CrossRefPubMed Ito H, Ibaraki M, Kanno I, Fukuda H, Miura S. Changes in the arterial fraction of human cerebral blood volume during hypercapnia and hypocapnia measured by positron emission tomography. J Cereb Blood Flow Metab. 2005;25(7):852–7.CrossRefPubMed
5.
Zurück zum Zitat Rivers E, Nguyen B, Havstad S, et al. Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med. 2001;345(19):1368–77.CrossRefPubMed Rivers E, Nguyen B, Havstad S, et al. Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med. 2001;345(19):1368–77.CrossRefPubMed
6.
7.
8.
Zurück zum Zitat Crabtree V, Echiadis A, Smith P, et al. Prospective venox feasibility study. Conf Proc IEEE Eng Med Biol Soc. 2005;2:1968–71.PubMed Crabtree V, Echiadis A, Smith P, et al. Prospective venox feasibility study. Conf Proc IEEE Eng Med Biol Soc. 2005;2:1968–71.PubMed
9.
Zurück zum Zitat Echiadis AS, Crabtree VP, Bence J, et al. Non-invasive measurement of peripheral venous oxygen saturation using a new venous oximetry method: evaluation during bypass in heart surgery. Physiol Meas. 2007;28(8):897–911.CrossRefPubMed Echiadis AS, Crabtree VP, Bence J, et al. Non-invasive measurement of peripheral venous oxygen saturation using a new venous oximetry method: evaluation during bypass in heart surgery. Physiol Meas. 2007;28(8):897–911.CrossRefPubMed
10.
Zurück zum Zitat Walton ZD, Kyriacou PA, Silverman DG, Shelley KH. Measuring venous oxygenation using the photoplethysmograph waveform. J Clin Monit Comput. 2010;24(4):295–303.CrossRefPubMed Walton ZD, Kyriacou PA, Silverman DG, Shelley KH. Measuring venous oxygenation using the photoplethysmograph waveform. J Clin Monit Comput. 2010;24(4):295–303.CrossRefPubMed
11.
Zurück zum Zitat Thiele RH, Tucker-Schwartz JM, Lu Y, Gillies GT, Durieux ME. Technical communication: transcutaneous regional venous oximetry: a feasibility study. Anesth Analg. 2011;112(6):1353–7.CrossRefPubMed Thiele RH, Tucker-Schwartz JM, Lu Y, Gillies GT, Durieux ME. Technical communication: transcutaneous regional venous oximetry: a feasibility study. Anesth Analg. 2011;112(6):1353–7.CrossRefPubMed
12.
Zurück zum Zitat Franceschini MA, Boas DA, Zourabian A, et al. Near-infrared spiroximetry: noninvasive measurements of venous saturation in piglets and human subjects. J Appl Physiol. 2002;92(1):372–84.PubMedPubMedCentral Franceschini MA, Boas DA, Zourabian A, et al. Near-infrared spiroximetry: noninvasive measurements of venous saturation in piglets and human subjects. J Appl Physiol. 2002;92(1):372–84.PubMedPubMedCentral
13.
Zurück zum Zitat Alian AA, Shelley KH. Respiratory physiology and the impact of different modes of ventilation on the photoplethysmographic waveform. Sensors (Basel). 2012;12(2):2236–54.CrossRef Alian AA, Shelley KH. Respiratory physiology and the impact of different modes of ventilation on the photoplethysmographic waveform. Sensors (Basel). 2012;12(2):2236–54.CrossRef
14.
Zurück zum Zitat Huffmyer JL, Groves DS, Scalzo DC, et al. The effect of the intrathoracic pressure regulator on hemodynamics and cardiac output. Shock. 2011;35(2):114–6.CrossRefPubMed Huffmyer JL, Groves DS, Scalzo DC, et al. The effect of the intrathoracic pressure regulator on hemodynamics and cardiac output. Shock. 2011;35(2):114–6.CrossRefPubMed
15.
Zurück zum Zitat Thiele RH, Colquhoun DA, Tucker-Schwartz JM, Gillies GT, Durieux ME. Radial-femoral concordance in time and frequency domain-based estimates of systemic arterial respiratory variation. J Clin Monit Comput. 2012;26(5):393–400.CrossRefPubMed Thiele RH, Colquhoun DA, Tucker-Schwartz JM, Gillies GT, Durieux ME. Radial-femoral concordance in time and frequency domain-based estimates of systemic arterial respiratory variation. J Clin Monit Comput. 2012;26(5):393–400.CrossRefPubMed
16.
Zurück zum Zitat Matcher J, Kirkpatrick P, Nahid K, Cope M, Delpy DT. Absolute quantification methods in tissue near infrared spectroscopy. Proc SPIE. 1995;1995(486):486–95.CrossRef Matcher J, Kirkpatrick P, Nahid K, Cope M, Delpy DT. Absolute quantification methods in tissue near infrared spectroscopy. Proc SPIE. 1995;1995(486):486–95.CrossRef
17.
Zurück zum Zitat Delpy DT, Cope M. Quantification in tissue near-infrared spectroscopy. Philos Trans R Soc B: Biol Sci. 1997;352:649–59.CrossRef Delpy DT, Cope M. Quantification in tissue near-infrared spectroscopy. Philos Trans R Soc B: Biol Sci. 1997;352:649–59.CrossRef
18.
Zurück zum Zitat Suzuki S, Takasaki S, Ozaki T, Kobayashi Y. A tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc SPIE. 1999;1999(3597):582–92.CrossRef Suzuki S, Takasaki S, Ozaki T, Kobayashi Y. A tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc SPIE. 1999;1999(3597):582–92.CrossRef
19.
Zurück zum Zitat Wray S, Cope M, Delpy DT, Wyatt JS, Reynolds EO. Characterization of the near infrared absorption spectra of cytochrome aa3 and haemoglobin for the non-invasive monitoring of cerebral oxygenation. Biochim Biophys Acta. 1988;933(1):184–92.CrossRefPubMed Wray S, Cope M, Delpy DT, Wyatt JS, Reynolds EO. Characterization of the near infrared absorption spectra of cytochrome aa3 and haemoglobin for the non-invasive monitoring of cerebral oxygenation. Biochim Biophys Acta. 1988;933(1):184–92.CrossRefPubMed
20.
Zurück zum Zitat Ryan KL, Cooke WH, Rickards CA, Lurie KG, Convertino VA. Breathing through an inspiratory threshold device improves stroke volume during central hypovolemia in humans. J Appl Physiol. 2008;104(5):1402–9.CrossRefPubMed Ryan KL, Cooke WH, Rickards CA, Lurie KG, Convertino VA. Breathing through an inspiratory threshold device improves stroke volume during central hypovolemia in humans. J Appl Physiol. 2008;104(5):1402–9.CrossRefPubMed
21.
Zurück zum Zitat Marino BS, Yannopoulos D, Sigurdsson G, et al. Spontaneous breathing through an inspiratory impedance threshold device augments cardiac index and stroke volume index in a pediatric porcine model of hemorrhagic hypovolemia. Crit Care Med. 2004;32(9 Suppl):S398–405.CrossRefPubMed Marino BS, Yannopoulos D, Sigurdsson G, et al. Spontaneous breathing through an inspiratory impedance threshold device augments cardiac index and stroke volume index in a pediatric porcine model of hemorrhagic hypovolemia. Crit Care Med. 2004;32(9 Suppl):S398–405.CrossRefPubMed
22.
Zurück zum Zitat Convertino VA, Ratliff DA, Ryan KL, et al. Hemodynamics associated with breathing through an inspiratory impedance threshold device in human volunteers. Crit Care Med. 2004;32(9 Suppl):S381–6.CrossRefPubMed Convertino VA, Ratliff DA, Ryan KL, et al. Hemodynamics associated with breathing through an inspiratory impedance threshold device in human volunteers. Crit Care Med. 2004;32(9 Suppl):S381–6.CrossRefPubMed
23.
Zurück zum Zitat Yannopoulos D, McKnite SH, Metzger A, Lurie KG. Intrathoracic pressure regulation for intracranial pressure management in normovolemic and hypovolemic pigs. Crit Care Med. 2006;34(12 Suppl):S495–500.CrossRefPubMed Yannopoulos D, McKnite SH, Metzger A, Lurie KG. Intrathoracic pressure regulation for intracranial pressure management in normovolemic and hypovolemic pigs. Crit Care Med. 2006;34(12 Suppl):S495–500.CrossRefPubMed
24.
Zurück zum Zitat Rickards CA, Ryan KL, Cooke WH, Lurie KG, Convertino VA. Inspiratory resistance delays the reporting of symptoms with central hypovolemia: association with cerebral blood flow. Am J Physiol Regul Integr Comp Physiol. 2007;293(1):R243–50.CrossRefPubMed Rickards CA, Ryan KL, Cooke WH, Lurie KG, Convertino VA. Inspiratory resistance delays the reporting of symptoms with central hypovolemia: association with cerebral blood flow. Am J Physiol Regul Integr Comp Physiol. 2007;293(1):R243–50.CrossRefPubMed
25.
Zurück zum Zitat Colquhoun DA, Tucker-Schwartz JM, Durieux ME, Thiele RH. Non-invasive estimation of jugular venous oxygen saturation: a comparison between near infrared spectroscopy and transcutaneous venous oximetry. J Clin Monit Comput. 2012;26(2):91–8.CrossRefPubMed Colquhoun DA, Tucker-Schwartz JM, Durieux ME, Thiele RH. Non-invasive estimation of jugular venous oxygen saturation: a comparison between near infrared spectroscopy and transcutaneous venous oximetry. J Clin Monit Comput. 2012;26(2):91–8.CrossRefPubMed
26.
Zurück zum Zitat Lynch JM, Buckley EM, Schwab PJ, et al. Noninvasive optical quantification of cerebral venous oxygen saturation in humans. Acad Radiol. 2014;21(2):162–7.CrossRefPubMedPubMedCentral Lynch JM, Buckley EM, Schwab PJ, et al. Noninvasive optical quantification of cerebral venous oxygen saturation in humans. Acad Radiol. 2014;21(2):162–7.CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat O’Rourke MF, Taylor MG. Vascular impedance of the femoral bed. Circ Res. 1966;18:126–39.CrossRef O’Rourke MF, Taylor MG. Vascular impedance of the femoral bed. Circ Res. 1966;18:126–39.CrossRef
28.
Zurück zum Zitat Ma HY, Xu Q, Ballesteros JR, Ntziachristos V, Zhang Q, Chance B. Quantitative study of hypoxia stress in piglet brain by IQ phase modulation oximetry. Proc SPIE. 1999;3597:642–9.CrossRef Ma HY, Xu Q, Ballesteros JR, Ntziachristos V, Zhang Q, Chance B. Quantitative study of hypoxia stress in piglet brain by IQ phase modulation oximetry. Proc SPIE. 1999;3597:642–9.CrossRef
29.
Zurück zum Zitat Hueber DM, Franceschini MA, Ma HY, et al. Non-invasive and quantitative near-infrared haemoglobin spectrometry in the piglet brain during hypoxic stress, using a frequency-domain multidistance instrument. Phys Med Biol. 2001;46(1):41–62.CrossRefPubMed Hueber DM, Franceschini MA, Ma HY, et al. Non-invasive and quantitative near-infrared haemoglobin spectrometry in the piglet brain during hypoxic stress, using a frequency-domain multidistance instrument. Phys Med Biol. 2001;46(1):41–62.CrossRefPubMed
30.
Zurück zum Zitat Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37(9):2642–7.CrossRefPubMed Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37(9):2642–7.CrossRefPubMed
31.
Zurück zum Zitat Cannesson M, Le Manach Y, Hofer CK, et al. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a “gray zone” approach. Anesthesiology. 2011;115(2):231–41.CrossRefPubMed Cannesson M, Le Manach Y, Hofer CK, et al. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a “gray zone” approach. Anesthesiology. 2011;115(2):231–41.CrossRefPubMed
32.
Zurück zum Zitat Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112(6):1392–402.CrossRefPubMed Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112(6):1392–402.CrossRefPubMed
33.
Zurück zum Zitat Wardhan R, Shelley K. Peripheral venous pressure waveform. Curr Opin Anaesthesiol. 2009;22(6):814–21.CrossRefPubMed Wardhan R, Shelley K. Peripheral venous pressure waveform. Curr Opin Anaesthesiol. 2009;22(6):814–21.CrossRefPubMed
34.
Zurück zum Zitat Cannesson M, Awad AA, Shelley K. Oscillations in the plethysmographic waveform amplitude: phenomenon hides behind artifacts. Anesthesiology. 2009;111(1):206–7 (author reply 207–208).CrossRefPubMed Cannesson M, Awad AA, Shelley K. Oscillations in the plethysmographic waveform amplitude: phenomenon hides behind artifacts. Anesthesiology. 2009;111(1):206–7 (author reply 207–208).CrossRefPubMed
35.
Zurück zum Zitat Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology. 2012;116(4):834–40.CrossRefPubMed Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology. 2012;116(4):834–40.CrossRefPubMed
36.
Zurück zum Zitat Broch O, Bein B, Gruenewald M, et al. Accuracy of the pleth variability index to predict fluid responsiveness depends on the perfusion index. Acta Anaesthesiol Scand. 2011;55(6):686–93.CrossRefPubMed Broch O, Bein B, Gruenewald M, et al. Accuracy of the pleth variability index to predict fluid responsiveness depends on the perfusion index. Acta Anaesthesiol Scand. 2011;55(6):686–93.CrossRefPubMed
37.
Zurück zum Zitat Johnston WE, DeWitt DS, Vinten-Johansen J, Stump DA, Prough DS. Phenylephrine does not reduce cerebral perfusion during canine cardiopulmonary bypass. Anesth Analg. 1994;79(1):14–8.CrossRefPubMed Johnston WE, DeWitt DS, Vinten-Johansen J, Stump DA, Prough DS. Phenylephrine does not reduce cerebral perfusion during canine cardiopulmonary bypass. Anesth Analg. 1994;79(1):14–8.CrossRefPubMed
38.
Zurück zum Zitat Matcher SJ, Elwell CE, Cooper CE, Cope M, Delpy DT. Performance comparison of several published tissue near-infrared spectroscopy algorithms. Anal Biochem. 1995;227(1):54–68.CrossRefPubMed Matcher SJ, Elwell CE, Cooper CE, Cope M, Delpy DT. Performance comparison of several published tissue near-infrared spectroscopy algorithms. Anal Biochem. 1995;227(1):54–68.CrossRefPubMed
Metadaten
Titel
Frequency domain analysis of cerebral near infrared spectroscopy signals during application of an impedance threshold device in spontaneously ventilating volunteers
verfasst von
Douglas A. Colquhoun
Kimberly Naden
Robert H. Thiele
Publikationsdatum
27.06.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 4/2016
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9729-0

Weitere Artikel der Ausgabe 4/2016

Journal of Clinical Monitoring and Computing 4/2016 Zur Ausgabe

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Beutel versus Maschine: Beste Beatmungstechnik bei Herzstillstand gesucht

02.05.2024 Kardiopulmonale Reanimation Nachrichten

Stehen die Chancen auf eine Rückkehr der Spontanzirkulation nach Herz-Kreislauf-Stillstand bei manueller oder maschineller Beatmung besser? Und unterscheidet sich das neurologische Outcome nach der Reanimation? Das belgische Herzstillstand-Register liefert die Daten für einen direkten Vergleich zwischen Beutel und Beatmungsgerät.

Tipps für den Umgang mit Behandlungsfehlern

01.05.2024 DGIM 2024 Kongressbericht

Es ist nur eine Frage der Zeit, bis es zu einem Zwischenfall kommt und ein Behandlungsfehler passiert. Doch wenn Ärztinnen und Ärzte gut vorbereitet sind, schaffen es alle Beteiligten den Umständen entsprechend gut durch diese Krise. 

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.