Image quality and diagnostic performance evaluation in transcatheter aortic valve implantation candidates with atrial fibrillation using a whole-heart coverage CT scanner

verfasst von: Yu Zhang, Zhenlin Li, Yongchun You, Liqing Peng, Jianying Li, Tao Shuai

Erschienen in: European Radiology | Ausgabe 2/2022

Einloggen, um Zugang zu erhalten

Abstract

Objective

To evaluate the image quality and diagnostic performance for obstructive coronary artery disease of transcatheter aortic valve implantation (TAVI) patients with atrial fibrillation (AF) during TAVI planning CT using a whole-heart coverage CT scanner.

Methods

Eighty-eight consecutive TAVI candidates with AF (50 men, 74 ± 6 years) who underwent both TAVI planning CT and invasive coronary catheter angiography (ICA) were retrospectively analyzed. With ICA results as the reference standard, the accuracy of TAVI planning CT for lesion detection on a per-vessel and per-patient level was calculated. Meanwhile, image quality, contrast volume, and effective dose (ED) were evaluated. A 5-point visual scale (1–5) was used to assess the subjective image quality. The CT value and signal-to-noise ratio were measured for the left main coronary artery (LM), left anterior descending (LAD), left circumflex (LCX), and right coronary arteries (RCA).

Results

The ED for CCTA was 3.25 ± 1.39 mSv and contrast volume was 58.14 ± 12.34 mL. A total of 1371 (1371/1408 = 97.4%) segments with diameter > 1.5 mm were analyzed. For the subjective evaluation, the mean score was 3.99 ± 0.96 for overall image quality. The mean CT values in LM, RCA, LCX, and LAD were all above 400 HU. For the detection of > 50% stenosis, TAVI planning CT provided on the per-vessel and per-patient basis 97.06% and 100% in sensitivity, 96.23% and 89.06% in specificity, 99.7% and 100% in negative predictive value, and 73.3% and 77.4% in positive predictive value, respectively.

Conclusion

TAVI planning CT with whole-heart coverage demonstrates good CCTA image quality and a high sensitivity and NPV in excluding obstructive CAD in TAVI candidates with AF.

Key Points

Transcatheter aortic valve implantation planning (TAVI) CT with whole-heart coverage enables good image quality of CCTA in TAVI candidates with atrial fibrillation.
Obstructive coronary artery disease may be excluded with high accuracy in transcatheter aortic valve implantation (TAVI) candidates with atrial fibrillation with the usage of whole-heart coverage TAVI planning CT.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Baumgartner H, Falk V, Bax JJ et al (2017) 2017 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J 38:2739–2791CrossRef Baumgartner H, Falk V, Bax JJ et al (2017) 2017 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J 38:2739–2791CrossRef
2.
Zurück zum Zitat Opolski MP, Staruch AD, Jakubczyk M et al (2016) CT Angiography for the detection of coronary artery stenoses in patients referred for cardiac valve surgery: systematic review and meta-analysis. JACC Cardiovasc Imaging 9:1059–1070CrossRef Opolski MP, Staruch AD, Jakubczyk M et al (2016) CT Angiography for the detection of coronary artery stenoses in patients referred for cardiac valve surgery: systematic review and meta-analysis. JACC Cardiovasc Imaging 9:1059–1070CrossRef
3.
Zurück zum Zitat Blanke P, Weir-McCall JR, Achenbach S et al (2019) Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI) / transcatheter aortic valve replacement (TAVR): an expert consensus document of the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr 13:1–20CrossRef Blanke P, Weir-McCall JR, Achenbach S et al (2019) Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI) / transcatheter aortic valve replacement (TAVR): an expert consensus document of the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr 13:1–20CrossRef
4.
Zurück zum Zitat van den Boogert TPW, Vendrik J, Claessen B et al (2018) CTCA for detection of significant coronary artery disease in routine TAVI work-up: a systematic review and meta-analysis. Neth Hear J 26:591–599CrossRef van den Boogert TPW, Vendrik J, Claessen B et al (2018) CTCA for detection of significant coronary artery disease in routine TAVI work-up: a systematic review and meta-analysis. Neth Hear J 26:591–599CrossRef
5.
Zurück zum Zitat Hamdan A, Wellnhofer E, Konen E et al (2015) Coronary CT angiography for the detection of coronary artery stenosis in patients referred for transcatheter aortic valve replacement. J Cardiovasc Comput Tomogr 9:31–41CrossRef Hamdan A, Wellnhofer E, Konen E et al (2015) Coronary CT angiography for the detection of coronary artery stenosis in patients referred for transcatheter aortic valve replacement. J Cardiovasc Comput Tomogr 9:31–41CrossRef
6.
Zurück zum Zitat Andreini D, Pontone G, Mushtaq S et al (2014) Diagnostic accuracy of multidetector computed tomography coronary angiography in 325 consecutive patients referred for transcatheter aortic valve replacement. Am Heart J 168:332–339CrossRef Andreini D, Pontone G, Mushtaq S et al (2014) Diagnostic accuracy of multidetector computed tomography coronary angiography in 325 consecutive patients referred for transcatheter aortic valve replacement. Am Heart J 168:332–339CrossRef
7.
Zurück zum Zitat Pontone G, Andreini D, Bartorelli AL et al (2011) Feasibility and accuracy of a comprehensive multidetector computed tomography acquisition for patients referred for balloon-expandable transcatheter aortic valve implantation. Am Heart J 161:1106–1113CrossRef Pontone G, Andreini D, Bartorelli AL et al (2011) Feasibility and accuracy of a comprehensive multidetector computed tomography acquisition for patients referred for balloon-expandable transcatheter aortic valve implantation. Am Heart J 161:1106–1113CrossRef
8.
Zurück zum Zitat Matsumoto S, Yamada Y, Hashimoto M et al (2017) CT imaging before transcatheter aortic valve implantation (TAVI) using variable helical pitch scanning and its diagnostic performance for coronary artery disease. Eur Radiol 27:1963–1970CrossRef Matsumoto S, Yamada Y, Hashimoto M et al (2017) CT imaging before transcatheter aortic valve implantation (TAVI) using variable helical pitch scanning and its diagnostic performance for coronary artery disease. Eur Radiol 27:1963–1970CrossRef
9.
Zurück zum Zitat Menke J, Kowalski J (2016) Diagnostic accuracy and utility of coronary CT angiography with consideration of unevaluable results: a systematic review and multivariate Bayesian random-effects meta-analysis with intention to diagnose. Eur Radiol 26:451–458CrossRef Menke J, Kowalski J (2016) Diagnostic accuracy and utility of coronary CT angiography with consideration of unevaluable results: a systematic review and multivariate Bayesian random-effects meta-analysis with intention to diagnose. Eur Radiol 26:451–458CrossRef
10.
Zurück zum Zitat Vorre MM, Abdulla J (2013) Diagnostic accuracy and radiation dose of CT coronary angiography in atrial fibrillation: systematic review and meta-analysis. Radiology 267:376–386CrossRef Vorre MM, Abdulla J (2013) Diagnostic accuracy and radiation dose of CT coronary angiography in atrial fibrillation: systematic review and meta-analysis. Radiology 267:376–386CrossRef
11.
Zurück zum Zitat Xu L, Yang L, Fan Z, Yu W, Lv B, Zhang Z (2011) Diagnostic performance of 320-detector CT coronary angiography in patients with atrial fibrillation: preliminary results. Eur Radiol 21:936–943CrossRef Xu L, Yang L, Fan Z, Yu W, Lv B, Zhang Z (2011) Diagnostic performance of 320-detector CT coronary angiography in patients with atrial fibrillation: preliminary results. Eur Radiol 21:936–943CrossRef
12.
Zurück zum Zitat Di Cesare E, Gennarelli A, Di Sibio A et al (2015) Image quality and radiation dose of single heartbeat 640-slice coronary CT angiography: a comparison between patients with chronic atrial fibrillation and subjects in normal sinus rhythm by propensity analysis. Eur J Radiol 84:631–636CrossRef Di Cesare E, Gennarelli A, Di Sibio A et al (2015) Image quality and radiation dose of single heartbeat 640-slice coronary CT angiography: a comparison between patients with chronic atrial fibrillation and subjects in normal sinus rhythm by propensity analysis. Eur J Radiol 84:631–636CrossRef
13.
Zurück zum Zitat Annoni AD, Andreini D, Pontone G et al (2018) CT angiography prior to TAVI procedure using third-generation scanner with wide volume coverage: feasibility, renal safety and diagnostic accuracy for coronary tree. Br J Radiol 91:20180196CrossRef Annoni AD, Andreini D, Pontone G et al (2018) CT angiography prior to TAVI procedure using third-generation scanner with wide volume coverage: feasibility, renal safety and diagnostic accuracy for coronary tree. Br J Radiol 91:20180196CrossRef
14.
Zurück zum Zitat Yamamoto M, Hayashida K, Mouillet G et al (2013) Renal function-based contrast dosing predicts acute kidney injury following transcatheter aortic valve implantation. JACC Cardiovasc Interv 6:479–486CrossRef Yamamoto M, Hayashida K, Mouillet G et al (2013) Renal function-based contrast dosing predicts acute kidney injury following transcatheter aortic valve implantation. JACC Cardiovasc Interv 6:479–486CrossRef
15.
Zurück zum Zitat Tavakol M, Ashraf S, Brener SJ (2012) Risks and complications of coronary angiography: a comprehensive review. Global J Health Sci 4:65–93 Tavakol M, Ashraf S, Brener SJ (2012) Risks and complications of coronary angiography: a comprehensive review. Global J Health Sci 4:65–93
16.
Zurück zum Zitat Blanke P, Weir-McCall JR, Achenbach S et al (2019) Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/transcatheter aortic valve replacement (TAVR): an expert consensus document of the Society of Cardiovascular Computed Tomography. JACC Cardiovasc Imaging 12:1–24CrossRef Blanke P, Weir-McCall JR, Achenbach S et al (2019) Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/transcatheter aortic valve replacement (TAVR): an expert consensus document of the Society of Cardiovascular Computed Tomography. JACC Cardiovasc Imaging 12:1–24CrossRef
17.
Zurück zum Zitat Harris BS, De Cecco CN, Schoepf UJ et al (2015) Dual-source CT imaging to plan transcatheter aortic valve replacement: accuracy for diagnosis of obstructive coronary artery disease. Radiology 275:80–88CrossRef Harris BS, De Cecco CN, Schoepf UJ et al (2015) Dual-source CT imaging to plan transcatheter aortic valve replacement: accuracy for diagnosis of obstructive coronary artery disease. Radiology 275:80–88CrossRef
18.
Zurück zum Zitat Shuai T, Deng L, Pan Y et al (2018) Free-breathing coronary CT angiography using 16-cm wide-detector for challenging patients: comparison with invasive coronary angiography. Clin Radiol 73:986.e981–986.e986CrossRef Shuai T, Deng L, Pan Y et al (2018) Free-breathing coronary CT angiography using 16-cm wide-detector for challenging patients: comparison with invasive coronary angiography. Clin Radiol 73:986.e981–986.e986CrossRef
19.
Zurück zum Zitat Andreini D, Pontone G, Mushtaq S et al (2015) Low-dose CT coronary angiography with a novel IntraCycle motion-correction algorithm in patients with high heart rate or heart rate variability. Eur Heart J Cardiovasc Imaging 16:1093–1100CrossRef Andreini D, Pontone G, Mushtaq S et al (2015) Low-dose CT coronary angiography with a novel IntraCycle motion-correction algorithm in patients with high heart rate or heart rate variability. Eur Heart J Cardiovasc Imaging 16:1093–1100CrossRef
20.
Zurück zum Zitat Halliburton SS, Abbara S, Chen MY et al (2011) SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT. J Cardiovasc Comput Tomogr 5:198–224CrossRef Halliburton SS, Abbara S, Chen MY et al (2011) SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT. J Cardiovasc Comput Tomogr 5:198–224CrossRef
21.
Zurück zum Zitat Fei X, Du X, Yang Q et al (2008) 64-MDCT coronary angiography: phantom study of effects of vascular attenuation on detection of coronary stenosis. AJR Am J Roentgenol 191:43–49CrossRef Fei X, Du X, Yang Q et al (2008) 64-MDCT coronary angiography: phantom study of effects of vascular attenuation on detection of coronary stenosis. AJR Am J Roentgenol 191:43–49CrossRef
22.
Zurück zum Zitat Hausleiter J, Meyer TS, Martuscelli E et al (2012) Image quality and radiation exposure with prospectively ECG-triggered axial scanning for coronary CT angiography: the multicenter, multivendor, randomized PROTECTION-III study. JACC Cardiovasc Imaging 5:484–493CrossRef Hausleiter J, Meyer TS, Martuscelli E et al (2012) Image quality and radiation exposure with prospectively ECG-triggered axial scanning for coronary CT angiography: the multicenter, multivendor, randomized PROTECTION-III study. JACC Cardiovasc Imaging 5:484–493CrossRef
23.
Zurück zum Zitat Andreini D, Pontone G, Mushtaq S et al (2018) Image quality and radiation dose of coronary CT angiography performed with whole-heart coverage CT scanner with intra-cycle motion correction algorithm in patients with atrial fibrillation. Eur Radiol 28:1383–1392CrossRef Andreini D, Pontone G, Mushtaq S et al (2018) Image quality and radiation dose of coronary CT angiography performed with whole-heart coverage CT scanner with intra-cycle motion correction algorithm in patients with atrial fibrillation. Eur Radiol 28:1383–1392CrossRef
24.
Zurück zum Zitat Liang J, Wang H, Xu L et al (2017) Diagnostic performance of 256-row detector coronary CT angiography in patients with high heart rates within a single cardiac cycle: a preliminary study. Clin Radiol 72:694.e697–694.e614CrossRef Liang J, Wang H, Xu L et al (2017) Diagnostic performance of 256-row detector coronary CT angiography in patients with high heart rates within a single cardiac cycle: a preliminary study. Clin Radiol 72:694.e697–694.e614CrossRef
25.
Zurück zum Zitat Liang J, Wang H, Xu L et al (2018) Impact of SSF on diagnostic performance of coronary computed tomography angiography within 1 heart beat in patients with high heart rate using a 256-row detector computed tomography. J Comput Assist Tomogr 42:54–61CrossRef Liang J, Wang H, Xu L et al (2018) Impact of SSF on diagnostic performance of coronary computed tomography angiography within 1 heart beat in patients with high heart rate using a 256-row detector computed tomography. J Comput Assist Tomogr 42:54–61CrossRef
26.
Zurück zum Zitat Rossi A, De Cecco CN, Kennon SRO et al (2017) CT angiography to evaluate coronary artery disease and revascularization requirement before trans-catheter aortic valve replacement. J Cardiovasc Comput Tomogr 11:338–346CrossRef Rossi A, De Cecco CN, Kennon SRO et al (2017) CT angiography to evaluate coronary artery disease and revascularization requirement before trans-catheter aortic valve replacement. J Cardiovasc Comput Tomogr 11:338–346CrossRef
27.
Zurück zum Zitat Pontone G, Andreini D, Quaglia C, Ballerini G, Nobili E, Pepi M (2007) Accuracy of multidetector spiral computed tomography in detecting significant coronary stenosis in patient populations with differing pre-test probabilities of disease. Clin Radiol 62:978–985CrossRef Pontone G, Andreini D, Quaglia C, Ballerini G, Nobili E, Pepi M (2007) Accuracy of multidetector spiral computed tomography in detecting significant coronary stenosis in patient populations with differing pre-test probabilities of disease. Clin Radiol 62:978–985CrossRef
28.
Zurück zum Zitat Andreini D, Pontone G, Mushtaq S et al (2017) Atrial fibrillation: diagnostic accuracy of coronary CT angiography performed with a whole-heart 230-μm spatial resolution CT scanner. Radiology 284:676–684CrossRef Andreini D, Pontone G, Mushtaq S et al (2017) Atrial fibrillation: diagnostic accuracy of coronary CT angiography performed with a whole-heart 230-μm spatial resolution CT scanner. Radiology 284:676–684CrossRef
29.
Zurück zum Zitat Renker M, Nance JW Jr, Schoepf UJ et al (2011) Evaluation of heavily calcified vessels with coronary CT angiography: comparison of iterative and filtered back projection image reconstruction. Radiology 260:390–399CrossRef Renker M, Nance JW Jr, Schoepf UJ et al (2011) Evaluation of heavily calcified vessels with coronary CT angiography: comparison of iterative and filtered back projection image reconstruction. Radiology 260:390–399CrossRef
30.
Zurück zum Zitat Meijs MF, Meijboom WB, Prokop M et al (2009) Is there a role for CT coronary angiography in patients with symptomatic angina? Effect of coronary calcium score on identification of stenosis. Int J Card Imaging 25:847–854CrossRef Meijs MF, Meijboom WB, Prokop M et al (2009) Is there a role for CT coronary angiography in patients with symptomatic angina? Effect of coronary calcium score on identification of stenosis. Int J Card Imaging 25:847–854CrossRef
31.
Zurück zum Zitat Ippolito D, Riva L, Talei Franzesi C et al (2020) Computed tomography angiography combined with knowledge-based iterative algorithm for transcatheter aortic valve implantation planning: image quality and radiation dose exposure with low-kV and low-contrast-medium protocol. J Comput Assist Tomogr 44:13–19CrossRef Ippolito D, Riva L, Talei Franzesi C et al (2020) Computed tomography angiography combined with knowledge-based iterative algorithm for transcatheter aortic valve implantation planning: image quality and radiation dose exposure with low-kV and low-contrast-medium protocol. J Comput Assist Tomogr 44:13–19CrossRef
32.
Zurück zum Zitat Opolski MP, Kim WK, Liebetrau C et al (2015) Diagnostic accuracy of computed tomography angiography for the detection of coronary artery disease in patients referred for transcatheter aortic valve implantation. Clin Res Cardiol 104:471–480CrossRef Opolski MP, Kim WK, Liebetrau C et al (2015) Diagnostic accuracy of computed tomography angiography for the detection of coronary artery disease in patients referred for transcatheter aortic valve implantation. Clin Res Cardiol 104:471–480CrossRef
33.
Zurück zum Zitat Mangold S, De Cecco CN, Schoepf UJ et al (2017) CT angiography for planning transcatheter aortic valve replacement using automated tube voltage selection: image quality and radiation exposure. Eur J Radiol 86:276–283CrossRef Mangold S, De Cecco CN, Schoepf UJ et al (2017) CT angiography for planning transcatheter aortic valve replacement using automated tube voltage selection: image quality and radiation exposure. Eur J Radiol 86:276–283CrossRef
34.
Zurück zum Zitat Mangold S, Wichmann JL, Schoepf UJ et al (2016) Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3(rd) generation dual-source CT. Eur Radiol 26:3608–3616CrossRef Mangold S, Wichmann JL, Schoepf UJ et al (2016) Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3(rd) generation dual-source CT. Eur Radiol 26:3608–3616CrossRef
Metadaten
Titel
Image quality and diagnostic performance evaluation in transcatheter aortic valve implantation candidates with atrial fibrillation using a whole-heart coverage CT scanner
verfasst von
Yu Zhang
Zhenlin Li
Yongchun You
Liqing Peng
Jianying Li
Tao Shuai
Publikationsdatum
02.08.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2022
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-021-08187-z

Weitere Artikel der Ausgabe 2/2022

European Radiology 2/2022 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.