Improvements in the application and reporting of advanced Bland–Altman methods of comparison

verfasst von: Erik Olofsen, Albert Dahan, Gerard Borsboom, Gordon Drummond

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2015

Einloggen, um Zugang zu erhalten

Abstract

Bland and Altman have developed a measure called “limits of agreement” to assess correspondence of two methods of clinical measurement. In many circumstances, comparisons are made using several paired measurements in each individual subject. If such measurements are considered as statistically independent pairs, rather than as sets of measurements from separate individuals, limits of agreement will be too narrow. In addition, the confidence intervals for these limits will also be too narrow. Suitable software to compute valid limits of agreement and their confidence intervals is not readily available. Therefore, we set out to provide a freely available implementation accompanied by a formal description of the more advanced Bland–Altman comparison methods. We validate the implementation using simulated data, and demonstrate the effects caused by failing to take the presence of multiple paired measurements per individual properly into account. We propose a standard format of reporting that would improve analysis and interpretation of comparison studies.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Altman DG, Bland JM. Measurement in medicine: the analysis of method comparison studies. Stat. 1983;32:307–17. Altman DG, Bland JM. Measurement in medicine: the analysis of method comparison studies. Stat. 1983;32:307–17.
2.
Zurück zum Zitat Biancofiore G, Critchley LAH, Lee A, Yang X, Bindi LM, Esposito M, Bisà M, Meacci L, Mozzo R, Filipponi F. Evaluation of a new software version of the Flotrac/Vigileo (version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery. Anesth Analg. 2011;113:515–22.PubMed Biancofiore G, Critchley LAH, Lee A, Yang X, Bindi LM, Esposito M, Bisà M, Meacci L, Mozzo R, Filipponi F. Evaluation of a new software version of the Flotrac/Vigileo (version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery. Anesth Analg. 2011;113:515–22.PubMed
3.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;i:307–10. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;i:307–10.
4.
Zurück zum Zitat Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–60.PubMedCrossRef Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–60.PubMedCrossRef
5.
Zurück zum Zitat Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.PubMedCrossRef Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.PubMedCrossRef
6.
Zurück zum Zitat Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.PubMedCrossRef Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.PubMedCrossRef
7.
Zurück zum Zitat Bland JM, Altman DG. Agreed statistics: measurement method comparison. Anesthesiology. 2012;116:182–5.PubMedCrossRef Bland JM, Altman DG. Agreed statistics: measurement method comparison. Anesthesiology. 2012;116:182–5.PubMedCrossRef
8.
Zurück zum Zitat Burdick RK, Graybill FA. Confidence intervals on linear combinations of variance components in the unbalanced one-way classification. Technometrics. 1984;26:131–6.CrossRef Burdick RK, Graybill FA. Confidence intervals on linear combinations of variance components in the unbalanced one-way classification. Technometrics. 1984;26:131–6.CrossRef
9.
Zurück zum Zitat Columb MO. Clinical measurement and assessing agreement. Curr Anaesth Crit Care. 2008;19:328–9.CrossRef Columb MO. Clinical measurement and assessing agreement. Curr Anaesth Crit Care. 2008;19:328–9.CrossRef
10.
Zurück zum Zitat Donner A, Zou GY. Closed-form confidence intervals for functions of the normal mean and standard deviation. Stat Methods Med Res. 2012;21:347–59.PubMedCrossRef Donner A, Zou GY. Closed-form confidence intervals for functions of the normal mean and standard deviation. Stat Methods Med Res. 2012;21:347–59.PubMedCrossRef
11.
Zurück zum Zitat Efron B, Tibshirani RJ. An introduction to the bootstrap. New York: Chapman and Hall; 1993.CrossRef Efron B, Tibshirani RJ. An introduction to the bootstrap. New York: Chapman and Hall; 1993.CrossRef
12.
Zurück zum Zitat Hamilton C, Lewis S. The importance of using the correct bounds on the Bland–Altman limits of agreement when multiple measurements are recorded per patient. J Clin Monit Comput. 2010;24:173–5.PubMedCrossRef Hamilton C, Lewis S. The importance of using the correct bounds on the Bland–Altman limits of agreement when multiple measurements are recorded per patient. J Clin Monit Comput. 2010;24:173–5.PubMedCrossRef
13.
Zurück zum Zitat Hamilton C, Stamey J. Using Bland–Altman to assess agreement between two medical devices—don’t forget the confidence intervals!. J Clin Monit Comput. 2007;21:331–3.PubMedCrossRef Hamilton C, Stamey J. Using Bland–Altman to assess agreement between two medical devices—don’t forget the confidence intervals!. J Clin Monit Comput. 2007;21:331–3.PubMedCrossRef
14.
Zurück zum Zitat Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J. Comparing methods of clinical measurement: reporting standards for Bland and Altman analysis. Anesth Analg. 2000;90:593–602.PubMedCrossRef Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J. Comparing methods of clinical measurement: reporting standards for Bland and Altman analysis. Anesth Analg. 2000;90:593–602.PubMedCrossRef
16.
Zurück zum Zitat Myles PS, Cui J. Using the Bland–Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11.PubMedCrossRef Myles PS, Cui J. Using the Bland–Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11.PubMedCrossRef
17.
Zurück zum Zitat Oldham PD. A note on the analysis of repeated measurements of the same subjects. J Chronic Dis. 1962;15:969–77.PubMedCrossRef Oldham PD. A note on the analysis of repeated measurements of the same subjects. J Chronic Dis. 1962;15:969–77.PubMedCrossRef
19.
Zurück zum Zitat Sjöstrand F, Rodhe P, Berglund E, Lundström N, Svensen C. The use of a noninvasive hemoglobin monitor for volume kinetic analysis in an emergency room setting. Anesth Analg. 2013;116:337–42.PubMedCrossRef Sjöstrand F, Rodhe P, Berglund E, Lundström N, Svensen C. The use of a noninvasive hemoglobin monitor for volume kinetic analysis in an emergency room setting. Anesth Analg. 2013;116:337–42.PubMedCrossRef
21.
Zurück zum Zitat Thomas JD, Hultquist RA. Interval estimation for the unbalanced case of the one-way random effects model. Ann Stat. 1978;6:582–7.CrossRef Thomas JD, Hultquist RA. Interval estimation for the unbalanced case of the one-way random effects model. Ann Stat. 1978;6:582–7.CrossRef
22.
Zurück zum Zitat Uemura K, Kawada T, Inagaki M, Sugimachi M. A minimally invasive monitoring system of cardiac output using aortic flow velocity and peripheral arterial pressure profile. Anesth Analg. 2013;116:1006–17.PubMedCrossRef Uemura K, Kawada T, Inagaki M, Sugimachi M. A minimally invasive monitoring system of cardiac output using aortic flow velocity and peripheral arterial pressure profile. Anesth Analg. 2013;116:1006–17.PubMedCrossRef
24.
Zurück zum Zitat Zou GY. Confidence interval estimation for the Bland–Altman limits of agreement with multiple observations per individual. Stat Methods Med Res. 2013;22:630–42.PubMedCrossRef Zou GY. Confidence interval estimation for the Bland–Altman limits of agreement with multiple observations per individual. Stat Methods Med Res. 2013;22:630–42.PubMedCrossRef
Metadaten
Titel
Improvements in the application and reporting of advanced Bland–Altman methods of comparison
verfasst von
Erik Olofsen
Albert Dahan
Gerard Borsboom
Gordon Drummond
Publikationsdatum
01.02.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2015
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-014-9577-3

Weitere Artikel der Ausgabe 1/2015

Journal of Clinical Monitoring and Computing 1/2015 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.