01.06.2012

Intraoperative MRI electrical noise and monitor ECG filters affect arrhythmia detection and identification

verfasst von: Melissa Bailey, Gwynne Kirchen, Bridget Bonaventura, Kelly Rosborough, Mahmoud Abdel-Rasoul, Roger Dzwonczyk

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 3/2012

Einloggen, um Zugang zu erhalten

Abstract

Most electrical equipment in the modern operating room (OR) radiates electrical noise (EN) that can interfere with patient monitors. We have described the EN that an intraoperative magnetic resonance imaging (iMRI) system emits and have shown that this high-energy EN diminishes the quality of the ECG waveform during iMRI scans in our neurosurgical OR. We have also shown that the ECG signal filters in our iMRI-compatible patient monitor reduce this interference but, in the process, disturb the true morphology of the displayed waveform. This simulation study evaluates how iMRI-generated EN affects the ability of the anesthetist to detect and identify ECG arrhythmias and whether the patient monitor’s ECG signal filters can improve arrhythmia recognition. Using an ECG simulator, we generated Lead II and V5 ECG signal segments that contained either no arrhythmia or one of four common cardiac arrhythmias. We filtered the ECG segments with four filters available on our iMRI-compatible monitor (Veris MR, MEDRAD Inc., Indianola, PA USA). We then digitized the segments and mixed simulated iMRI EN into the resultant tracings. With institutional approval and written informed consent, board-certified anesthesiologists reviewed the tracings, determined if an arrhythmia was present and identified the arrhythmia. We conducted the study anonymously. We reported the data as percent correct arrhythmia detection and correct arrhythmia identification. Thirty-one anesthesiologists completed the study. Overall, the participants correctly detected 79.5% (95% CI: 77.2, 81.7%) of the arrhythmias and correctly identified 62.5% (95% CI: 59.8, 65.3%) of the arrhythmias, regardless of EN presence. Although the proportions among monitor noise filters studied were not significant, the manufacturer-designated MR5 Veris MR filter optimized arrhythmia detection and arrhythmia identification for our participants, regardless if EN was present in the ECG tracings. In the neurosurgical OR, the anesthetist must be able to effectively monitor a patient in the presence of iMRI-generated EN. Depending on the OR design, the patient may be out of the anesthetist’s direct view during a scan procedure. The anesthetist must rely on monitored physiologic parameters to assess patient status during this time. He/she should be familiar with his/her monitor’s filtering capabilities and routinely adjust the ECG filters to achieve the best compromise between minimized EN effects and maximized displayed ECG signal quality.
Literatur
1.
Zurück zum Zitat Dzwonczyk R, Fujji J, Simonetti L, Nieves-Ramos R, Bergese S. Electrical noise in the intraoperative magnetic resonance imaging setting. Anesth Analg 2009;108:181–6.PubMedCrossRef Dzwonczyk R, Fujji J, Simonetti L, Nieves-Ramos R, Bergese S. Electrical noise in the intraoperative magnetic resonance imaging setting. Anesth Analg 2009;108:181–6.PubMedCrossRef
2.
Zurück zum Zitat Biomedical Signal Processing. Volume II. Compression and automatic recognition, Arnon Cohen. 1986 CRC Press, Coca Raton, Florida. Appendix A, Characteristics of some dynamic biomedical signals, p 124. Biomedical Signal Processing. Volume II. Compression and automatic recognition, Arnon Cohen. 1986 CRC Press, Coca Raton, Florida. Appendix A, Characteristics of some dynamic biomedical signals, p 124.
3.
Zurück zum Zitat American Society of Anesthesiologists, House of Delegates, Standards for Basic Anesthetic Monitoring, 25 October 2005. American Society of Anesthesiologists, House of Delegates, Standards for Basic Anesthetic Monitoring, 25 October 2005.
4.
Zurück zum Zitat Bertrand CA, Steiner NV, Jameson G, Lopez M. Disturbances or cardiac rhythm during anesthesia and surgery. JAMA 1971;216:1615–17. Bertrand CA, Steiner NV, Jameson G, Lopez M. Disturbances or cardiac rhythm during anesthesia and surgery. JAMA 1971;216:1615–17.
5.
Zurück zum Zitat Cohen MM, Duncan PG, Pope WDB, Wolkenstein C. A survey of 112,000 anaesthetics at one teaching hospital (1975–1983). Can J Anaesth 1986;33:22–31. Cohen MM, Duncan PG, Pope WDB, Wolkenstein C. A survey of 112,000 anaesthetics at one teaching hospital (1975–1983). Can J Anaesth 1986;33:22–31.
Metadaten
Titel
Intraoperative MRI electrical noise and monitor ECG filters affect arrhythmia detection and identification
verfasst von
Melissa Bailey
Gwynne Kirchen
Bridget Bonaventura
Kelly Rosborough
Mahmoud Abdel-Rasoul
Roger Dzwonczyk
Publikationsdatum
01.06.2012
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 3/2012
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-012-9346-0

Weitere Artikel der Ausgabe 3/2012

Journal of Clinical Monitoring and Computing 3/2012 Zur Ausgabe

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.