Skip to main content
Erschienen in: Surgical Endoscopy 11/2013

01.11.2013

Blend mode reduces unintended thermal injury by laparoscopic monopolar instruments: a randomized controlled trial

verfasst von: Edward L. Jones, Christina L. Dunn, Nicole T. Townsend, Teresa S. Jones, J. Bruce Dunne, Paul N. Montero, Henry R. Govekar, Greg V. Stiegmann, Thomas N. Robinson

Erschienen in: Surgical Endoscopy | Ausgabe 11/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

The purpose of this study was to compare histologic evidence of thermal injury at the epigastric and umbilical incisions after elective laparoscopic cholecystectomy performed using the monopolar “Bovie” instrument set on the higher voltage coag mode versus the lower voltage blend mode. We hypothesized that the higher voltage coag mode would create more unintended thermal tissue injury at the epigastric trocar’s incision.

Methods

A prospective blinded randomized controlled trial of patients undergoing elective laparoscopic cholecystectomy was performed. Patients were randomized to have their operation performed with the monopolar instrument set at 30 W on either the coag mode or the blend mode. Immediately at the end of the operation, a biopsy sample of skin was obtained from the lower edge of the epigastric incision (through which the monopolar instrument was inserted) and the umbilical incision (through which the camera/telescope was inserted). The outcomes measured were histologic evidence of thermal injury at the epigastric and umbilical incisions (determined by a blinded pathologist).

Results

Forty patients were randomized (20 per group). Baseline demographics in the two groups were similar for age, gender, body mass index, preoperative diagnosis, operative time, and blood loss. Unintentional thermal injury was found at 20 % of epigastric incisions and 35 % of umbilical incisions in the total group. The incidence of thermal injury was higher after operations using the coag mode compared to the blend mode at both the epigastric (35 vs. 5 %; p = 0.044) and umbilical (55 vs. 15 %; p = 0.019) trocar incisions.

Conclusions

Radiofrequency energy from the monopolar Bovie instrument causes unintentional thermal injury to skin adjacent to the epigastric and umbilical trocar incisions. The incidence of thermal injury was reduced by using the lower voltage blend mode compared to the coag mode at both the epigastric and umbilical trocar incisions. Registration number: NCT016648060 (www.​clinicaltrials.​gov).
Literatur
1.
Zurück zum Zitat Hulka JF, Levy BS, Parker WH, Phillips JM (1997) Laparoscopic-assisted vaginal hysterectomy: American association of gynecologic laparoscopists’ 1995 membership survey. J Am Assoc Gynecol Laparosc 4:167–171PubMedCrossRef Hulka JF, Levy BS, Parker WH, Phillips JM (1997) Laparoscopic-assisted vaginal hysterectomy: American association of gynecologic laparoscopists’ 1995 membership survey. J Am Assoc Gynecol Laparosc 4:167–171PubMedCrossRef
2.
Zurück zum Zitat Nduka CC, Super PA, Monson JR, Darzi AW (1994) Cause and prevention of electrosurgical injuries in laparoscopy. J Am Coll Surg 179:161–170PubMed Nduka CC, Super PA, Monson JR, Darzi AW (1994) Cause and prevention of electrosurgical injuries in laparoscopy. J Am Coll Surg 179:161–170PubMed
3.
Zurück zum Zitat Wu MP, Ou CS, Chen SL, Yen EY, Rowbotham R (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179:67–73PubMedCrossRef Wu MP, Ou CS, Chen SL, Yen EY, Rowbotham R (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179:67–73PubMedCrossRef
4.
Zurück zum Zitat IEEE (2009) American national standard dictionary of electromagnetic compatibility (EMC) including electromagnetic environmental effects (E3). Institute of Electrical and Electronics Engineers (IEEE), Washington IEEE (2009) American national standard dictionary of electromagnetic compatibility (EMC) including electromagnetic environmental effects (E3). Institute of Electrical and Electronics Engineers (IEEE), Washington
5.
Zurück zum Zitat Willson PD, van der Walt JD, Moxon D, Rogers J (1997) Port site electrosurgical (diathermy) burns during surgical laparoscopy. Surg Endosc 11:653–654PubMedCrossRef Willson PD, van der Walt JD, Moxon D, Rogers J (1997) Port site electrosurgical (diathermy) burns during surgical laparoscopy. Surg Endosc 11:653–654PubMedCrossRef
6.
Zurück zum Zitat IEEE (2005) IEEE guide for instrumentation and control equipment grounding in generating stations. Institute of Electrical and Electronics Engineers (IEEE), Washington IEEE (2005) IEEE guide for instrumentation and control equipment grounding in generating stations. Institute of Electrical and Electronics Engineers (IEEE), Washington
7.
Zurück zum Zitat Robinson TN, Pavlovsky KR, Looney H, Stiegmann GV, McGreevy FT (2010) Surgeon-controlled factors that reduce monopolar electrosurgery capacitive coupling during laparoscopy. Surg Laparosc Endosc Percutan Tech 20:317–320PubMedCrossRef Robinson TN, Pavlovsky KR, Looney H, Stiegmann GV, McGreevy FT (2010) Surgeon-controlled factors that reduce monopolar electrosurgery capacitive coupling during laparoscopy. Surg Laparosc Endosc Percutan Tech 20:317–320PubMedCrossRef
8.
Zurück zum Zitat Schulz KF, Altman DG, Moher D (2010) CONSORT 2010 statement: updated guidelines for reporting parallel group randomized trials. Ann Intern Med 152:726–732PubMedCrossRef Schulz KF, Altman DG, Moher D (2010) CONSORT 2010 statement: updated guidelines for reporting parallel group randomized trials. Ann Intern Med 152:726–732PubMedCrossRef
9.
Zurück zum Zitat Strasberg SM, Brunt LM (2010) Rationale and use of the critical view of safety in laparoscopic cholecystectomy. J Am Coll Surg 211:132–138PubMedCrossRef Strasberg SM, Brunt LM (2010) Rationale and use of the critical view of safety in laparoscopic cholecystectomy. J Am Coll Surg 211:132–138PubMedCrossRef
10.
Zurück zum Zitat Voyles CR, Tucker RD (1992) Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. Am J Surg 164:57–62PubMedCrossRef Voyles CR, Tucker RD (1992) Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. Am J Surg 164:57–62PubMedCrossRef
11.
Zurück zum Zitat Robinson TN, Barnes K, Govekar HR, Stiegmann GV, Dunn C, McGreevy FT (2012) Antenna coupling: a novel mechanism of radiofrequency electrosurgery complication: practical implications. Ann Surg 256(2):213–218PubMedCrossRef Robinson TN, Barnes K, Govekar HR, Stiegmann GV, Dunn C, McGreevy FT (2012) Antenna coupling: a novel mechanism of radiofrequency electrosurgery complication: practical implications. Ann Surg 256(2):213–218PubMedCrossRef
12.
Zurück zum Zitat Jones EL, Robinson TN, McHenry JR, Dunn CL, Montero PN, Govekar HR, Stiegmann GV (2012) Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications. Surg Endosc 26:3053–3057PubMedCrossRef Jones EL, Robinson TN, McHenry JR, Dunn CL, Montero PN, Govekar HR, Stiegmann GV (2012) Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications. Surg Endosc 26:3053–3057PubMedCrossRef
13.
Zurück zum Zitat Feder BJ (2006) Surgical device poses a rare but serious peril. Blackwell, Malden Feder BJ (2006) Surgical device poses a rare but serious peril. Blackwell, Malden
Metadaten
Titel
Blend mode reduces unintended thermal injury by laparoscopic monopolar instruments: a randomized controlled trial
verfasst von
Edward L. Jones
Christina L. Dunn
Nicole T. Townsend
Teresa S. Jones
J. Bruce Dunne
Paul N. Montero
Henry R. Govekar
Greg V. Stiegmann
Thomas N. Robinson
Publikationsdatum
01.11.2013
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 11/2013
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-013-3032-2

Weitere Artikel der Ausgabe 11/2013

Surgical Endoscopy 11/2013 Zur Ausgabe

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Was nützt die Kraniektomie bei schwerer tiefer Hirnblutung?

17.05.2024 Hirnblutung Nachrichten

Eine Studie zum Nutzen der druckentlastenden Kraniektomie nach schwerer tiefer supratentorieller Hirnblutung deutet einen Nutzen der Operation an. Für überlebende Patienten ist das dennoch nur eine bedingt gute Nachricht.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.