Skip to main content
Erschienen in: Intensive Care Medicine 5/2005

01.05.2005 | Technical Note

Intra- and interobserver variability during in vitro validation of two novel methods for intra-abdominal pressure monitoring

verfasst von: Tom J. R. De Potter, Hilde Dits, Manu L. N. G. Malbrain

Erschienen in: Intensive Care Medicine | Ausgabe 5/2005

Einloggen, um Zugang zu erhalten

Abstract

Objective

Intra-abdominal pressure (IAP) measurement techniques vary in automaticity and reproducibility. This study evaluated the intra- and interobserver variability of two new IAP measurement techniques.

Methods

A one-half open, 30-l container was used with two IAP catheters (Foley Manometer, balloon-tipped catheter) contained in a 100-ml infusion bag. To simulate intra-abdominal hypertension the container was filled with water using 5-cm increments (0–25 cmH2O). Pressure was determined by observers using the Foley Manometer and simultaneously recorded using an IAP monitor. Observers were blinded to the reference levels.

Results

Fifteen observers conducted three pressure readings at each of the six pressure levels with the Foley technique, giving 270 readings. These were paired with the automated monitor readings and the height of the water column. The intra- and interobserver coefficients of variation were low for both methods. The Spearman correlation coefficient was higher than 0.9 for all paired measurements and Bland-Altman analysis comparing the reference H2O column to both measurement techniques showed a very good agreement at all pressure intervals (bias 0.1±0.6 cmH2O) and a consistent, low underestimation of the reference water column pressure by both techniques.

Conclusions

Both the Foley Manometer and the IAP monitor are reliable and reproducible methods to measure IAP in this in vitro model. The coefficient of variation for each technique is low and decreases with increasing IAP, the monitor giving more reproducible results than the Foley Manometer.
Literatur
1.
Zurück zum Zitat Malbrain ML (2004) Is it wise not to think about intraabdominal hypertension in the ICU? Curr Opin Crit Care 10:132–145 Malbrain ML (2004) Is it wise not to think about intraabdominal hypertension in the ICU? Curr Opin Crit Care 10:132–145
2.
Zurück zum Zitat Quintel M, Pelosi P, Caironi P, Meinhardt JP, Lücke T, Herrmann P, Taccone P, Rylander C, Valenza F, Carlesso E, Gattinoni L (2004) An increase of abdominal pressure increases pulmonary edema in oleic acid induced lung injury. Am J Respir Crit Care Med 169:534–541CrossRefPubMed Quintel M, Pelosi P, Caironi P, Meinhardt JP, Lücke T, Herrmann P, Taccone P, Rylander C, Valenza F, Carlesso E, Gattinoni L (2004) An increase of abdominal pressure increases pulmonary edema in oleic acid induced lung injury. Am J Respir Crit Care Med 169:534–541CrossRefPubMed
3.
Zurück zum Zitat Balogh Z, McKinley BA, Holcomb JB, Miller CC, Cocanour CS, Kozar RA, Valdivia A, Ware DN, Moore FA (2003) Both primary and secondary abdominal compartment syndrome can be predicted early and are harbingers of multiple organ failure. J Trauma 54:848–859PubMed Balogh Z, McKinley BA, Holcomb JB, Miller CC, Cocanour CS, Kozar RA, Valdivia A, Ware DN, Moore FA (2003) Both primary and secondary abdominal compartment syndrome can be predicted early and are harbingers of multiple organ failure. J Trauma 54:848–859PubMed
4.
Zurück zum Zitat Richardson JD Trinkle JK (1976) Hemodynamic and respiratory alterations with increased intra-abdominal pressure. J Surg Res 20:401–404CrossRefPubMed Richardson JD Trinkle JK (1976) Hemodynamic and respiratory alterations with increased intra-abdominal pressure. J Surg Res 20:401–404CrossRefPubMed
5.
Zurück zum Zitat Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L (2004) Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 30:822–829CrossRefPubMed Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L (2004) Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 30:822–829CrossRefPubMed
6.
Zurück zum Zitat Diebel LN, Wilson RF, Dulchavsky SA, Saxe J (1992) Effect of increased intra-abdominal pressure on hepatic arterial, portal venous, and hepatic microcirculatory blood flow. J Trauma 33:279–282PubMed Diebel LN, Wilson RF, Dulchavsky SA, Saxe J (1992) Effect of increased intra-abdominal pressure on hepatic arterial, portal venous, and hepatic microcirculatory blood flow. J Trauma 33:279–282PubMed
7.
Zurück zum Zitat Malbrain ML (1999) Abdominal pressure in the critically ill: measurement and clinical relevance. Intensive Care Med 25:1453–1458CrossRefPubMed Malbrain ML (1999) Abdominal pressure in the critically ill: measurement and clinical relevance. Intensive Care Med 25:1453–1458CrossRefPubMed
8.
Zurück zum Zitat Cheatham ML, White MW, Sagraves SG, Johnson JL, Block EF (2000) Abdominal perfusion pressure: a superior parameter in the assessment of intra-abdominal hypertension. J Trauma 49:621–626 Cheatham ML, White MW, Sagraves SG, Johnson JL, Block EF (2000) Abdominal perfusion pressure: a superior parameter in the assessment of intra-abdominal hypertension. J Trauma 49:621–626
9.
Zurück zum Zitat Malbrain ML (2004) Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 30:357–371CrossRef Malbrain ML (2004) Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 30:357–371CrossRef
10.
Zurück zum Zitat Sugrue M, Bauman A, Jones F, Bishop G, Flabouris A, Parr M, Stewart A, Hillman K, Deane SA (2002) Clinical examination is an inaccurate predictor of intraabdominal pressure. World J Surg 26:1428–1431 Sugrue M, Bauman A, Jones F, Bishop G, Flabouris A, Parr M, Stewart A, Hillman K, Deane SA (2002) Clinical examination is an inaccurate predictor of intraabdominal pressure. World J Surg 26:1428–1431
11.
Zurück zum Zitat Kirkpatrick AW, Brenneman FD, McLean RF, Rapanos T, Boulanger BR (2000) Is clinical examination an accurate indicator of raised intra-abdominal pressure in critically injured patients? Can J Surg 43:207–211 Kirkpatrick AW, Brenneman FD, McLean RF, Rapanos T, Boulanger BR (2000) Is clinical examination an accurate indicator of raised intra-abdominal pressure in critically injured patients? Can J Surg 43:207–211
12.
Zurück zum Zitat Cheatham ML Safcsak K (1998) Intraabdominal pressure: a revised method for measurement. J Am Coll Surg 186:594–595 Cheatham ML Safcsak K (1998) Intraabdominal pressure: a revised method for measurement. J Am Coll Surg 186:594–595
13.
Zurück zum Zitat Malbrain M (2004) The assumed problem of air bubbles in the tubing during intra-abdominal pressure measurement. Reply to Dr. De waele. Intensive Care Med 30:1693 Malbrain M (2004) The assumed problem of air bubbles in the tubing during intra-abdominal pressure measurement. Reply to Dr. De waele. Intensive Care Med 30:1693
14.
Zurück zum Zitat Schachtrupp A, Tons C, Fackeldey V, Hoer J, Reinges M, Schumpelick V (2003) Evaluation of two novel methods for the direct and continuous measurement of the intra-abdominal pressure in a porcine model. Intensive Care Med 29:1605–1608 Schachtrupp A, Tons C, Fackeldey V, Hoer J, Reinges M, Schumpelick V (2003) Evaluation of two novel methods for the direct and continuous measurement of the intra-abdominal pressure in a porcine model. Intensive Care Med 29:1605–1608
15.
Zurück zum Zitat De Potter T, Dits H, Deeren D, Libeer C, Malbrain MLNG (2004) Inter and intra-observer variability during intermittent and continuous intraabdominal pressure monitoring. WCACS abstract book, A72 De Potter T, Dits H, Deeren D, Libeer C, Malbrain MLNG (2004) Inter and intra-observer variability during intermittent and continuous intraabdominal pressure monitoring. WCACS abstract book, A72
16.
Zurück zum Zitat Balogh Z, Jones F, D’Amours S, Parr M, Sugrue M (2004) Continuous intra-abdominal pressure measurement technique. Am J Surg 188:679–684 Balogh Z, Jones F, D’Amours S, Parr M, Sugrue M (2004) Continuous intra-abdominal pressure measurement technique. Am J Surg 188:679–684
17.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310 Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310
18.
Zurück zum Zitat Schein M, Wittmann DH, Aprahamian CC, Condon RE (1995) The abdominal compartment syndrome: the physiological and clinical consequences of elevated intra-abdominal pressure. J Am Coll Surg 180:745–753 Schein M, Wittmann DH, Aprahamian CC, Condon RE (1995) The abdominal compartment syndrome: the physiological and clinical consequences of elevated intra-abdominal pressure. J Am Coll Surg 180:745–753
19.
Zurück zum Zitat Ivatury RR, Diebel L, Porter JM, Simon RJ (1997) Intra-abdominal hypertension and the abdominal compartment syndrome. Surg Clin North Am 77:783–800 Ivatury RR, Diebel L, Porter JM, Simon RJ (1997) Intra-abdominal hypertension and the abdominal compartment syndrome. Surg Clin North Am 77:783–800
20.
Zurück zum Zitat Collee GG, Lomax DM, Ferguson C, Hanson GC (1993) Bedside measurement of intra-abdominal pressure (IAP) via an indwelling naso-gastric tube: clinical validation of the technique. Intensive Care Med 19:478–480 Collee GG, Lomax DM, Ferguson C, Hanson GC (1993) Bedside measurement of intra-abdominal pressure (IAP) via an indwelling naso-gastric tube: clinical validation of the technique. Intensive Care Med 19:478–480
21.
Zurück zum Zitat Lacey SR, Bruce J, Brooks SP, Griswald J, Ferguson W, Allen JE, Jewett TC Jr, Karp MP, Cooney DR (1987) The relative merits of various methods of indirect measurement of intraabdominal pressure as a guide to closure of abdominal wall defects. J Pediatr Surg 22:1207–1211 Lacey SR, Bruce J, Brooks SP, Griswald J, Ferguson W, Allen JE, Jewett TC Jr, Karp MP, Cooney DR (1987) The relative merits of various methods of indirect measurement of intraabdominal pressure as a guide to closure of abdominal wall defects. J Pediatr Surg 22:1207–1211
22.
Zurück zum Zitat Obeid F, Saba A, Fath J, Guslits B, Chung R, Sorensen V, Buck J, Horst M (1995) Increases in intra-abdominal pressure affect pulmonary compliance. Arch Surg 130:544–547 Obeid F, Saba A, Fath J, Guslits B, Chung R, Sorensen V, Buck J, Horst M (1995) Increases in intra-abdominal pressure affect pulmonary compliance. Arch Surg 130:544–547
Metadaten
Titel
Intra- and interobserver variability during in vitro validation of two novel methods for intra-abdominal pressure monitoring
verfasst von
Tom J. R. De Potter
Hilde Dits
Manu L. N. G. Malbrain
Publikationsdatum
01.05.2005
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 5/2005
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-005-2597-1

Weitere Artikel der Ausgabe 5/2005

Intensive Care Medicine 5/2005 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.