01.05.2013 | Computed Tomography

Optimisation of vascular input and output functions in CT-perfusion imaging using 256(or more)-slice multidetector CT

Erschienen in: European Radiology | Ausgabe 5/2013

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the accuracy and reproducibility of CT-perfusion (CTP) by finding the optimal artery for the arterial input function (AIF) and re-evaluating the necessity of the venous output function (VOF).

Methods

Forty-four acute ischaemic stroke patients who underwent non-enhanced CT, CTP and CT-angiography using 256-slice multidetector computed tomography (MDCT) were evaluated. The anterior cerebral artery (ACA), middle cerebral artery (MCA), internal carotid artery (ICA) and basilar artery were selected as the AIF. Subsequently the resulting area under the time–enhancement curve of the AIF (AUCAIF) and quantitative perfusion measurements were analysed by repeated measures ANOVA and subsequently the paired t test. To evaluate reproducibility we examined if the VOF could be deleted by comparing the perfusion measurements using versus not using the VOF (paired t test).

Results

The AUCAIF and perfusion measurements resulting from the different AIFs showed significant group differences (all P < 0.0001). The ICA had the largest AUCAIF and resulted in the highest mean transient time (MTT) and lowest cerebral blood flow (CBF), whereas the basilar artery showed the lowest cerebral blood volume (CBV). Not using the VOF showed significantly higher CBV and CBF in 66 % of patients on the ipsilateral (P < 0.0001 and P = 0.007, respectively) and contralateral hemisphere (P < 0.0001 and P = 0.019, respectively).

Conclusion

Selecting the ICA as the AIF and continuing the use of the VOF would improve the accuracy of CTP.

Key Points

Perfusion imaging is an increasingly important aspect of multidetector computed tomography (MDCT).
Vascular input functions were evaluated for CT-perfusion using 256-slice MDCT.
Selecting different arterial input functions (AIFs) leads to variation in quantitative values.
Using the internal carotid artery for AIF provides optimal perfusion values.
Deleting the venous output function would be detrimental for validity.
Literatur
1.
Zurück zum Zitat Wintermark M, Reichhart M, Thiran JP et al (2002) Prognostic accuracy of cerebral blood flow measurement by perfusion computed tomography, at the time of emergency room admission, in acute stroke patients. Ann Neurol 51:417–432PubMedCrossRef Wintermark M, Reichhart M, Thiran JP et al (2002) Prognostic accuracy of cerebral blood flow measurement by perfusion computed tomography, at the time of emergency room admission, in acute stroke patients. Ann Neurol 51:417–432PubMedCrossRef
2.
Zurück zum Zitat Miles KA, Lee TY, Goh V et al (2012) Current status and guidelines for the assessment of tumour vascular support with dynamic contrast-enhanced computed tomography. Eur Radiol 22:1430–1441PubMedCrossRef Miles KA, Lee TY, Goh V et al (2012) Current status and guidelines for the assessment of tumour vascular support with dynamic contrast-enhanced computed tomography. Eur Radiol 22:1430–1441PubMedCrossRef
3.
Zurück zum Zitat Bisdas S, Donnerstag F, Ahl B, Bohrer I, Weissenborn K, Becker H (2004) Comparison of perfusion computed tomography with diffusion-weighted magnetic resonance imaging in hyperacute ischemic stroke. J Comput Assist Tomogr 28:747–755PubMedCrossRef Bisdas S, Donnerstag F, Ahl B, Bohrer I, Weissenborn K, Becker H (2004) Comparison of perfusion computed tomography with diffusion-weighted magnetic resonance imaging in hyperacute ischemic stroke. J Comput Assist Tomogr 28:747–755PubMedCrossRef
4.
Zurück zum Zitat Nabavi DG, Cenic A, Henderson S, Gelb AW, Lee TY (2001) Perfusion mapping using computed tomography allows accurate prediction of cerebral infarction in experimental brain ischemia. Stroke 32:175–183PubMedCrossRef Nabavi DG, Cenic A, Henderson S, Gelb AW, Lee TY (2001) Perfusion mapping using computed tomography allows accurate prediction of cerebral infarction in experimental brain ischemia. Stroke 32:175–183PubMedCrossRef
5.
Zurück zum Zitat Rother J, Jonetz-Mentzel L, Fiala A et al (2000) Hemodynamic assessment of acute stroke using dynamic single-slice computed tomographic perfusion imaging. Arch Neurol 57:1161–1166PubMedCrossRef Rother J, Jonetz-Mentzel L, Fiala A et al (2000) Hemodynamic assessment of acute stroke using dynamic single-slice computed tomographic perfusion imaging. Arch Neurol 57:1161–1166PubMedCrossRef
6.
Zurück zum Zitat Wintermark M, Meuli R, Browaeys P et al (2007) Comparison of CT perfusion and angiography and MRI in selecting stroke patients for acute treatment. Neurology 68:694–697PubMedCrossRef Wintermark M, Meuli R, Browaeys P et al (2007) Comparison of CT perfusion and angiography and MRI in selecting stroke patients for acute treatment. Neurology 68:694–697PubMedCrossRef
7.
Zurück zum Zitat Eastwood JD, Lev MH, Azhari T et al (2002) CT perfusion scanning with deconvolution analysis: pilot study in patients with acute middle cerebral artery stroke. Radiology 222:227–236PubMedCrossRef Eastwood JD, Lev MH, Azhari T et al (2002) CT perfusion scanning with deconvolution analysis: pilot study in patients with acute middle cerebral artery stroke. Radiology 222:227–236PubMedCrossRef
8.
Zurück zum Zitat Turk AS, Grayev A, Rowley HA et al (2007) Variability of clinical CT perfusion measurements in patients with carotid stenosis. Neuroradiology 49:955–961PubMedCrossRef Turk AS, Grayev A, Rowley HA et al (2007) Variability of clinical CT perfusion measurements in patients with carotid stenosis. Neuroradiology 49:955–961PubMedCrossRef
9.
Zurück zum Zitat Zussman B, Jabbour P, Talekar K, Gorniak R, Flanders AE (2011) Sources of variability in computed tomography perfusion: implications for acute stroke management. Neurosurg Focus 30:E8PubMedCrossRef Zussman B, Jabbour P, Talekar K, Gorniak R, Flanders AE (2011) Sources of variability in computed tomography perfusion: implications for acute stroke management. Neurosurg Focus 30:E8PubMedCrossRef
10.
Zurück zum Zitat Soares BP, Dankbaar JW, Bredno J et al (2009) Automated versus manual post-processing of perfusion-CT data in patients with acute cerebral ischemia: influence on interobserver variability. Neuroradiology 51:445–451PubMedCrossRef Soares BP, Dankbaar JW, Bredno J et al (2009) Automated versus manual post-processing of perfusion-CT data in patients with acute cerebral ischemia: influence on interobserver variability. Neuroradiology 51:445–451PubMedCrossRef
11.
Zurück zum Zitat Wintermark M, Ko NU, Smith WS, Liu S, Higashida RT, Dillon WP (2006) Vasospasm after subarachnoid hemorrhage: utility of perfusion CT and CT angiography on diagnosis and management. AJNR Am J Neuroradiol 27:26–34PubMed Wintermark M, Ko NU, Smith WS, Liu S, Higashida RT, Dillon WP (2006) Vasospasm after subarachnoid hemorrhage: utility of perfusion CT and CT angiography on diagnosis and management. AJNR Am J Neuroradiol 27:26–34PubMed
12.
Zurück zum Zitat Dankbaar JW, de Rooij NK, Rijsdijk M et al (2010) Diagnostic threshold values of cerebral perfusion measured with computed tomography for delayed cerebral ischemia after aneurysmal subarachnoid hemorrhage. Stroke 41:1927–1932PubMedCrossRef Dankbaar JW, de Rooij NK, Rijsdijk M et al (2010) Diagnostic threshold values of cerebral perfusion measured with computed tomography for delayed cerebral ischemia after aneurysmal subarachnoid hemorrhage. Stroke 41:1927–1932PubMedCrossRef
13.
Zurück zum Zitat Kealey SM, Loving VA, Delong DM, Eastwood JD (2004) User-defined vascular input function curves: influence on mean perfusion parameter values and signal-to-noise ratio. Radiology 231:587–593PubMedCrossRef Kealey SM, Loving VA, Delong DM, Eastwood JD (2004) User-defined vascular input function curves: influence on mean perfusion parameter values and signal-to-noise ratio. Radiology 231:587–593PubMedCrossRef
14.
Zurück zum Zitat Sakai F, Nakazawa K, Tazaki Y et al (1985) Regional cerebral blood volume and hematocrit measured in normal human volunteers by single-photon emission computed tomography. J Cereb Blood Flow Metab 5:207–213PubMedCrossRef Sakai F, Nakazawa K, Tazaki Y et al (1985) Regional cerebral blood volume and hematocrit measured in normal human volunteers by single-photon emission computed tomography. J Cereb Blood Flow Metab 5:207–213PubMedCrossRef
15.
Zurück zum Zitat van der Schaaf I, Vonken EJ, Waaijer A, Velthuis B, Quist M, van Osch T (2006) Influence of partial volume on venous output and arterial input function. AJNR Am J Neuroradiol 27:46–50PubMed van der Schaaf I, Vonken EJ, Waaijer A, Velthuis B, Quist M, van Osch T (2006) Influence of partial volume on venous output and arterial input function. AJNR Am J Neuroradiol 27:46–50PubMed
16.
Zurück zum Zitat Axel L (1983) Tissue mean transit time from dynamic computed tomography by a simple deconvolution technique. Investig Radiol 18:94–99CrossRef Axel L (1983) Tissue mean transit time from dynamic computed tomography by a simple deconvolution technique. Investig Radiol 18:94–99CrossRef
17.
Zurück zum Zitat Ferreira RM, Lev MH, Goldmakher GV et al (2010) Arterial input function placement for accurate CT perfusion map construction in acute stroke. AJR Am J Roentgenol 194:1330–1336PubMedCrossRef Ferreira RM, Lev MH, Goldmakher GV et al (2010) Arterial input function placement for accurate CT perfusion map construction in acute stroke. AJR Am J Roentgenol 194:1330–1336PubMedCrossRef
18.
Zurück zum Zitat Sanelli PC, Lev MH, Eastwood JD, Gonzalez RG, Lee TY (2004) The effect of varying user-selected input parameters on quantitative values in CT perfusion maps. Acad Radiol 11:1085–1092PubMedCrossRef Sanelli PC, Lev MH, Eastwood JD, Gonzalez RG, Lee TY (2004) The effect of varying user-selected input parameters on quantitative values in CT perfusion maps. Acad Radiol 11:1085–1092PubMedCrossRef
19.
Zurück zum Zitat Dorn F, Muenzel D, Meier R, Poppert H, Rummeny EJ, Huber A (2012) Brain perfusion CT for acute stroke using a 256-slice CT: improvement of diagnostic information by large volume coverage. Eur Radiol 21:1803–1810CrossRef Dorn F, Muenzel D, Meier R, Poppert H, Rummeny EJ, Huber A (2012) Brain perfusion CT for acute stroke using a 256-slice CT: improvement of diagnostic information by large volume coverage. Eur Radiol 21:1803–1810CrossRef
20.
Zurück zum Zitat Wintermark M, Flanders AE, Velthuis B et al (2006) Perfusion-CT assessment of infarct core and penumbra: receiver operating characteristic curve analysis in 130 patients suspected of acute hemispheric stroke. Stroke 37:979–985PubMedCrossRef Wintermark M, Flanders AE, Velthuis B et al (2006) Perfusion-CT assessment of infarct core and penumbra: receiver operating characteristic curve analysis in 130 patients suspected of acute hemispheric stroke. Stroke 37:979–985PubMedCrossRef
21.
Zurück zum Zitat Wintermark M, Lau BC, Chien J, Arora S (2008) The anterior cerebral artery is an appropriate arterial input function for perfusion-CT processing in patients with acute stroke. Neuroradiology 50:227–236PubMedCrossRef Wintermark M, Lau BC, Chien J, Arora S (2008) The anterior cerebral artery is an appropriate arterial input function for perfusion-CT processing in patients with acute stroke. Neuroradiology 50:227–236PubMedCrossRef
22.
Zurück zum Zitat Kamath A, Smith WS, Powers WJ et al (2008) Perfusion CT compared to H(2) (15)O/O (15)O PET in patients with chronic cervical carotid artery occlusion. Neuroradiology 50:745–751PubMedCrossRef Kamath A, Smith WS, Powers WJ et al (2008) Perfusion CT compared to H(2) (15)O/O (15)O PET in patients with chronic cervical carotid artery occlusion. Neuroradiology 50:745–751PubMedCrossRef
23.
Zurück zum Zitat Calamante F, Morup M, Hansen LK (2004) Defining a local arterial input function for perfusion MRI using independent component analysis. Magn Reson Med 789–797 Calamante F, Morup M, Hansen LK (2004) Defining a local arterial input function for perfusion MRI using independent component analysis. Magn Reson Med 789–797
24.
Zurück zum Zitat Sheikh K, Schipper MJ, Hoeffner EG (2009) Feasibility of superficial temporal artery as the input artery for cerebral perfusion CT. AJR Am J Roentgenol 192:W321–W329PubMedCrossRef Sheikh K, Schipper MJ, Hoeffner EG (2009) Feasibility of superficial temporal artery as the input artery for cerebral perfusion CT. AJR Am J Roentgenol 192:W321–W329PubMedCrossRef
25.
Zurück zum Zitat Lee TY (2002) Functional CT: psychological models. Trends Biotechnol 20:s3–s10CrossRef Lee TY (2002) Functional CT: psychological models. Trends Biotechnol 20:s3–s10CrossRef
26.
Zurück zum Zitat Bisdas S, Konstantinou GN, Gurung J et al (2007) Effect of the arterial input function on the measured perfusion values and infarct volumetric in acute cerebral ischemia evaluated by perfusion computed tomography. Investig Radiol 42:147–156CrossRef Bisdas S, Konstantinou GN, Gurung J et al (2007) Effect of the arterial input function on the measured perfusion values and infarct volumetric in acute cerebral ischemia evaluated by perfusion computed tomography. Investig Radiol 42:147–156CrossRef
27.
Zurück zum Zitat Chiu FY, Teng MM, Kao YH, et al (2012) Selection of arterial input function for postprocessing of cerebral ct perfusion in chronic unilateral high-grade stenosis or occlusion of the carotid or middle cerebral artery. Acad Radiol 19:8–16 Chiu FY, Teng MM, Kao YH, et al (2012) Selection of arterial input function for postprocessing of cerebral ct perfusion in chronic unilateral high-grade stenosis or occlusion of the carotid or middle cerebral artery. Acad Radiol 19:8–16
28.
Zurück zum Zitat Wu O, Ostergaard L, Weisskoff RM, Benner T, Rosen BR, Sorensen AG (2003) Tracer arrival timing-insensitive technique for estimating flow in MR perfusion-weighted imaging using singular value decomposition with a block-circulant deconvolution matrix. Magn Reson Med 50:164–174PubMedCrossRef Wu O, Ostergaard L, Weisskoff RM, Benner T, Rosen BR, Sorensen AG (2003) Tracer arrival timing-insensitive technique for estimating flow in MR perfusion-weighted imaging using singular value decomposition with a block-circulant deconvolution matrix. Magn Reson Med 50:164–174PubMedCrossRef
29.
Zurück zum Zitat Sasaki M, Kudo K, Ogasawara K, Fujiwara S (2009) Tracer delay-insensitive algorithm can improve reliability of CT perfusion imaging for cerebrovascular steno-occlusive disease: comparison with quantitative single-photon emission CT. AJNR Am J Neuroradiol 30:188–193PubMedCrossRef Sasaki M, Kudo K, Ogasawara K, Fujiwara S (2009) Tracer delay-insensitive algorithm can improve reliability of CT perfusion imaging for cerebrovascular steno-occlusive disease: comparison with quantitative single-photon emission CT. AJNR Am J Neuroradiol 30:188–193PubMedCrossRef
30.
Zurück zum Zitat Smith MR, Lu H, Trochet S, Frayne R (2004) Removing the effect of SVD algorithmic artifacts present in quantitative MR perfusion studies. Magn Reson Med 51:631–634PubMedCrossRef Smith MR, Lu H, Trochet S, Frayne R (2004) Removing the effect of SVD algorithmic artifacts present in quantitative MR perfusion studies. Magn Reson Med 51:631–634PubMedCrossRef
31.
Zurück zum Zitat Abels B, Villablanca JP, Tomandl BF, Uder M, Lell MM (2012) Acute stroke: a comparison of different CT perfusion algorithms and validation of ischaemic lesions by follow-up imaging. Eur Radiol. doi:10.1007/s00330-012-2529-8 Abels B, Villablanca JP, Tomandl BF, Uder M, Lell MM (2012) Acute stroke: a comparison of different CT perfusion algorithms and validation of ischaemic lesions by follow-up imaging. Eur Radiol. doi:10.​1007/​s00330-012-2529-8
Metadaten
Titel
Optimisation of vascular input and output functions in CT-perfusion imaging using 256(or more)-slice multidetector CT
Publikationsdatum
01.05.2013
Erschienen in
European Radiology / Ausgabe 5/2013
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-012-2731-8

Weitere Artikel der Ausgabe 5/2013

European Radiology 5/2013 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.