Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 2/2018

02.05.2017 | Original Research

Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: a comparison between EQUANOX and O3

verfasst von: Arnaud Ferraris, Matthias Jacquet-Lagrèze, Jean-Luc Fellahi

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 2/2018

Einloggen, um Zugang zu erhalten

Abstract

Near-infrared spectroscopy (NIRS) is a continuous and noninvasive technology that measures regional tissue oxygen saturation (rSO2). A new 4-wavelength generation of NIRS monitors is now available. We aimed to compare peripheral somatic rSO2 values given by the 4-wavelength EQUANOX™ 7600 device (Nonin Medical Inc., Plymouth, Mn) and O3™ device (Masimo Corporation, Irvine, CA). Twenty adult patients scheduled for conventional elective cardiac surgery with cardiopulmonary bypass over a 4-month period were included after local Ethics Committee approval. For each patient, 2 NIRS sensors (EQUANOX and O3) were placed over the medial part of the forearm. Thirteen couples of measurements were performed at predefined intraoperative time points. We compared 260 couples of absolute intraoperative rSO2 values. No significant difference was found between both monitors: EQUANOX median rSO2 60% (95% CI 57–62) versus O3 median rSO2 62% (95% CI 61–64), P = 0.103. Bias was 4.0% and limits of agreement were ±26.3%. Significant correlations were evidenced between EQUANOX and O3 rSO2 absolute values: rho = 0.758 (95% CI 0.701–0.806), P < 0.0001, and rSO2 percent maximum difference versus baseline: rho = 0.582 (95% CI 0.188–0.815), P = 0.007. While absolute values of rSO2 given by both devices were equivalent and well correlated, the clinical agreement is probably not acceptable, meaning that EQUANOX and O3 are not interchangeable in routine practice.
Literatur
1.
Zurück zum Zitat Murkin JM, Arango M. Near-infrared spectroscopy as an index of brain and tissue oxygenation. Br J Anaesth. 2009;103(Suppl 1):i3–i13.CrossRefPubMed Murkin JM, Arango M. Near-infrared spectroscopy as an index of brain and tissue oxygenation. Br J Anaesth. 2009;103(Suppl 1):i3–i13.CrossRefPubMed
2.
Zurück zum Zitat Bartels SA, Bezemer R, de Vries FJW, Milstein DMJ, Lima A, Cherpanath TGV, Anton Van den Meiracker H, Jasper Van Bommel J, Heger M, Karemaker JM, Ince C. Multi-site and multi-depth near-infrared spectroscopy in a model of simulated (central) hypovolemia: lower body negative pressure. Intensive Care Med. 2011;37(4):671–7.CrossRefPubMedPubMedCentral Bartels SA, Bezemer R, de Vries FJW, Milstein DMJ, Lima A, Cherpanath TGV, Anton Van den Meiracker H, Jasper Van Bommel J, Heger M, Karemaker JM, Ince C. Multi-site and multi-depth near-infrared spectroscopy in a model of simulated (central) hypovolemia: lower body negative pressure. Intensive Care Med. 2011;37(4):671–7.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Cole AL, Herman RA, Heimlich JB, Ahsan S, Freedman BA, Shuler MS. Ability of near infrared spectroscopy to measure oxygenation in isolated upper extremity muscle compartments. J Hand Surg. 2012;37(2):297–302.CrossRef Cole AL, Herman RA, Heimlich JB, Ahsan S, Freedman BA, Shuler MS. Ability of near infrared spectroscopy to measure oxygenation in isolated upper extremity muscle compartments. J Hand Surg. 2012;37(2):297–302.CrossRef
4.
Zurück zum Zitat Siegenthaler N, Giraud R, Piriou V, Bendjelid K. Near-infrared spectroscopy monitoring during cardiac surgery: a promising concept? Ann Fr Anesth Réanim. 2011;30(7–8):531–2.CrossRefPubMed Siegenthaler N, Giraud R, Piriou V, Bendjelid K. Near-infrared spectroscopy monitoring during cardiac surgery: a promising concept? Ann Fr Anesth Réanim. 2011;30(7–8):531–2.CrossRefPubMed
5.
Zurück zum Zitat Scheeren TWL. Journal of Clinical Monitoring and Computing 2015 end of year summary: tissue oxygenation and microcirculation. J Clin Monit Comput. 2016;30:141–6.CrossRefPubMedPubMedCentral Scheeren TWL. Journal of Clinical Monitoring and Computing 2015 end of year summary: tissue oxygenation and microcirculation. J Clin Monit Comput. 2016;30:141–6.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Scheeren TWL, Schober P, Schwarte LA. Monitoring tissue oxygenation by near infrared spectroscopy (NIRS): background and current applications. J Clin Monit Comput. 2012;26(4):279–87.CrossRefPubMedPubMedCentral Scheeren TWL, Schober P, Schwarte LA. Monitoring tissue oxygenation by near infrared spectroscopy (NIRS): background and current applications. J Clin Monit Comput. 2012;26(4):279–87.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Fellahi J-L, Butin G, Fischer M-O, Zamparini G, Gérard J-L, Hanouz J-L. Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: a comparison between INVOS and EQUANOX. J Crit Care. 2013;28(5):881.e1-881.e6.CrossRef Fellahi J-L, Butin G, Fischer M-O, Zamparini G, Gérard J-L, Hanouz J-L. Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: a comparison between INVOS and EQUANOX. J Crit Care. 2013;28(5):881.e1-881.e6.CrossRef
8.
Zurück zum Zitat Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology. 2012;116(4):834–40.CrossRefPubMed Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology. 2012;116(4):834–40.CrossRefPubMed
9.
Zurück zum Zitat Thielmann M, Kottenberg E, Kleinbongard P, Wendt D, Gedik N, Pasa S, Price V, Tsagakis K, Neuhäuser M, Peters J, Jakob H, Heusch G. Cardioprotective and prognostic effects of remote ischaemic preconditioning in patients undergoing coronary artery bypass surgery: a single-centre randomised, double-blind, controlled trial. Lancet Lond Engl. 2013;382(9892):597–604.CrossRef Thielmann M, Kottenberg E, Kleinbongard P, Wendt D, Gedik N, Pasa S, Price V, Tsagakis K, Neuhäuser M, Peters J, Jakob H, Heusch G. Cardioprotective and prognostic effects of remote ischaemic preconditioning in patients undergoing coronary artery bypass surgery: a single-centre randomised, double-blind, controlled trial. Lancet Lond Engl. 2013;382(9892):597–604.CrossRef
10.
Zurück zum Zitat Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.CrossRefPubMed Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.CrossRefPubMed
11.
Zurück zum Zitat Bickler PE, Feiner JR, Rollins MD. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers. Anesth Analg. 2013;117(4):813–23.CrossRefPubMed Bickler PE, Feiner JR, Rollins MD. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers. Anesth Analg. 2013;117(4):813–23.CrossRefPubMed
12.
Zurück zum Zitat Redford D, Paidy S, Kashif F. Absolute and trend accuracy of a new regional oximeter in healthy volunteers during controlled hypoxia. Anesth Analg. 2014;119(6):1315–9.CrossRefPubMedPubMedCentral Redford D, Paidy S, Kashif F. Absolute and trend accuracy of a new regional oximeter in healthy volunteers during controlled hypoxia. Anesth Analg. 2014;119(6):1315–9.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Pisano A, Galdieri N, Iovino TP, Angelone M, Corcione A. Direct comparison between cerebral oximetry by INVOSTM and EQUANOXTM during cardiac surgery: a pilot study. Heart Lung Vessels. 2014;6(3):197–203.PubMedPubMedCentral Pisano A, Galdieri N, Iovino TP, Angelone M, Corcione A. Direct comparison between cerebral oximetry by INVOSTM and EQUANOXTM during cardiac surgery: a pilot study. Heart Lung Vessels. 2014;6(3):197–203.PubMedPubMedCentral
14.
Zurück zum Zitat Cournoyer A, Denault A, Cossette S, Fortier A, Daoust R, Iseppon M, Chauny JM, Notebaert E. Reproducibility, interchangeability of measures, time to measure stabilization, and reference values of two tissue oximeters in healthy volunteers. J Biomed Opt. 2016;21(9):97003.CrossRefPubMed Cournoyer A, Denault A, Cossette S, Fortier A, Daoust R, Iseppon M, Chauny JM, Notebaert E. Reproducibility, interchangeability of measures, time to measure stabilization, and reference values of two tissue oximeters in healthy volunteers. J Biomed Opt. 2016;21(9):97003.CrossRefPubMed
15.
Zurück zum Zitat Douds MT, Straub EJ, Kent AC, Bistrick CH, Sistino JJ. A systematic review of cerebral oxygenation-monitoring devices in cardiac surgery. Perfusion. 2014;29(6):545–52.CrossRefPubMed Douds MT, Straub EJ, Kent AC, Bistrick CH, Sistino JJ. A systematic review of cerebral oxygenation-monitoring devices in cardiac surgery. Perfusion. 2014;29(6):545–52.CrossRefPubMed
16.
Zurück zum Zitat Steenhaut K, Lapage K, Bové T, De Hert S, Moerman A. Evaluation of different near-infrared spectroscopy technologies for assessment of tissue oxygen saturation during a vascular occlusion test. J Clin Monit Comput. 2016. doi:10.1007/s10877-016-9962-1. Steenhaut K, Lapage K, Bové T, De Hert S, Moerman A. Evaluation of different near-infrared spectroscopy technologies for assessment of tissue oxygen saturation during a vascular occlusion test. J Clin Monit Comput. 2016. doi:10.​1007/​s10877-016-9962-1.
17.
Zurück zum Zitat Gómez H, Torres A, Polanco P, Kim HK, Zenker S, Puyana JC, Pinsky MR. Use of non-invasive NIRS during a vascular occlusion test to assess dynamic tissue O(2) saturation response. Intensive Care Med. 2008;34(9):1600–7.CrossRefPubMed Gómez H, Torres A, Polanco P, Kim HK, Zenker S, Puyana JC, Pinsky MR. Use of non-invasive NIRS during a vascular occlusion test to assess dynamic tissue O(2) saturation response. Intensive Care Med. 2008;34(9):1600–7.CrossRefPubMed
18.
Zurück zum Zitat Mayeur C, Campard S, Richard C, Teboul J-L. Comparison of four different vascular occlusion tests for assessing reactive hyperemia using near-infrared spectroscopy. Crit Care Med. 2011;39(4):695–701.CrossRefPubMed Mayeur C, Campard S, Richard C, Teboul J-L. Comparison of four different vascular occlusion tests for assessing reactive hyperemia using near-infrared spectroscopy. Crit Care Med. 2011;39(4):695–701.CrossRefPubMed
19.
Zurück zum Zitat Fellahi J-L, Butin G, Zamparini G, Fischer M-O, Gérard J-L, Hanouz J-L. Lower limb peripheral NIRS parameters during a vascular occlusion test: an experimental study in healthy volunteers. Ann Fr Anesth Réanim. 2014;33(1):e9–e14.CrossRefPubMed Fellahi J-L, Butin G, Zamparini G, Fischer M-O, Gérard J-L, Hanouz J-L. Lower limb peripheral NIRS parameters during a vascular occlusion test: an experimental study in healthy volunteers. Ann Fr Anesth Réanim. 2014;33(1):e9–e14.CrossRefPubMed
20.
Zurück zum Zitat Hausenloy DJ, Mwamure PK, Venugopal V, Harris J, Barnard M, Grundy E, Ashley E, Vichare S, Di Salvo C, Kolvekar S, Hayward M, Keogh B, MacAllister RJ, Yellon DM. Effect of remote ischaemic preconditioning on myocardial injury in patients undergoing coronary artery bypass graft surgery: a randomised controlled trial. Lancet Lond Engl. 2007;370(9587):575–9.CrossRef Hausenloy DJ, Mwamure PK, Venugopal V, Harris J, Barnard M, Grundy E, Ashley E, Vichare S, Di Salvo C, Kolvekar S, Hayward M, Keogh B, MacAllister RJ, Yellon DM. Effect of remote ischaemic preconditioning on myocardial injury in patients undergoing coronary artery bypass graft surgery: a randomised controlled trial. Lancet Lond Engl. 2007;370(9587):575–9.CrossRef
21.
Zurück zum Zitat Kim MB, Ward DS, Cartwright CR, Kolano J, Chlebowski S, Henson LC. Estimation of jugular venous O2 saturation from cerebral oximetry or arterial O2 saturation during isocapnic hypoxia. J Clin Monit Comput. 2000;16(3):191–9.CrossRefPubMed Kim MB, Ward DS, Cartwright CR, Kolano J, Chlebowski S, Henson LC. Estimation of jugular venous O2 saturation from cerebral oximetry or arterial O2 saturation during isocapnic hypoxia. J Clin Monit Comput. 2000;16(3):191–9.CrossRefPubMed
Metadaten
Titel
Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: a comparison between EQUANOX and O3
verfasst von
Arnaud Ferraris
Matthias Jacquet-Lagrèze
Jean-Luc Fellahi
Publikationsdatum
02.05.2017
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 2/2018
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-017-0025-z

Weitere Artikel der Ausgabe 2/2018

Journal of Clinical Monitoring and Computing 2/2018 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.