Skip to main content
Erschienen in: Critical Care 1/2015

Open Access 01.12.2015 | Review

Perioperative cardiovascular monitoring of high-risk patients: a consensus of 12

verfasst von: Jean-Louis Vincent, Paolo Pelosi, Rupert Pearse, Didier Payen, Azriel Perel, Andreas Hoeft, Stefano Romagnoli, V Marco Ranieri, Carole Ichai, Patrice Forget, Giorgio Della Rocca, Andrew Rhodes

Erschienen in: Critical Care | Ausgabe 1/2015

Abstract

A significant number of surgical patients are at risk of intra- or post-operative complications or both, which are associated with increased lengths of stay, costs, and mortality. Reducing these risks is important for the individual patient but also for health-care planners and managers. Insufficient tissue perfusion and cellular oxygenation due to hypovolemia, heart dysfunction or both is one of the leading causes of perioperative complications. Adequate perioperative management guided by effective and timely hemodynamic monitoring can help reduce the risk of complications and thus potentially improve outcomes. In this review, we describe the various available hemodynamic monitoring systems and how they can best be used to guide cardiovascular and fluid management in the perioperative period in high-risk surgical patients.
Hinweise

Competing interests

RP has received equipment loans from LiDCO Ltd and has performed consultancy work for Edwards Lifesciences, Covidien (Dublin, Ireland), and Masimo. AP has received advisory board fees from Pulsion Medical Systems. AH has received lecture honoraria from Edwards Lifesciences and is an advisory board member without fees for UPmed (Munich, Germany). PF has received honoraria from Masimo for presentations in congresses. AR has received lecture and advisory board fees from LiDCO, Edwards Lifesciences, and Masimo. J-LV, PP, DP, SR, VMR, CI, and GDR declare that they have no competing interests.
Abkürzungen
CO
Cardiac output
CVC
Central venous catheter
CVP
Central venous pressure
DO2
Oxygen delivery
EVLW
Extravascular lung water
GEDV
Global end-diastolic volume
Hb
Hemoglobin
ICG
Indocyanine green
ICU
Intensive care unit
LV
Left ventricular
OR
Operating room
PAC
Pulmonary artery catheter
PACU
Post-anesthesia care unit
PetCO2
Partial pressure of end-tidal carbon dioxide
PLR
Passive leg raising
PPV
Pulse pressure variation
PVI
Pleth variability index
RV
Right ventricular
ScvO2
Central venous oxygen saturation
SV
Stroke volume
SvO2
Mixed venous oxygen saturation
SVV
Stroke volume variation
TEE
Transesophagel echocardiography
TTE
Transthoracic echocardiography
VO2
Oxygen consumption
Zt
Vascular impedance

Introduction

An estimated 230 million surgical procedures are performed each year around the world [1], and a significant number are in patients at risk of intra- or post-operative complications or both. Although less than 15% of in-patient procedures are performed in high-risk patients, such patients account for 80% of deaths [2-4]. Even for those patients who survive to leave hospital, post-operative complications remain an important determinant of functional recovery, long-term survival [5], and health-care costs. Thus, mitigation of these risks is important not only for the individual patient but also for health-care managers.
The risk of perioperative complications is related to various factors, including patient status and comorbidities, the type of surgery performed and its duration, the degree of urgency, the skills and experience of the operating and anesthetic teams, and the post-operative management. Insufficient tissue perfusion and cellular oxygenation due to hypovolemia, heart dysfunction or both is one of the leading causes of perioperative complications and poor outcomes [6-9]. Thus, effective fluid management to prevent and treat hypo/hypervolemia and titration of vasoactive drugs for heart dysfunction are crucial to maintain adequate oxygen delivery (DO2) and prevent fluid overload and its consequences [10-12]. Therefore, selecting the most appropriate hemodynamic monitoring device (for diagnosis and to guide therapies) may be an important first step in reducing the risk of complications. The aims of this review are to describe the available hemodynamic monitoring systems and to evaluate the most appropriate clinical setting for each.

Basic hemodynamic monitoring

Clinical examination remains an important initial step in the hemodynamic assessment of high-risk surgical patients. However, individual vital signs often lack the specificity and sensitivity that are needed to guide hemodynamic management. For example, blood pressure is a variable influenced by both cardiac output (CO) and vascular tone; hence, blood pressure can remain within the normal range in the presence of low-flow states, including hypovolemia, as a result of increased peripheral vascular resistance. Similarly, heart rate may fail to reflect the development of hypovolemia under anesthesia [13].
Combining and integrating parameters from various hemodynamic monitoring systems may help improve our understanding of hemodynamic status [14]. For example, the combination of arterial pressure and the partial pressure of end-tidal carbon dioxide (PetCO2) can help differentiate between vasodilation and low CO as a cause of hypotension (PetCO2 transiently decreases when CO decreases) and may prevent ‘reflex’ fluid administration whenever blood pressure decreases. Similarly, a reduction in the PetCO2 value for the same minute ventilation (in the absence of hypothermia) suggests decreased pulmonary blood flow (and thus CO) and may serve as a trigger for more advanced hemodynamic monitoring.

Arterial pressure

Continuous invasive measurement of arterial pressure helps identify the rapid fluctuations in arterial pressure that may occur in high-risk patients. Artifacts (over- or under-damping) should be carefully identified and eliminated, especially when systolic-diastolic components and waveform have to be analyzed. Non-invasive techniques for continuous measurement of blood pressure are usually performed in peripheral arteries and may become unreliable in case of vasoconstriction or low peripheral flow. Non-invasive assessment of pressure waveforms from more central measurement sites, such as the brachial artery, may be a valuable option in the future.

Central venous pressure

A central venous catheter (CVC) is often used for administration of fluids, vasopressors, and inotropes and for measurement of central venous pressure (CVP). Since transmural CVP is the only value related to right ventricular (RV) preload but is not commonly monitored, interpretation of CVP values must take into account intrathoracic pressure changes, which are largely influenced by mechanical ventilation. Thus, changes in CVP with concomitant CO variations give an indication of RV function and potential peripheral venous congestion, the latter of which is an important factor for organ perfusion [15]. In addition, careful checking of the CVP wave may help to diagnose tricuspid regurgitation with a ‘v’ wave during systole. When the CVP is low (<6 mm Hg) with a concomitant low CO, there is almost certainly some degree of hypovolemia. Although changes in CVP correlate poorly with changes in CO [16] (as for pulmonary artery occlusion pressure), they can be used to assess the dynamic response to a fluid challenge [17] and to diagnose severe hypovolemia or cardiac dysfunction or both, especially where other monitoring systems are not available.

Cardiac output monitoring

The perioperative period is characterized by large variations in whole body oxygen consumption (VO2). The main goal in this period is to maintain an adequate DO2 to meet the fluctuating tissue oxygen requirements. Global DO2 is determined by CO and the oxygen content of arterial blood, and so after correction of hypoxemia and anemia (topics that will not be dealt with here), maintenance of an adequate CO is the next logical step to improve DO2. There are various methods available for monitoring CO [18], although a survey indicated that CO is routinely monitored in high-risk surgical patients by only about 35% of practitioners in Europe and North America [19] (Table 1).
Table 1
What hemodynamic monitoring do you routinely use for the management of high-risk surgery patients? (Please mark all that apply)
 
ASA respondents
ESA respondents
(n = 237)
(n = 195)
Answer options
Response percentage
Response percentage
Invasive arterial pressure
95.4%
89.7%
Central venous pressure
72.6%
83.6%
Non-invasive arterial pressure
51.9%
53.8%
Cardiac output
35.4%
34.9%
Pulmonary capillary wedge pressure
30.8%
14.4%
Transesophageal echocardiography
28.3%
19.0%
Systolic pressure variation
20.3%
23.6%
Plethysmographic waveform variation
17.3%
17.9%
Pulse pressure variation
15.2%
25.6%
Mixed venous saturation (ScvO2)
14.3%
15.9%
Central venous saturation (SvO2)
12.7%
33.3%
Oxygen delivery (DO2)
6.3%
14.4%
Stroke volume variation
6.3%
21.5%
Near-infrared spectroscopy
4.6%
5.1%
Global end-diastolic volume
2.1%
8.2%
From [19]. ASA, American Society of Anesthesiology; ESA, European Society of Anaesthesiology.

Doppler echocardiography

Though difficult to use as a continuous monitor of CO with conventional probes, transthoracic (TTE) or transesophageal (TEE) echocardiography can provide immediate point-of-care assessment of acute hemodynamic changes in selected patients. Echo techniques can also help to visualize the lungs, but this is beyond the scope of this review. Obviously, it is not possible to use TEE in all types of surgery. In addition to the estimation of CO (usually easier with TEE than with TTE), Doppler echocardiographic examination can provide an indication of cardiac function because it allows visualization of the cardiac chambers, valves, and pericardium [20]. It also allows measurement of the ejected stroke volume (SV) and derived left ventricular (LV) function parameters.
TEE provides several views, including the following:
  • The LV short-axis view, which can be used to evaluate LV function. Calculation of the LV fractional area contraction, or the simpler ‘eyeballing method’, informs about the kinetic (contractile) state and the shape (volume) of the heart. Poor contractility may indicate that inotropic support could help, and ‘kissing’ of the papillary muscle may indicate the need for fluids if the right heart is functioning normally. The short-axis view may also be used to identify septal dyskinesia. The finding of an RV D-shape may suggest the presence of RV dysfunction/failure, indicating a non-adaptation to an acute increase in RV afterload (pulmonary embolism) or RV myocardial ischemia.
  • The four-chamber view, which can help in assessing LV and RV function by evaluation of the right-to-left size ratio (normal <0.6).
In more advanced echocardiographic evaluation, fluid status and fluid responsiveness can also be assessed in mechanically ventilated patients by means of the superior vena cava collapsibility index (TEE bicaval view) or inferior vena cava distensibility index (TTE subcostal view). In addition, echocardiography allows the rapid and reliable estimation of SV. Finally, there are particular and specific conditions in which diagnosis and treatment are strictly related to the echocardiographic examination (for example, pericardial effusion, valve disruptions, aortic dissection, and systolic anterior motion of the mitral valve).
A miniaturized, disposable monoplane TEE probe that can be left in place for up to 72 hours (ClariTEE™, ImaCor Inc., Garden City, NY, USA) was recently introduced and has the potential to provide ongoing qualitative cardiac assessment [21]. We believe that, where expert echocardiography skills are not available, training programs should be developed to ensure that clinicians taking care of the high-risk patient are familiar with at least the basic applications of TTE and TEE.

Pulmonary artery catheter

Though criticized in recent years for its intrinsic invasiveness and no clear evidence of improved outcomes [22-25], the pulmonary artery catheter (PAC) is the only tool that provides continuous monitoring of pulmonary artery pressure, right-sided and left-sided filling pressures, CO, and mixed venous oxygen saturation (SvO2). Although the PAC can now be largely replaced by less invasive hemodynamic monitoring techniques in many cases, in some complex clinical situations (for example, cardiac surgery, organ transplant surgery, and surgery associated with major fluid shifts or high risk of respiratory failure or in patients with compromised RV function), the PAC still represents a valuable tool when used by physicians adequately trained to correctly interpret and apply the data provided [26,27]. In such patients, the PAC can be inserted for limited periods of time and removed when no longer necessary.

Other cardiac output monitoring devices

Pulse contour analysis

SV can be estimated continuously by analysis of the arterial pressure waveform, usually derived from an indwelling arterial catheter or by a non-invasive finger pressure cuff. To calculate SV from a pressure trace, the algorithms used by these devices have to compensate for the overall impedance of the system on the basis of the estimation of compliance and resistance of the cardiovascular tree. In this regard, optimization of the input signal is imperative, and severe distortions of the arterial waveform (for example, severe arrhythmias and multiple ectopic beats) and inadequate response of fluid-filled transducer systems (that is, over- and under-damping) [28] can result in unreliable CO measurement.
Calibrated devices
  • The PiCCOplus™/PiCCO2™ system (Pulsion Medical Systems, Munich, Germany) consists of a thermistor-tipped catheter which is usually placed in the femoral artery, although catheters for radial, axillary, or brachial applications are also available. The PiCCO™ device measures CO by transpulmonary thermodilution, which additionally provides the computation of volumetric preload parameters—global end-diastolic volume (GEDV) and intrathoracic blood volume—and extravascular lung water (EVLW). The CO measured by the Stewart-Hamilton principle from the thermodilution curve is used to calibrate a pulse contour algorithm, which measures the area under the systolic pulse pressure curve and calculates the SV in order to provide beat-by-beat CO measurement. The system has to be frequently recalibrated, at least every 8 hours in hemodynamically stable patients and more often if changes in vasoactive support are provided [29]. The system has been validated in a variety of clinical settings [30].
  • The EV1000™/VolumeView™ system (Edwards Lifesciences, Irvine, CA, USA) has been more recently introduced and is analogous to the PiCCO™ monitor, using pulse wave analysis to calculate CO. A proprietary thermistor-tipped femoral artery catheter and a separate sensor are the main components of the system. This system requires calibration by transpulmonary thermodilution. It has been validated against the PiCCO™ and transpulmonary thermodilution in critically ill patients [31].
  • The LiDCO™plus system (LiDCO Ltd, Cambridge, UK) uses pulse power analysis to calculate SV and therefore is not technically a pulse contour device. The algorithm is based on the principle of conservation of mass (power), assuming a linear relationship between the net power change and the net flow in the vascular system. This system requires correction for vascular compliance, with calibration using a transpulmonary lithium indicator dilution technique performed via an indwelling arterial catheter. It has been validated in critically ill patients [32,33].
Uncalibrated devices (without external calibration)
With preloaded data
  • The PulsioFlex™ system (Pulsion Medical Systems) displays trends of estimated CO by using the patient’s anthropometric and demographic characteristics (necessary for internal calibration), analysis of the arterial pressure tracing, and a proprietary algorithm for data analysis. The system requires a dedicated additional sensor, which can be connected to a regular arterial pressure catheter. Based on the same pulse contour algorithm used by the PiCCO™, the device can be calibrated by entering a CO obtained from an external source (for example, Doppler echocardiography) or by the system’s own internal algorithm.
  • The LiDCO™rapid (LiDCO Ltd) device uses the same algorithm as the LiDCO™plus system, but instead of lithium dilution, nomograms based on the patient’s age, weight, and height are used to estimate SV and CO (so-called ‘nominal’ SV and CO). An externally estimated CO can be used to calibrate the device.
  • The FloTrac™/Vigileo™ system (Edwards Lifesciences) consists of a proprietary transducer (FloTrac™) connected to a standard (radial or femoral) arterial catheter. Individual demographic variables (age, sex, height, and weight) and a database containing CO variables derived by using the PAC are used to calculate impedance and a ‘normal’ SV against which the standard deviation of the pulse pressure sampled during a 20-second interval is correlated to estimate CO. Arterial waveform analysis is used to calculate vascular resistance and compliance. The algorithm used by the Vigileo™ device has been modified over time, and recent studies evaluating the device in the perioperative setting have shown an improved performance and a significant reduction in the time needed to adapt to vascular dynamics. In the intensive care unit (ICU) setting, concerns remain regarding the accuracy in situations of acute hemodynamic instability as well as hyperdynamic conditions, although recent software modifications seem to improve the reliability of CO measurements. The FloTrac™/Vigileo™ system has been shown to be suitable for integration into perioperative optimization protocols, resulting in improved clinical outcomes [34,35].
Without preloaded data
  • The MostCare system (Vytech, Padua, Italy), powered by the pressure recording analytical method, performs a beat-to-beat estimation of SV and CO by analyzing the pressure waveform, sampled at high resolution (1,000 points per second = 1 kHz). The area under the pressure wave is determined during the whole cardiac cycle. In each phase, the method identifies specific points (‘points of instability’) characterized by modifications in velocity and acceleration in relationship to the previous and the subsequent point. All of these ‘points of instability’, mainly caused by reflected waves from the periphery (backward travelling waves), give the arterial pulse its specific profile, which is analyzed by MostCare for estimation of the vascular impedance (Zt). The contribution of the reflected waves to the forward travelling wave can be accurately identified only with a very high sampling rate. The ability to update the Zt during each heart beat makes the system extremely reactive when abrupt changes in impedance occur (for example, changes in vascular tone) [36,37]. Although some promising clinical data are available [38], larger validation studies are needed to confirm these observations. A multicenter study comparing MostCare with echo-Doppler for CO measurement was recently completed (ClinicalTrials.gov identifier: NCT01678950).
Non-invasive pulse contour analysis
  • Recently, several newly marketed monitors have purported to track changes in arterial pressure non-invasively from finger probes. These include the continuous non-invasive arterial pressure probe (used with the LiDCO™ system) and the ClearSight device (Edwards Lifesciences). Based on the volume-clamp technique of Penaz [39] incorporated into a finger cuff, a brachial artery pressure waveform is reconstructed. The ClearSight monitor estimates CO on a beat-to-beat basis by dividing the integrated pulsatile systolic area by the aortic input impedance, which in turn is derived from a three-element Windkessel model incorporating patient-specific aortic mechanical characteristics [40]. Preliminary clinical validation studies, currently limited to a critical care setting and cardiac surgery, have shown acceptable agreements in CO when compared with thermodilution measurements performed with a PAC [41-44]. These monitors have the potential to track SV and CO in situations requiring early hemodynamic intervention when more invasive monitoring modalities are not readily available.

Doppler monitoring devices

  • Esophageal Doppler offers a minimally invasive determination of CO. The CardioQ™/CardioQ-ODM™ (Deltex Medical Ltd, Chichester, UK) is the most commonly used device. Esophageal probes measure blood flow in the descending part of the aorta. SV is calculated by multiplying the cross-sectional area of the aorta (from nomograms based on height, weight, and age) by the blood flow velocity. Technical and methodological concerns regarding probe positioning and the use of nomograms have been raised.
  • A transthoracic continuous Doppler CO monitor, the USCOM (Uscom, Sydney, Australia), is also available for intermittent SV and CO assessment. Using supra- or parasternal windows, flow velocity at the level of the aortic or pulmonary valves can be assessed non-invasively. Aortic diameter can be loaded from another measurement (two-dimensional echocardiography) or from a nomogram as mentioned above. The technique is rapid and can be used in adult and pediatric patients. Clinical validation studies have shown conflicting results, partially due to variations in signal acquisition inherent in the technique [45].

Applied Fick principle and dye dilution

  • The NICO™ System (Novametrix Medical Systems, Wallingford, CT, USA), using the partial CO2 rebreathing method, applies the Fick principle to CO2 in patients who are intubated and mechanically ventilated via a disposable rebreathing circuit that is added to the ventilator tubing. CO2 production is calculated as the product of CO2 concentration and airflow during a breathing cycle, and arterial CO2 content is derived from end-tidal CO2 and its corresponding dissociation curve. The rebreathing loop can induce an intermittent partial rebreathing state in 3-minute intervals. This rebreathing cycle results in an increased end-tidal CO2, mimicking a decrease in CO2 production. The differences in these values can be used to calculate CO. The system is not really clinically acceptable, because the assessment of CO is possible only in patients with fixed ventilator settings and good respiratory function with no relevant pulmonary shunt [46,47].
  • DDG-30® pulsed dye densitometry (Nihon Kohden, Tokyo, Japan) is based on the transpulmonary dye dilution technique using indocyanine green (ICG). Signal detection in the arterial blood is performed by transcutaneous optical absorbance measurements similar to pulse oximetry. CO is calculated from the ICG-dye dilution curve according to the Stewart-Hamilton principle. Factors that compromise signal detection, such as vasoconstriction, interstitial edema, movement, or ambient light artifacts, all limit the reliability of CO assessment using this method [48].

Bioimpedance and bioreactance

  • Electrical bioimpedance estimates CO continuously by detecting variations in thoracic or whole body impedance induced by cyclic changes in blood flow. Electrodes attached to the skin (BioZ, CardioDynamics, San Diego, CA, USA) or an endotracheal tube (ECOM™, Conmed Corporation, Utica, NY, USA) provide electric current stimulation, and signal variations are analyzed by using mathematical algorithms. The reliability of these systems is poor [49,50].
  • The Bioreactance® technique (NICOM®, Cheetah Medical Ltd, Maidenhead, Berkshire, UK) analyzes the variations in the frequency spectra associated with delivery of an oscillating electrical current. This technique may be somewhat superior to the bioimpedance technique [51,52] but is dependent on body size [53] and the aeration of the lung. It is less accurate in diseased lungs in which reactance may be affected by the amount of EVLW and alveolar collapse or consolidation or both.

Pitfalls in the interpretation of cardiac output

Although CO can be measured with reasonable accuracy and precision with some of these systems, it is difficult to assess the optimal CO for an individual patient. A ‘normal’ or even high CO does not preclude the presence of inadequate regional and microcirculatory flow, and a low CO may be adequate in a context of low metabolic demand, especially during surgery under general anesthesia. Moreover, simple identification of a low CO does not tell us what to do about it. To correctly interpret the data acquired by any of the described devices, we need to combine/integrate several variables to help decide whether the CO/SV is adequate and how it can be optimized in the most effective manner.

How to select the best system

All monitoring systems have unique characteristics in terms of accuracy, precision, validity, stability, and reliability [18]. Not all monitoring devices have been evaluated against the same set of criteria, and uncertainty remains regarding acceptance thresholds for the performance of CO monitors and the used reference techniques [54-57]. Clinicians must consider the technical limitations of each monitoring system and the potential trade-off between more invasive but highly accurate measurements of CO and less invasive but also less accurate modalities.
Several questions can be raised when considering choice of CO monitoring in the perioperative period:
1.
Are we ready to accept a less accurate measurement in order to limit invasiveness? (Figure 1). A less accurate measurement may be acceptable if the trend analysis is reliable. Cost may also be an important issue.
 
2.
Do we need continuous, semi-continuous, or intermittent measurements? Most complications after surgery do not have a sudden onset (except sudden cardiac failure due, for example, to myocardial infarction or pulmonary embolism) or an obvious cause (for example, massive bleeding during surgery) but develop slowly; therefore, semi-continuous or intermittent measurements may be acceptable. However, it should be noted that only beat-by-beat measurement of SV allows assessment of the response to preload-modifying maneuvers, such as a fluid challenge or passive leg raising (PLR) test.
 
3.
Are calibrated or uncalibrated systems preferable? Non-calibrated systems are acceptable for the operating room (OR) or the post-anesthesia care unit (PACU) but may not be suitable for more complex cases, especially in the ICU. In unstable patients, there is a necessity to ‘re-calibrate’ more often because of frequent changes in vascular tone and also because derived variables (for example, EVLW and GEDV) need to be re-calculated. A practical option may be to use an uncalibrated system in the OR/PACU and replace it with a calibrated system in the ICU.
 
4.
What alarms do we need? A major problem for patient surveillance by telemetric monitoring is artifact robustness. Any system with too many false alarms is prone to failure as personnel become desensitized.
 
5.
What kind of monitoring for what kind of patient? This is not a ‘one size fits all’ decision; rather, the optimal monitoring technique for each patient will vary depending on the degree of risk and the extent of the surgical procedure (Figure 2).
 

Fluid management

Inadequate fluid management may lead to reduced CO and DO2 to injured tissues, which is associated with an increased incidence of post-operative complications. Moreover, the systemic inflammatory response associated with tissue injury results in capillary leak and tissue edema. Fluid restriction and diuresis may decrease edema in patients with poor ventricular function but may also increase the incidence of acute kidney injury. Meanwhile, excessive fluid administration may lead to a range of adverse effects, including coagulopathy and edema of lungs, gut, and peripheral tissues (Figure 3). Retention of sodium and water following surgery may reduce fluid requirements. Once the patient is stabilized, fluids should be given only to correct deficit or continuing losses. Unfortunately, estimates of fluid deficit that are based on traditional physiological parameters, such as heart rate, blood pressure, and cardiac filling pressures, are not sufficient.

Static indicators of preload

  • CVP: Many high-risk surgical patients have a CVC in place, and a CVC is a requirement for some devices needing calibration by thermodilution. Despite its limitations (vide supra), changes in CVP over time may be helpful to guide fluid therapy, especially when the CVP is low and associated with low flow. A CVP of more than 8 mm Hg might also be considered an ‘alarm’ for potential venous congestion associated (or not) with fluid overload [15].
  • GEDV and EVLW: These are volumetric parameters derived from transpulmonary thermodilution and are integrated into the PiCCO™plus, PiCCO2™, and EV1000™ monitors. EVLW can help in the identification of (cardiogenic or non-cardiogenic) pulmonary edema and has the potential to increase the safety of fluid therapy in patients with structural lung disease, acute respiratory distress syndrome, or congestive heart failure.
  • The end-diastolic area of the left ventricle may be the most reliable static parameter of preload but is largely dependent on LV diastolic compliance. Its ability to accurately predict fluid responsiveness is limited.

Functional hemodynamic parameters

Positive pressure ventilation induces cyclical changes in intrathoracic pressure, which affect preload by decreasing venous return to the right heart and increasing venous return to the left ventricle. The degree of the resulting changes in LV SV (stroke volume variation; SVV) and pulse pressure (pulse pressure variation; PPV) better predict fluid responsiveness than do static parameters, when RV function is not a limitation and for a fixed tidal volume. Most devices using pulse contour analysis, including the current version of the non-invasive ClearSight monitor, display SVV and PPV. Despite the numerous validity criteria required to interpret such variations, these variables may help predict fluid responsiveness at different thresholds and have been integrated into hemodynamic optimization protocols [58].
Respiratory variations in the pulse oximeter plethysmographic waveform have been shown to predict fluid responsiveness in mechanically ventilated patients, similar to changes in the arterial pressure waveform [59]. The Masimo™ device (Masimo Corporation, Irvine, CA, USA) provides automated calculation of the pleth variability index (PVI) by measuring changes in perfusion index over a time interval including at least one complete respiratory cycle. The PVI has been shown to predict fluid responsiveness in various perioperative settings and has been integrated into fluid optimization algorithms. However, the PVI has the same limitations as the other dynamic parameters and has limited accuracy in the presence of vasoconstriction with or without the use of vasopressors [60-62].

Limitations

It is important to note that all of the dynamic variables have significant confounding factors [58]. The reliability of these indices is affected by spontaneous breathing activity, arrhythmias, right heart failure, decreased chest wall compliance, and increased intra-abdominal pressure, although most of these limitations are uncommon in the OR. Nevertheless, in the ICU, a relatively small proportion of patients present suitable criteria for these indices [63]. Another major limitation of dynamic parameters is that they are dependent on the size of the tidal volume. Some authors have suggested that they require a tidal volume of at least 8 mL/kg body weight [64], although they have been successfully used with tidal volumes of 6 to 8 mL/kg body weight [61,62]. A recent study and meta-analysis have indicated a decreased rate of post-operative complications when low tidal volumes are applied during anesthesia [65,66], and increased use of protective ventilation (lower tidal volumes) in the OR may reduce the usefulness of dynamic parameters or at least require new interpretation rules. Finally, within a range of PPV values of 9% to 13%, fluid responsiveness cannot always be reliably predicted; there is a ‘gray zone’ in which prediction of fluid responsiveness is difficult. One study [67] indicated that fluid responsiveness could not be reliably predicted by using dynamic measures in as many as 25% of anesthetized patients.
A PLR test has been suggested to overcome some of these limitations in dynamic evaluation but should be performed rigorously with simultaneous analysis of continuous CO monitoring. It is obviously impractical during most operative conditions [68]. In addition, the blood volume shift from the leg to the central compartment is non-predictable. In a hypovolemic state, it is reasonable to consider this volume shift less than that generated in ‘normal’ volemic conditions.
Despite these limitations and confounding factors, whenever possible, one is advised to assess fluid responsiveness by using the available functional hemodynamic parameters before attempting to increase CO with fluid administration. This approach can indicate if and when CO can be further increased by fluids and can identify when the flat portion of the cardiac function curve has been reached, thus preventing unnecessary fluid loading [58]. It is important to remember that, generally speaking, fluid responsiveness is not an (absolute) indication to give fluids. Decisions about fluid administration should be based not only on dynamic parameters but also on the likely risk associated with fluid administration. During surgery, systematic fluid administration in the presence of fluid responsiveness may improve post-operative outcomes [69].

Mixed venous oxygen saturation

Changes in SvO2 may reflect important pathophysiological changes in the relationship between DO2 and VO2, both of which may fluctuate significantly during the perioperative period.
Reorganization of the Fick equation shows that:
$$ \mathrm{S}\mathrm{v}{\mathrm{O}}_2 = \mathrm{S}\mathrm{a}{\mathrm{O}}_2\hbox{--}\ \left(\mathrm{V}{\mathrm{O}}_2/\left[\mathrm{C}\mathrm{O}\times \mathrm{H}\mathrm{b}\times \mathrm{C}\right]\right), $$
where C is the amount of oxygen bound to 1 g of hemoglobin (Hb). From this equation, it is clear that SvO2 will decrease in the presence of hypoxemia, hypermetabolic states (increased VO2), a decrease in CO, or anemia. Therefore, changes in SvO2 are directly proportional to those in CO but only when arterial oxygen saturation (SaO2), VO2, and Hb concentration remain constant. The SvO2 is around 75% in healthy patients but is closer to 70% in acutely ill patients who have a somewhat lower Hb concentration.
Central venous oxygen saturation (ScvO2) is used as a surrogate for SvO2 when a PAC is not in situ, but has some limitations. Although the determinants of ScvO2 and SvO2 are similar, they cannot be used interchangeably [70-73]. Regional variations in the balance between DO2 and VO2 result in differences in the Hb saturation of blood in the superior and inferior vena cavae [74]. ScvO2 is affected disproportionately by changes in the upper body and does not reflect the SvO2 of coronary sinus blood [74]. In healthy individuals, ScvO2 may be slightly less than SvO2 [75] because of the high oxygen content of effluent venous blood from the kidneys [76], but this relationship is reversed during periods of hemodynamic instability as blood is redistributed to the upper body at the expense of the splanchnic and renal circulations [77]. In shock states, therefore, ScvO2 may exceed SvO2 by up to 20% [72]. This lack of equivalence has been demonstrated in various groups of acutely ill patients, including not only those with shock [70,71,78] but also patients undergoing general anesthesia for cardiac [73,79] and non-cardiac [71,80] surgery. Even trends in ScvO2 do not closely reflect those of SvO2 [70,73,78].
Lower values of ScvO2 have been associated with more complications in patients undergoing cardiothoracic surgery [81]. Therefore, some authors have proposed to maintain SvO2 or ScvO2 above a cutoff value. In patients undergoing elective cardiac surgery, administration of intravenous fluid and inotropic therapy to attain a target SvO2 of at least 70% in the first 8 hours after surgery was associated with fewer complications and a shorter hospital stay [82]. In patients undergoing major abdominal (including aortic) surgery, achieving an oxygen extraction ratio of less than 27% (from intermittent measurements of ScvO2) was associated with a shorter hospital stay [83].
During surgery, this measurement is less informative: firstly, hypoxemia is generally corrected; secondly, under anesthesia, especially with neuromuscular paralysis, oxygen use decreases in all tissues, so that reductions in ScvO2 are uncommon [84]. Nevertheless, low ScvO2 values imply first and foremost that CO may be inadequate. At the same time, very high ScvO2 values may imply that oxygen extraction is low, purporting a worse prognosis, at least during cardiac surgery [85].

Blood lactate concentrations

Lactate is a physiological substrate (carbohydrate) produced from pyruvate reduction during cytosolic glycolysis. In stable conditions, lactate production and elimination are equivalent (that is, 1,200 to 1,500 mmol per day), leading to a stable blood lactate concentration of 0.8 to 1.2 mmol/L. The net flux of lactate depends on the difference between release and uptake and varies among organs and with their energetic conditions [86]. Hyperlactatemia is associated with increased morbidity and mortality in critically ill patients [87-90]. Persistent hyperlactatemia is a more relevant indicator of poor outcome than an isolated elevated lactate value is. Hyperlactatemia is not always a consequence of tissue hypoxia; sometimes, it stems from an accelerated ‘aerobic’ glycolysis resulting from cytokine influence and catecholamine stimulation, a situation termed ‘stress hyperlactatemia’. In practice, irrespective of the different metabolic modifications, an elevated lactate level indicates the presence of shock, and a decrease in lactate levels over time is a good indicator of effective treatment. Accordingly, repeated blood lactate measurements are recommended to monitor lactate production and clearance over time during surgery in high-risk patients.

Management strategies based on perioperative monitoring

There is good evidence that use of flow-based hemodynamic monitoring combined with hemodynamic manipulation in the perioperative period can reduce morbidity and sometimes mortality [83,91-97]. For a variety of reasons, however, this approach has not been adopted everywhere and has even been challenged [98]. Indeed, there have been some important problems with many clinical trials in the field, such as lack of blinding and suboptimal management of the control group.
There are basically two options to optimize perioperative cardiovascular management (Table 2), both of which aim to increase SV/CO by means of fluid loading (increase in cardiac preload) or inotrope administration (increase in contractility) or both:
  • One option is reactive, by applying a rapid intervention only when a hemodynamic change occurs. One should then individualize treatment with fluid challenge techniques. The response to the rapid administration of a fluid bolus (for example, 250 mL) can be evaluated during surgery (especially in the presence of signs of fluid responsiveness). The response can be monitored by evaluating the blood pressure or heart rate, but the CO/SV response is much more accurate. Inotropic agents are added in the absence of an adequate response.
  • The other option is pro-active and is based on a strategy of hemodynamic manipulation targeting supranormal CO or DO2 values to minimize the risk of tissue hypoperfusion. Adequate fluid administration is the first element of this strategy. Several studies have indicated that fluid management based on PPV, SVV, and SV optimization may decrease post-operative wound infections and possibly post-operative organ dysfunction [99,100]. Inotropic agents may be added if fluids alone are not sufficient for this purpose. There is a risk of overtreatment as excessive use of dobutamine has been associated with increased rates of complications [101]. The use of dopexamine as an alternative has given controversial results [102,103].
Table 2
Options to optimize perioperative hemodynamic management in high-risk patients
Reactive
Correct hypotension, tachycardia.
Give fluids in the presence of suspected hypovolemia with increased pulse pressure variation (PPV), systolic pressure variation, stroke volume variation (SVV), or pleth variability index (PVI).
Identify a reduction in cardiac output and react promptly with fluid challenge.
Identify a reduction in central venous oxygen saturation (ScvO2) and react promptly with fluid challenge.
Pro-active
Maintain arterial pressure and heart rate within acceptable ranges.
Maximize stroke volume.
Maintain PPV or SVV at less than 12% or PVI at less than 14%.
Maintain cardiac index (CI) or oxygen delivery (DO2) in a desired range (for example, CI of more than 4.5 L/minute/m2 and DO2 of more than 600 mL/minute/m2).
Maintain ScvO2 at more than 65%.

Conclusions

Cardiovascular monitoring systems play an important role in optimizing perioperative hemodynamic management. Use of hemodynamic monitoring devices per se in the perioperative setting has not been linked to improved outcomes; however, appropriate measurement and interpretation of cardiovascular variables may help guide therapeutic interventions, which in turn can improve patient outcomes. The most appropriate system must be selected for the individual patient prior to surgery, taking into consideration the individual risks of the patient and the procedure. Appropriate interpretation of the information offered by hemodynamic monitoring requires the integration of several variables. Echocardiography is increasingly used as a first tool to identify a problem and help select initial treatment. To improve patient management and outcome, the clinician must understand the advantages and the limitations of the various tools and parameters used during perioperative care.

Competing interests

RP has received equipment loans from LiDCO Ltd and has performed consultancy work for Edwards Lifesciences, Covidien (Dublin, Ireland), and Masimo. AP has received advisory board fees from Pulsion Medical Systems. AH has received lecture honoraria from Edwards Lifesciences and is an advisory board member without fees for UPmed (Munich, Germany). PF has received honoraria from Masimo for presentations in congresses. AR has received lecture and advisory board fees from LiDCO, Edwards Lifesciences, and Masimo. J-LV, PP, DP, SR, VMR, CI, and GDR declare that they have no competing interests.
Literatur
1.
Zurück zum Zitat Weiser TG, Regenbogen SE, Thompson KD, Haynes AB, Lipsitz SR, Berry WR, et al. An estimation of the global volume of surgery: a modelling strategy based on available data. Lancet. 2008;372:139–44.PubMed Weiser TG, Regenbogen SE, Thompson KD, Haynes AB, Lipsitz SR, Berry WR, et al. An estimation of the global volume of surgery: a modelling strategy based on available data. Lancet. 2008;372:139–44.PubMed
2.
Zurück zum Zitat Jhanji S, Thomas B, Ely A, Watson D, Hinds CJ, Pearse RM. Mortality and utilisation of critical care resources amongst high-risk surgical patients in a large NHS trust. Anaesthesia. 2008;63:695–700.PubMed Jhanji S, Thomas B, Ely A, Watson D, Hinds CJ, Pearse RM. Mortality and utilisation of critical care resources amongst high-risk surgical patients in a large NHS trust. Anaesthesia. 2008;63:695–700.PubMed
3.
Zurück zum Zitat Pearse RM, Harrison DA, James P, Watson D, Hinds C, Rhodes A, et al. Identification and characterisation of the high-risk surgical population in the UK. Crit Care. 2006;10:R81.PubMedPubMedCentral Pearse RM, Harrison DA, James P, Watson D, Hinds C, Rhodes A, et al. Identification and characterisation of the high-risk surgical population in the UK. Crit Care. 2006;10:R81.PubMedPubMedCentral
4.
Zurück zum Zitat Lobo SM, de Oliveira NE. Clinical review: What are the best hemodynamic targets for noncardiac surgical patients? Crit Care. 2013;17:210.PubMedPubMedCentral Lobo SM, de Oliveira NE. Clinical review: What are the best hemodynamic targets for noncardiac surgical patients? Crit Care. 2013;17:210.PubMedPubMedCentral
5.
Zurück zum Zitat Khuri SF, Henderson WG, DePalma RG, Mosca C, Healey NA, Kumbhani DJ. Determinants of long-term survival after major surgery and the adverse effect of postoperative complications. Ann Surg. 2005;242:326–41.PubMedPubMedCentral Khuri SF, Henderson WG, DePalma RG, Mosca C, Healey NA, Kumbhani DJ. Determinants of long-term survival after major surgery and the adverse effect of postoperative complications. Ann Surg. 2005;242:326–41.PubMedPubMedCentral
6.
Zurück zum Zitat Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112:1392–402.PubMed Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112:1392–402.PubMed
7.
Zurück zum Zitat Gurgel ST. do Nascimento P Jr. Maintaining tissue perfusion in high-risk surgical patients: a systematic review of randomized clinical trials. Anesth Analg. 2011;112:1384–91.PubMed Gurgel ST. do Nascimento P Jr. Maintaining tissue perfusion in high-risk surgical patients: a systematic review of randomized clinical trials. Anesth Analg. 2011;112:1384–91.PubMed
8.
Zurück zum Zitat Cecconi M, Corredor C, Arulkumaran N, Abuella G, Ball J, Grounds RM, et al. Clinical review: Goal-directed therapy-what is the evidence in surgical patients? The effect on different risk groups. Crit Care. 2013;17:209.PubMedPubMedCentral Cecconi M, Corredor C, Arulkumaran N, Abuella G, Ball J, Grounds RM, et al. Clinical review: Goal-directed therapy-what is the evidence in surgical patients? The effect on different risk groups. Crit Care. 2013;17:209.PubMedPubMedCentral
9.
Zurück zum Zitat Jhanji S, Lee C, Watson D, Hinds C, Pearse RM. Microvascular flow and tissue oxygenation after major abdominal surgery: association with post-operative complications. Intensive Care Med. 2009;35:671–7.PubMed Jhanji S, Lee C, Watson D, Hinds C, Pearse RM. Microvascular flow and tissue oxygenation after major abdominal surgery: association with post-operative complications. Intensive Care Med. 2009;35:671–7.PubMed
10.
Zurück zum Zitat Marjanovic G, Villain C, Juettner E, zur Hausen A, Hoeppner J, Hopt UT, et al. Impact of different crystalloid volume regimes on intestinal anastomotic stability. Ann Surg. 2009;249:181–5.PubMed Marjanovic G, Villain C, Juettner E, zur Hausen A, Hoeppner J, Hopt UT, et al. Impact of different crystalloid volume regimes on intestinal anastomotic stability. Ann Surg. 2009;249:181–5.PubMed
11.
Zurück zum Zitat Kulemann B, Timme S, Seifert G, Holzner PA, Glatz T, Sick O, et al. Intraoperative crystalloid overload leads to substantial inflammatory infiltration of intestinal anastomoses - a histomorphological analysis. Surgery. 2013;154:596–603.PubMed Kulemann B, Timme S, Seifert G, Holzner PA, Glatz T, Sick O, et al. Intraoperative crystalloid overload leads to substantial inflammatory infiltration of intestinal anastomoses - a histomorphological analysis. Surgery. 2013;154:596–603.PubMed
12.
Zurück zum Zitat Nessim C, Sideris L, Turcotte S, Vafiadis P, Lapostole AC, Simard S, et al. The effect of fluid overload in the presence of an epidural on the strength of colonic anastomoses. J Surg Res. 2013;183:567–73.PubMed Nessim C, Sideris L, Turcotte S, Vafiadis P, Lapostole AC, Simard S, et al. The effect of fluid overload in the presence of an epidural on the strength of colonic anastomoses. J Surg Res. 2013;183:567–73.PubMed
13.
Zurück zum Zitat Pizov R, Eden A, Bystritski D, Kalina E, Tamir A, Gelman S. Hypotension during gradual blood loss: waveform variables response and absence of tachycardia. Br J Anaesth. 2012;109:911–8.PubMed Pizov R, Eden A, Bystritski D, Kalina E, Tamir A, Gelman S. Hypotension during gradual blood loss: waveform variables response and absence of tachycardia. Br J Anaesth. 2012;109:911–8.PubMed
14.
Zurück zum Zitat Vincent JL, Rhodes A, Perel A, Martin GS, Della Rocca G, Vallet B, et al. Clinical review: update on hemodynamic monitoring - a consensus of 16. Crit Care. 2011;15:229.PubMedPubMedCentral Vincent JL, Rhodes A, Perel A, Martin GS, Della Rocca G, Vallet B, et al. Clinical review: update on hemodynamic monitoring - a consensus of 16. Crit Care. 2011;15:229.PubMedPubMedCentral
15.
Zurück zum Zitat Legrand M, Dupuis C, Simon C, Gayat E, Mateo J, Lukaszewicz AC, et al. Association between systemic hemodynamics and septic acute kidney injury in critically ill patients: a retrospective observational study. Crit Care. 2013;17:R278.PubMedPubMedCentral Legrand M, Dupuis C, Simon C, Gayat E, Mateo J, Lukaszewicz AC, et al. Association between systemic hemodynamics and septic acute kidney injury in critically ill patients: a retrospective observational study. Crit Care. 2013;17:R278.PubMedPubMedCentral
16.
Zurück zum Zitat Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134:172–8.PubMed Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134:172–8.PubMed
17.
Zurück zum Zitat Vincent JL, Weil MH. Fluid challenge revisited. Crit Care Med. 2006;34:1333–7.PubMed Vincent JL, Weil MH. Fluid challenge revisited. Crit Care Med. 2006;34:1333–7.PubMed
18.
Zurück zum Zitat Thiele RH, Bartels K, Gan TJ. Cardiac output monitoring: a contemporary assessment and review. Crit Care Med. 2015;43:177–85.PubMed Thiele RH, Bartels K, Gan TJ. Cardiac output monitoring: a contemporary assessment and review. Crit Care Med. 2015;43:177–85.PubMed
19.
Zurück zum Zitat Cannesson M, Pestel G, Ricks C, Hoeft A, Perel A. Hemodynamic monitoring and management in patients undergoing high risk surgery: a survey among North American and European anesthesiologists. Crit Care. 2011;15:R197.PubMedPubMedCentral Cannesson M, Pestel G, Ricks C, Hoeft A, Perel A. Hemodynamic monitoring and management in patients undergoing high risk surgery: a survey among North American and European anesthesiologists. Crit Care. 2011;15:R197.PubMedPubMedCentral
20.
Zurück zum Zitat Repesse X, Bodson L, Vieillard-Baron A. Doppler echocardiography in shocked patients. Curr Opin Crit Care. 2013;19:221–7.PubMed Repesse X, Bodson L, Vieillard-Baron A. Doppler echocardiography in shocked patients. Curr Opin Crit Care. 2013;19:221–7.PubMed
21.
Zurück zum Zitat Maltais S, Costello WT, Billings FT, Bick JS, Byrne JG, Ahmad RM, et al. Episodic monoplane transesophageal echocardiography impacts postoperative management of the cardiac surgery patient. J Cardiothorac Vasc Anesth. 2013;27:665–9.PubMed Maltais S, Costello WT, Billings FT, Bick JS, Byrne JG, Ahmad RM, et al. Episodic monoplane transesophageal echocardiography impacts postoperative management of the cardiac surgery patient. J Cardiothorac Vasc Anesth. 2013;27:665–9.PubMed
22.
Zurück zum Zitat Rhodes A, Cusack RJ, Newman PJ, Grounds RM, Bennett ED. A randomised, controlled trial of the pulmonary artery catheter in critically ill patients. Intensive Care Med. 2002;28:256–64.PubMed Rhodes A, Cusack RJ, Newman PJ, Grounds RM, Bennett ED. A randomised, controlled trial of the pulmonary artery catheter in critically ill patients. Intensive Care Med. 2002;28:256–64.PubMed
23.
Zurück zum Zitat Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, et al. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet. 2005;366:472–7.PubMed Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, et al. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet. 2005;366:472–7.PubMed
24.
Zurück zum Zitat Harvey S, Young D, Brampton W, Cooper AB, Doig G, Sibbald W, et al. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev. 2006;3:CD003408. Harvey S, Young D, Brampton W, Cooper AB, Doig G, Sibbald W, et al. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev. 2006;3:CD003408.
25.
Zurück zum Zitat Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G, et al. Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA. 2005;294:1664–70.PubMed Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G, et al. Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA. 2005;294:1664–70.PubMed
26.
Zurück zum Zitat Vincent JL, Pinsky MR, Sprung CL, Levy M, Marini JJ, Payen D, et al. The pulmonary artery catheter: in medio virtus. Crit Care Med. 2008;36:3093–6.PubMed Vincent JL, Pinsky MR, Sprung CL, Levy M, Marini JJ, Payen D, et al. The pulmonary artery catheter: in medio virtus. Crit Care Med. 2008;36:3093–6.PubMed
27.
Zurück zum Zitat Vincent JL. The pulmonary artery catheter. J Clin Monit Comput. 2012;26:341–5.PubMed Vincent JL. The pulmonary artery catheter. J Clin Monit Comput. 2012;26:341–5.PubMed
28.
Zurück zum Zitat Gardner RM. Direct blood pressure measurement - dynamic response requirements. Anesthesiology. 1981;54:227–36.PubMed Gardner RM. Direct blood pressure measurement - dynamic response requirements. Anesthesiology. 1981;54:227–36.PubMed
29.
Zurück zum Zitat Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL. Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med. 2008;36:434–40.PubMed Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL. Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med. 2008;36:434–40.PubMed
30.
Zurück zum Zitat Oren-Grinberg A. The PiCCO Monitor. Int Anesthesiol Clin. 2010;48:57–85.PubMed Oren-Grinberg A. The PiCCO Monitor. Int Anesthesiol Clin. 2010;48:57–85.PubMed
31.
Zurück zum Zitat Bendjelid K, Marx G, Kiefer N, Simon TP, Geisen M, Hoeft A, et al. Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients. Br J Anaesth. 2013;111:573–9.PubMed Bendjelid K, Marx G, Kiefer N, Simon TP, Geisen M, Hoeft A, et al. Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients. Br J Anaesth. 2013;111:573–9.PubMed
32.
Zurück zum Zitat Cecconi M, Fawcett J, Grounds RM, Rhodes A. A prospective study to evaluate the accuracy of pulse power analysis to monitor cardiac output in critically ill patients. BMC Anesthesiol. 2008;8:3.PubMedPubMedCentral Cecconi M, Fawcett J, Grounds RM, Rhodes A. A prospective study to evaluate the accuracy of pulse power analysis to monitor cardiac output in critically ill patients. BMC Anesthesiol. 2008;8:3.PubMedPubMedCentral
33.
Zurück zum Zitat Cecconi M, Dawson D, Grounds RM, Rhodes A. Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med. 2009;35:498–504.PubMed Cecconi M, Dawson D, Grounds RM, Rhodes A. Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med. 2009;35:498–504.PubMed
34.
Zurück zum Zitat Senn A, Button D, Zollinger A, Hofer CK. Assessment of cardiac output changes using a modified FloTrac/Vigileo algorithm in cardiac surgery patients. Crit Care. 2009;13:R32.PubMedPubMedCentral Senn A, Button D, Zollinger A, Hofer CK. Assessment of cardiac output changes using a modified FloTrac/Vigileo algorithm in cardiac surgery patients. Crit Care. 2009;13:R32.PubMedPubMedCentral
35.
Zurück zum Zitat Cecconi M, Fasano N, Langiano N, Divella M, Costa MG, Rhodes A, et al. Goal-directed haemodynamic therapy during elective total hip arthroplasty under regional anaesthesia. Crit Care. 2011;15:R132.PubMedPubMedCentral Cecconi M, Fasano N, Langiano N, Divella M, Costa MG, Rhodes A, et al. Goal-directed haemodynamic therapy during elective total hip arthroplasty under regional anaesthesia. Crit Care. 2011;15:R132.PubMedPubMedCentral
36.
Zurück zum Zitat Romano SM, Pistolesi M. Assessment of cardiac output from systemic arterial pressure in humans. Crit Care Med. 2002;30:1834–41.PubMed Romano SM, Pistolesi M. Assessment of cardiac output from systemic arterial pressure in humans. Crit Care Med. 2002;30:1834–41.PubMed
37.
Zurück zum Zitat Scolletta S, Bodson L, Donadello K, Taccone FS, Devigili A, Vincent JL, et al. Assessment of left ventricular function by pulse wave analysis in critically ill patients. Intensive Care Med. 2013;39:1025–33.PubMed Scolletta S, Bodson L, Donadello K, Taccone FS, Devigili A, Vincent JL, et al. Assessment of left ventricular function by pulse wave analysis in critically ill patients. Intensive Care Med. 2013;39:1025–33.PubMed
38.
Zurück zum Zitat Romagnoli S, Romano SM, Bevilacqua S, Ciappi F, Lazzeri C, Peris A, et al. Cardiac output by arterial pulse contour: reliability under hemodynamic derangements. Interact Cardiovasc Thorac Surg. 2009;8:642–6.PubMed Romagnoli S, Romano SM, Bevilacqua S, Ciappi F, Lazzeri C, Peris A, et al. Cardiac output by arterial pulse contour: reliability under hemodynamic derangements. Interact Cardiovasc Thorac Surg. 2009;8:642–6.PubMed
39.
Zurück zum Zitat Penaz J. Criteria for set point estimation in the volume clamp method of blood pressure measurement. Physiol Res. 1992;41:5–10.PubMed Penaz J. Criteria for set point estimation in the volume clamp method of blood pressure measurement. Physiol Res. 1992;41:5–10.PubMed
40.
Zurück zum Zitat Westerhof N, Lankhaar JW, Westerhof BE. The arterial Windkessel. Med Biol Eng Comput. 2009;47:131–41.PubMed Westerhof N, Lankhaar JW, Westerhof BE. The arterial Windkessel. Med Biol Eng Comput. 2009;47:131–41.PubMed
41.
Zurück zum Zitat Stover JF, Stocker R, Lenherr R, Neff TA, Cottini SR, Zoller B, et al. Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients. BMC Anesthesiol. 2009;9:6.PubMedPubMedCentral Stover JF, Stocker R, Lenherr R, Neff TA, Cottini SR, Zoller B, et al. Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients. BMC Anesthesiol. 2009;9:6.PubMedPubMedCentral
42.
Zurück zum Zitat Bogert LW, Wesseling KH, Schraa O, Van Lieshout EJ, de Mol BA, van Goudoever J, et al. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia. 2010;65:1119–25.PubMed Bogert LW, Wesseling KH, Schraa O, Van Lieshout EJ, de Mol BA, van Goudoever J, et al. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia. 2010;65:1119–25.PubMed
43.
Zurück zum Zitat Broch O, Renner J, Gruenewald M, Meybohm P, Schottler J, Caliebe A, et al. A comparison of the Nexfin(R) and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgery. Anaesthesia. 2012;67:377–83.PubMed Broch O, Renner J, Gruenewald M, Meybohm P, Schottler J, Caliebe A, et al. A comparison of the Nexfin(R) and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgery. Anaesthesia. 2012;67:377–83.PubMed
44.
Zurück zum Zitat Bubenek-Turconi SI, Craciun M, Miclea I, Perel A. Noninvasive continuous cardiac output by the Nexfin before and after preload-modifying maneuvers: a comparison with intermittent thermodilution cardiac output. Anesth Analg. 2013;117:366–72.PubMed Bubenek-Turconi SI, Craciun M, Miclea I, Perel A. Noninvasive continuous cardiac output by the Nexfin before and after preload-modifying maneuvers: a comparison with intermittent thermodilution cardiac output. Anesth Analg. 2013;117:366–72.PubMed
45.
Zurück zum Zitat Thom O, Taylor DM, Wolfe RE, Cade J, Myles P, Krum H, et al. Comparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter. Br J Anaesth. 2009;103:800–4.PubMed Thom O, Taylor DM, Wolfe RE, Cade J, Myles P, Krum H, et al. Comparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter. Br J Anaesth. 2009;103:800–4.PubMed
46.
Zurück zum Zitat Gueret G, Kiss G, Rossignol B, Bezon E, Wargnier JP, Miossec A, et al. Cardiac output measurements in off-pump coronary surgery: comparison between NICO and the Swan-Ganz catheter. Eur J Anaesthesiol. 2006;23:848–54.PubMed Gueret G, Kiss G, Rossignol B, Bezon E, Wargnier JP, Miossec A, et al. Cardiac output measurements in off-pump coronary surgery: comparison between NICO and the Swan-Ganz catheter. Eur J Anaesthesiol. 2006;23:848–54.PubMed
47.
Zurück zum Zitat Tachibana K, Imanaka H, Takeuchi M, Takauchi Y, Miyano H, Nishimura M. Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present. Anesthesiology. 2003;98:830–7.PubMed Tachibana K, Imanaka H, Takeuchi M, Takauchi Y, Miyano H, Nishimura M. Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present. Anesthesiology. 2003;98:830–7.PubMed
48.
Zurück zum Zitat Hofer CK, Buhlmann S, Klaghofer R, Genoni M, Zollinger A. Pulsed dye densitometry with two different sensor types for cardiac output measurement after cardiac surgery: a comparison with the thermodilution technique. Acta Anaesthesiol Scand. 2004;48:653–7.PubMed Hofer CK, Buhlmann S, Klaghofer R, Genoni M, Zollinger A. Pulsed dye densitometry with two different sensor types for cardiac output measurement after cardiac surgery: a comparison with the thermodilution technique. Acta Anaesthesiol Scand. 2004;48:653–7.PubMed
49.
Zurück zum Zitat Ball TR, Culp BC, Patel V, Gloyna DF, Ciceri DP, Culp Jr WC. Comparison of the endotracheal cardiac output monitor to thermodilution in cardiac surgery patients. J Cardiothorac Vasc Anesth. 2010;24:762–6.PubMed Ball TR, Culp BC, Patel V, Gloyna DF, Ciceri DP, Culp Jr WC. Comparison of the endotracheal cardiac output monitor to thermodilution in cardiac surgery patients. J Cardiothorac Vasc Anesth. 2010;24:762–6.PubMed
50.
Zurück zum Zitat Gujjar AR, Muralidhar K, Banakal S, Gupta R, Sathyaprabha TN, Jairaj PS. Non-invasive cardiac output by transthoracic electrical bioimpedence in post-cardiac surgery patients: comparison with thermodilution method. J Clin Monit Comput. 2008;22:175–80.PubMed Gujjar AR, Muralidhar K, Banakal S, Gupta R, Sathyaprabha TN, Jairaj PS. Non-invasive cardiac output by transthoracic electrical bioimpedence in post-cardiac surgery patients: comparison with thermodilution method. J Clin Monit Comput. 2008;22:175–80.PubMed
51.
Zurück zum Zitat Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C. Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med. 2007;33:1191–4.PubMed Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C. Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med. 2007;33:1191–4.PubMed
52.
Zurück zum Zitat Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22:113–9.PubMed Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22:113–9.PubMed
53.
Zurück zum Zitat Garisto C, Favia I, Ricci Z, Romagnoli S, Haiberger R, Polito A, et al. Pressure recording analytical method and bioreactance for stroke volume index monitoring during pediatric cardiac surgery. Paediatr Anaesth. 2015;25:143–9.PubMed Garisto C, Favia I, Ricci Z, Romagnoli S, Haiberger R, Polito A, et al. Pressure recording analytical method and bioreactance for stroke volume index monitoring during pediatric cardiac surgery. Paediatr Anaesth. 2015;25:143–9.PubMed
54.
Zurück zum Zitat Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.PubMed Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.PubMed
55.
Zurück zum Zitat Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies - with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201.PubMedPubMedCentral Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies - with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201.PubMedPubMedCentral
56.
Zurück zum Zitat Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD. Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009;35:1801–8.PubMed Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD. Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009;35:1801–8.PubMed
57.
Zurück zum Zitat Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.PubMed Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.PubMed
58.
Zurück zum Zitat Perel A, Habicher M, Sander M. Bench-to-bedside review: Functional hemodynamics during surgery - should it be used for all high-risk cases? Crit Care. 2013;17:203.PubMedPubMedCentral Perel A, Habicher M, Sander M. Bench-to-bedside review: Functional hemodynamics during surgery - should it be used for all high-risk cases? Crit Care. 2013;17:203.PubMedPubMedCentral
59.
Zurück zum Zitat Desebbe O, Cannesson M. Using ventilation-induced plethysmographic variations to optimize patient fluid status. Curr Opin Anaesthesiol. 2008;21:772–8.PubMed Desebbe O, Cannesson M. Using ventilation-induced plethysmographic variations to optimize patient fluid status. Curr Opin Anaesthesiol. 2008;21:772–8.PubMed
60.
Zurück zum Zitat Sandroni C, Cavallaro F, Marano C, Falcone C, De Santis P, Antonelli M. Accuracy of plethysmographic indices as predictors of fluid responsiveness in mechanically ventilated adults: a systematic review and meta-analysis. Intensive Care Med. 2012;38:1429–37.PubMed Sandroni C, Cavallaro F, Marano C, Falcone C, De Santis P, Antonelli M. Accuracy of plethysmographic indices as predictors of fluid responsiveness in mechanically ventilated adults: a systematic review and meta-analysis. Intensive Care Med. 2012;38:1429–37.PubMed
61.
Zurück zum Zitat Forget P, Lois F, de Kock M. Goal-directed fluid management based on the pulse oximeter-derived pleth variability index reduces lactate levels and improves fluid management. Anesth Analg. 2010;111:910–4.PubMed Forget P, Lois F, de Kock M. Goal-directed fluid management based on the pulse oximeter-derived pleth variability index reduces lactate levels and improves fluid management. Anesth Analg. 2010;111:910–4.PubMed
62.
Zurück zum Zitat Forget P, Lois F, Kartheuser A, Leonard D, Remue C, de Kock M. The concept of titration can be transposed to fluid management but does is change the volumes? Randomised trial on pleth variability index during fast-track colonic surgery. Curr Clin Pharmacol. 2013;8:110–4.PubMed Forget P, Lois F, Kartheuser A, Leonard D, Remue C, de Kock M. The concept of titration can be transposed to fluid management but does is change the volumes? Randomised trial on pleth variability index during fast-track colonic surgery. Curr Clin Pharmacol. 2013;8:110–4.PubMed
63.
Zurück zum Zitat Mahjoub Y, Lejeune V, Muller L, Perbet S, Zieleskiewicz L, Bart F, et al. Evaluation of pulse pressure variation validity criteria in critically ill patients: a prospective observational multicentre point-prevalence study. Br J Anaesth. 2014;112:681–5.PubMed Mahjoub Y, Lejeune V, Muller L, Perbet S, Zieleskiewicz L, Bart F, et al. Evaluation of pulse pressure variation validity criteria in critically ill patients: a prospective observational multicentre point-prevalence study. Br J Anaesth. 2014;112:681–5.PubMed
64.
Zurück zum Zitat Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37:2642–7.PubMed Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37:2642–7.PubMed
65.
Zurück zum Zitat Futier E, Constantin JM, Paugam-Burtz C, Pascal J, Eurin M, Neuschwander A, et al. A trial of intraoperative low-tidal-volume ventilation in abdominal surgery. N Engl J Med. 2013;369:428–37.PubMed Futier E, Constantin JM, Paugam-Burtz C, Pascal J, Eurin M, Neuschwander A, et al. A trial of intraoperative low-tidal-volume ventilation in abdominal surgery. N Engl J Med. 2013;369:428–37.PubMed
66.
Zurück zum Zitat Serpa Neto A, Cardoso SO, Manetta JA, Pereira VG, Esposito DC, Pasqualucci Mde O, et al. Association between use of lung-protective ventilation with lower tidal volumes and clinical outcomes among patients without acute respiratory distress syndrome: a meta-analysis. JAMA. 2012;308:1651–9.PubMed Serpa Neto A, Cardoso SO, Manetta JA, Pereira VG, Esposito DC, Pasqualucci Mde O, et al. Association between use of lung-protective ventilation with lower tidal volumes and clinical outcomes among patients without acute respiratory distress syndrome: a meta-analysis. JAMA. 2012;308:1651–9.PubMed
67.
Zurück zum Zitat Cannesson M, Le Manach Y, Hofer CK, Goarin JP, Lehot JJ, Vallet B, et al. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a ‘gray zone’ approach. Anesthesiology. 2011;115:231–41.PubMed Cannesson M, Le Manach Y, Hofer CK, Goarin JP, Lehot JJ, Vallet B, et al. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a ‘gray zone’ approach. Anesthesiology. 2011;115:231–41.PubMed
68.
Zurück zum Zitat Monnet X, Teboul JL. Passive leg raising. Intensive Care Med. 2008;34:659–63.PubMed Monnet X, Teboul JL. Passive leg raising. Intensive Care Med. 2008;34:659–63.PubMed
70.
Zurück zum Zitat Chawla LS, Zia H, Gutierrez G, Katz NM, Seneff MG, Shah M. Lack of equivalence between central and mixed venous oxygen saturation. Chest. 2004;126:1891–6.PubMed Chawla LS, Zia H, Gutierrez G, Katz NM, Seneff MG, Shah M. Lack of equivalence between central and mixed venous oxygen saturation. Chest. 2004;126:1891–6.PubMed
71.
Zurück zum Zitat Dueck MH, Klimek M, Appenrodt S, Weigand C, Boerner U. Trends but not individual values of central venous oxygen saturation agree with mixed venous oxygen saturation during varying hemodynamic conditions. Anesthesiology. 2005;103:249–57.PubMed Dueck MH, Klimek M, Appenrodt S, Weigand C, Boerner U. Trends but not individual values of central venous oxygen saturation agree with mixed venous oxygen saturation during varying hemodynamic conditions. Anesthesiology. 2005;103:249–57.PubMed
72.
Zurück zum Zitat Reinhart K, Rudolph T, Bredle DL, Hannemann L, Cain SM. Comparison of central-venous to mixed-venous oxygen saturation during changes in oxygen supply/demand. Chest. 1989;95:1216–21.PubMed Reinhart K, Rudolph T, Bredle DL, Hannemann L, Cain SM. Comparison of central-venous to mixed-venous oxygen saturation during changes in oxygen supply/demand. Chest. 1989;95:1216–21.PubMed
73.
Zurück zum Zitat Lorentzen AG, Lindskov C, Sloth E, Jakobsen CJ. Central venous oxygen saturation cannot replace mixed venous saturation in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth. 2008;22:853–7.PubMed Lorentzen AG, Lindskov C, Sloth E, Jakobsen CJ. Central venous oxygen saturation cannot replace mixed venous saturation in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth. 2008;22:853–7.PubMed
74.
Zurück zum Zitat Glamann DB, Lange RA, Hillis LD. Incidence and significance of a ‘step-down’ in oxygen saturation from superior vena cava to pulmonary artery. Am J Cardiol. 1991;68:695–7.PubMed Glamann DB, Lange RA, Hillis LD. Incidence and significance of a ‘step-down’ in oxygen saturation from superior vena cava to pulmonary artery. Am J Cardiol. 1991;68:695–7.PubMed
75.
Zurück zum Zitat Barratt-Boyes BG, Wood EH. The oxygen saturation of blood in the venae cavae, right-heart chambers, and pulmonary vessels of healthy subjects. J Lab Clin Med. 1957;50:93–106.PubMed Barratt-Boyes BG, Wood EH. The oxygen saturation of blood in the venae cavae, right-heart chambers, and pulmonary vessels of healthy subjects. J Lab Clin Med. 1957;50:93–106.PubMed
76.
Zurück zum Zitat Dahn MS, Lange MP, Jacobs LA. Central mixed and splanchnic venous oxygen saturation monitoring. Intensive Care Med. 1988;14:373–8.PubMed Dahn MS, Lange MP, Jacobs LA. Central mixed and splanchnic venous oxygen saturation monitoring. Intensive Care Med. 1988;14:373–8.PubMed
77.
Zurück zum Zitat Lee J, Wright F, Barber R, Stanley L. Central venous oxygen saturation in shock: a study in man. Anesthesiology. 1972;36:472–8.PubMed Lee J, Wright F, Barber R, Stanley L. Central venous oxygen saturation in shock: a study in man. Anesthesiology. 1972;36:472–8.PubMed
78.
Zurück zum Zitat Ho KM, Harding R, Chamberlain J, Bulsara M. A comparison of central and mixed venous oxygen saturation in circulatory failure. J Cardiothorac Vasc Anesth. 2010;24:434–9.PubMed Ho KM, Harding R, Chamberlain J, Bulsara M. A comparison of central and mixed venous oxygen saturation in circulatory failure. J Cardiothorac Vasc Anesth. 2010;24:434–9.PubMed
79.
Zurück zum Zitat Turnaoglu S, Tugrul M, Camci E, Cakar N, Akinci O, Ergin P. Clinical applicability of the substitution of mixed venous oxygen saturation with central venous oxygen saturation. J Cardiothorac Vasc Anesth. 2001;15:574–9.PubMed Turnaoglu S, Tugrul M, Camci E, Cakar N, Akinci O, Ergin P. Clinical applicability of the substitution of mixed venous oxygen saturation with central venous oxygen saturation. J Cardiothorac Vasc Anesth. 2001;15:574–9.PubMed
80.
Zurück zum Zitat Reinhart K, Kersting T, Fohring U, Schafer M. Can central-venous replace mixed-venous oxygen saturation measurements during anesthesia? Adv Exp Med Biol. 1986;200:67–72.PubMed Reinhart K, Kersting T, Fohring U, Schafer M. Can central-venous replace mixed-venous oxygen saturation measurements during anesthesia? Adv Exp Med Biol. 1986;200:67–72.PubMed
81.
Zurück zum Zitat Collaborative Study Group on Perioperative ScvO2 Monitoring. Multicentre study on peri- and postoperative central venous oxygen saturation in high-risk surgical patients. Crit Care. 2006;10:R158.PubMedCentral Collaborative Study Group on Perioperative ScvO2 Monitoring. Multicentre study on peri- and postoperative central venous oxygen saturation in high-risk surgical patients. Crit Care. 2006;10:R158.PubMedCentral
82.
Zurück zum Zitat Polonen P, Ruokonen E, Hippelainen M, Poyhonen M, Takala J. A prospective, randomized study of goal-oriented hemodynamic therapy in cardiac surgical patients. Anesth Analg. 2000;90:1052–9.PubMed Polonen P, Ruokonen E, Hippelainen M, Poyhonen M, Takala J. A prospective, randomized study of goal-oriented hemodynamic therapy in cardiac surgical patients. Anesth Analg. 2000;90:1052–9.PubMed
83.
Zurück zum Zitat Donati A, Loggi S, Preiser JC, Orsetti G, Munch C, Gabbanelli V, et al. Goal-directed intraoperative therapy reduces morbidity and length of hospital stay in high-risk surgical patients. Chest. 2007;132:1817–24.PubMed Donati A, Loggi S, Preiser JC, Orsetti G, Munch C, Gabbanelli V, et al. Goal-directed intraoperative therapy reduces morbidity and length of hospital stay in high-risk surgical patients. Chest. 2007;132:1817–24.PubMed
84.
Zurück zum Zitat Van der Linden P, Schmartz D, Gilbart E, Engelman E, Vincent JL. Effects of propofol, etomidate, and pentobarbital on critical oxygen delivery. Crit Care Med. 2000;28:2492–9.PubMed Van der Linden P, Schmartz D, Gilbart E, Engelman E, Vincent JL. Effects of propofol, etomidate, and pentobarbital on critical oxygen delivery. Crit Care Med. 2000;28:2492–9.PubMed
85.
Zurück zum Zitat Perz S, Uhlig T, Kohl M, Bredle DL, Reinhart K, Bauer M, et al. Low and ‘supranormal’ central venous oxygen saturation and markers of tissue hypoxia in cardiac surgery patients: a prospective observational study. Intensive Care Med. 2011;37:52–9.PubMed Perz S, Uhlig T, Kohl M, Bredle DL, Reinhart K, Bauer M, et al. Low and ‘supranormal’ central venous oxygen saturation and markers of tissue hypoxia in cardiac surgery patients: a prospective observational study. Intensive Care Med. 2011;37:52–9.PubMed
86.
Zurück zum Zitat Fuller BM, Dellinger RP. Lactate as a hemodynamic marker in the critically ill. Curr Opin Crit Care. 2012;18:267–72.PubMedPubMedCentral Fuller BM, Dellinger RP. Lactate as a hemodynamic marker in the critically ill. Curr Opin Crit Care. 2012;18:267–72.PubMedPubMedCentral
87.
Zurück zum Zitat Meregalli A, Oliveira RP, Friedman G. Occult hypoperfusion is associated with increased mortality in hemodynamically stable, high-risk, surgical patients. Crit Care. 2004;8:R60–5.PubMedPubMedCentral Meregalli A, Oliveira RP, Friedman G. Occult hypoperfusion is associated with increased mortality in hemodynamically stable, high-risk, surgical patients. Crit Care. 2004;8:R60–5.PubMedPubMedCentral
88.
Zurück zum Zitat Bakker J, Coffernils M, Leon M, Gris P, Vincent JL. Blood lactate levels are superior to oxygen-derived variables in predicting outcome in human septic shock. Chest. 1991;99:956–62.PubMed Bakker J, Coffernils M, Leon M, Gris P, Vincent JL. Blood lactate levels are superior to oxygen-derived variables in predicting outcome in human septic shock. Chest. 1991;99:956–62.PubMed
89.
Zurück zum Zitat Jansen TC, van Bommel J, Schoonderbeek FJ, Sleeswijk Visser SJ, van der Klooster JM, Lima AP, et al. Early lactate-guided therapy in intensive care unit patients: a multicenter, open-label, randomized controlled trial. Am J Respir Crit Care Med. 2010;182:752–61.PubMed Jansen TC, van Bommel J, Schoonderbeek FJ, Sleeswijk Visser SJ, van der Klooster JM, Lima AP, et al. Early lactate-guided therapy in intensive care unit patients: a multicenter, open-label, randomized controlled trial. Am J Respir Crit Care Med. 2010;182:752–61.PubMed
90.
Zurück zum Zitat Jansen TC, van Bommel J, Woodward R, Mulder PG, Bakker J. Association between blood lactate levels, Sequential Organ Failure Assessment subscores, and 28-day mortality during early and late intensive care unit stay: a retrospective observational study. Crit Care Med. 2009;37:2369–74.PubMed Jansen TC, van Bommel J, Woodward R, Mulder PG, Bakker J. Association between blood lactate levels, Sequential Organ Failure Assessment subscores, and 28-day mortality during early and late intensive care unit stay: a retrospective observational study. Crit Care Med. 2009;37:2369–74.PubMed
91.
Zurück zum Zitat McKendry M, McGloin H, Saberi D, Caudwell L, Brady AR, Singer M. Randomised controlled trial assessing the impact of a nurse delivered, flow monitored protocol for optimisation of circulatory status after cardiac surgery. BMJ. 2004;329:258.PubMedPubMedCentral McKendry M, McGloin H, Saberi D, Caudwell L, Brady AR, Singer M. Randomised controlled trial assessing the impact of a nurse delivered, flow monitored protocol for optimisation of circulatory status after cardiac surgery. BMJ. 2004;329:258.PubMedPubMedCentral
92.
Zurück zum Zitat Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED. Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care. 2005;9:R687–93.PubMedPubMedCentral Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED. Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care. 2005;9:R687–93.PubMedPubMedCentral
93.
Zurück zum Zitat Bundgaard-Nielsen M, Holte K, Secher NH, Kehlet H. Monitoring of peri-operative fluid administration by individualized goal-directed therapy. Acta Anaesthesiol Scand. 2007;51:331–40.PubMed Bundgaard-Nielsen M, Holte K, Secher NH, Kehlet H. Monitoring of peri-operative fluid administration by individualized goal-directed therapy. Acta Anaesthesiol Scand. 2007;51:331–40.PubMed
94.
Zurück zum Zitat Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, et al. Reducing the risk of major elective surgery: randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ. 1999;318:1099–103.PubMedPubMedCentral Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, et al. Reducing the risk of major elective surgery: randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ. 1999;318:1099–103.PubMedPubMedCentral
95.
Zurück zum Zitat Lobo SM, Salgado PF, Castillo VG, Borim AA, Polachini CA, Palchetti JC, et al. Effects of maximizing oxygen delivery on morbidity and mortality in high-risk surgical patients. Crit Care Med. 2000;28:3396–404.PubMed Lobo SM, Salgado PF, Castillo VG, Borim AA, Polachini CA, Palchetti JC, et al. Effects of maximizing oxygen delivery on morbidity and mortality in high-risk surgical patients. Crit Care Med. 2000;28:3396–404.PubMed
96.
Zurück zum Zitat Lopes MR, Oliveira MA, Pereira VO, Lemos IP, Auler Jr JO, Michard F. Goal-directed fluid management based on pulse pressure variation monitoring during high-risk surgery: a pilot randomized controlled trial. Crit Care. 2007;11:R100.PubMedPubMedCentral Lopes MR, Oliveira MA, Pereira VO, Lemos IP, Auler Jr JO, Michard F. Goal-directed fluid management based on pulse pressure variation monitoring during high-risk surgery: a pilot randomized controlled trial. Crit Care. 2007;11:R100.PubMedPubMedCentral
97.
Zurück zum Zitat Pearse RM, Harrison DA, MacDonald N, Gillies MA, Blunt M, Ackland G, et al. Effect of a perioperative, cardiac output-guided hemodynamic therapy algorithm on outcomes following major gastrointestinal surgery: a randomized clinical trial and systematic review. JAMA. 2014;311:2181–90.PubMed Pearse RM, Harrison DA, MacDonald N, Gillies MA, Blunt M, Ackland G, et al. Effect of a perioperative, cardiac output-guided hemodynamic therapy algorithm on outcomes following major gastrointestinal surgery: a randomized clinical trial and systematic review. JAMA. 2014;311:2181–90.PubMed
98.
Zurück zum Zitat Morris C. Oesophageal Doppler monitoring, doubt and equipoise: evidence based medicine means change. Anaesthesia. 2013;68:684–8.PubMed Morris C. Oesophageal Doppler monitoring, doubt and equipoise: evidence based medicine means change. Anaesthesia. 2013;68:684–8.PubMed
99.
Zurück zum Zitat Scheeren TW, Wiesenack C, Gerlach H, Marx G. Goal-directed intraoperative fluid therapy guided by stroke volume and its variation in high-risk surgical patients: a prospective randomized multicentre study. J Clin Monit Comput. 2013;27:225–33.PubMed Scheeren TW, Wiesenack C, Gerlach H, Marx G. Goal-directed intraoperative fluid therapy guided by stroke volume and its variation in high-risk surgical patients: a prospective randomized multicentre study. J Clin Monit Comput. 2013;27:225–33.PubMed
100.
Zurück zum Zitat Goepfert MS, Richter HP, Zu EC, Gruetzmacher J, Rafflenbeul E, Roeher K, et al. Individually optimized hemodynamic therapy reduces complications and length of stay in the intensive care unit: a prospective, randomized controlled trial. Anesthesiology. 2013;119:824–36.PubMed Goepfert MS, Richter HP, Zu EC, Gruetzmacher J, Rafflenbeul E, Roeher K, et al. Individually optimized hemodynamic therapy reduces complications and length of stay in the intensive care unit: a prospective, randomized controlled trial. Anesthesiology. 2013;119:824–36.PubMed
101.
Zurück zum Zitat Fellahi JL, Parienti JJ, Hanouz JL, Plaud B, Riou B, Ouattara A. Perioperative use of dobutamine in cardiac surgery and adverse cardiac outcome: propensity-adjusted analyses. Anesthesiology. 2008;108:979–87.PubMed Fellahi JL, Parienti JJ, Hanouz JL, Plaud B, Riou B, Ouattara A. Perioperative use of dobutamine in cardiac surgery and adverse cardiac outcome: propensity-adjusted analyses. Anesthesiology. 2008;108:979–87.PubMed
102.
Zurück zum Zitat Pearse RM, Belsey JD, Cole JN, Bennett ED. Effect of dopexamine infusion on mortality following major surgery: individual patient data meta-regression analysis of published clinical trials. Crit Care Med. 2008;36:1323–9.PubMed Pearse RM, Belsey JD, Cole JN, Bennett ED. Effect of dopexamine infusion on mortality following major surgery: individual patient data meta-regression analysis of published clinical trials. Crit Care Med. 2008;36:1323–9.PubMed
103.
Zurück zum Zitat Takala J, Meier-Hellmann A, Eddleston J, Hulstaert P, Sramek V. Effect of dopexamine on outcome after major abdominal surgery: a prospective, randomized, controlled multicenter study. European Multicenter Study Group on Dopexamine in Major Abdominal Surgery. Crit Care Med. 2000;28:3417–23.PubMed Takala J, Meier-Hellmann A, Eddleston J, Hulstaert P, Sramek V. Effect of dopexamine on outcome after major abdominal surgery: a prospective, randomized, controlled multicenter study. European Multicenter Study Group on Dopexamine in Major Abdominal Surgery. Crit Care Med. 2000;28:3417–23.PubMed
Metadaten
Titel
Perioperative cardiovascular monitoring of high-risk patients: a consensus of 12
verfasst von
Jean-Louis Vincent
Paolo Pelosi
Rupert Pearse
Didier Payen
Azriel Perel
Andreas Hoeft
Stefano Romagnoli
V Marco Ranieri
Carole Ichai
Patrice Forget
Giorgio Della Rocca
Andrew Rhodes
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 1/2015
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-015-0932-7

Weitere Artikel der Ausgabe 1/2015

Critical Care 1/2015 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.