Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software

verfasst von: Young Han Lee, Kwan Kyu Park, Ho-Taek Song, Sungjun Kim, Jin-Suck Suh

Erschienen in: European Radiology | Ausgabe 6/2012

Einloggen, um Zugang zu erhalten

Abstract

Objective

To assess the usefulness of gemstone spectral imaging (GSI) dual-energy CT (DECT) with/without metal artefact reduction software (MARs).

Methods

The DECTs were performed using fast kV-switching GSI between 80 and 140 kV. The CT data were retro-reconstructed with/without MARs, by different displayed fields-of-view (DFOV), and with synthesised monochromatic energy in the range 40–140 keV. A phantom study of size and CT numbers was performed in a titanium plate and a stainless steel plate. A clinical study was performed in 26 patients with metallic hardware. All images were retrospectively reviewed in terms of the visualisation of periprosthetic regions and the severity of beam-hardening artefacts by using a five-point scale.

Results

The GSI-MARs reconstruction can markedly reduce the metal-related artefacts, and the image quality was affected by the prosthesis composition and DFOV. The spectral CT numbers of the prosthesis and periprosthetic regions showed different patterns on stainless steel and titanium plates.

Conclusion

Dual-energy CT with GSI-MARs can reduce metal-related artefacts and improve the delineation of the prosthesis and periprosthetic region. We should be cautious when using GSI-MARs because the image quality was affected by the prosthesis composition, energy (in keV) and DFOV. The metallic composition and size should be considered in metallic imaging with GSI-MARs reconstruction.

Key Points

• Metal-related artefacts can be troublesome on musculoskeletal computed tomography (CT).
• Gemstone spectral imaging (GSI) with dual-energy CT (DECT) offers a novel solution
• GSI and metallic artefact reduction software (GSI-MAR) can markedly reduce these artefacts.
• However image quality is influenced by the prosthesis composition and other parameters.
• We should be aware about potential overcorrection when using GSI-MARs.
Literatur
2.
Zurück zum Zitat White LM, Buckwalter KA (2002) Technical considerations: CT and MR imaging in the postoperative orthopedic patient. Semin Musculoskelet Radiol 6:5–17PubMedCrossRef White LM, Buckwalter KA (2002) Technical considerations: CT and MR imaging in the postoperative orthopedic patient. Semin Musculoskelet Radiol 6:5–17PubMedCrossRef
3.
Zurück zum Zitat Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691PubMedCrossRef Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691PubMedCrossRef
4.
Zurück zum Zitat Haramati N, Staron RB, Mazel-Sperling K et al (1994) CT scans through metal scanning technique versus hardware composition. Comput Med Imaging Graph 18:429–434PubMedCrossRef Haramati N, Staron RB, Mazel-Sperling K et al (1994) CT scans through metal scanning technique versus hardware composition. Comput Med Imaging Graph 18:429–434PubMedCrossRef
5.
Zurück zum Zitat Love C, Marwin SE, Palestro CJ (2009) Nuclear medicine and the infected joint replacement. Semin Nucl Med 39:66–78PubMedCrossRef Love C, Marwin SE, Palestro CJ (2009) Nuclear medicine and the infected joint replacement. Semin Nucl Med 39:66–78PubMedCrossRef
7.
Zurück zum Zitat Gelman MI, Coleman RE, Stevens PM, Davey BW (1978) Radiography, radionuclide imaging, and arthrography in the evaluation of total hip and knee replacement. Radiology 128:677–682PubMed Gelman MI, Coleman RE, Stevens PM, Davey BW (1978) Radiography, radionuclide imaging, and arthrography in the evaluation of total hip and knee replacement. Radiology 128:677–682PubMed
8.
Zurück zum Zitat Phillips WC, Kattapuram SV (1983) Efficacy of preoperative hip aspiration performed in the radiology department. Clin Orthop Relat Res 141–146 Phillips WC, Kattapuram SV (1983) Efficacy of preoperative hip aspiration performed in the radiology department. Clin Orthop Relat Res 141–146
9.
Zurück zum Zitat Young SW, Muller HH, Marshall WH (1983) Computed tomography: beam hardening and environmental density artifact. Radiology 148:279–283PubMed Young SW, Muller HH, Marshall WH (1983) Computed tomography: beam hardening and environmental density artifact. Radiology 148:279–283PubMed
10.
Zurück zum Zitat Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803PubMedCrossRef Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803PubMedCrossRef
11.
Zurück zum Zitat Watzke O, Kalender WA (2004) A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images. Eur Radiol 14:849–856PubMedCrossRef Watzke O, Kalender WA (2004) A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images. Eur Radiol 14:849–856PubMedCrossRef
12.
Zurück zum Zitat Li H, Yu L, Liu X, Fletcher JG, McCollough CH (2010) Metal artifact suppression from reformatted projections in multislice helical CT using dual-front active contours. Med Phys 37:5155–5164PubMedCrossRef Li H, Yu L, Liu X, Fletcher JG, McCollough CH (2010) Metal artifact suppression from reformatted projections in multislice helical CT using dual-front active contours. Med Phys 37:5155–5164PubMedCrossRef
13.
Zurück zum Zitat Li H, Yu L, Liu X, McCollough CH (2009) Metal artifact suppression from reformatted projections in multi-slice helical CT using dual-front active contours. Conf Proc IEEE Eng Med Biol Soc 2009:993–996PubMed Li H, Yu L, Liu X, McCollough CH (2009) Metal artifact suppression from reformatted projections in multi-slice helical CT using dual-front active contours. Conf Proc IEEE Eng Med Biol Soc 2009:993–996PubMed
14.
Zurück zum Zitat Yu L, Li H, Mueller J et al (2009) Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography: techniques and initial clinical results. Invest Radiol 44:691–696PubMedCrossRef Yu L, Li H, Mueller J et al (2009) Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography: techniques and initial clinical results. Invest Radiol 44:691–696PubMedCrossRef
15.
Zurück zum Zitat Link TM, Berning W, Scherf S et al (2000) CT of metal implants: reduction of artifacts using an extended CT scale technique. J Comput Assist Tomogr 24:165–172PubMedCrossRef Link TM, Berning W, Scherf S et al (2000) CT of metal implants: reduction of artifacts using an extended CT scale technique. J Comput Assist Tomogr 24:165–172PubMedCrossRef
16.
Zurück zum Zitat Montner SM, Lehr JL, Oravez WT (1987) Quantitative evaluation of a dual energy CT system. J Comput Assist Tomogr 11:144–150PubMedCrossRef Montner SM, Lehr JL, Oravez WT (1987) Quantitative evaluation of a dual energy CT system. J Comput Assist Tomogr 11:144–150PubMedCrossRef
17.
Zurück zum Zitat Dilmanian FA (1992) Computed tomography with monochromatic x rays. Am J Physiol Imaging 7:175–193PubMed Dilmanian FA (1992) Computed tomography with monochromatic x rays. Am J Physiol Imaging 7:175–193PubMed
18.
Zurück zum Zitat Dilmanian FA, Wu XY, Parsons EC et al (1997) Single-and dual-energy CT with monochromatic synchrotron x-rays. Phys Med Biol 42:371–387PubMedCrossRef Dilmanian FA, Wu XY, Parsons EC et al (1997) Single-and dual-energy CT with monochromatic synchrotron x-rays. Phys Med Biol 42:371–387PubMedCrossRef
19.
Zurück zum Zitat Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol 21:1424–1429PubMedCrossRef Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol 21:1424–1429PubMedCrossRef
20.
Zurück zum Zitat Boas FE, Fleischmann D (2011) Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259:894–902PubMedCrossRef Boas FE, Fleischmann D (2011) Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259:894–902PubMedCrossRef
21.
Zurück zum Zitat Lin XZ, Miao F, Li JY, Dong HP, Shen Y, Chen KM (2011) High-definition CT Gemstone spectral imaging of the brain: initial results of selecting optimal monochromatic image for beam-hardening artifacts and image noise reduction. J Comput Assist Tomogr 35:294–297PubMedCrossRef Lin XZ, Miao F, Li JY, Dong HP, Shen Y, Chen KM (2011) High-definition CT Gemstone spectral imaging of the brain: initial results of selecting optimal monochromatic image for beam-hardening artifacts and image noise reduction. J Comput Assist Tomogr 35:294–297PubMedCrossRef
22.
Zurück zum Zitat Zhang D, Li X, Liu B (2011) Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode. Med Phys 38:1178–1188PubMedCrossRef Zhang D, Li X, Liu B (2011) Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode. Med Phys 38:1178–1188PubMedCrossRef
23.
Zurück zum Zitat Douglas-Akinwande AC, Buckwalter KA, Rydberg J, Rankin JL, Choplin RH (2006) Multichannel CT: evaluating the spine in postoperative patients with orthopedic hardware. Radiographics 26:S97–S110PubMedCrossRef Douglas-Akinwande AC, Buckwalter KA, Rydberg J, Rankin JL, Choplin RH (2006) Multichannel CT: evaluating the spine in postoperative patients with orthopedic hardware. Radiographics 26:S97–S110PubMedCrossRef
Metadaten
Titel
Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software
verfasst von
Young Han Lee
Kwan Kyu Park
Ho-Taek Song
Sungjun Kim
Jin-Suck Suh
Publikationsdatum
01.06.2012
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 6/2012
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-011-2370-5

Weitere Artikel der Ausgabe 6/2012

European Radiology 6/2012 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.