Pediatric chest computed tomography at 100 kVp with tin filtration: comparison of image quality with 70-kVp imaging at comparable radiation dose

verfasst von: Sophie Vivier, Valérie Deken, Younes Arous, Jean-Baptiste Faivre, Alain Duhamel, Antoine Deschildre, Thomas Flohr, Jacques Remy, Martine Remy-Jardin

Erschienen in: Pediatric Radiology | Ausgabe 2/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

Radiation dose reduction is a primary objective in pediatric populations owing to the well-known risks of radiation-induced cancers. Low-energy photons participate in the radiation dose without significantly contributing to image formation. Their suppression by means of tin filtration should decrease the image noise, anticipating a subsequent application to dose saving.

Objective

To investigate the level of noise reduction achievable with tin (Sn) filtration at 100 kVp for chest computed tomography (CT) in comparison with a standard scanning mode at 70 kVp with comparable radiation dose.

Materials and methods

Fifty consecutive children (Group 1) underwent non-contrast chest CT examinations on a third-generation dual-source CT system at tin-filtered 100 kVp and pitch 2. The tube-current time product (mAs) was adjusted to maintain the predicted dose length product (DLP) value at 70 kVp for the respective patient. Each child was then paired by weight and age to a child scanned at 70 kVp on the same CT unit (Group 2); Group 2 patients were consecutive patients, retrospectively selected from our database of children prospectively scanned at 70 kVp. Objective and subjective image quality were compared between the two groups of patients to investigate the overall image quality and level of noise reduction that could be subsequently achievable with tin filtration in clinical practice.

Results

The mean image noise was significantly lower in Group 1 compared to Group 2 when measured in the air (P<0.0001) and inside the aorta (P<0.001). The mean noise reduction was 21.6% (standard deviation [SD] 16.1) around the thorax and 12.0% (SD 32.7) inside the thorax. There was no significant difference in subjective image quality of lung and mediastinal images with excellent overall subjective scores in both groups.

Conclusion

At comparable radiation dose, the image noise was found to be reduced by 21.6% compared to the 70-kVp protocol, providing basis for dose reduction without altering image quality in further investigations.
Literatur
1.
Zurück zum Zitat Kleinerman RA (2006) Cancer risks following diagnostic and therapeutic radiation exposure in children. Pediatr Radiol 36:121–125CrossRef Kleinerman RA (2006) Cancer risks following diagnostic and therapeutic radiation exposure in children. Pediatr Radiol 36:121–125CrossRef
3.
Zurück zum Zitat Lucaya J, Piqueras J, García-Peña P et al (2000) Low-dose high-resolution CT of the chest in children and young adults: dose, cooperation, artifact incidence and image quality. AJR Am J Roentgenol 175:985–992CrossRef Lucaya J, Piqueras J, García-Peña P et al (2000) Low-dose high-resolution CT of the chest in children and young adults: dose, cooperation, artifact incidence and image quality. AJR Am J Roentgenol 175:985–992CrossRef
4.
Zurück zum Zitat Kim JE, Newman B (2010) Evaluation of a radiation dose reduction strategy for pediatric chest CT. AJR Am J Roentgenol 194:1188–1193CrossRef Kim JE, Newman B (2010) Evaluation of a radiation dose reduction strategy for pediatric chest CT. AJR Am J Roentgenol 194:1188–1193CrossRef
5.
Zurück zum Zitat Siegel MJ, Hildebolt C, Bradley D (2013) Effects of automated kilovoltage selection technology on contrast-enhanced pediatric CT and CT angiography. Radiology 268:538–547CrossRef Siegel MJ, Hildebolt C, Bradley D (2013) Effects of automated kilovoltage selection technology on contrast-enhanced pediatric CT and CT angiography. Radiology 268:538–547CrossRef
6.
Zurück zum Zitat Singh S, Kalra MK, Moore MA et al (2009) Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies. Radiology 252:200–208CrossRef Singh S, Kalra MK, Moore MA et al (2009) Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies. Radiology 252:200–208CrossRef
7.
Zurück zum Zitat Kroft LJ, Roelofs JJ, Geleijns J (2010) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294–300CrossRef Kroft LJ, Roelofs JJ, Geleijns J (2010) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294–300CrossRef
8.
Zurück zum Zitat Remy JM, Santangelo T, Colas L et al (2017) Radiation dose levels in pediatric chest CT: experience in 499 children evaluated with dual-source single-energy CT. Pediatr Radiol 47:161–168 Remy JM, Santangelo T, Colas L et al (2017) Radiation dose levels in pediatric chest CT: experience in 499 children evaluated with dual-source single-energy CT. Pediatr Radiol 47:161–168
9.
Zurück zum Zitat Niemann T, Henry S, Duhamel A et al (2014) Pediatric chest CT at 70 kVp: a feasibility study in 129 children. Pediatr Radiol 44:1347–1357CrossRef Niemann T, Henry S, Duhamel A et al (2014) Pediatric chest CT at 70 kVp: a feasibility study in 129 children. Pediatr Radiol 44:1347–1357CrossRef
10.
Zurück zum Zitat Weis M, Henzler T, Nance JW Jr et al (2017) Radiation dose comparison between 70 kVp and 100 kVp with spectral beam shaping for non-contrast-enhanced pediatric chest computed tomography: a prospective randomized controlled study. Invest Radiol 52:155–162CrossRef Weis M, Henzler T, Nance JW Jr et al (2017) Radiation dose comparison between 70 kVp and 100 kVp with spectral beam shaping for non-contrast-enhanced pediatric chest computed tomography: a prospective randomized controlled study. Invest Radiol 52:155–162CrossRef
11.
Zurück zum Zitat Singh S, Kalra MK, Shenoy-Bhangle AS et al (2012) Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 263:537–546CrossRef Singh S, Kalra MK, Shenoy-Bhangle AS et al (2012) Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 263:537–546CrossRef
12.
Zurück zum Zitat Lell MM, May M, Deak P et al (2011) High-pitch spiral computed tomography. Effect on image quality and radiation dose in pediatric chest computed tomography. Invest Radiol 46:116–123CrossRef Lell MM, May M, Deak P et al (2011) High-pitch spiral computed tomography. Effect on image quality and radiation dose in pediatric chest computed tomography. Invest Radiol 46:116–123CrossRef
13.
Zurück zum Zitat Gordic S, Morsbach F, Schmidt B et al (2014) Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation with single energy scanning with spectral shaping. Invest Radiol 49:465–473CrossRef Gordic S, Morsbach F, Schmidt B et al (2014) Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation with single energy scanning with spectral shaping. Invest Radiol 49:465–473CrossRef
14.
Zurück zum Zitat Newell JD Jr, Fuld MK, Allmendinger T et al (2015) Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third generation dual-source CT scanner with corresponding third-generation iterative reconstruction software. Invest Radiol 50:40–50CrossRef Newell JD Jr, Fuld MK, Allmendinger T et al (2015) Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third generation dual-source CT scanner with corresponding third-generation iterative reconstruction software. Invest Radiol 50:40–50CrossRef
15.
Zurück zum Zitat Braun FM, Johnson TR, Sommer WH et al (2015) Chest CT using spectral filtration: radiation dose, image quality and spectrum of clinical utility. Eur Radiol 25:1598–1606CrossRef Braun FM, Johnson TR, Sommer WH et al (2015) Chest CT using spectral filtration: radiation dose, image quality and spectrum of clinical utility. Eur Radiol 25:1598–1606CrossRef
16.
Zurück zum Zitat Haubenreisser H, Meyer M, Sudarski S et al (2015) Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100kVp. Eur J Radiol 84:1608–1613CrossRef Haubenreisser H, Meyer M, Sudarski S et al (2015) Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100kVp. Eur J Radiol 84:1608–1613CrossRef
17.
Zurück zum Zitat Bodelle B, Fischbach C, Booz C et al (2017) Single-energy pediatric chest computed tomography with spectral filtration at 100 kVp: effects on radiation parameters and image quality. Pediatr Radiol 47:831–837CrossRef Bodelle B, Fischbach C, Booz C et al (2017) Single-energy pediatric chest computed tomography with spectral filtration at 100 kVp: effects on radiation parameters and image quality. Pediatr Radiol 47:831–837CrossRef
18.
Zurück zum Zitat Suntharalingam S, Allmendinger T, Blex S et al (2018) Spectral beam shaping in unenhanced chest CT examinations: a phantom study on dose reduction and image quality. Acad Radiol 25:153–158CrossRef Suntharalingam S, Allmendinger T, Blex S et al (2018) Spectral beam shaping in unenhanced chest CT examinations: a phantom study on dose reduction and image quality. Acad Radiol 25:153–158CrossRef
20.
Zurück zum Zitat Hagelstein C, Henzler T, Haubenreisser H et al (2016) Ultra-high pitch chest computed tomography at 70 kVp tube voltage in an anthropomorphic pediatric phantom and non-sedated pediatric patients: initial experience with third-generation dual-source CT. Z Med Phys 26:349–361CrossRef Hagelstein C, Henzler T, Haubenreisser H et al (2016) Ultra-high pitch chest computed tomography at 70 kVp tube voltage in an anthropomorphic pediatric phantom and non-sedated pediatric patients: initial experience with third-generation dual-source CT. Z Med Phys 26:349–361CrossRef
21.
Zurück zum Zitat O'Connor OJ, Vandeleur M, McGarrigle AM et al (2010) Development of low-dose protocols for thin-section CT assessment of cystic fibrosis in pediatric patients. Radiology 257:820–829CrossRef O'Connor OJ, Vandeleur M, McGarrigle AM et al (2010) Development of low-dose protocols for thin-section CT assessment of cystic fibrosis in pediatric patients. Radiology 257:820–829CrossRef
Metadaten
Titel
Pediatric chest computed tomography at 100 kVp with tin filtration: comparison of image quality with 70-kVp imaging at comparable radiation dose
verfasst von
Sophie Vivier
Valérie Deken
Younes Arous
Jean-Baptiste Faivre
Alain Duhamel
Antoine Deschildre
Thomas Flohr
Jacques Remy
Martine Remy-Jardin
Publikationsdatum
01.02.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 2/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-019-04543-w

Weitere Artikel der Ausgabe 2/2020

Pediatric Radiology 2/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.