Erschienen in: Skeletal Radiology 7/2015

01.07.2015 | Scientific Article

Reliability of measures used in radiographic evaluation of the adult hip

verfasst von: J. A. Bjarnason, A. H. Pripp, O. Reikeras

Erschienen in: Skeletal Radiology | Ausgabe 7/2015

Einloggen, um Zugang zu erhalten

Abstract

Objective

The reliability of radiographic measurements has been studied in pediatric hips, but less has been published on the adult hip, and none have examined the reliability of measurements for the location of the center of rotation (COR) of the hip joint. We have investigated the reliability of various radiographic variables with a focus on the COR.

Materials and methods

The study was carried out on a standardized format for anterior-posterior radiographs of the pelvis. The measured variables were; (A) the distance from a sagittal reference line to the COR, (B) the distance from the sagittal reference line to the proximal end of the lateral cortical line of the femur, (C) the distance from the sagittal reference line to the medial rim of the acetabulum, (D) the distance from the horizontal reference line to the roof of the acetabulum, and (E) the distance from the horizontal reference line to the COR. One observer (JAB) conducted the measurements twice separated by a time interval of 45–60 days to assess intra-observer reliability, and the first measurements of JAB were compared to those performed by another observer (OR) to assess inter-observer reliability.

Results

Intraclass correlation coefficients were above 0.98 for all measurements, and the minimum and maximum values that statistically include 95 % of the observer differences were all within −3 to +3 mm.

Conclusions

These measurements proved to have high reliability and agreement of both within the same observer and between two observers. They should therefore be reproducible in a clinical setting.
Literatur
1.
Zurück zum Zitat Doehring TC, Rubash HE, Shelley FJ, et al. Effect of superior and superolateral relocation of the hip center on hip joint forces. J Arthroplasty. 1996;11:693–703.CrossRefPubMed Doehring TC, Rubash HE, Shelley FJ, et al. Effect of superior and superolateral relocation of the hip center on hip joint forces. J Arthroplasty. 1996;11:693–703.CrossRefPubMed
2.
Zurück zum Zitat Pagnano W, Hanssen AD, Lewallen DG. The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty. J Bone Joint Surg Am. 1996;78:1004–14.PubMed Pagnano W, Hanssen AD, Lewallen DG. The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty. J Bone Joint Surg Am. 1996;78:1004–14.PubMed
3.
Zurück zum Zitat Sexton SA, Yeung E, Jackson MP, et al. The role of patient factors and implant position in squeaking of ceramic-on-ceramic total hip replacements. J Bone Joint Surg Br. 2011;93:439–42.CrossRefPubMed Sexton SA, Yeung E, Jackson MP, et al. The role of patient factors and implant position in squeaking of ceramic-on-ceramic total hip replacements. J Bone Joint Surg Br. 2011;93:439–42.CrossRefPubMed
4.
Zurück zum Zitat Yoder SA, Brand RA, Pedersen DR, O’Gorman TW. Total hip acetabular component position affects component loosening rates. Clin Orthop Relat Res. 1988;228:79–87.PubMed Yoder SA, Brand RA, Pedersen DR, O’Gorman TW. Total hip acetabular component position affects component loosening rates. Clin Orthop Relat Res. 1988;228:79–87.PubMed
5.
Zurück zum Zitat Callaghan JJ, Salvati EA, Pelicci PM, Wilson Jr PD, Ranawat CS. Results of revision for mechanical failure after cemented total hip replacement 1979–1982, a two to five—year follow up. J Bone Joint Surg Am. 1985;67:1074–85.PubMed Callaghan JJ, Salvati EA, Pelicci PM, Wilson Jr PD, Ranawat CS. Results of revision for mechanical failure after cemented total hip replacement 1979–1982, a two to five—year follow up. J Bone Joint Surg Am. 1985;67:1074–85.PubMed
6.
Zurück zum Zitat Boniforti FG, Fujii G, Angliss RD, Benson MK. The reliability of measurements of pelvic radiographs in infants. J Bone Joint Surg Br. 1997;79:570–5.CrossRefPubMed Boniforti FG, Fujii G, Angliss RD, Benson MK. The reliability of measurements of pelvic radiographs in infants. J Bone Joint Surg Br. 1997;79:570–5.CrossRefPubMed
7.
Zurück zum Zitat Broughton NS, Brougham DI, Cole WG, Menelaus MB. Reliability of radiological measurements in the assessment of the child’s hip. J Bone Joint Surg Br. 1989;71:6–8.PubMed Broughton NS, Brougham DI, Cole WG, Menelaus MB. Reliability of radiological measurements in the assessment of the child’s hip. J Bone Joint Surg Br. 1989;71:6–8.PubMed
8.
Zurück zum Zitat Clohisy JC, Carlisle JC, Beaule PE, Kim YJ, Trousdale RT, Sierra RJ, et al. A systematic approach to the plain radiographic evaluation of the young adult hip. J Bone Joint Surg Am. 2008;90:47–66.CrossRefPubMedCentralPubMed Clohisy JC, Carlisle JC, Beaule PE, Kim YJ, Trousdale RT, Sierra RJ, et al. A systematic approach to the plain radiographic evaluation of the young adult hip. J Bone Joint Surg Am. 2008;90:47–66.CrossRefPubMedCentralPubMed
9.
Zurück zum Zitat Kay RM, Watts HG, Dorey FJ. Variability in the assessment of acetabular index. J Pediatr Orthop. 1997;17:170–3.CrossRefPubMed Kay RM, Watts HG, Dorey FJ. Variability in the assessment of acetabular index. J Pediatr Orthop. 1997;17:170–3.CrossRefPubMed
10.
Zurück zum Zitat Spatz DK, Reiger M, Klaumann M, Miller F, Stanton RP, Lipton GE. Measurement of acetabular index intraobserver and interobserver variation. J Pediatr Orthop. 1997;17:174–5.CrossRefPubMed Spatz DK, Reiger M, Klaumann M, Miller F, Stanton RP, Lipton GE. Measurement of acetabular index intraobserver and interobserver variation. J Pediatr Orthop. 1997;17:174–5.CrossRefPubMed
11.
Zurück zum Zitat Pons C, Rémy-Neris O, Médée B, Brochard S. Validity and reliability of radiological methods to assess proximal hip geometry in children with cerebral palsy: a systematic review. Dev Med Child Neurol. 2013;55(12):1089–102.CrossRefPubMed Pons C, Rémy-Neris O, Médée B, Brochard S. Validity and reliability of radiological methods to assess proximal hip geometry in children with cerebral palsy: a systematic review. Dev Med Child Neurol. 2013;55(12):1089–102.CrossRefPubMed
12.
Zurück zum Zitat Mast NH, Impellizzeri F, Keller S, Leunig M. Reliability and agreement of measures used in radiographic evaluation of the adult hip. Clin Orthop Relat Res. 2011;469:188–99.CrossRefPubMedCentralPubMed Mast NH, Impellizzeri F, Keller S, Leunig M. Reliability and agreement of measures used in radiographic evaluation of the adult hip. Clin Orthop Relat Res. 2011;469:188–99.CrossRefPubMedCentralPubMed
13.
Zurück zum Zitat Nelitz M, Guenther KP, Gunkel S, Puhl W. Reliability of radiological measurements in the assessment of hip dysplasia in adults. Br J Radiol. 1999;72:331–4.CrossRefPubMed Nelitz M, Guenther KP, Gunkel S, Puhl W. Reliability of radiological measurements in the assessment of hip dysplasia in adults. Br J Radiol. 1999;72:331–4.CrossRefPubMed
14.
Zurück zum Zitat Tannast M, Mistry S, Steppacher SD, Reichenbach S, Langlotz F, Siebenrock KA, et al. Radiographic analysis of femoroacetabular impingement with Hip2Norm-reliable and validated. J Orthop Res. 2008;26:1199–205.CrossRefPubMed Tannast M, Mistry S, Steppacher SD, Reichenbach S, Langlotz F, Siebenrock KA, et al. Radiographic analysis of femoroacetabular impingement with Hip2Norm-reliable and validated. J Orthop Res. 2008;26:1199–205.CrossRefPubMed
15.
Zurück zum Zitat Siebenrock KA, Kalbermatten DF, Ganz R. Effect of pelvic tilt on acetabular retroversion: a study of pelves from cadavers. Clin Orthop. 2003;407:241–8.CrossRefPubMed Siebenrock KA, Kalbermatten DF, Ganz R. Effect of pelvic tilt on acetabular retroversion: a study of pelves from cadavers. Clin Orthop. 2003;407:241–8.CrossRefPubMed
16.
Zurück zum Zitat Walter SD, Eiasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med. 1998;17:101–10.CrossRefPubMed Walter SD, Eiasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med. 1998;17:101–10.CrossRefPubMed
17.
Zurück zum Zitat Sutherland CJ, Bresina SJ. Measurement of acetabular component migration using two-dimensional radiography. J Arthroplasty. 1992;7:377–9.CrossRefPubMed Sutherland CJ, Bresina SJ. Measurement of acetabular component migration using two-dimensional radiography. J Arthroplasty. 1992;7:377–9.CrossRefPubMed
18.
Zurück zum Zitat Tannast M, Zheng G, Anderegg C, Burckhardt K, Langlotz F, Ganz R, et al. Tilt and rotation correction of acetabular version on pelvic radiographs. Clin Orthop Relat Res. 2005;438:182–90.CrossRefPubMed Tannast M, Zheng G, Anderegg C, Burckhardt K, Langlotz F, Ganz R, et al. Tilt and rotation correction of acetabular version on pelvic radiographs. Clin Orthop Relat Res. 2005;438:182–90.CrossRefPubMed
19.
Zurück zum Zitat Wegner A, Kauther MD, Landgraeber S, von Knoch M. Fixation method does not affect restoration of rotation center in hip replacements: a single-site retrospective study. J Orthop Surg Res. 2012;7–25. Wegner A, Kauther MD, Landgraeber S, von Knoch M. Fixation method does not affect restoration of rotation center in hip replacements: a single-site retrospective study. J Orthop Surg Res. 2012;7–25.
20.
Zurück zum Zitat Cassidy KA, Noticewala MS, Macaulay W, Lee JH, Geller JA. Effect of femoral offset on pain and function after total hip arthroplasty. J Arthroplasty. 2012;27:1863–9.CrossRefPubMed Cassidy KA, Noticewala MS, Macaulay W, Lee JH, Geller JA. Effect of femoral offset on pain and function after total hip arthroplasty. J Arthroplasty. 2012;27:1863–9.CrossRefPubMed
21.
Zurück zum Zitat Liebs TR, Nasser L, Herzberg W, Rüther W, Hassenpflug J. The influence of femoral offset on health-related quality of life after total hip replacement. Bone Joint J. 2014;96-B:36–42.CrossRefPubMed Liebs TR, Nasser L, Herzberg W, Rüther W, Hassenpflug J. The influence of femoral offset on health-related quality of life after total hip replacement. Bone Joint J. 2014;96-B:36–42.CrossRefPubMed
Metadaten
Titel
Reliability of measures used in radiographic evaluation of the adult hip
verfasst von
J. A. Bjarnason
A. H. Pripp
O. Reikeras
Publikationsdatum
01.07.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 7/2015
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-015-2101-0

Weitere Artikel der Ausgabe 7/2015

Skeletal Radiology 7/2015 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.