Skip to main content
Erschienen in: Skeletal Radiology 4/2015

01.04.2015 | Technical Report

Reproducibility of trabecular bone score with different scan modes using dual-energy X-ray absorptiometry: a phantom study

verfasst von: Michele Bandirali, Giovanni Di Leo, Carmelo Messina, Maria Juana Pastor Lopez, Alessandro Mai, Fabio M. Ulivieri, Francesco Sardanelli

Erschienen in: Skeletal Radiology | Ausgabe 4/2015

Einloggen, um Zugang zu erhalten

Abstract

Objective

The trabecular bone score (TBS) accounts for the bone microarchitecture and is calculated on dual-energy X-ray absorptiometry (DXA). We estimated the reproducibility of the TBS using different scan modes compared to the reproducibility bone mineral density (BMD).

Materials and methods

A spine phantom was used with a Hologic QDR-Discovery A densitometer. For each scan mode [fast array, array, high definition (HD)], 25 scans were automatically performed without phantom repositioning; a further 25 scans were performed with phantom repositioning. For each scan, the TBS was obtained. The coefficient of variation (CoV) was calculated as the ratio between standard deviation and mean; percent least significant change (LSC%) as 2.8×CoV; reproducibility as the complement to 100 % of LSC%. Differences among scan modes were assessed using ANOVA.

Results

Without phantom repositioning, the mean TBS (mm−1) was: 1.352 (fast array), 1.321 (array), and 1.360 (HD); with phantom repositioning, it was 1.345, 1.332, and 1.362, respectively. Reproducibility of the TBS without phantom repositioning was 97.7 % (fast array), 98.3 % (array), and 98.2 % (HD); with phantom repositioning, it was 97.9 %, 98.7 %, and 98.4 %, respectively. LSC% was ≤2.26 %. Differences among scan modes were all statistically significant (p ≤ 0.019). Reproducibility of BMD was 99.1 % with all scan modes, while LSC% was from 0.86 % to 0.91 %.

Conclusion

Reproducibility error of the TBS was 2–3-fold higher than that of BMD. Although statistically significant, differences in TBS among scan modes were within the highest LSC%. Thus, the three scan modes can be considered interchangeable.
Literatur
1.
Zurück zum Zitat NIH Consensus Development Panel on Osteoporosis Prevention, Diagnosis and Therapy. JAMA. 2001;285:785–95. NIH Consensus Development Panel on Osteoporosis Prevention, Diagnosis and Therapy. JAMA. 2001;285:785–95.
2.
Zurück zum Zitat Bouxsein ML, Seeman E. Quantifying the material and structural determinants of bone strength. Best Pract Res Clin Rheumatol. 2009;23:741–53.CrossRefPubMed Bouxsein ML, Seeman E. Quantifying the material and structural determinants of bone strength. Best Pract Res Clin Rheumatol. 2009;23:741–53.CrossRefPubMed
3.
Zurück zum Zitat Kanis JA, Borgstrom F, De Laet C, et al. Assessment of fracture risk. Osteoporos Int. 2005;16:581–9.CrossRefPubMed Kanis JA, Borgstrom F, De Laet C, et al. Assessment of fracture risk. Osteoporos Int. 2005;16:581–9.CrossRefPubMed
4.
Zurück zum Zitat Brandi ML. Microarchitecture, the key to bone quality. Rheumatology. 2009;48 Suppl 4:iv3–8.CrossRefPubMed Brandi ML. Microarchitecture, the key to bone quality. Rheumatology. 2009;48 Suppl 4:iv3–8.CrossRefPubMed
5.
Zurück zum Zitat Wainwright SA, Marshall LM, Ensrud KE, et al. Hip fracture in women without osteoporosis. J Clin Endocrinol Metab. 2005;90:2787–93.CrossRefPubMed Wainwright SA, Marshall LM, Ensrud KE, et al. Hip fracture in women without osteoporosis. J Clin Endocrinol Metab. 2005;90:2787–93.CrossRefPubMed
7.
Zurück zum Zitat Pothuaud L, Carceller P, Hans D. Correlations between grey-level variations in 2D projection images (TBS) and 3D microarchitecture: applications in the study of human trabecular bone microarchitecture. Bone. 2008;42:775–87.CrossRefPubMed Pothuaud L, Carceller P, Hans D. Correlations between grey-level variations in 2D projection images (TBS) and 3D microarchitecture: applications in the study of human trabecular bone microarchitecture. Bone. 2008;42:775–87.CrossRefPubMed
8.
Zurück zum Zitat Pothuaud L. Process of interpretation of two-dimensional densitometry images for the prediction of bone mechanical strength. Lecture notes in computer science. Springer-Verlag GmbH; 2004. p. 1079–80. Pothuaud L. Process of interpretation of two-dimensional densitometry images for the prediction of bone mechanical strength. Lecture notes in computer science. Springer-Verlag GmbH; 2004. p. 1079–80.
9.
Zurück zum Zitat Hans D, Barthe N, Boutroy S, Pothuaud L, Winzenrieth R, Krieg MA. Correlations between trabecular bone score, measured using anteroposterior dual-energy X-ray absorptiometry acquisition, and 3-dimensional parameters of bone microarchitecture: an experimental study on human cadaver vertebrae. J Clin Densitom. 2011;14(3):302–12. doi:10.1016/j.jocd.2011.05.005.CrossRefPubMed Hans D, Barthe N, Boutroy S, Pothuaud L, Winzenrieth R, Krieg MA. Correlations between trabecular bone score, measured using anteroposterior dual-energy X-ray absorptiometry acquisition, and 3-dimensional parameters of bone microarchitecture: an experimental study on human cadaver vertebrae. J Clin Densitom. 2011;14(3):302–12. doi:10.​1016/​j.​jocd.​2011.​05.​005.CrossRefPubMed
10.
Zurück zum Zitat Delnevo A, Bandirali M, Di Leo G, et al. Differences among array, fast array, and high-definition scan modes in bone mineral density measurement at dual-energy x-ray absorptiometry on a phantom. Clin Radiol. 2013;68(6):616–9. doi:10.1016/j.crad.2012.11.017.CrossRefPubMed Delnevo A, Bandirali M, Di Leo G, et al. Differences among array, fast array, and high-definition scan modes in bone mineral density measurement at dual-energy x-ray absorptiometry on a phantom. Clin Radiol. 2013;68(6):616–9. doi:10.​1016/​j.​crad.​2012.​11.​017.CrossRefPubMed
11.
Zurück zum Zitat Bousson V et al. Trabecular bone score (TBS): available knowledge, clinical relevance, and future prospects. Osteoporos Int. 2012;23:1489–501.CrossRefPubMed Bousson V et al. Trabecular bone score (TBS): available knowledge, clinical relevance, and future prospects. Osteoporos Int. 2012;23:1489–501.CrossRefPubMed
12.
Zurück zum Zitat Lewiecki EM, Gordon CM, Baim S, et al. Special report on the 2007 adult and pediatric Position Development Conferences of the International Society for clinical densitometry. Osteoporos Int. 2008;19(10):1369–78. doi:10.1007/s00198-008-0689-9.CrossRefPubMed Lewiecki EM, Gordon CM, Baim S, et al. Special report on the 2007 adult and pediatric Position Development Conferences of the International Society for clinical densitometry. Osteoporos Int. 2008;19(10):1369–78. doi:10.​1007/​s00198-008-0689-9.CrossRefPubMed
13.
Zurück zum Zitat Blake GM, Fogelman I. Role of dual-energy X-ray absorptiometry in the diagnosis and treatment of osteoporosis. J Clin Densitom. 2007;10(1):102–10.CrossRefPubMed Blake GM, Fogelman I. Role of dual-energy X-ray absorptiometry in the diagnosis and treatment of osteoporosis. J Clin Densitom. 2007;10(1):102–10.CrossRefPubMed
14.
15.
Zurück zum Zitat Bandirali M, Lanza E, Messina C et al. Dose Absorption in Lumbar and Femoral Dual Energy X-ray absorptiometry examinations using three different scan modalities: an anthropomorphic phantom study. J Clin Densitom. 2013;24. doi:10.1016/j.jocd.2013.02.005. Bandirali M, Lanza E, Messina C et al. Dose Absorption in Lumbar and Femoral Dual Energy X-ray absorptiometry examinations using three different scan modalities: an anthropomorphic phantom study. J Clin Densitom. 2013;24. doi:10.​1016/​j.​jocd.​2013.​02.​005.
Metadaten
Titel
Reproducibility of trabecular bone score with different scan modes using dual-energy X-ray absorptiometry: a phantom study
verfasst von
Michele Bandirali
Giovanni Di Leo
Carmelo Messina
Maria Juana Pastor Lopez
Alessandro Mai
Fabio M. Ulivieri
Francesco Sardanelli
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 4/2015
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-014-1980-9

Weitere Artikel der Ausgabe 4/2015

Skeletal Radiology 4/2015 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.