Stent graft visualization and planning tool for endovascular surgery using finite element analysis

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 4/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

A new approach to optimize stent graft selection for endovascular aortic repair is the use of finite element analysis. Once the finite element model is created and solved, a software module is needed to view the simulation results in the clinical work environment. A new tool for interpretation of simulation results, named Medical Postprocessor, that enables comparison of different stent graft configurations and products was designed, implemented and tested.

Methods

 Aortic endovascular stent graft ring forces and sealing states in the vessel landing zone of three different configurations were provided in a surgical planning software using the Medical Imaging Interaction Tool Kit (MITK) software system. For data interpretation, software modules for 2D and 3D presentations were implemented. Ten surgeons evaluated the software features of the Medical Postprocessor. These surgeons performed usability tests and answered questionnaires based on their experience with the system.

Results

The Medical Postprocessor visualization system enabled vascular surgeons to determine the configuration with the highest overall fixation force in \(16 \pm 6\) s, best proximal sealing in \(56 \pm 24\) s and highest proximal fixation force in \(38\pm 12\) s. The majority considered the multiformat data provided helpful and found the Medical Postprocessor to be an efficient decision support system for stent graft selection. The evaluation of the user interface results in an ISONORM-conform user interface (113.5 points).

Conclusion

The Medical Postprocessor visualization software tool for analyzing stent graft properties was evaluated by vascular surgeons. The results show that the software can assist the interpretation of simulation results to optimize stent graft configuration and sizing.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Volodos NL, Shekhanin VE, Karpovich IP, Troian VI, Gur’ev IA (1986) A self-fixing synthetic blood vessel endoprosthesis. Vestn Khir Im I I Grek 137:123–125PubMed Volodos NL, Shekhanin VE, Karpovich IP, Troian VI, Gur’ev IA (1986) A self-fixing synthetic blood vessel endoprosthesis. Vestn Khir Im I I Grek 137:123–125PubMed
2.
Zurück zum Zitat Parodi JC, Palmaz JC, Barone HD (1991) Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg 5:491–499PubMedCrossRef Parodi JC, Palmaz JC, Barone HD (1991) Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg 5:491–499PubMedCrossRef
3.
Zurück zum Zitat Dake MD, Miller DC, Semba CP (1994) Transluminal placement of endovascular stent-grafts for the treatment of descending thoracic aortic aneurysms. N Engl J Med 331(26):1729–1734PubMedCrossRef Dake MD, Miller DC, Semba CP (1994) Transluminal placement of endovascular stent-grafts for the treatment of descending thoracic aortic aneurysms. N Engl J Med 331(26):1729–1734PubMedCrossRef
4.
Zurück zum Zitat Stelter WJ, Umscheid T, Ziegler P (1997) Three-year experience with modular stent graft devices for endovascular AAA treatment. J Endovasc surg 4:362–369PubMedCrossRef Stelter WJ, Umscheid T, Ziegler P (1997) Three-year experience with modular stent graft devices for endovascular AAA treatment. J Endovasc surg 4:362–369PubMedCrossRef
5.
Zurück zum Zitat Deutsche Gesellschaft für Gefäßchirurgie (DGG) (2008) 8. Qualitätsmanagement Bauchaortenaneurysma Deutsche Gesellschaft für Gefäßchirurgie (DGG) (2008) 8. Qualitätsmanagement Bauchaortenaneurysma
7.
Zurück zum Zitat Kortmann H (2011) Bauchaortenaneurysmen (Kapitel 11). In: Kompaktwissen Gefäßchirurgie, B.L.P. Luther (Hrsg.). Springer, Heidelberg, pp 199–219 Kortmann H (2011) Bauchaortenaneurysmen (Kapitel 11). In: Kompaktwissen Gefäßchirurgie, B.L.P. Luther (Hrsg.). Springer, Heidelberg, pp 199–219
10.
Zurück zum Zitat Raghavan ML, Vorp DA (2000) Toward a biomechanical tool to evaluate rupture potential of abdominal aortic aneurysm: identification of a finite strain constitutive model and valuation of its applicability. J Biomech 33:475–482PubMedCrossRef Raghavan ML, Vorp DA (2000) Toward a biomechanical tool to evaluate rupture potential of abdominal aortic aneurysm: identification of a finite strain constitutive model and valuation of its applicability. J Biomech 33:475–482PubMedCrossRef
12.
Zurück zum Zitat Molony DS, Broderick S, Callanan A, McGloughlin TM, Walsh MT (2011) Fluid-structure interaction in healthy, diseased and endovascularly treated abdominal aortic aneurysms. Stud Mech Anobiol Tissue Eng Biomater, vol 7, Springer, Berlin, pp. 163–179. doi:10.1007/8415_2011_85. ISBN:978-3-642-18094-1 Molony DS, Broderick S, Callanan A, McGloughlin TM, Walsh MT (2011) Fluid-structure interaction in healthy, diseased and endovascularly treated abdominal aortic aneurysms. Stud Mech Anobiol Tissue Eng Biomater, vol 7, Springer, Berlin, pp. 163–179. doi:10.​1007/​8415_​2011_​85. ISBN:978-3-642-18094-1
13.
Zurück zum Zitat Sun Z, Chaichana T, Sangworasil M, Tungjitkusolmun S (2008) Computational fluid analysis of blood flow characteristics in abdominal aortic aneurysms treated with suprarenal endovascular grafts. ICBME Proc 23:1728–1732 Sun Z, Chaichana T, Sangworasil M, Tungjitkusolmun S (2008) Computational fluid analysis of blood flow characteristics in abdominal aortic aneurysms treated with suprarenal endovascular grafts. ICBME Proc 23:1728–1732
14.
Zurück zum Zitat Figueroa CA, Taylor CA, Yeh V, Chiou AJ, Zarins CK (2009) Effect of curvature on displacement forces acting on aortic endografts: a 3-dimensional computational analysis. J Endovasc Ther 16:284–294 Figueroa CA, Taylor CA, Yeh V, Chiou AJ, Zarins CK (2009) Effect of curvature on displacement forces acting on aortic endografts: a 3-dimensional computational analysis. J Endovasc Ther 16:284–294
15.
Zurück zum Zitat Scherer S, Treichel T, Ritter N, Triebel G, Drossel, WG, Burgert O (2011) Surgical stent planning—simulation parameter study for models based on DICOM standards. Int J CARS, 6, 3. Springer, Heidelberg, pp 319–327 Scherer S, Treichel T, Ritter N, Triebel G, Drossel, WG, Burgert O (2011) Surgical stent planning—simulation parameter study for models based on DICOM standards. Int J CARS, 6, 3. Springer, Heidelberg, pp 319–327
16.
Zurück zum Zitat de Bock S, de Beule M, de Santis G, Vermassen F, Segers P, Verhegghe B (2011) A simulation tool for virtual stent graft deployment in patient-specific abdominal aortic aneurysms. In: Proceedings of the ASME, summer engineering conference de Bock S, de Beule M, de Santis G, Vermassen F, Segers P, Verhegghe B (2011) A simulation tool for virtual stent graft deployment in patient-specific abdominal aortic aneurysms. In: Proceedings of the ASME, summer engineering conference
17.
Zurück zum Zitat von Sachsen S, Etz CD, Mohr FW, Senf B, Neugebauer R, Florek HJ (2011) A method for integrating finite element results in a surgical planning software for evaluating stent graft properties in endovascular surgery. In: IEEE proceedings 4th international conference on biomedical engineering and informatics, Shanghai (China), pp 1349–1354. ISBN:978-1-4244-9350-0 von Sachsen S, Etz CD, Mohr FW, Senf B, Neugebauer R, Florek HJ (2011) A method for integrating finite element results in a surgical planning software for evaluating stent graft properties in endovascular surgery. In: IEEE proceedings 4th international conference on biomedical engineering and informatics, Shanghai (China), pp 1349–1354. ISBN:978-1-4244-9350-0
18.
Zurück zum Zitat Gebert de Uhlenbrock (2009) A. Dissertation Technische Universität Hamburg-Harburg, Designanalyse von endovaskulären Aortenprothesen Gebert de Uhlenbrock (2009) A. Dissertation Technische Universität Hamburg-Harburg, Designanalyse von endovaskulären Aortenprothesen
19.
Zurück zum Zitat Senf B, von Sachsen S, Neugebauer R, Drossel WG, Florek HJ, Mohr FW, Etz CD (2013) The effect of stent graft oversizing on radial forces considering Nitinol wire behavior and vessel characteristics. J Med Eng (accepted) Senf B, von Sachsen S, Neugebauer R, Drossel WG, Florek HJ, Mohr FW, Etz CD (2013) The effect of stent graft oversizing on radial forces considering Nitinol wire behavior and vessel characteristics. J Med Eng (accepted)
20.
Zurück zum Zitat Sinha Roy A, Westt K, Rontala RS, Greenberg RK, Banerjee RK (2007) In vitro measurement and calculation of drag force on iliac limb stentgraft in a compliant arterial wall model. Mol Cell Biomech 4(4):211–226PubMed Sinha Roy A, Westt K, Rontala RS, Greenberg RK, Banerjee RK (2007) In vitro measurement and calculation of drag force on iliac limb stentgraft in a compliant arterial wall model. Mol Cell Biomech 4(4):211–226PubMed
21.
Zurück zum Zitat Corbett TJ, Molony DS, Callanan A, Mc Gloughlin TM (2011) The effect of vessel material properties and pulsatile wall motion on the fixation of a proximal stent of an endovscular graft. Med Eng Phys 33(1):106–111PubMedCrossRef Corbett TJ, Molony DS, Callanan A, Mc Gloughlin TM (2011) The effect of vessel material properties and pulsatile wall motion on the fixation of a proximal stent of an endovscular graft. Med Eng Phys 33(1):106–111PubMedCrossRef
22.
Zurück zum Zitat Arko FR, Heikkinen M, Lee ES, Bass A, Alsac JM, Zarins CK (2004) Iliac fixation length and resistance to in-vivo stentgraft displacement. In: Proceedings of the 19. Annual meeting of the western vascular society, vol 41, number 4, pp 664–671. doi:10.1016/j.jvs.2004.12.050 Arko FR, Heikkinen M, Lee ES, Bass A, Alsac JM, Zarins CK (2004) Iliac fixation length and resistance to in-vivo stentgraft displacement. In: Proceedings of the 19. Annual meeting of the western vascular society, vol 41, number 4, pp 664–671. doi:10.​1016/​j.​jvs.​2004.​12.​050
23.
Zurück zum Zitat Kleinstreuer C, Li Z, Basciano CA, Seelecke S, Farber MA (2008) Computational mechanics of Nitinol stent grafts. J Biomech 41:2370–2378 Kleinstreuer C, Li Z, Basciano CA, Seelecke S, Farber MA (2008) Computational mechanics of Nitinol stent grafts. J Biomech 41:2370–2378
24.
Zurück zum Zitat Gao F, Watanabem M, Matsuzawa T (2006) Stress analysis in a layered aortic arch model under pulsatile blood flow. BioMed Eng Online 5. doi:10.1186/1475-925X-5-25 Gao F, Watanabem M, Matsuzawa T (2006) Stress analysis in a layered aortic arch model under pulsatile blood flow. BioMed Eng Online 5. doi:10.​1186/​1475-925X-5-25
25.
Zurück zum Zitat Holzapfel GA, Sommer G, Regitnig P (2004) Anisotropic mechanical properties of tissue components in human atherosclerotic plaques. J Biomech Eng H 126:657–665. doi:10.1115/1.1800557 Holzapfel GA, Sommer G, Regitnig P (2004) Anisotropic mechanical properties of tissue components in human atherosclerotic plaques. J Biomech Eng H 126:657–665. doi:10.​1115/​1.​1800557
28.
Zurück zum Zitat ANSYS, Inc. (2009) ANSYS manual; Version 11 ANSYS, Inc. (2009) ANSYS manual; Version 11
31.
Zurück zum Zitat Murphy EH, Johnson D, Arko FR (2007) Device-specific resistance to in vivo displacement of stent-grafts implanted with maximum iliac fixation. J Endovasc Ther 14:585–592PubMedCrossRef Murphy EH, Johnson D, Arko FR (2007) Device-specific resistance to in vivo displacement of stent-grafts implanted with maximum iliac fixation. J Endovasc Ther 14:585–592PubMedCrossRef
32.
Zurück zum Zitat Bosman WMPF, Steenhoven vd, Steenhoven TJ, Suárez DR, Hinnen JW, Valstar ER, Hamming JF (2010) The proximal fixation strength of modern EVAR grafts in a short aneurysm neck. An in vitro study. Eur J Vasc Endovasc Surg 39:187–192. doi:10.1016/j.ejvs2009.10.019 Bosman WMPF, Steenhoven vd, Steenhoven TJ, Suárez DR, Hinnen JW, Valstar ER, Hamming JF (2010) The proximal fixation strength of modern EVAR grafts in a short aneurysm neck. An in vitro study. Eur J Vasc Endovasc Surg 39:187–192. doi:10.​1016/​j.​ejvs2009.​10.​019
34.
Zurück zum Zitat Prümper J (1997) Der Benutzungsfragebogen ISONORM 9241/10: Ergebnisse zur Reliabilität und Validität. In Liskkowsky R, Velichkovsky BM, Wünsschmann W (Hrsg) Software-Ergonomie ’97-usability engineering: integration von Mensch-Computer-Interaktion und Software-Entwicklung, pp 253–262 Prümper J (1997) Der Benutzungsfragebogen ISONORM 9241/10: Ergebnisse zur Reliabilität und Validität. In Liskkowsky R, Velichkovsky BM, Wünsschmann W (Hrsg) Software-Ergonomie ’97-usability engineering: integration von Mensch-Computer-Interaktion und Software-Entwicklung, pp 253–262
Metadaten
Titel
Stent graft visualization and planning tool for endovascular surgery using finite element analysis
Publikationsdatum
01.07.2014
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 4/2014
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-013-0943-2

Weitere Artikel der Ausgabe 4/2014

International Journal of Computer Assisted Radiology and Surgery 4/2014 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.