Skip to main content
Erschienen in: Journal of Digital Imaging 3/2013

01.06.2013

Using a High-Speed Movie Camera to Evaluate Slice Dropping in Clinical Image Interpretation with Stack Mode Viewers

verfasst von: Masahiro Yakami, Akira Yamamoto, Morio Yanagisawa, Hiroyuki Sekiguchi, Takeshi Kubo, Kaori Togashi

Erschienen in: Journal of Imaging Informatics in Medicine | Ausgabe 3/2013

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study is to verify objectively the rate of slice omission during paging on picture archiving and communication system (PACS) viewers by recording the images shown on the computer displays of these viewers with a high-speed movie camera. This study was approved by the institutional review board. A sequential number from 1 to 250 was superimposed on each slice of a series of clinical Digital Imaging and Communication in Medicine (DICOM) data. The slices were displayed using several DICOM viewers, including in-house developed freeware and clinical PACS viewers. The freeware viewer and one of the clinical PACS viewers included functions to prevent slice dropping. The series was displayed in stack mode and paged in both automatic and manual paging modes. The display was recorded with a high-speed movie camera and played back at a slow speed to check whether slices were dropped. The paging speeds were also measured. With a paging speed faster than half the refresh rate of the display, some viewers dropped up to 52.4 % of the slices, while other well-designed viewers did not, if used with the correct settings. Slice dropping during paging was objectively confirmed using a high-speed movie camera. To prevent slice dropping, the viewer must be specially designed for the purpose and must be used with the correct settings, or the paging speed must be slower than half of the display refresh rate.
Literatur
1.
Zurück zum Zitat Mathie AG, Strickland NH: Interpretation of CT scans with PACS image display in stack-mode. Radiology 203(1):207–209, 1997PubMed Mathie AG, Strickland NH: Interpretation of CT scans with PACS image display in stack-mode. Radiology 203(1):207–209, 1997PubMed
2.
Zurück zum Zitat Jung JI, Kim KJ, Ahn MI, Kim HR, Park HJ, Jung S, Lim HW, Park SH: Detection of pulmonary embolism using 64-slice multidetector-row computed tomography: accuracy and reproducibility on different image reconstruction parameters. Acta Radiol 52(5):417–421, 2011PubMedCrossRef Jung JI, Kim KJ, Ahn MI, Kim HR, Park HJ, Jung S, Lim HW, Park SH: Detection of pulmonary embolism using 64-slice multidetector-row computed tomography: accuracy and reproducibility on different image reconstruction parameters. Acta Radiol 52(5):417–421, 2011PubMedCrossRef
3.
Zurück zum Zitat Meenman C, Daly B, Toland C, Nagy P: Use of a thin-section archive and enterprise 3D software for long-term storage of thin-slice CT data sets. J Digit Imaging 19(suppl 1):84–88, 2006CrossRef Meenman C, Daly B, Toland C, Nagy P: Use of a thin-section archive and enterprise 3D software for long-term storage of thin-slice CT data sets. J Digit Imaging 19(suppl 1):84–88, 2006CrossRef
4.
Zurück zum Zitat Holcombe AO: Seeing slow and seeing fast: two limits on perception. Trends Cogn Sci 13(5):216–221, 2009PubMedCrossRef Holcombe AO: Seeing slow and seeing fast: two limits on perception. Trends Cogn Sci 13(5):216–221, 2009PubMedCrossRef
12.
Zurück zum Zitat Yakami M, Ishizu K, Kubo T, Okada T, Togashi K: Development and evaluation of a low-cost and high-capacity DICOM image data storage system for research. J Digit Imaging 24(2):190–195, 2011PubMedCrossRef Yakami M, Ishizu K, Kubo T, Okada T, Togashi K: Development and evaluation of a low-cost and high-capacity DICOM image data storage system for research. J Digit Imaging 24(2):190–195, 2011PubMedCrossRef
13.
Zurück zum Zitat Bhargavan M, Kaye AH, Forman HP, Sunshine JH: Workload of radiologists in United States in 2006–2007 and trends since 1991–1992. Radiology 252(2):458–467, 2009PubMedCrossRef Bhargavan M, Kaye AH, Forman HP, Sunshine JH: Workload of radiologists in United States in 2006–2007 and trends since 1991–1992. Radiology 252(2):458–467, 2009PubMedCrossRef
15.
Zurück zum Zitat Siegel EL, Reiner BI, Cadogan M: Frequency and impact of high-resolution monitor failure in a filmless imaging department. J Digit Imaging 13(3):114–118, 2000PubMedCrossRef Siegel EL, Reiner BI, Cadogan M: Frequency and impact of high-resolution monitor failure in a filmless imaging department. J Digit Imaging 13(3):114–118, 2000PubMedCrossRef
Metadaten
Titel
Using a High-Speed Movie Camera to Evaluate Slice Dropping in Clinical Image Interpretation with Stack Mode Viewers
verfasst von
Masahiro Yakami
Akira Yamamoto
Morio Yanagisawa
Hiroyuki Sekiguchi
Takeshi Kubo
Kaori Togashi
Publikationsdatum
01.06.2013
Verlag
Springer-Verlag
Erschienen in
Journal of Imaging Informatics in Medicine / Ausgabe 3/2013
Print ISSN: 2948-2925
Elektronische ISSN: 2948-2933
DOI
https://doi.org/10.1007/s10278-012-9534-y

Weitere Artikel der Ausgabe 3/2013

Journal of Digital Imaging 3/2013 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.