Using aluminum for scatter control in mammography: preliminary work using measurements of CNR and FOM

verfasst von: Khaled Al Khalifah, Rob Davidson, Abel Zhou

Erschienen in: Radiological Physics and Technology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

Full-field digital mammography (FFDM) systems provide the current gold standard in mammographic examinations. Although FFDM provides the lowest mammographic doses, the radiation dose to the breast during mammographic examinations is still a concern. Thus, image quality optimization at the lowest dose is a major goal. In planar X-ray imaging, thin sheets of aluminum (Al) are used as filtration to reduce the number of low-energy X-ray photons reaching the patient. The goal of this work was to evaluate whether Al can be used in FFDM to remove scatter radiation from reaching the image detector, hence improving image quality. Doses were compared with the use of a grid. A Hologic Selenia mammographic unit was used to acquire images of two phantoms, namely, the ACR phantom and a Perspex phantom of 5 cm. Images were acquired using two tube voltages (kVp) and filter combinations under two exposure/dose conditions. Al sheets of various thicknesses were placed between the phantom and the image detector. Contrast-to-noise ratio (CNR) and figure of merit (FOM) values were measured and compared with images acquired using a grid. When a constant dose was delivered to the image detector, the highest CNR was achieved using a grid; however, the highest FOM values were achieved when using 0.05 mm thick Al sheets. This study successfully demonstrates that thin sheets of Al can be used in mammography examinations to reduce scattered radiation and improve image quality, as indicated by the measured CNR values. Given the limitations of this work, further kVp and target/filter combinations and various methods of image quality measurement need to be studied.
Literatur
1.
Zurück zum Zitat The Swedish Organised Service Screening Evaluation Group. Reduction in breast cancer mortality from organized service screening with mammography with extended data. Cancer Epidemiol Biomark Prev. 2006;15:45–51.CrossRef The Swedish Organised Service Screening Evaluation Group. Reduction in breast cancer mortality from organized service screening with mammography with extended data. Cancer Epidemiol Biomark Prev. 2006;15:45–51.CrossRef
2.
Zurück zum Zitat Tabar L, Vitak B, Chen HH, Yen MF, Duffy SW, Smith RA. Beyond randomized controlled trials: organized mammographic screening substantially reduces breast carcinoma mortality. Cancer. 2001;91(9):1724–31.CrossRef Tabar L, Vitak B, Chen HH, Yen MF, Duffy SW, Smith RA. Beyond randomized controlled trials: organized mammographic screening substantially reduces breast carcinoma mortality. Cancer. 2001;91(9):1724–31.CrossRef
3.
Zurück zum Zitat Tabar L, Yen MF, Vitak B, Chen HH, Smith RA, Duffy SW. Mammography service screening and mortality in breast cancer patients: 20-year follow-up before and after introduction of screening. Lancet (London, England). 2003;361(9367):1405–10.CrossRef Tabar L, Yen MF, Vitak B, Chen HH, Smith RA, Duffy SW. Mammography service screening and mortality in breast cancer patients: 20-year follow-up before and after introduction of screening. Lancet (London, England). 2003;361(9367):1405–10.CrossRef
4.
Zurück zum Zitat Chawla AS, Samei E, Saunders R, Abbey C, Delong D. Effect of dose reduction on the detection of mammographic lesions: a mathematical observer model analysis. Med Phys. 2007;34(8):3385–98.CrossRef Chawla AS, Samei E, Saunders R, Abbey C, Delong D. Effect of dose reduction on the detection of mammographic lesions: a mathematical observer model analysis. Med Phys. 2007;34(8):3385–98.CrossRef
5.
Zurück zum Zitat Bushberg J, Seibert J, Leidholdt J, Boone J. The essential physics of medical imaging. 3rd ed. Philadelphia: Lippincott Williams & Wilkins; 2012. Bushberg J, Seibert J, Leidholdt J, Boone J. The essential physics of medical imaging. 3rd ed. Philadelphia: Lippincott Williams & Wilkins; 2012.
6.
Zurück zum Zitat Aminah M, Ng KH, Abdullah BJJ, Jamal N. Optimal beam quality selection based on contrast-to-noise ratio and mean glandular dose in digital mammography. Australas Phys Eng Sci Med. 2010;33(4):329–34.CrossRef Aminah M, Ng KH, Abdullah BJJ, Jamal N. Optimal beam quality selection based on contrast-to-noise ratio and mean glandular dose in digital mammography. Australas Phys Eng Sci Med. 2010;33(4):329–34.CrossRef
7.
Zurück zum Zitat Kanal KM, Krupinski E, Berns EA, Geiser WR, Karellas A, Mainiero MB, et al. ACR–AAPM–SIIM practice guideline for determinants of image quality in digital mammography. J Digit Imaging. 2013;26(1):10–25.CrossRef Kanal KM, Krupinski E, Berns EA, Geiser WR, Karellas A, Mainiero MB, et al. ACR–AAPM–SIIM practice guideline for determinants of image quality in digital mammography. J Digit Imaging. 2013;26(1):10–25.CrossRef
8.
Zurück zum Zitat Yaffe MJ, Bunch PC, Desponds L, Jong RA, Nishikawa RM, Tapiovaara MJ, et al. Technical aspects of image quality in mammography. J Int Comm Radiat Units Meas. 2009;9(2):33–51. Yaffe MJ, Bunch PC, Desponds L, Jong RA, Nishikawa RM, Tapiovaara MJ, et al. Technical aspects of image quality in mammography. J Int Comm Radiat Units Meas. 2009;9(2):33–51.
9.
Zurück zum Zitat Varjonen M, Strömmer P. Optimizing the anode-filter combination in the sense of image quality and average glandular dose in digital mammography. In: Proc. SPIE 6913, Medical imaging 2008, San Diego, California; 2008. Varjonen M, Strömmer P. Optimizing the anode-filter combination in the sense of image quality and average glandular dose in digital mammography. In: Proc. SPIE 6913, Medical imaging 2008, San Diego, California; 2008.
10.
Zurück zum Zitat Williams MB, Raghunathan P, More MJ, Seibert AJ, Kwan A, Lo JY, et al. Optimization of exposure parameters in full field digital mammography. Med Phys. 2008;35(6Part1):2414–23.CrossRef Williams MB, Raghunathan P, More MJ, Seibert AJ, Kwan A, Lo JY, et al. Optimization of exposure parameters in full field digital mammography. Med Phys. 2008;35(6Part1):2414–23.CrossRef
11.
Zurück zum Zitat The Royal Australian and New Zealand College of Radiologists. Guidelines for quality control testing for digital (CR & DR) mammography. V4.0. 2018. https://www.ranzcr.com. The Royal Australian and New Zealand College of Radiologists. Guidelines for quality control testing for digital (CR & DR) mammography. V4.0. 2018. https://​www.​ranzcr.​com.
12.
Zurück zum Zitat Xinming L, Chao-Jen L, Whitman GJ, Geiser WR, Youtao S, Ying Y, et al. Effects of exposure equalization on image signal-to-noise ratios in digital mammography: a simulation study with an anthropomorphic breast phantom. Med Phys. 2011;38(12):6489–501.CrossRef Xinming L, Chao-Jen L, Whitman GJ, Geiser WR, Youtao S, Ying Y, et al. Effects of exposure equalization on image signal-to-noise ratios in digital mammography: a simulation study with an anthropomorphic breast phantom. Med Phys. 2011;38(12):6489–501.CrossRef
13.
Zurück zum Zitat Borg M, Badr I, Royle G. The use of a figure-of-merit (FOM) for optimisation in digital mammography: a literature review. Radiat Prot Dosim. 2012;151(1):81–8.CrossRef Borg M, Badr I, Royle G. The use of a figure-of-merit (FOM) for optimisation in digital mammography: a literature review. Radiat Prot Dosim. 2012;151(1):81–8.CrossRef
14.
Zurück zum Zitat Delis H, Spyrou G, Costaridou L, Tzanakos G, Panayiotakis G. Evaluating the Figure of Merit in mammography utilizing Monte Carlo simulation. Nucl Instrum Methods Phys Res Sect A. 2007;580(1):493–6.CrossRef Delis H, Spyrou G, Costaridou L, Tzanakos G, Panayiotakis G. Evaluating the Figure of Merit in mammography utilizing Monte Carlo simulation. Nucl Instrum Methods Phys Res Sect A. 2007;580(1):493–6.CrossRef
15.
Zurück zum Zitat Shepard SJ, Wang J, Flynn M, Gingold E, Goldman L, Krugh K, et al. An exposure indicator for digital radiography: AAPM Task Group 116 (Executive Summary). Med Phys. 2009;36(7):2898–914.CrossRef Shepard SJ, Wang J, Flynn M, Gingold E, Goldman L, Krugh K, et al. An exposure indicator for digital radiography: AAPM Task Group 116 (Executive Summary). Med Phys. 2009;36(7):2898–914.CrossRef
16.
Zurück zum Zitat Gress DA, Dickinson RL, Erwin WD, Jordan DW, Kobistek RJ, Stevens DM, et al. AAPM medical physics practice guideline 6a: performance characteristics of radiation dose index monitoring systems. J Appl Clin Med Phys. 2017;18(4):12–22.CrossRef Gress DA, Dickinson RL, Erwin WD, Jordan DW, Kobistek RJ, Stevens DM, et al. AAPM medical physics practice guideline 6a: performance characteristics of radiation dose index monitoring systems. J Appl Clin Med Phys. 2017;18(4):12–22.CrossRef
17.
Zurück zum Zitat Koen L, Herbst C, Rae W. Computed radiography exposure indices in mammography. SA J Radiol. 2008;12(2):28–31.CrossRef Koen L, Herbst C, Rae W. Computed radiography exposure indices in mammography. SA J Radiol. 2008;12(2):28–31.CrossRef
Metadaten
Titel
Using aluminum for scatter control in mammography: preliminary work using measurements of CNR and FOM
verfasst von
Khaled Al Khalifah
Rob Davidson
Abel Zhou
Publikationsdatum
20.11.2019
Verlag
Springer Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 1/2020
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-019-00545-3

Weitere Artikel der Ausgabe 1/2020

Radiological Physics and Technology 1/2020 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.