Validation of the contralateral side as reference for selecting radial head implant sizes

verfasst von: Paul W. L. Ten Berg, Johannes G. G. Dobbe, Gerhard van Wolfswinkel, Simon D. Strackee, Geert J. Streekstra

Erschienen in: Surgical and Radiologic Anatomy | Ausgabe 7/2016

Abstract

Purpose

In arthroplasty of comminuted radial head fractures, the contralateral radial head diameter can be used as reference for implant selection. However, potential bilateral asymmetry may result in a mismatch of the implant with the native bone. Therefore, our purpose was to evaluate anatomical right-to-left differences of radial head diameters. We also compared conventional two-dimensional (2D) with three-dimensional (3D) measurements.

Methods

We used bilateral CT-scans from 25 intact proximal radius pairs of right-handed adult subjects to obtain 50 3D radial head models. After contralateral matching, diameters were calculated using a 3D-based method using an automated circle-fit in standardized cross-sections at the widest level midway through the radial head. The 3D-based diameters were compared to orthogonal line measurements in standard axial CT-slices.

Results

Three-dimensional analysis yielded a radial head diameter of 23.0 ± 1.7 mm. The dominant right side was significantly wider, with right-to-left differences of 0.2 ± 0.4 mm, with a maximum of 0.9 mm. The 2D-based diameter was 22.9 ± 1.7 mm, which was 0.1 ± 0.3 mm smaller compared to corresponding 3D-based diameter.

Conclusions

In healthy radial heads, the diameter was biased to the dominant right side, but individual differences were not larger than 1 mm. Compared to implant designs, in which diameter increments are usually 2 mm, this right-bias is not clinically relevant, as it would not affect implant selection. Therefore, the contralateral side can be considered a suitable reference. In clinical practice, the surgeon could estimate this diameter using standard axial CT slices, since its difference with the 3D-based evaluation was also relatively small compared to implant sizing increments.

Introduction

Radial head arthroplasty is a well-accepted procedure in the treatment of comminuted unreconstructable fractures or post-traumatic arthritis [1]. The implant should approximate the size of the native radial head to replicate the native joint kinematics and avoid postoperative pain, decreased range of motion, and eventually osteoarthritis of the capitulum [2, 3]. Adequate implant sizing is therefore an important aspect of arthroplasty.
Recent research focused on methods for predicting the native diameter of injured radial heads, as reference for implant selection [46]. Alolabi et al. [5] stated that the excised radial head, when available, should be used to select the implant. However, assessing the native bone shape becomes more difficult in cases of a high degree of comminution, or open fractures with missing bone. In these injuries, the surgeon can use the opposite healthy bone as reference to estimate the native head diameter. A prerequisite for this approach is the existence of sufficient bilateral anatomical symmetry. One cadaveric study showed that left-sided radial heads are similar to right-sided heads [7]. However, this study was limited by having a relatively small sample size (eight radial head pairs), and not taking dominance into account. Dominance is an important aspect in anatomic studies. One study, for example, analyzed bilateral symmetry of the radius, and showed that the dominant right side was generally longer [8].
The main purpose of this anatomic imaging study was to investigate the bilateral symmetry of normal radial head pairs obtained from 20 healthy right-handed volunteers. To this end, we used three-dimensional (3D) computed tomography (CT) analysis, providing detailed 3D information of bony anatomy, using standardized measurements [911]. We quantified right-to-left differences of the outer diameter of the paired 3D radial head models, and hypothesized that there was no bias between right and left. Our second purpose was to evaluate to what extent conventional measurements in standard axial CT slices, comparable with common practice, are in agreement with the aforementioned 3D-based measurements of the radial head.

Materials and methods

Data acquisition

In this study we used bilateral CT-scans of intact forearms including the proximal radius obtained from a historical group available from previously conducted experiments [8]. This group included 20 healthy right-handed volunteers (14 women and 6 men; average age 28 years; range 22–56 years). The volunteers had no history of elbow injury or other musculoskeletal disorders. The volunteers confirmed to be right-handed. To increase the number of male subjects, we added bilateral forearm CT scans from five male patients (average age 31 years; range 18–45 years) which were treated for a unilateral distal or midshaft radius malunion. The patients had also no history of elbow injury. For all proximal radii pathology was ruled out based on radiological reports and on reviewing the images again. All patients also confirmed to be right-handed. High-resolution bilateral CT scans (Philips Brilliance 64 CT scanner, Amsterdam, The Netherlands) were made using standardized methods (voxel size 0.45 × 0.45 × 0.45 mm, 120 kV, 150 mAs, pitch 0.6, Slice thickness 0.67 mm). This study was approved by our Human Research Committee. Informed consent of each volunteer was obtained prior to participation.

3D bone modeling

Twenty-five bilateral CT-scans were used to obtain 40 virtual 3D models of left and mirrored right radii based on custom made 3D image segmentation software [12]. Next, we selected the proximal radial head of the left side, and matched each to the opposite side in a semi-automated fashion based on image registration (Fig. 1a) [12, 13]. This matching enabled visual inspection of shape symmetry. It further enables selecting the same level for the cross-sectional diameter calculation of right and left radius models.

3D-based determination of the radial head diameter

A plane was used to cross-sect the widest region of the radial head and to subsequently determine the outer diameter. Since the right and left radial head models were aligned by registration, only one plane was sufficient to cross-sect both models from each volunteer. The 3D orientation of this plane was set in a standardized fashion. First, we fitted automatically an axial reference plane onto the most proximal part of the left radial head using three tangential points based on a custom written algorithm (Fig. 1b, c). A plane parallel to this reference plane was manually shifted to the widest point midway through the radial head. By cross-secting the paired radial head models, right and left 2D contours of the cortical bone were obtained (Fig. 2a). Next, we fitted automatically a 2D circle through each contour using a least-squares circle fit (Fig. 2b). The diameter of this best-fitting circle served as measure of the outer diameter of the radial head.
As additional measure, we used a second plane parallel to the reference plane to cross-sect the radial head at a more proximal level of the trough of the radial head (Fig. 1c) [14]. A similar best-fitting circle was used to obtain a diameter at this level. The 3D-based measurements were independently performed by two observers (research fellows; P. W. B and G. W).

2D-based determination of the radial head diameter

In the second phase, we measured the on-screen radial head diameter interactively at the widest level using orthogonal line measurements in standard axial CT slices, and compared them with the corresponding 3D-based diameters. First, the slice showing the widest part of the radial head was chosen. In this slice, we visually selected the minimum and maximum length (i.e., minimal and maximum 2D-based diameter) [14], and calculated the average of these two, which served as average 2-D based diameter (Fig. 3). The 2D-based measurements were independently performed by two observers (research fellows; P. W. B and G. W).

Statistical analysis

Statistical analyses of the measurements included the Shapiro Wilks W test as normality test, and determining the mean, and standard deviation (SD) for normally distributed data. The correlations between diameters of right and left sides and a best-fitting line through the points were calculated using linear regression modelling. Right-to-left differences were calculated by subtracting the 3-D based diameter obtained from the left side from the diameter obtained from the right side (right minus left diameter). A one-sample Student t test was used to test if right-to-left differences were significantly different from zero, indicating a left- or right-bias. A post hoc power analysis for one-sample t-test was used to calculate what mean right-to-left differences of diameters at the widest level could have been tested on significance with sufficient power. This power analysis requires input of the sample size (N = 25), comparison mean (=0), and standard deviation, while using an α-level of 0.05 and a power of 0.80.
Regarding the 2D-based diameters, minimum and maximum diameters were compared with a paired Student t test. Next, for each radius, we calculated the differences between the average 2D-based diameter and the corresponding 3D-based diameters (3D-based minus 2D-based diameter).
Interobserver agreement was calculated using the intraclass correlation coefficient (ICC) through a two-way mixed effects model with absolute agreement [15]. An ICC above 0.8 indicates very high interobserver agreement. A 5 % significance level was used for all analyses.

Results

In this section all anatomical measurements are expressed as mean ± standard deviation (SD) unless stated otherwise. All evaluation parameters were normally distributed. Based on the total sample, the 3D-based diameter at the widest level of the radial head was 23.0 ± 1.7 mm (males 24.5 ± 0.9 mm; females 21.9 ± 1.2 mm). The right–left correlation coefficient of radial head diameters at the widest level was very strong [r = 0.98 (p < 0.001)]. The best-fitting line through the points ran parallel to the line of identity (right diameter = left diameter), but with a slight bias towards the right side indicating that the dominant right side was generally wider (Fig. 4). The right-to-left difference of the diameter was 0.24 ± 0.4 mm, with a maximum of 0.9 mm (Table 1). Based on this latter standard deviation, there was sufficient power to detect a significant right–left bias if the mean difference was >0.18 mm. Our reported mean was larger than this cut-off, and statistically different from zero (p = 0.003). This confirmed that, in our sample, right-to-left differences were biased to the right side. Regarding anatomical measurements at level of the trough, the 3D-based diameter was slightly, but consistently smaller (22.6 ± 1.7 mm), with comparable right-to-left differences (0.31 ± 0.4 mm) which were also significantly different from zero (p = 0.001). A subgroup analysis including only the healthy volunteer data did not alter significance of right-to-left differences. The ICC between the two observers for the diameters as measured at the widest level and at level of the trough were both 0.99, indicating very high interobserver agreement.
Table 1
Right and left radial head diameters measured using 3D-based evaluation at the widest level in intact proximal radius pairs (N = 25)
Side
Diameter
Difference (R minus L)
Mean ± SD
Mean ± SD
|maximum|*
Right (R)
23.2 ± 1.7
  
Left (L)
22.9 ± 1.7
0.2 ± 0.4
0.9
* Absolute maximum
Significantly different from zero (p = 0.003)
Regarding the 2D-based diameters, the minimum diameter was 22.2 ± 1.5 mm, the maximum diameter 23.6 ± 1.8 mm (difference: p < 0.001), and the average diameter 22.9 ± 1.7 mm. The ICC between the two observers for the average diameter was also 0.99, indicating very high interobserver agreement. The average 2D-based diameter was significantly smaller than the corresponding 3D-based diameter (p = 0.013), with a difference of 0.1 ± 0.3 mm, with a maximum of 1.0 mm (Table 2).
Table 2
Radial head diameters measured using on 3D-based and 2D-based evaluation at the widest level in intact proximal radii (N = 50)
Evaluation type
Diameter
Difference (3D minus 2D)
Mean ± SD
Mean ± SD
|maximum|*
3D-based
23.0 ± 1.7
  
2D-based (minimum)
22.2 ± 1.5
0.8 ± 0.5
2.0
2D-based (maximum)
23.6 ± 1.8
−0.5 ± 0.4
1.8
2D-based (average)
22.9 ± 1.7
0.1 ± 0.3
1.0
* Absolute maximum
Significantly different from zero (all p < 0.01)

Discussion

In arthroplasty of the radial head, most radial head implant designs are circular and come in fixed ratios of height and diameter. They usually change with diameter increments of 2 mm [4, 16]. An important aspect in choosing implant size is estimating the native radial head diameter [4]. When using the contralateral side as reference, presence of sufficient anatomical bilateral symmetry is a prerequisite. Potential right-to-left differences may result in a mismatch between the implant and the native anatomy. In this article we evaluated whether the contralateral side is an appropriate reference for estimating the radial head diameter by using 3D-based measurements and comparing right and left radial head diameters in healthy individuals. Second, we evaluated to what extent conventional 2D-based measurements in the standard axial CT slices are in agreement with 3D-based measurements.
Two morphological studies of Swieszkowski et al. [7] and Koslowsky et al. [17] investigated right-to-left differences of radial head diameters within respectively 8 and 18 cadaveric elbow pairs. They did not found significant differences. Contrary, our data demonstrated small but significantly larger diameters for the dominant right side. This difference in findings between our study and the latter studies may be explained by the difference in applied measurements techniques. As there is no consensus about the optimal selection reference in radial head arthroplasty, some surgeons prefer measuring the radial head diameter at, e.g., the articular facet [5]. We demonstrated also a larger diameter for the dominant right side at this level.
Two studies investigated methods for selecting the diameter of a radial head implant using anatomical landmarks of an ipsilateral bone, instead of using the contralateral radial head [5, 6]. Alalobi et al. [5] used the curvature of the lesser sigmoid notch in the proximal ulna as reference. They found that the reliability of this landmark to estimate radial head diameter was only moderate. Leclerc et al. [6] used the width from the lateral aspect of the capitulum to the lateral trochlear ridge in the distal humerus to estimate the radial head diameter. Measurements showed very strong correlations, based on a best-fitting line. In some individuals, however, the radial diameter deviated approximately 3 mm from this best-fitting line. In the current 3D CT study, we showed that right-to-left diameter differences as observed in single individuals were not larger than 0.9 mm, suggesting that the contralateral side is a better reference. Compared to the commonly available implant sizing increments, the anatomical right-to-left differences can be considered minimal, as it would not affect selecting implant size. The uninjured contralateral side can therefore be considered a suitable reference for selecting the implant diameter in radial head arthroplasty, without the need to correct for hand dominance.
Considering 2D-based diameter assessment, we showed that the radial head is more elliptical rather than perfectly circular, since the minimum and maximum diameters within radial heads differed significantly. This is consistent with previous findings, which showed that radial head is not always circular, but often oval-shaped [1719]. Most radial head implants, however, are still circular, since elliptical implants would increase the technical difficulty in placing radial head implants although this may result in poor replication of the physiological kinematics of the radial head [14, 20]. Based on our results, if only the minimum diameter in an axial slice would be used as selection reference, the implant size could deviate up to 2.0 mm from the 3D-based diameter. This deviation is similar to the 2 mm diameter increments of implants, which may result in a difference between sizing up and sizing down. By averaging minimum and maximum diameters, the radial head diameter can be better approximated with a maximum difference of 1.0 mm compared to 3D-based diameter. The manual measurements of minimum and maximum diameters showed very high interobserver agreement, which is consistent with previous findings [4].
A limitation of a technique that uses CT images is lack of information of the articular cartilage because this is not visible. Therefore, our estimates could deviate slightly from true radial head size. Another important limitation of this study is that all participants were right-handed, which does not provide information about the right-to-left differences in left-handed individuals. Although not proven in this study, we do not expect larger right-to-left differences in left-handed individuals. In this study, we showed that our proposed 2D-based method for estimating the radial head diameter was comparable with the 3D-based method, and therefore suitable for clinical practice. However, future biomechanical and clinical studies are needed to evaluate to what extent the selected implant size actually restores the native elbow kinematics when using these measurement methods.
Besides choosing the appropriate implant diameter, the surgeon also has to restore bone length. The height of the implant relative to the surrounding bones can be altered intra-operatively by either removing additional native bone or by adjusting the collar size of the implant. Previous studies already assessed anatomical landmarks that guide the height of the radial head [2124]. One of these studies concluded that measurements based on contralateral images of the healthy elbow were accurate in predicting radial head implant length to avoid over lengthening [21].
In conclusion, our study showed that the right radial head diameters is slightly larger in right dominant individuals. However, in radial head arthroplasty, this right-bias is not clinically relevant, as it would not affect choosing implant size based on the contralateral side. The uninjured contralateral side can therefore be considered a suitable reference for selecting the implant diameter. For standard clinical practice where 3D-based evaluation may not be available, we recommend using standard axial CT slices to measure the average of the minimum and maximum diameter, for estimating the radial head outer diameter.

Compliance with ethical standards

Ethical standards

The procedures followed to obtain CT scans from healthy volunteers were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national). The ethics committee of Academic Medical Center (University of Amsterdam, Amsterdam, The Netherlands) gave approval (MEC 08/372).

Conflict of interest

The authors declare that they have no conflict of interest.

Funding

PWB received a PhD Grant (2014) from the Academic Medical Center (Amsterdam, Netherlands) supporting this research.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Weitere Produktempfehlungen anzeigen
Literatur
1.
Zurück zum Zitat Van Glabbeek F, Van Riet R, Verstreken J (2001) Current concepts in the treatment of radial head fractures in the adult. A clinical and biomechanical approach. Acta Orthop Belg 67:430–441PubMed Van Glabbeek F, Van Riet R, Verstreken J (2001) Current concepts in the treatment of radial head fractures in the adult. A clinical and biomechanical approach. Acta Orthop Belg 67:430–441PubMed
2.
Zurück zum Zitat Beingessner DM, Dunning CE, Gordon KD, Johnson JA, King GJ (2004) The effect of radial head excision and arthroplasty on elbow kinematics and stability. J Bone Joint Surg Am 86-A:1730–1739PubMed Beingessner DM, Dunning CE, Gordon KD, Johnson JA, King GJ (2004) The effect of radial head excision and arthroplasty on elbow kinematics and stability. J Bone Joint Surg Am 86-A:1730–1739PubMed
3.
Zurück zum Zitat Gupta GG, Lucas G, Hahn DL (1997) Biomechanical and computer analysis of radial head prostheses. J Shoulder Elbow Surg 6:37–48CrossRefPubMed Gupta GG, Lucas G, Hahn DL (1997) Biomechanical and computer analysis of radial head prostheses. J Shoulder Elbow Surg 6:37–48CrossRefPubMed
7.
Zurück zum Zitat Swieszkowski W, Skalski K, Pomianowski S, Kedzior K (2001) The anatomic features of the radial head and their implication for prosthesis design. Clin Biomech (Bristol, Avon) 16:880–887CrossRef Swieszkowski W, Skalski K, Pomianowski S, Kedzior K (2001) The anatomic features of the radial head and their implication for prosthesis design. Clin Biomech (Bristol, Avon) 16:880–887CrossRef
12.
Zurück zum Zitat Dobbe JG, Strackee SD, Schreurs AW, Jonges R, Carelsen B, Vroemen JC, Grimbergen CA, Streekstra GJ (2011) Computer-assisted planning and navigation for corrective distal radius osteotomy, based on pre- and intraoperative imaging. IEEE Trans Biomed Eng 58:182–190. doi:10.1109/TBME.2010.2084576 CrossRefPubMed Dobbe JG, Strackee SD, Schreurs AW, Jonges R, Carelsen B, Vroemen JC, Grimbergen CA, Streekstra GJ (2011) Computer-assisted planning and navigation for corrective distal radius osteotomy, based on pre- and intraoperative imaging. IEEE Trans Biomed Eng 58:182–190. doi:10.​1109/​TBME.​2010.​2084576 CrossRefPubMed
20.
Zurück zum Zitat Wegmann K, Hain MK, Ries C, Neiss WF, Muller LP, Burkhart KJ (2015) Do the radial head prosthesis components fit with the anatomical structures of the proximal radioulnar joint? Surg Radiol Anat 37:743–747. doi:10.1007/s00276-014-1407-2 CrossRefPubMed Wegmann K, Hain MK, Ries C, Neiss WF, Muller LP, Burkhart KJ (2015) Do the radial head prosthesis components fit with the anatomical structures of the proximal radioulnar joint? Surg Radiol Anat 37:743–747. doi:10.​1007/​s00276-014-1407-2 CrossRefPubMed
21.
22.
Zurück zum Zitat Beredjiklian PK, Nalbantoglu U, Potter HG, Hotchkiss RN (1999) Prosthetic radial head components and proximal radial morphology: a mismatch. J Shoulder Elbow Surg 8:471–475CrossRefPubMed Beredjiklian PK, Nalbantoglu U, Potter HG, Hotchkiss RN (1999) Prosthetic radial head components and proximal radial morphology: a mismatch. J Shoulder Elbow Surg 8:471–475CrossRefPubMed
24.
Zurück zum Zitat Frank SG, Grewal R, Johnson J, Faber KJ, King GJ, Athwal GS (2009) Determination of correct implant size in radial head arthroplasty to avoid overlengthening. J Bone Joint Surg Am 91:1738–1746. doi:10.2106/JBJS.H.01161 CrossRefPubMed Frank SG, Grewal R, Johnson J, Faber KJ, King GJ, Athwal GS (2009) Determination of correct implant size in radial head arthroplasty to avoid overlengthening. J Bone Joint Surg Am 91:1738–1746. doi:10.​2106/​JBJS.​H.​01161 CrossRefPubMed
Metadaten
Titel
Validation of the contralateral side as reference for selecting radial head implant sizes
verfasst von
Paul W. L. Ten Berg
Johannes G. G. Dobbe
Gerhard van Wolfswinkel
Simon D. Strackee
Geert J. Streekstra
Publikationsdatum
21.01.2016
Verlag
Springer Paris
Erschienen in
Surgical and Radiologic Anatomy / Ausgabe 7/2016
Print ISSN: 0930-1038
Elektronische ISSN: 1279-8517
DOI
https://doi.org/10.1007/s00276-016-1625-x

Weitere Artikel der Ausgabe 7/2016

Surgical and Radiologic Anatomy 7/2016 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.