This work studied the influence of PET cardiac gating on the LV cardiac function assessed from hybrid PET/MR exams of patients with known coronary disease. Correlation and agreement analyses were performed on LV contractile parameters obtained by cardiac MRI and PET images reconstructed with three gating methods. Using MR values as reference, PET-based values showed high correlation, slight to moderate trend toward underestimation, and wide limits of agreement. The most significant differences were found in the gating method that did not perform well in the setting of large R-R intervals variability (STD), while a similar performance was observed in the methods using uniform adjustment of gates width per beat with the beat acceptance window (BR-STD), and fixed gate width along all the beats (FW).
Cardiac gating
Results from the STD gating approach showed the importance of the beat acceptance window. The inclusion of irregular R-R intervals in the gating process drives to merging PET counts from dissimilar cardiac phases, leading to an effective smoothing of the myocardium and subsequently affecting the estimation of left ventricular volumes. The similar performance of BR-STD and FW in this study indicated that the effects associated with the selection of the gate width could not be clearly discerned, potentially due to the higher impact of differences linked to the multimodal comparison PET-MRI, as discussed below.
Even though the STD-BR method avoids abnormal R-R intervals, the rejection of PET counts increases image noise, affecting the accuracy of the measurement. For instance, in one of the subjects, only 12 out of 40 minutes was used to reconstruct the PET image due to 40% of abnormal R-R intervals (equivalent to rejection of 66% of the total time)—an effect which would be even more detrimental in shorter exams. Extra efforts are needed to reduce this loss in counts: exhaustive quality controls to the ECG signal and R peak detected during acquisition, improvements of R peak detection algorithms (in particular, incrementing robustness to MR artefacts
11,12), generalization of ECG gating taking thoughtfully into account arrhythmic heartbeats as well.
In particular, it is important to emphasize the need for a strict quality control of the ECG signal and its processing in the clinical environment, since these artefacts affect not only the cardiac PET gating method, but also the acquisition of the cardiac MRI sequences that use this signal as a trigger.
12 Further investigations for alternatives to ECG-based cardiac gating are also encouraged.
13,14
Multimodal cross-validation
Comparable works
15‐20 studied FDG-PET and MRI-based values of EDV, ESV and EF obtained using PET/MR, PET/CT, PET and MRI systems. Table
3 summarizes the main results. In general, we can see high correlations, but varied biases: (− 1.1-28.4)% for EDV, (− 5.9-29.6)% for ESV, and (− 12.0-13.6)% for EF, and principally wide limits of agreement: (44.3-95.4)% for EDV, (25.5-153.9)% for ESV, and (26.1-142.5)% for EF. Another comparative study
21 between nitrogen-13 ammonia PET and MRI-based values reported higher correlations and lower biases, but the limits of agreements are still wide (33.3% for EDV, 48.2% for ESV, and 29.7% for EF). These comparisons support the evidence that cardiac function measurements are not interchangeable between modalities.
Table 3
Comparison of correlation and agreement analyses in similar studies
Khorsand 2003 | 20 | PET | In-house | Phillips | 8 | 0.6 | 0.92 | 10.1 | 95.4 |
Schäfer 2004 | 42 | PET | QGS | Phillips | 8 | 1.0 | 0.94 | 0.3 | 44.3 |
4D-MSPECT | 1.0 | 0.94 | − 1.1 | 46.6 |
Slart 2004 | 38 | PET/CT | QGS | MASS | 16 | 0.9 | 0.91 | 15.0 | 56.0 |
Li 2014 | 89 | PET/CT | QGS | MASS | 8 | 0.9 | 0.92 | 12.7 | 65.9 |
4D-MSPECT | 1.1 | 0.93 | 1.0 | 66.6 |
Lücke 2017 | 29 | PET/MR | Corridor4DM | cmr42 | 16 | 1.0 | 0.95 | 16.5 | 62.3 |
Yao 2019 | 76 | PET/CT | QGS | MASS | 8 | 0.8 | 0.91 | 28.4 | 70.6 |
ECTB | 0.8 | 0.86 | 25.5 | 88.4 |
4D-MSPECT | 0.9 | 0.89 | 18.5 | 80.2 |
Our study STD | 30 | PET/MR | Munich Heart | Munich Heart | 8 | 1.2 | 0.75 | 12.8 | 67.8 |
Our study STD-BR | 1.2 | 0.81 | 9.5 | 59.3 |
Our study FW | 1.3 | 0.80 | 9.4 | 63.9 |
Khorsand 2003 | 0.6 | 0.93 | 9.8 | 153.9 | 0.6 | 0.85 | 6.5 | 85.2 |
Schäfer 2004 | 1.0 | 0.95 | − 5.9 | 25.5 | 0.7 | 0.94 | 11.4 | 51.4 |
1.0 | 0.95 | − 4.2 | 55.1 | 0.7 | 0.90 | 2.9 | 57.1 |
Slart 2004 | 0.9 | 0.94 | 13.7 | 58.2 | 1.0 | 0.96 | 10.3 | 26.1 |
Li 2014 | 0.9 | 0.92 | 11.7 | 83.4 | 0.9 | 0.76 | 0.3 | 115.3 |
1.0 | 0.94 | − 4.4 | 80.7 | 0.7 | 0.75 | 12.5 | 108.6 |
Lücke 2017 | 1.2 | 0.97 | 19.6 | 104.8 | 1.0 | 0.91 | − 4.1 | 68.8 |
Yao 2019 | 0.9 | 0.93 | 28.1 | 77.8 | 1.0 | 0.79 | − 3.6 | 113.8 |
0.7 | 0.85 | 29.6 | 108.0 | 0.7 | 0.62 | − 12.0 | 142.5 |
0.9 | 0.91 | 18.1 | 92.6 | 0.8 | 0.76 | 0.7 | 105.8 |
Our study STD | 1.0 | 0.92 | 2.0 | 65.1 | 0.7 | 0.79 | 13.6 | 70.6 |
Our study STD-BR | 1.0 | 0.92 | 4.3 | 67.1 | 0.8 | 0.91 | 5.0 | 49.2 |
Our study FW | 1.1 | 0.92 | 0.6 | 69.9 | 0.7 | 0.87 | 8.8 | 57.8 |
Although such a setting should ideally be the perfect scenario for multimodal comparisons with comparable physiological conditions,
22 several multiple factors contribute to the observed variability. In addition to potential intra-/inter-observer variability, the discrepancies between the two modalities might be also explained by temporal and spatial resolutions, differences in the geometrical model of the heart between MRI and PET, and volume variation with heart rate changes.
Since the effective number of PET gates is three times less than in MRI, the measured PET-derived LV volumes are potentially undersampled (smoothing effect), affecting the accuracy of the parameters. Studies compared the volumes obtained with different numbers of gates (8, 16, and 32) in SPECT and PET images
23‐25: when using only 8 gates, smaller EDV, larger ESV, and a corresponding lower EF was observed, with the highest impact found in ESV measurement (median changes among those studies of 3% vs − 10%, in EDV and ESV, respectively). Nevertheless, temporal resolution issues might not be sufficient to explain the discrepancies in our cohort since the highest differences were found in EDV values.
The delineation of short and long axes, of base and apex of the heart in both modalities, and the lack of compatibility between them may also increase variability in the definition of the geometric model used to measure LV volume. In MRI, the axes are defined prospectively before the acquisition of the cine images, while in PET, the task is retrospectively carried out by physicians as a part of the data post-processing, adding to the inter-observer variability from two fully different directions. Another plausible factor of the discrepancies is an inaccurate FDG uptake-based PET contouring due to the severity myocardial damage, however, most of the LV myocardial segments of this cohort were classified as viable by FDG-PET assessment.
The lack of a perfect time matching (different time and duration) between PET and MRI acquisition in the exam (total scan time in PET vs several minutes in MRI) potentially leads to parameters that may not fully coincide. Two studies
26,27 analyzed the heart rate dependency of LV volumes of cardiac phantoms in SPECT/CT and CT images and a range 40-100 beats per minute. They found differences in EF of up to 2.5% in the CT images with low temporal resolution (175 ms per gate) between 60 and 80 beats per minute, but on average of only 1% with higher temporal resolutions (75 ms in CT and 38 ms in SPECT per gate). Regarding our study, even though PET temporal resolution was on average 120 ms (62 beats per minute), the spatial resolution was lower than with CT images, and thus intra-scan heart rate variation might be considered as part of the source for the discrepancies.
Even though the current clinical CINE MRI sequences provide images with the highest in-plane spatial and temporal resolutions, clinical 2-dimensional acquisition schemes limit the accuracy of inter-plane information. Due to the sequential, slice-by-slice acquisition, 2-dimensional images are more susceptible to motion, and the thickness of slices is acquired with worse spatial resolution than in the in-plane situation to cover the entire LV chamber in a feasible clinical acquisition time. Consequently, the use of cine MR values as the reference should be re-evaluated. In this sense, 3-dimensional acquisition for cine images seems to be the next step. However, clinically suitable schemes for its implementation are still in development and might hold other limitations in store.
28