Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 2/2016

01.04.2016 | Original Research

Comparison of foot finding methods for deriving instantaneous pulse rates from photoplethysmographic signals

verfasst von: Mathilde C. Hemon, Justin P. Phillips

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

The suitability of different methods of finding the foot point of a pulse as measured using earlobe photoplethysmography during stationary conditions was investigated. Instantaneous pulse period (PP) values from PPG signals recorded from the ear in healthy volunteer subjects were compared with simultaneous ECG-derived cardiac periods (RR interval). Six methods of deriving pulse period were used, each based on a different method of finding specific landmark points on the PPG waveform. These methods included maximum and minimum value, maximum first and second derivative, ‘intersecting tangents’ and ‘diastole patching’ methods. Selected time domain HRV variables were also calculated from the PPG signals obtained using multiple methods and compared with ECG-derived HRV variables. The correlation between PPG and ECG was greatest for the intersecting tangents method compared to the other methods (RMSE = 5.69 ms, r 2 = 0.997). No significant differences between PP and RR were seen for all PPG methods, however the PRV variables derived using all methods showed significant differences to HRV, attributable to the sensitivity of PRV parameters to pulse transients and artifacts. The results suggest that the intersecting tangents method shows the most promise for extracting accurate pulse rate variability data from PPG datasets. This work has applications in other areas where pulse arrival time is a key measurement including pulse wave velocity assessment.
Literatur
1.
Zurück zum Zitat Shelley KH. Photoplethysmography: beyond the calculation of arterial oxygen saturation and heart rate. Anesth Analg. 2007;105:S31–6.CrossRefPubMed Shelley KH. Photoplethysmography: beyond the calculation of arterial oxygen saturation and heart rate. Anesth Analg. 2007;105:S31–6.CrossRefPubMed
2.
Zurück zum Zitat Tamura T, Maeda Y, Sekine M, Yoshida M. Wearable photoplethysmographic sensors—past and present. Electronics. 2014;3:282–302.CrossRef Tamura T, Maeda Y, Sekine M, Yoshida M. Wearable photoplethysmographic sensors—past and present. Electronics. 2014;3:282–302.CrossRef
3.
Zurück zum Zitat Acharya UR, Joseph KP, Kannathal M, Lim CM, Suri JS. Heart rate variability: a review. Med Biol Eng Comput. 2006;44:1031–51.CrossRef Acharya UR, Joseph KP, Kannathal M, Lim CM, Suri JS. Heart rate variability: a review. Med Biol Eng Comput. 2006;44:1031–51.CrossRef
4.
Zurück zum Zitat Task force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Heart rate variability: standards of measurement, physiological interpretation and clinical use. Circulation. 1996;93:1043–65.CrossRef Task force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Heart rate variability: standards of measurement, physiological interpretation and clinical use. Circulation. 1996;93:1043–65.CrossRef
5.
Zurück zum Zitat Weippert M, Kumar M, Kreuzfeld S, Arndt D, Rieger A, Stoll R. Comparison of three mobile devices for measuring R–R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system. Eur J Appl Physiol. 2010;109:779–86.CrossRefPubMed Weippert M, Kumar M, Kreuzfeld S, Arndt D, Rieger A, Stoll R. Comparison of three mobile devices for measuring R–R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system. Eur J Appl Physiol. 2010;109:779–86.CrossRefPubMed
6.
Zurück zum Zitat Lu S, Zhao H, Ju K, Shin K, Lee M, Shelley KH, Chon KH. Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information? J Clin Monit Comput. 2008;22:23–9.CrossRefPubMed Lu S, Zhao H, Ju K, Shin K, Lee M, Shelley KH, Chon KH. Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information? J Clin Monit Comput. 2008;22:23–9.CrossRefPubMed
7.
Zurück zum Zitat Bolanos M, Nazeran H, Haltiwanger E. Comparison of heart rate variability signal features derived from electrocardiography and photoplethysmography in healthy individuals. Conf Proc IEEE Eng Med Biol Soc. 2006;1:4289–94.PubMed Bolanos M, Nazeran H, Haltiwanger E. Comparison of heart rate variability signal features derived from electrocardiography and photoplethysmography in healthy individuals. Conf Proc IEEE Eng Med Biol Soc. 2006;1:4289–94.PubMed
8.
Zurück zum Zitat Constant I, Laude D, Murat I, Elghozi J. Pulse rate variability is not a surrogate for heart rate variability. Clin Sci. 1999;97:391–7.CrossRefPubMed Constant I, Laude D, Murat I, Elghozi J. Pulse rate variability is not a surrogate for heart rate variability. Clin Sci. 1999;97:391–7.CrossRefPubMed
9.
Zurück zum Zitat Gil E, Orini M, Bailón R, Vergara JM, Mainardi L, Laguna P. Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol Meas. 2010;31:1271–90.CrossRefPubMed Gil E, Orini M, Bailón R, Vergara JM, Mainardi L, Laguna P. Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol Meas. 2010;31:1271–90.CrossRefPubMed
10.
Zurück zum Zitat Khandoker AH, Karmakar CK, Palaniswami M. Comparison of pulse rate variability with heart rate variability during obstructive sleep apnea. Med Eng Phys. 2011;33:204–9.CrossRefPubMed Khandoker AH, Karmakar CK, Palaniswami M. Comparison of pulse rate variability with heart rate variability during obstructive sleep apnea. Med Eng Phys. 2011;33:204–9.CrossRefPubMed
11.
Zurück zum Zitat Lu G, Yang F, Taylor JA, Stein JF. A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjects. J Med Eng Technol. 2009;33:634–41.CrossRefPubMed Lu G, Yang F, Taylor JA, Stein JF. A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjects. J Med Eng Technol. 2009;33:634–41.CrossRefPubMed
12.
Zurück zum Zitat Selvaraj N, Jaryal A, Santhosh J, Deepak K, Anand S. Assessment of heart rate variability derived from finger-tip photoplethysmography as compared to electrocardiography. J Med Eng Technol. 2008;32:479–84.CrossRefPubMed Selvaraj N, Jaryal A, Santhosh J, Deepak K, Anand S. Assessment of heart rate variability derived from finger-tip photoplethysmography as compared to electrocardiography. J Med Eng Technol. 2008;32:479–84.CrossRefPubMed
13.
Zurück zum Zitat Wong J, Lu W, Wu K, Liu M, Chen G, Kuo C. A comparative study of pulse rate variability and heart rate variability in healthy subjects. J Clin Monit Comput. 2012;26:107–14.CrossRefPubMed Wong J, Lu W, Wu K, Liu M, Chen G, Kuo C. A comparative study of pulse rate variability and heart rate variability in healthy subjects. J Clin Monit Comput. 2012;26:107–14.CrossRefPubMed
14.
Zurück zum Zitat Chui YC, Arand PW, Shroff SG, Feldman T, Carroll JD. Determination of pulse wave velocities with computerized algorithms. Am Heart J. 1991;121:1460–70.CrossRef Chui YC, Arand PW, Shroff SG, Feldman T, Carroll JD. Determination of pulse wave velocities with computerized algorithms. Am Heart J. 1991;121:1460–70.CrossRef
15.
Zurück zum Zitat Kazanavicius E, Gircys R, Vrubliauskas A. Mathematical methods for determining the foot point of the arterial pulse wave and evaluation of proposed methods. Inf Technol Control. 2005;34:29–36. Kazanavicius E, Gircys R, Vrubliauskas A. Mathematical methods for determining the foot point of the arterial pulse wave and evaluation of proposed methods. Inf Technol Control. 2005;34:29–36.
16.
Zurück zum Zitat Millasseau SC, Stewart AD, Patel SJ, Redwood SR, Chowienczyk PJ. Evaluation of carotid-femoral pulse wave velocity: influence of timing algorithm and heart rate. Hypertension. 2005;45:222–6.CrossRefPubMed Millasseau SC, Stewart AD, Patel SJ, Redwood SR, Chowienczyk PJ. Evaluation of carotid-femoral pulse wave velocity: influence of timing algorithm and heart rate. Hypertension. 2005;45:222–6.CrossRefPubMed
17.
Zurück zum Zitat Phillips JP, Kyriacou PA. Comparison of methods for determining pulse arrival time from Doppler and photoplethysmography signals. Conf Proc IEEE Eng Med Biol Soc. 2014;1:3809–12. Phillips JP, Kyriacou PA. Comparison of methods for determining pulse arrival time from Doppler and photoplethysmography signals. Conf Proc IEEE Eng Med Biol Soc. 2014;1:3809–12.
18.
Zurück zum Zitat Vardoulis O, Papaioannou TG, Stergiopulos N. Validation of a novel and existing algorithms for the estimation of pulse transit time: advancing the accuracy in pulse wave velocity measurement. Am J Physiol Heart Circ Physiol. 2013;304:H1558–67.CrossRefPubMed Vardoulis O, Papaioannou TG, Stergiopulos N. Validation of a novel and existing algorithms for the estimation of pulse transit time: advancing the accuracy in pulse wave velocity measurement. Am J Physiol Heart Circ Physiol. 2013;304:H1558–67.CrossRefPubMed
19.
Zurück zum Zitat Nichols W, O’Rourke M, Vlachopoulos C. McDonald’s blood flow in arteries, sixth edition: theoretical, experimental and clinical principles. Boca Raton: CRC Press; 2011. Nichols W, O’Rourke M, Vlachopoulos C. McDonald’s blood flow in arteries, sixth edition: theoretical, experimental and clinical principles. Boca Raton: CRC Press; 2011.
20.
Zurück zum Zitat Rybynok V, May JM, Budidha K, Kyriacou PA. Design and development of a novel multi-channel photoplethysmographic research system. In: IEEE point-of-care healthcare technologies. 2013. Rybynok V, May JM, Budidha K, Kyriacou PA. Design and development of a novel multi-channel photoplethysmographic research system. In: IEEE point-of-care healthcare technologies. 2013.
21.
Zurück zum Zitat O’Haver TC, Begley T. Signal-to-noise ratio in higher order derivative spectrometry. Anal Chem. 1981;53:1876–8.CrossRef O’Haver TC, Begley T. Signal-to-noise ratio in higher order derivative spectrometry. Anal Chem. 1981;53:1876–8.CrossRef
22.
Zurück zum Zitat Phillips JP, Kyriacou PA, Jones DP, Shelley KH, Langford RM. Pulse oximetry and photoplethysmographic waveform analysis of the esophagus and bowel. Curr Opin Anesthesiol. 2008;21:779–83.CrossRef Phillips JP, Kyriacou PA, Jones DP, Shelley KH, Langford RM. Pulse oximetry and photoplethysmographic waveform analysis of the esophagus and bowel. Curr Opin Anesthesiol. 2008;21:779–83.CrossRef
Metadaten
Titel
Comparison of foot finding methods for deriving instantaneous pulse rates from photoplethysmographic signals
verfasst von
Mathilde C. Hemon
Justin P. Phillips
Publikationsdatum
01.04.2016
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 2/2016
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9695-6

Weitere Artikel der Ausgabe 2/2016

Journal of Clinical Monitoring and Computing 2/2016 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.