Erschienen in: Oral and Maxillofacial Surgery 4/2013

01.12.2013 | Review Article

Fixation of mandibular angle fractures: in vitro biomechanical assessments and computer-based studies

verfasst von: Bruno Ramos Chrcanovic

Erschienen in: Oral and Maxillofacial Surgery | Ausgabe 4/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of this study was to review the literature regarding the evolution of current thoughts on fixation of mandibular angle fractures (MAFs), based on in vitro biomechanical assessments and computer-based studies.

Methods

An electronic search in PubMed was undertaken in August 2012. The titles and abstracts from these results were read to identify studies within the selection criteria. Eligibility criteria included studies from the last 30 years (from 1983 onwards).

Results

The search strategy initially identified 767 studies. Thirty-one studies were identified without repetition within the selection criteria. Two articles showing significance in the development of treatment techniques was included. Additional hand searching yielded five additional papers. Thus, a total of 38 studies were included.

Conclusions

The osteosynthesis positions as well as the plating technique play important roles in the stability of MAF repair. The only in vitro study evaluating the use of wire osteosynthesis concluded that wires placed through the lower border approach would provide greater stability than those at the upper border. Many studies indicate that the use of two miniplates avoids (or decreases) lateral displacement of the lower mandibular border and opening of the inferior fracture gap. Some studies even suggest that the use of two miniplates may be considered a more “rigid” fixation technique for MAFs than the use of a reconstruction plate. When using two miniplates, the biplanar plate orientation provides greater biomechanical stability than the monoplanar one. However, despite its greater biomechanical stability, the two-miniplate technique has some disadvantages that should also be taken into account. Studies with biodegradable plates suggest the use of at least two plates for each MAF. There are few studies with compression plates, and they have not yet reached a consensus. The solitary lag screw proved to withstand the functional loading of the mandible; however, only few biomechanical assessments were performed. In vitro studies have shown good biomechanical stability with the use of 3-D grid plates. The use of malleable miniplates alone is not sufficient to withstand the early postoperative bite force. Some studies suggest that the segment of the tension band miniplate located at the distal fragment of the MAF should be fixed with three screws. The studies also showed some limitations. None considered the stabilization of the fracture site afforded by the masseter–pterygoid muscle pouch. Most of the studies did not evaluate plating system strength in the long term and therefore did not observe the effect of resorption on the strength of the different biodegradable plating systems. Another limitation of many studies is the absence of a control group. A confounding factor that could not be tested in in vitro investigations is the additional resistance to displacement of jagged fracture margins present in the human fracture.
Literatur
1.
Zurück zum Zitat Chrcanovic BR, Freire-Maia B, Souza LN, Araújo VO, Abreu MHNG (2004) Facial fractures: a 1-year retrospective study in a hospital in Belo Horizonte. Braz Oral Res 18:322–328PubMedCrossRef Chrcanovic BR, Freire-Maia B, Souza LN, Araújo VO, Abreu MHNG (2004) Facial fractures: a 1-year retrospective study in a hospital in Belo Horizonte. Braz Oral Res 18:322–328PubMedCrossRef
2.
Zurück zum Zitat Chrcanovic BR, Abreu MH, Freire-Maia B, Souza LN (2010) Facial fractures in children and adolescents: a retrospective study of 3 years in a hospital in Belo Horizonte, Brazil. Dent Traumatol 26:262–270PubMedCrossRef Chrcanovic BR, Abreu MH, Freire-Maia B, Souza LN (2010) Facial fractures in children and adolescents: a retrospective study of 3 years in a hospital in Belo Horizonte, Brazil. Dent Traumatol 26:262–270PubMedCrossRef
3.
Zurück zum Zitat Chrcanovic BR, Souza LN, Freire-Maia B, Abreu MH (2010) Facial fractures in the elderly: a retrospective study in a hospital in Belo Horizonte, Brazil. J Trauma 69:E73–E78PubMedCrossRef Chrcanovic BR, Souza LN, Freire-Maia B, Abreu MH (2010) Facial fractures in the elderly: a retrospective study in a hospital in Belo Horizonte, Brazil. J Trauma 69:E73–E78PubMedCrossRef
4.
Zurück zum Zitat Zandi M, Khayati A, Lamei A, Zarei H (2011) Maxillofacial injuries in western Iran: a prospective study. Oral Maxillofac Surg 15:201–209PubMedCrossRef Zandi M, Khayati A, Lamei A, Zarei H (2011) Maxillofacial injuries in western Iran: a prospective study. Oral Maxillofac Surg 15:201–209PubMedCrossRef
5.
Zurück zum Zitat Chrcanovic BR (2012) Factors influencing the incidence of maxillofacial fractures. Oral Maxillofac Surg 16:3–17PubMedCrossRef Chrcanovic BR (2012) Factors influencing the incidence of maxillofacial fractures. Oral Maxillofac Surg 16:3–17PubMedCrossRef
6.
Zurück zum Zitat Chrcanovic BR, Abreu MH, Freire-Maia B, Souza LN (2012) 1,454 mandibular fractures: a 3-year study in a hospital in Belo Horizonte, Brazil. J Craniomaxillofac Surg 40:116–123PubMedCrossRef Chrcanovic BR, Abreu MH, Freire-Maia B, Souza LN (2012) 1,454 mandibular fractures: a 3-year study in a hospital in Belo Horizonte, Brazil. J Craniomaxillofac Surg 40:116–123PubMedCrossRef
7.
Zurück zum Zitat Chrcanovic BR, Custódio AL (2010) Considerations of mandibular angle fractures during and after surgery for removal of third molars: a review of the literature. Oral Maxillofac Surg 14:71–80PubMedCrossRef Chrcanovic BR, Custódio AL (2010) Considerations of mandibular angle fractures during and after surgery for removal of third molars: a review of the literature. Oral Maxillofac Surg 14:71–80PubMedCrossRef
8.
Zurück zum Zitat Marciani RD, Anderson GE, Gonty AA (1994) Treatment of mandibular angle fractures: transoral internal wire fixation. J Oral Maxillofac Surg 52:752–756PubMedCrossRef Marciani RD, Anderson GE, Gonty AA (1994) Treatment of mandibular angle fractures: transoral internal wire fixation. J Oral Maxillofac Surg 52:752–756PubMedCrossRef
9.
Zurück zum Zitat Esen A, Ataoğlu H, Gemi L (2008) Comparison of stability of titanium and absorbable plate and screw fixation for mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:806–811PubMedCrossRef Esen A, Ataoğlu H, Gemi L (2008) Comparison of stability of titanium and absorbable plate and screw fixation for mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:806–811PubMedCrossRef
10.
Zurück zum Zitat Wittenberg J, Mukherjee D, Smith B, Kruse R (1997) Biomechanical evaluation of new fixation devices for mandibular angle fractures. Int J Oral Maxillofac Surg 26:68–73PubMedCrossRef Wittenberg J, Mukherjee D, Smith B, Kruse R (1997) Biomechanical evaluation of new fixation devices for mandibular angle fractures. Int J Oral Maxillofac Surg 26:68–73PubMedCrossRef
11.
Zurück zum Zitat Champy M, Loddé JP, Schmitt R, Jaeger JH, Muster D (1978) Mandibular osteosynthesis by miniature screwed plates via a buccal approach. J Maxillofac Surg 6:14–21PubMedCrossRef Champy M, Loddé JP, Schmitt R, Jaeger JH, Muster D (1978) Mandibular osteosynthesis by miniature screwed plates via a buccal approach. J Maxillofac Surg 6:14–21PubMedCrossRef
12.
Zurück zum Zitat Schierle HP, Schmelzeisen R, Rahn B, Pytlik C (1997) One- or two-plate fixation of mandibular angle fractures? J Craniomaxillofac Surg 25:162–168PubMedCrossRef Schierle HP, Schmelzeisen R, Rahn B, Pytlik C (1997) One- or two-plate fixation of mandibular angle fractures? J Craniomaxillofac Surg 25:162–168PubMedCrossRef
13.
Zurück zum Zitat Perren SM, Rahn B, Cordey J (1975) Mechanik und Biologie der Frakturheilung. Fortschr Kiefer Gesichtschir 19:33–37PubMed Perren SM, Rahn B, Cordey J (1975) Mechanik und Biologie der Frakturheilung. Fortschr Kiefer Gesichtschir 19:33–37PubMed
14.
Zurück zum Zitat Shetty V, McBrearty D, Fourney M, Caputo AA (1995) Fracture line stability as a function of the internal fixation system: an in vitro comparison using a mandibular angle fracture model. J Oral Maxillofac Surg 53:791–801PubMedCrossRef Shetty V, McBrearty D, Fourney M, Caputo AA (1995) Fracture line stability as a function of the internal fixation system: an in vitro comparison using a mandibular angle fracture model. J Oral Maxillofac Surg 53:791–801PubMedCrossRef
15.
Zurück zum Zitat Lovald ST, Wagner JD, Baack B (2009) Biomechanical optimization of bone plates used in rigid fixation of mandibular fractures. J Oral Maxillofac Surg 67:973–985PubMedCrossRef Lovald ST, Wagner JD, Baack B (2009) Biomechanical optimization of bone plates used in rigid fixation of mandibular fractures. J Oral Maxillofac Surg 67:973–985PubMedCrossRef
16.
Zurück zum Zitat Kimsal J, Baack B, Candelaria L, Khraishi T, Lovald S (2011) Biomechanical analysis of mandibular angle fractures. J Oral Maxillofac Surg 69:3010–3014PubMedCrossRef Kimsal J, Baack B, Candelaria L, Khraishi T, Lovald S (2011) Biomechanical analysis of mandibular angle fractures. J Oral Maxillofac Surg 69:3010–3014PubMedCrossRef
17.
Zurück zum Zitat Tams J, Otten B, Van Loon J-P, Bos RRM (1999) A computer study of fracture mobility and plate strains of biodegradable plates for fixation of mandibular fractures. J Oral Maxillofac Surg 57:973–981PubMedCrossRef Tams J, Otten B, Van Loon J-P, Bos RRM (1999) A computer study of fracture mobility and plate strains of biodegradable plates for fixation of mandibular fractures. J Oral Maxillofac Surg 57:973–981PubMedCrossRef
18.
Zurück zum Zitat Tams J, Van Loon JP, Otten B, Bos RR (2001) A computer study of biodegradable plates for internal fixation of mandibular angle fractures. J Oral Maxillofac Surg 59:404–407PubMedCrossRef Tams J, Van Loon JP, Otten B, Bos RR (2001) A computer study of biodegradable plates for internal fixation of mandibular angle fractures. J Oral Maxillofac Surg 59:404–407PubMedCrossRef
19.
Zurück zum Zitat Cox T, Kohn MW, Impelluso T (2003) Computerized analysis of resorbable polymer plates and screws for the rigid fixation of mandibular angle fractures. J Oral Maxillofac Surg 61:481–487PubMedCrossRef Cox T, Kohn MW, Impelluso T (2003) Computerized analysis of resorbable polymer plates and screws for the rigid fixation of mandibular angle fractures. J Oral Maxillofac Surg 61:481–487PubMedCrossRef
20.
Zurück zum Zitat Feller KU, Schneider M, Hlawitschka M, Pfeifer G, Lauer G, Eckelt U (2003) Analysis of complications in fractures of the mandibular angle: a study with finite element computation and evaluation of data of 277 patients. J Craniomaxillofac Surg 31:290–295PubMedCrossRef Feller KU, Schneider M, Hlawitschka M, Pfeifer G, Lauer G, Eckelt U (2003) Analysis of complications in fractures of the mandibular angle: a study with finite element computation and evaluation of data of 277 patients. J Craniomaxillofac Surg 31:290–295PubMedCrossRef
21.
Zurück zum Zitat Michelet FX, Deymes J, Dessus B (1973) Osteosynthesis with miniaturized screwed plates in maxillo-facial surgery. J Maxillofac Surg 1:79–84PubMedCrossRef Michelet FX, Deymes J, Dessus B (1973) Osteosynthesis with miniaturized screwed plates in maxillo-facial surgery. J Maxillofac Surg 1:79–84PubMedCrossRef
22.
Zurück zum Zitat Fisher IT, Cleaton-Jones PE, Lownie JF (1990) Relative efficiencies of various wiring configurations commonly used in open reductions of fractures of the angle of the mandible. Oral Surg Oral Med Oral Pathol 70:10–17PubMedCrossRef Fisher IT, Cleaton-Jones PE, Lownie JF (1990) Relative efficiencies of various wiring configurations commonly used in open reductions of fractures of the angle of the mandible. Oral Surg Oral Med Oral Pathol 70:10–17PubMedCrossRef
23.
Zurück zum Zitat Kroon FHM, Mathisson M, Cordey JA, Rahn BA (1991) The use of miniplates in mandibular fractures. An in vitro study. J Cranio Maxillofac Surg 19:199–204CrossRef Kroon FHM, Mathisson M, Cordey JA, Rahn BA (1991) The use of miniplates in mandibular fractures. An in vitro study. J Cranio Maxillofac Surg 19:199–204CrossRef
24.
Zurück zum Zitat Shetty V, Caputo A (1992) Biomechanical validation of the solitary lag screw technique for reduction mandibular angle fractures. J Oral Maxillofac Surg 50:603–607PubMedCrossRef Shetty V, Caputo A (1992) Biomechanical validation of the solitary lag screw technique for reduction mandibular angle fractures. J Oral Maxillofac Surg 50:603–607PubMedCrossRef
25.
Zurück zum Zitat Dichard A, Klotch DW (1994) Testing biomechanical strength of repairs for the mandibular angle fracture. Laryngoscope 104:201–208PubMed Dichard A, Klotch DW (1994) Testing biomechanical strength of repairs for the mandibular angle fracture. Laryngoscope 104:201–208PubMed
26.
Zurück zum Zitat Choi BH, Kim KN, Kang HS (1995) Clinical and in vitro evaluation of mandibular angle fracture fixation with two-miniplate system. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 79:692–695PubMedCrossRef Choi BH, Kim KN, Kang HS (1995) Clinical and in vitro evaluation of mandibular angle fracture fixation with two-miniplate system. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 79:692–695PubMedCrossRef
27.
Zurück zum Zitat Choi BH, Yoo JH, Kim KN, Kang HS (1995) Stability testing of a two-miniplate fixation technique for mandibular angle fractures: an in vitro study. J Craniomaxillofac Surg 23:123–125PubMedCrossRef Choi BH, Yoo JH, Kim KN, Kang HS (1995) Stability testing of a two-miniplate fixation technique for mandibular angle fractures: an in vitro study. J Craniomaxillofac Surg 23:123–125PubMedCrossRef
28.
Zurück zum Zitat Haug RH, Barber JE, Reifeis R (1996) A comparison of mandibular angle fracture plating techniques. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 82:257–263PubMedCrossRef Haug RH, Barber JE, Reifeis R (1996) A comparison of mandibular angle fracture plating techniques. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 82:257–263PubMedCrossRef
29.
Zurück zum Zitat Schierle HP, Schmelzeisen R, Rahn B (1996) Experimentelle Untersuchungen zur biomechanischen Stabilität unterschiedlicher Miniplattenkonfigurationen am Kieferwinkel. Fortschr Kiefer Gesichtschir 41:166–170PubMed Schierle HP, Schmelzeisen R, Rahn B (1996) Experimentelle Untersuchungen zur biomechanischen Stabilität unterschiedlicher Miniplattenkonfigurationen am Kieferwinkel. Fortschr Kiefer Gesichtschir 41:166–170PubMed
30.
Zurück zum Zitat Nissenbaum M, Lownie M, Cleaton-Jones P (1997) Relative displacement resistance of standard and low-profile bone plates in experimental mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 83:427–432PubMedCrossRef Nissenbaum M, Lownie M, Cleaton-Jones P (1997) Relative displacement resistance of standard and low-profile bone plates in experimental mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 83:427–432PubMedCrossRef
31.
Zurück zum Zitat Rudman RA, Rosenthal SC, Shen C, Ruskin JD, Ifju PG (1997) Photoelastic analysis of miniplate osteosynthesis for mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 84:129–136PubMedCrossRef Rudman RA, Rosenthal SC, Shen C, Ruskin JD, Ifju PG (1997) Photoelastic analysis of miniplate osteosynthesis for mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 84:129–136PubMedCrossRef
32.
Zurück zum Zitat Fedok FG, Van Kooten DW, DeJoseph LM, McGinn JD, Sobota B, Levin RJ, Jacobs CR (1998) Plating technique and plate orientation in repair of mandibular angle fractures: an in vitro study. Laryngoscope 108:1218–1224PubMedCrossRef Fedok FG, Van Kooten DW, DeJoseph LM, McGinn JD, Sobota B, Levin RJ, Jacobs CR (1998) Plating technique and plate orientation in repair of mandibular angle fractures: an in vitro study. Laryngoscope 108:1218–1224PubMedCrossRef
33.
Zurück zum Zitat Gutwald R, Büscher P, Schramm A, Schmelzeisen R (1999) Biomechanical stability of an internal mini-fixation-system in maxillofacial osteosynthesis. Med Biol Eng Comp 37:280–281 Gutwald R, Büscher P, Schramm A, Schmelzeisen R (1999) Biomechanical stability of an internal mini-fixation-system in maxillofacial osteosynthesis. Med Biol Eng Comp 37:280–281
34.
Zurück zum Zitat Haug RH, Fattahi TT, Goltz M (2001) A biomechanical evaluation of mandibular angle fracture plating techniques. J Oral Maxillofac Surg 59:1199–1210PubMedCrossRef Haug RH, Fattahi TT, Goltz M (2001) A biomechanical evaluation of mandibular angle fracture plating techniques. J Oral Maxillofac Surg 59:1199–1210PubMedCrossRef
35.
Zurück zum Zitat Haug RH, Street CC, Goltz M (2002) Does plate adaptation affect stability? A biomechanical comparison of locking and nonlocking plates. J Oral Maxillofac Surg 60:1319–1326PubMedCrossRef Haug RH, Street CC, Goltz M (2002) Does plate adaptation affect stability? A biomechanical comparison of locking and nonlocking plates. J Oral Maxillofac Surg 60:1319–1326PubMedCrossRef
36.
Zurück zum Zitat Feledy J, Caterson EJ, Steger S, Stal S, Hollier L (2004) Treatment of mandibular angle fractures with a matrix miniplate: a preliminary report. Plast Reconstr Surg 114:1711–1716PubMedCrossRef Feledy J, Caterson EJ, Steger S, Stal S, Hollier L (2004) Treatment of mandibular angle fractures with a matrix miniplate: a preliminary report. Plast Reconstr Surg 114:1711–1716PubMedCrossRef
37.
Zurück zum Zitat Chacon GE, Dillard FM, Clelland N, Rashid R (2005) Comparison of strains produced by titanium and poly d, l-lactide acid plating systems to in vitro forces. J Oral Maxillofac Surg 63:968–972PubMedCrossRef Chacon GE, Dillard FM, Clelland N, Rashid R (2005) Comparison of strains produced by titanium and poly d, l-lactide acid plating systems to in vitro forces. J Oral Maxillofac Surg 63:968–972PubMedCrossRef
38.
Zurück zum Zitat Jain SA, Simmons OP, Hanemann MS, Baratta RV, Guerra AB, Metzinger SE (2006) Comparison of tensile strength of resorbable plating systems used in monocortical mandible angle osteotomy repair. Arch Facial Plast Surg 8:390–395PubMedCrossRef Jain SA, Simmons OP, Hanemann MS, Baratta RV, Guerra AB, Metzinger SE (2006) Comparison of tensile strength of resorbable plating systems used in monocortical mandible angle osteotomy repair. Arch Facial Plast Surg 8:390–395PubMedCrossRef
39.
Zurück zum Zitat Wang H, Chen MS, Tang W, Tian WD (2006) Biomechanical study of rigid internal fixation for mandibular angle fracture. Hua Xi Kou Qiang Yi Xue Za Zhi 24:393–396PubMed Wang H, Chen MS, Tang W, Tian WD (2006) Biomechanical study of rigid internal fixation for mandibular angle fracture. Hua Xi Kou Qiang Yi Xue Za Zhi 24:393–396PubMed
40.
Zurück zum Zitat Alkan A, Celebi N, Ozden B, Baş B, Inal S (2007) Biomechanical comparison of different plating techniques in repair of mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:752–756PubMedCrossRef Alkan A, Celebi N, Ozden B, Baş B, Inal S (2007) Biomechanical comparison of different plating techniques in repair of mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:752–756PubMedCrossRef
41.
Zurück zum Zitat Feichtinger M, Schultes G, Kärcher H (2008) The use of a 3D navigation system in the treatment of mandibular angle fractures by minimally invasive insertion of Herbert screws for osteosynthesis. Comput Aided Surg 13:47–54PubMed Feichtinger M, Schultes G, Kärcher H (2008) The use of a 3D navigation system in the treatment of mandibular angle fractures by minimally invasive insertion of Herbert screws for osteosynthesis. Comput Aided Surg 13:47–54PubMed
42.
Zurück zum Zitat Turgut G, Kayali MU, Soydan AT, Gündüz O, Salman S, Oktar F, Baş L (2008) Biomechanical comparison of a new technique of mandibular angle fractures: biplanar and bicortical superior proximal 3 holes and bicortical inferior plate fixation. J Craniofac Surg 19:428–432PubMedCrossRef Turgut G, Kayali MU, Soydan AT, Gündüz O, Salman S, Oktar F, Baş L (2008) Biomechanical comparison of a new technique of mandibular angle fractures: biplanar and bicortical superior proximal 3 holes and bicortical inferior plate fixation. J Craniofac Surg 19:428–432PubMedCrossRef
43.
Zurück zum Zitat Bayram B, Araz K, Uckan S, Balcik C (2009) Comparison of fixation stability of resorbable versus titanium plate and screws in mandibular angle fractures. J Oral Maxillofac Surg 67:1644–1648PubMedCrossRef Bayram B, Araz K, Uckan S, Balcik C (2009) Comparison of fixation stability of resorbable versus titanium plate and screws in mandibular angle fractures. J Oral Maxillofac Surg 67:1644–1648PubMedCrossRef
44.
Zurück zum Zitat Kalfarentzos EF, Deligianni D, Mitros G, Tyllianakis M (2009) Biomechanical evaluation of plating techniques for fixing mandibular angle fractures: the introduction of a new 3D plate approach. Oral Maxillofac Surg 13:139–144PubMedCrossRef Kalfarentzos EF, Deligianni D, Mitros G, Tyllianakis M (2009) Biomechanical evaluation of plating techniques for fixing mandibular angle fractures: the introduction of a new 3D plate approach. Oral Maxillofac Surg 13:139–144PubMedCrossRef
45.
Zurück zum Zitat Ribeiro-Junior PD, Magro-Filho O, Shastri KA, Papageorge MB (2010) In vitro evaluation of conventional and locking miniplate/screw systems for the treatment of mandibular angle fractures. Int J Oral Maxillofac Surg 39:1109–1114PubMedCrossRef Ribeiro-Junior PD, Magro-Filho O, Shastri KA, Papageorge MB (2010) In vitro evaluation of conventional and locking miniplate/screw systems for the treatment of mandibular angle fractures. Int J Oral Maxillofac Surg 39:1109–1114PubMedCrossRef
46.
Zurück zum Zitat Bregagnolo LA, Bertelli PF, Ribeiro MC, Sverzut CE, Trivellato AE (2011) Evaluation of in vitro resistance of titanium and resorbable (poly-l-dl-lactic acid) fixation systems on the mandibular angle fracture. Int J Oral Maxillofac Surg 40:316–321PubMedCrossRef Bregagnolo LA, Bertelli PF, Ribeiro MC, Sverzut CE, Trivellato AE (2011) Evaluation of in vitro resistance of titanium and resorbable (poly-l-dl-lactic acid) fixation systems on the mandibular angle fracture. Int J Oral Maxillofac Surg 40:316–321PubMedCrossRef
47.
Zurück zum Zitat Esen A, Dolanmaz D, Tüz HH (2012) Biomechanical evaluation of malleable noncompression miniplates in mandibular angle fractures: an experimental study. Br J Oral Maxillofac Surg 50:e65–e68PubMedCrossRef Esen A, Dolanmaz D, Tüz HH (2012) Biomechanical evaluation of malleable noncompression miniplates in mandibular angle fractures: an experimental study. Br J Oral Maxillofac Surg 50:e65–e68PubMedCrossRef
48.
Zurück zum Zitat Pektas ZO, Bayram B, Balcik C, Develi T, Uckan S (2012) Effects of different mandibular fracture patterns on the stability of miniplate screw fixation in angle mandibular fractures. Int J Oral Maxillofac Surg 41:339–343PubMedCrossRef Pektas ZO, Bayram B, Balcik C, Develi T, Uckan S (2012) Effects of different mandibular fracture patterns on the stability of miniplate screw fixation in angle mandibular fractures. Int J Oral Maxillofac Surg 41:339–343PubMedCrossRef
49.
Zurück zum Zitat Bredbenner TL, Haug RH (2000) Substitutes for human cadaveric bone in maxillofacial rigid fixation research. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 90:574–580PubMedCrossRef Bredbenner TL, Haug RH (2000) Substitutes for human cadaveric bone in maxillofacial rigid fixation research. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 90:574–580PubMedCrossRef
50.
Zurück zum Zitat Ellis E 3rd (1993) Treatment of mandibular angle fractures using the AO reconstruction plate. J Oral Maxillofac Surg 51:250–254PubMedCrossRef Ellis E 3rd (1993) Treatment of mandibular angle fractures using the AO reconstruction plate. J Oral Maxillofac Surg 51:250–254PubMedCrossRef
51.
Zurück zum Zitat Chrcanovic BR (2012) Open versus closed reduction: comminuted mandibular fractures. Oral Maxillofac Surg (in press) Chrcanovic BR (2012) Open versus closed reduction: comminuted mandibular fractures. Oral Maxillofac Surg (in press)
52.
Zurück zum Zitat Uppal N, Baliga M, Ramanathan A (2009) Improper reduction and fixation of bilateral anterior mandibular fractures: a case report. Oral Maxillofac Surg 13:177–179PubMedCrossRef Uppal N, Baliga M, Ramanathan A (2009) Improper reduction and fixation of bilateral anterior mandibular fractures: a case report. Oral Maxillofac Surg 13:177–179PubMedCrossRef
53.
Zurück zum Zitat Rozema FR, Otten E, Bos RR, Boering G, van Willigen JD (1992) Computer-aided optimization of choice and positioning of bone plates and screws used for internal fixation of mandibular fractures. Int J Oral Maxlllofat Surg 21:373–377CrossRef Rozema FR, Otten E, Bos RR, Boering G, van Willigen JD (1992) Computer-aided optimization of choice and positioning of bone plates and screws used for internal fixation of mandibular fractures. Int J Oral Maxlllofat Surg 21:373–377CrossRef
54.
Zurück zum Zitat Haug R (1993) The effects of screw number and length on two methods of tension band plating. J Oral Maxillofac Surg 51:159–162PubMedCrossRef Haug R (1993) The effects of screw number and length on two methods of tension band plating. J Oral Maxillofac Surg 51:159–162PubMedCrossRef
55.
Zurück zum Zitat Perrott DH (1996) Discussion: treatment of mandibular angle fractures using one noncompression miniplate. J Oral Maxillofac Surg 54:871–872CrossRef Perrott DH (1996) Discussion: treatment of mandibular angle fractures using one noncompression miniplate. J Oral Maxillofac Surg 54:871–872CrossRef
56.
Zurück zum Zitat Perren SM, Russenberger M, Steinemann S, Müller ME, Allgöwer M (1969) A dynamic compression plate. Acta Orthop Stand Suppl 125:31–41 Perren SM, Russenberger M, Steinemann S, Müller ME, Allgöwer M (1969) A dynamic compression plate. Acta Orthop Stand Suppl 125:31–41
57.
Zurück zum Zitat Matsushita K, Inoue N, Totsuka Y (2011) In vitro biomechanical evaluation of the effect of an additional L-shaped plate on straight or box plate fixation in sagittal split ramus osteotomy using a bioabsorbable plate system. Oral Maxillofac Surg 15:139–146PubMedCrossRef Matsushita K, Inoue N, Totsuka Y (2011) In vitro biomechanical evaluation of the effect of an additional L-shaped plate on straight or box plate fixation in sagittal split ramus osteotomy using a bioabsorbable plate system. Oral Maxillofac Surg 15:139–146PubMedCrossRef
58.
Zurück zum Zitat Gerlach KL, Schwarz A (2002) Bite forces in patients after treatment of mandibular angle fractures with miniplate osteosynthesis according to Champy. Int J Oral Maxillofac Surg 31:345–348PubMedCrossRef Gerlach KL, Schwarz A (2002) Bite forces in patients after treatment of mandibular angle fractures with miniplate osteosynthesis according to Champy. Int J Oral Maxillofac Surg 31:345–348PubMedCrossRef
Metadaten
Titel
Fixation of mandibular angle fractures: in vitro biomechanical assessments and computer-based studies
verfasst von
Bruno Ramos Chrcanovic
Publikationsdatum
01.12.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
Oral and Maxillofacial Surgery / Ausgabe 4/2013
Print ISSN: 1865-1550
Elektronische ISSN: 1865-1569
DOI
https://doi.org/10.1007/s10006-012-0367-0

Weitere Artikel der Ausgabe 4/2013

Oral and Maxillofacial Surgery 4/2013 Zur Ausgabe

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Deutlich weniger Infektionen: Wundprotektoren schützen!

08.05.2024 Postoperative Wundinfektion Nachrichten

Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.

Chirurginnen und Chirurgen sind stark suizidgefährdet

07.05.2024 Suizid Nachrichten

Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.