Diagnostic accuracy of low-dose versus ultra-low-dose CT for lumbar disc disease and facet joint osteoarthritis in patients with low back pain with MRI correlation

verfasst von: Sun Hwa Lee, Seong Jong Yun, Hyeon Hwan Jo, Dong Hyeon Kim, Jae Gwang Song, Yong Sung Park

Erschienen in: Skeletal Radiology | Ausgabe 4/2018

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare the image quality, radiation dose, and diagnostic performance between low-dose (LD) and ultra-low-dose (ULD) lumbar-spine (L-spine) CT with iterative reconstruction (IR) for patients with chronic low back pain (LBP).

Methods

In total, 260 patients with chronic LBP who underwent L-spine CT between November 2015 and September 2016 were prospectively enrolled. Of these, 143 underwent LD-CT with IR and 117 underwent ULD-CT with IR. The patients were divided according to their body mass index (BMI) into BMI1 (<22.9 kg/m2), BMI2 (23.0–24.9 kg/m2), and BMI3 (≥25 kg/m2) groups. Two blinded radiologists independently evaluated the signal-to-noise ratio (SNR), qualitative image quality, and final diagnoses (lumbar disc disease and facet joint osteoarthritis). L-spine MRIs interpreted by consensus were used as the reference standard. All data were statistically analyzed.

Results

ULD protocol showed significantly lower SNR for all patients (p < 0.001) except the vertebral bodies and lower qualitative image quality for BMI3 patients (p ≤ 0.033). There was no statistically significant difference between ULD (sensitivity, 95.1–98.1%; specificity, 92.5–98.7%; accuracy, 94.6–98.0%) and LD protocols (sensitivity, 95.6–100%; specificity, 95.5–98.9%; accuracy, 97.4–98.1%), (all p≥0.1) in the BMI1 and BMI2; while dose was 60–68% lower with the ULD protocol. Interobserver agreements were excellent or good with regard to image quality and final diagnoses.

Conclusions

For the BM1 and BMI2 groups, ULD-CT provided an acceptable image quality and exhibited a diagnostic accuracy similar to that of LD-CT. These findings suggest that it is a useful diagnostic tool for patients with chronic LBP who exhibit a BMI of <25 kg/m2.
Literatur
1.
Zurück zum Zitat Patel ND, Broderick DF, Burns J, et al. ACR Appropriateness Criteria. Low Back Pain. J Am Coll Radiol. 2016;13:1069–78.CrossRef Patel ND, Broderick DF, Burns J, et al. ACR Appropriateness Criteria. Low Back Pain. J Am Coll Radiol. 2016;13:1069–78.CrossRef
2.
Zurück zum Zitat Van Rijn RM1, Wassenaar M, Verhagen AP, et al. Computed tomography for the diagnosis of lumbar spinal pathology in adult patients with low back pain or sciatica: a diagnostic systematic review. Eur Spine J. 2012;21:228–39.CrossRef Van Rijn RM1, Wassenaar M, Verhagen AP, et al. Computed tomography for the diagnosis of lumbar spinal pathology in adult patients with low back pain or sciatica: a diagnostic systematic review. Eur Spine J. 2012;21:228–39.CrossRef
3.
Zurück zum Zitat Brenner DJ, Hall EJ. Computed tomography—an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84.CrossRef Brenner DJ, Hall EJ. Computed tomography—an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84.CrossRef
4.
Zurück zum Zitat Hall EJ, Brenner DJ. Cancer risks from diagnostic radiology. Br J Radiol. 2008;81:362–78.CrossRef Hall EJ, Brenner DJ. Cancer risks from diagnostic radiology. Br J Radiol. 2008;81:362–78.CrossRef
5.
Zurück zum Zitat Omoumi P, Verdun FR, Becce F. Optimization of radiation dose and image quality in musculoskeletal CT: emphasis on iterative reconstruction techniques (Part 2). Semin Musculoskelet Radiol. 2015;19:422–30.CrossRef Omoumi P, Verdun FR, Becce F. Optimization of radiation dose and image quality in musculoskeletal CT: emphasis on iterative reconstruction techniques (Part 2). Semin Musculoskelet Radiol. 2015;19:422–30.CrossRef
6.
Zurück zum Zitat Omoumi P, Becce F, Ott JG, Racine D, Verdun FR. Optimization of radiation dose and image quality in musculoskeletal CT: emphasis on iterative reconstruction techniques (Part 1). Semin Musculoskelet Radiol. 2015;19:415–21.CrossRef Omoumi P, Becce F, Ott JG, Racine D, Verdun FR. Optimization of radiation dose and image quality in musculoskeletal CT: emphasis on iterative reconstruction techniques (Part 1). Semin Musculoskelet Radiol. 2015;19:415–21.CrossRef
7.
Zurück zum Zitat McCollough CH, Chen GH, KalenderW, et al. Achieving routine ultra-low dose CT scanning: report from the summit on management of radiation dose in CT. Radiology. 2012;264:567–80.CrossRef McCollough CH, Chen GH, KalenderW, et al. Achieving routine ultra-low dose CT scanning: report from the summit on management of radiation dose in CT. Radiology. 2012;264:567–80.CrossRef
8.
Zurück zum Zitat Boone JM, Hendee WR, McNitt-Gray MF, et al. Radiation exposure from CT scans: how to close our knowledge gaps, monitor and safeguard exposure—proceedings and recommendations of the Radiation Dose Summit, sponsored by NIBIB, February 24–25, 2011. Radiology. 2012;265:544–54.CrossRef Boone JM, Hendee WR, McNitt-Gray MF, et al. Radiation exposure from CT scans: how to close our knowledge gaps, monitor and safeguard exposure—proceedings and recommendations of the Radiation Dose Summit, sponsored by NIBIB, February 24–25, 2011. Radiology. 2012;265:544–54.CrossRef
9.
Zurück zum Zitat Shin CI, Kim SH, Im JP, et al. One-mSv CT colonography: effect of different iterative reconstruction algorithms on radiologists’ performance. Eur J Radiol. 2016;85:641–8.CrossRef Shin CI, Kim SH, Im JP, et al. One-mSv CT colonography: effect of different iterative reconstruction algorithms on radiologists’ performance. Eur J Radiol. 2016;85:641–8.CrossRef
10.
Zurück zum Zitat Kalra MK, Quick P, Singh S, et al. Whole spine CT for evaluation of scoliosis in children: feasibility of sub-milliSievert scanning protocol. Acta Radiol. 2013;54:226–30.CrossRef Kalra MK, Quick P, Singh S, et al. Whole spine CT for evaluation of scoliosis in children: feasibility of sub-milliSievert scanning protocol. Acta Radiol. 2013;54:226–30.CrossRef
11.
Zurück zum Zitat Cademartiri F, Maffei E, Arcadi T, Catalano O, Midiri M. CT coronary angiography at an ultra-low radiation dose (<0.1 mSv): feasible and viable in times of constraint on healthcare costs. Eur Radiol. 2013;23:607–13.CrossRef Cademartiri F, Maffei E, Arcadi T, Catalano O, Midiri M. CT coronary angiography at an ultra-low radiation dose (<0.1 mSv): feasible and viable in times of constraint on healthcare costs. Eur Radiol. 2013;23:607–13.CrossRef
12.
Zurück zum Zitat Kim Y, Kim YK, Lee BE, et al. Ultra-low-dose CT of the thorax using iterative reconstruction: evaluation of image quality and radiation dose reduction. AJR Am J Roentgenol. 2015;204:1197–202.CrossRef Kim Y, Kim YK, Lee BE, et al. Ultra-low-dose CT of the thorax using iterative reconstruction: evaluation of image quality and radiation dose reduction. AJR Am J Roentgenol. 2015;204:1197–202.CrossRef
13.
Zurück zum Zitat Rob S, Bryant T, Wilson I, Somani BK. Ultra-low-dose, low-dose, and standard-dose CT of the kidney, ureters, and bladder: is there a difference? Results from a systematic review of the literature. Clin Radiol. 2017;72:11–5.CrossRef Rob S, Bryant T, Wilson I, Somani BK. Ultra-low-dose, low-dose, and standard-dose CT of the kidney, ureters, and bladder: is there a difference? Results from a systematic review of the literature. Clin Radiol. 2017;72:11–5.CrossRef
14.
Zurück zum Zitat Willemink MJ, de Jong PA, Leiner T, et al. Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol. 2013;23:1623–1631.CrossRef Willemink MJ, de Jong PA, Leiner T, et al. Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol. 2013;23:1623–1631.CrossRef
15.
Zurück zum Zitat Geyer LL, Schoepf UJ, Meinel FG, et al. State of the art: iterative CT reconstruction techniques. Radiology. 2015;276:339–57.CrossRef Geyer LL, Schoepf UJ, Meinel FG, et al. State of the art: iterative CT reconstruction techniques. Radiology. 2015;276:339–57.CrossRef
16.
Zurück zum Zitat Gervaise A, Osemont B, Lecocq S, et al. CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol. 2012;22:295–301.CrossRef Gervaise A, Osemont B, Lecocq S, et al. CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol. 2012;22:295–301.CrossRef
17.
Zurück zum Zitat Bohy P, De Maertelaer V, Roquigny A, Keyzer C, Tack D, Gevenois PA. Multidetector CT in patients suspected of having lumbar disk herniation: comparison of standard-dose and simulated low-dose techniques. Radiology. 2007;244:524–31.CrossRef Bohy P, De Maertelaer V, Roquigny A, Keyzer C, Tack D, Gevenois PA. Multidetector CT in patients suspected of having lumbar disk herniation: comparison of standard-dose and simulated low-dose techniques. Radiology. 2007;244:524–31.CrossRef
18.
Zurück zum Zitat Alshamari M, Geijer M, Norrman E, et al. Low dose CT of the lumbar spine compared with radiography: a study on image quality with implications for clinical practice. Acta Radiol. 2016;57:602–11.CrossRef Alshamari M, Geijer M, Norrman E, et al. Low dose CT of the lumbar spine compared with radiography: a study on image quality with implications for clinical practice. Acta Radiol. 2016;57:602–11.CrossRef
19.
Zurück zum Zitat Alshamari M, Geijer M, Norrman E, Geijer H. Low-dose computed tomography of the lumbar spine: a phantom study on imaging parameters and image quality. Acta Radiol. 2014;55:824–32.CrossRef Alshamari M, Geijer M, Norrman E, Geijer H. Low-dose computed tomography of the lumbar spine: a phantom study on imaging parameters and image quality. Acta Radiol. 2014;55:824–32.CrossRef
20.
Zurück zum Zitat Yang CH, Wu TH, Chiou YY, et al. Imaging quality and diagnostic reliability of low-dose computed tomography lumbar spine for evaluating patients with spinal disorders. Spine J. 2014;14:2682–90.CrossRef Yang CH, Wu TH, Chiou YY, et al. Imaging quality and diagnostic reliability of low-dose computed tomography lumbar spine for evaluating patients with spinal disorders. Spine J. 2014;14:2682–90.CrossRef
21.
Zurück zum Zitat Yang CH, Wu TH, Lin CJ, et al. Knowledge-based iterative model reconstruction technique in computed tomography of lumbar spine lowers radiation dose and improves tissue differentiation for patients with lower back pain. Eur J Radiol. 2016;85:1757–64.CrossRef Yang CH, Wu TH, Lin CJ, et al. Knowledge-based iterative model reconstruction technique in computed tomography of lumbar spine lowers radiation dose and improves tissue differentiation for patients with lower back pain. Eur J Radiol. 2016;85:1757–64.CrossRef
22.
Zurück zum Zitat World Health Organization. Obesity: preventing and managing the global epidemic— report of a WHO consultation of obesity. Geneva, Switzerland: World Health Organization; 1997. World Health Organization. Obesity: preventing and managing the global epidemic— report of a WHO consultation of obesity. Geneva, Switzerland: World Health Organization; 1997.
23.
Zurück zum Zitat Kilic K, Erbas G, Guryildirim M, Arac M, Ilgit E, Coskun B. Lowering the dose in head CT using adaptive statistical iterative reconstruction. AJNR Am J Neuroradiol. 2011;32:1578–82.CrossRef Kilic K, Erbas G, Guryildirim M, Arac M, Ilgit E, Coskun B. Lowering the dose in head CT using adaptive statistical iterative reconstruction. AJNR Am J Neuroradiol. 2011;32:1578–82.CrossRef
24.
Zurück zum Zitat Patro SN, Chakraborty S, Sheikh A. The use of adaptive statistical iterative reconstruction (ASiR) technique in evaluation of patients with cervical spine trauma: impact on radiation dose reduction and image quality. Br J Radiol. 2016;89:20150082.CrossRef Patro SN, Chakraborty S, Sheikh A. The use of adaptive statistical iterative reconstruction (ASiR) technique in evaluation of patients with cervical spine trauma: impact on radiation dose reduction and image quality. Br J Radiol. 2016;89:20150082.CrossRef
25.
Zurück zum Zitat Bongartz G, Golding SJ, Jurik AJ, et al. European guidelines for multislice computed tomography: report EUR 16262 EN 2004. European Commission: Luxembourg; 2004. Bongartz G, Golding SJ, Jurik AJ, et al. European guidelines for multislice computed tomography: report EUR 16262 EN 2004. European Commission: Luxembourg; 2004.
26.
Zurück zum Zitat Weishaupt D, Zanetti M, Boos N, Hodler J. MR imaging and CT in osteoarthritis of the lumbar facet joints. Skeletal Radiol. 1999;28:215–9.CrossRef Weishaupt D, Zanetti M, Boos N, Hodler J. MR imaging and CT in osteoarthritis of the lumbar facet joints. Skeletal Radiol. 1999;28:215–9.CrossRef
27.
Zurück zum Zitat Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology. 2010;257:158–66.CrossRef Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology. 2010;257:158–66.CrossRef
28.
Zurück zum Zitat Fazel R, Krumholz HM, Wang Y, et al. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med. 2009;361:849–57.CrossRef Fazel R, Krumholz HM, Wang Y, et al. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med. 2009;361:849–57.CrossRef
29.
Zurück zum Zitat Martin CJ. Effective dose: how should it be applied to medical exposures? Br J Radiol. 2007;80:639–47.CrossRef Martin CJ. Effective dose: how should it be applied to medical exposures? Br J Radiol. 2007;80:639–47.CrossRef
30.
Zurück zum Zitat von Falck C, Galanski M, Shin HO. Informatics in radiology: sliding-thin-slab averaging for improved depiction of low-contrast lesions with radiation dose savings at thin-section CT. Radiographics. 2010;30:317–326.CrossRef von Falck C, Galanski M, Shin HO. Informatics in radiology: sliding-thin-slab averaging for improved depiction of low-contrast lesions with radiation dose savings at thin-section CT. Radiographics. 2010;30:317–326.CrossRef
31.
Zurück zum Zitat Omoumi P, Verdun FR, Ben Salah Y, et al. Low-dose multidetector computed tomography of the cervical spine: optimization of iterative reconstruction strength levels. Acta Radiol. 2014;55:335–44.CrossRef Omoumi P, Verdun FR, Ben Salah Y, et al. Low-dose multidetector computed tomography of the cervical spine: optimization of iterative reconstruction strength levels. Acta Radiol. 2014;55:335–44.CrossRef
Metadaten
Titel
Diagnostic accuracy of low-dose versus ultra-low-dose CT for lumbar disc disease and facet joint osteoarthritis in patients with low back pain with MRI correlation
verfasst von
Sun Hwa Lee
Seong Jong Yun
Hyeon Hwan Jo
Dong Hyeon Kim
Jae Gwang Song
Yong Sung Park
Publikationsdatum
06.11.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 4/2018
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-017-2811-6

Weitere Artikel der Ausgabe 4/2018

Skeletal Radiology 4/2018 Zur Ausgabe

ACKNOWLEDGEMENT TO REFEREES

Thank you to our reviewers of 2017

PET kann infarktgefährdete Koronararterien entdecken

04.06.2024 Koronare Herzerkrankung Nachrichten

Der Nachweis aktiver Plaques mittels 18F-Natriumfluorid-PET hilft nicht nur, infarktgefährdete Patienten, sondern auch infarktgefährdete Koronararterien zu erkennen. Von einer gezielten Behandlung vulnerabler Plaques ist man trotzdem weit entfernt.

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.