A comparison of the effectiveness of open and closed endotracheal suction

verfasst von: Beverley Copnell, David G. Tingay, Nicholas J. Kiraly, Magdy Sourial, Michael J. Gordon, John F. Mills, Colin J. Morley, Peter A. Dargaville

Erschienen in: Intensive Care Medicine | Ausgabe 9/2007

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare the effectiveness of open and closed endotracheal suction in recovering thin and thick secretions in normal and injured lungs during conventional and high frequency ventilation.

Design and setting

Randomised study in a paediatric intensive care model in the animal research laboratory of a tertiary paediatric hospital.

Subjects

16 New Zealand White rabbits.

Interventions

Anaesthetised animals were intubated with a 3.5-mm endotracheal tube. Simulated thin and thick secretions (iopamidol 2 ml, a watery radio-opaque fluid, and fluorescent mucin 1 ml) were instilled in turn 1 cm below the tube tip through a catheter placed via a tracheostomy. Open or closed suction, randomly assigned, was applied for 6 s at −140 mmHg using a 6-F gauge catheter. Following lung injury with repeated saline lavage the procedure was repeated on conventional and high frequency ventilation.

Measurements and results

Iopamidol recovery was determined by digitally subtracting the post-contrast and post-suction radiographic images. Mucin recovery was determined by fluorescence assay of the aspirate. In the normal lung similar amounts were recovered by both suction methods. In the lavaged lung closed suction recovered less iopamidol during conventional (22 ± 7.5%) and high frequency ventilation (11 ± 2.4%) than open suction (36 ± 12% and 22 ± 8.1%, respectively). Mucin recovery was less with closed suction during conventional (32 ± 28 μl) and high frequency ventilation (30 ± 31 μl) than with open suction (382 ± 235 μl and 241 ± 153 μl).

Conclusions

In the injured lung closed suction was less effective than open suction at recovering thin and thick simulated secretions, irrespective of ventilation mode.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Landa JF, Kwoka MA, Chapman GA, Brito M, Sackner MA (1980) Effects of suctioning on mucociliary transport. Chest 77:202–207PubMed Landa JF, Kwoka MA, Chapman GA, Brito M, Sackner MA (1980) Effects of suctioning on mucociliary transport. Chest 77:202–207PubMed
2.
Zurück zum Zitat Guglielminotti J, Desmonts JM, Dureuil B (1998) Effects of tracheal suctioning on respiratory resistances in mechanically ventilated patients. Chest 113:1335–1338PubMed Guglielminotti J, Desmonts JM, Dureuil B (1998) Effects of tracheal suctioning on respiratory resistances in mechanically ventilated patients. Chest 113:1335–1338PubMed
3.
Zurück zum Zitat Woodgate PG, Flenady V (2003) Tracheal suctioning without disconnection in intubated ventilated neonates (Cochrane Review). In: The Cochrane Library. Wiley, Chichester, UK Woodgate PG, Flenady V (2003) Tracheal suctioning without disconnection in intubated ventilated neonates (Cochrane Review). In: The Cochrane Library. Wiley, Chichester, UK
4.
Zurück zum Zitat Brandstater B, Muallem M (1969) Atelectasis following tracheal suction in infants. Anesthesiology 31:468–473PubMedCrossRef Brandstater B, Muallem M (1969) Atelectasis following tracheal suction in infants. Anesthesiology 31:468–473PubMedCrossRef
5.
Zurück zum Zitat Cabal L, Devaskar S, Siassi B, Plajstek C, Waffarn F, Blanco C, Hodgman J (1979) New endotracheal tube adaptor reducing cardiopulmonary effects of suctioning. Crit Care Med 7:552–555PubMedCrossRef Cabal L, Devaskar S, Siassi B, Plajstek C, Waffarn F, Blanco C, Hodgman J (1979) New endotracheal tube adaptor reducing cardiopulmonary effects of suctioning. Crit Care Med 7:552–555PubMedCrossRef
6.
Zurück zum Zitat Simbruner G, Coradello H, Fodor M, Havelec L, Lubec G, Pollak A (1981) Effect of tracheal suction on oxygenation, circulation, and lung mechanics in newborn infants. Arch Dis Child 56:326–330PubMedCrossRef Simbruner G, Coradello H, Fodor M, Havelec L, Lubec G, Pollak A (1981) Effect of tracheal suction on oxygenation, circulation, and lung mechanics in newborn infants. Arch Dis Child 56:326–330PubMedCrossRef
7.
Zurück zum Zitat Kerem E, Yatsiv I, Goitein KJ (1990) Effect of endotracheal suctioning on arterial blood gases in children. Intensive Care Med 16:95–99PubMed Kerem E, Yatsiv I, Goitein KJ (1990) Effect of endotracheal suctioning on arterial blood gases in children. Intensive Care Med 16:95–99PubMed
8.
Zurück zum Zitat Singh NC, Kissoon N, Frewen T, Tiffin N (1991) Physiological responses to endotracheal and oral suctioning in paediatric patients: the influence of endotracheal tube sizes and suction pressures. Clin Intensive Care 2:345–350PubMed Singh NC, Kissoon N, Frewen T, Tiffin N (1991) Physiological responses to endotracheal and oral suctioning in paediatric patients: the influence of endotracheal tube sizes and suction pressures. Clin Intensive Care 2:345–350PubMed
9.
Zurück zum Zitat Shah AR, Kurth CD, Gwiazdowski SG, Chance B, Delivoria-Papadopoulos M (1992) Fluctuations in cerebral oxygenation and blood volume during endotracheal suctioning in premature infants. J Pediatr 120:769–774PubMedCrossRef Shah AR, Kurth CD, Gwiazdowski SG, Chance B, Delivoria-Papadopoulos M (1992) Fluctuations in cerebral oxygenation and blood volume during endotracheal suctioning in premature infants. J Pediatr 120:769–774PubMedCrossRef
10.
Zurück zum Zitat Skov L, Ryding J, Pryds O, Greisen G (1992) Changes in cerebral oxygenation and cerebral blood volume during endotracheal suctioning in ventilated neonates. Acta Paediatr 81:389–393PubMed Skov L, Ryding J, Pryds O, Greisen G (1992) Changes in cerebral oxygenation and cerebral blood volume during endotracheal suctioning in ventilated neonates. Acta Paediatr 81:389–393PubMed
11.
Zurück zum Zitat Weidner I (2001) Closed endotracheal suction in pediatric and neonatal intensive care nursing. Kinderkrankenschwester 20:251–253PubMed Weidner I (2001) Closed endotracheal suction in pediatric and neonatal intensive care nursing. Kinderkrankenschwester 20:251–253PubMed
12.
Zurück zum Zitat Kalyn A, Blatz S, Feuerstake S, Paes B, Bautista C (2003) Closed suctioning of intubated neonates maintains better physiologic stability: a randomized trial. J Perinatol 23:218–222PubMedCrossRef Kalyn A, Blatz S, Feuerstake S, Paes B, Bautista C (2003) Closed suctioning of intubated neonates maintains better physiologic stability: a randomized trial. J Perinatol 23:218–222PubMedCrossRef
13.
Zurück zum Zitat Tan AM, Gomez JM, Mathews J, Williams M, Paratz J, Rajadurai VS (2005) Closed versus partially ventilated endotracheal suction in extremely preterm neonates: physiologic consequences. Intensive Crit Care Nurs 21:234–242PubMedCrossRef Tan AM, Gomez JM, Mathews J, Williams M, Paratz J, Rajadurai VS (2005) Closed versus partially ventilated endotracheal suction in extremely preterm neonates: physiologic consequences. Intensive Crit Care Nurs 21:234–242PubMedCrossRef
14.
Zurück zum Zitat Carlon GC, Fox SJ, Ackerman NJ (1987) Evaluation of a closed-tracheal suction system. Crit Care Med 15:522–525PubMedCrossRef Carlon GC, Fox SJ, Ackerman NJ (1987) Evaluation of a closed-tracheal suction system. Crit Care Med 15:522–525PubMedCrossRef
15.
Zurück zum Zitat Noll ML, Hix CD, Scott G (1990) Closed tracheal suction systems: effectiveness and nursing implications. AACN Clin Issues 1:318–328 Noll ML, Hix CD, Scott G (1990) Closed tracheal suction systems: effectiveness and nursing implications. AACN Clin Issues 1:318–328
16.
Zurück zum Zitat Crimlisk J, Paris R, McGonagle E, Calcutt J, Farber H (1994) The closed tracheal suction system: implications for critical care nursing. Dimens Crit Care Nurs 13:292–300PubMedCrossRef Crimlisk J, Paris R, McGonagle E, Calcutt J, Farber H (1994) The closed tracheal suction system: implications for critical care nursing. Dimens Crit Care Nurs 13:292–300PubMedCrossRef
17.
Zurück zum Zitat DePew CL, Noll ML (1994) Inline closed-system suctioning: a research analysis. Dimens Crit Care Nurs 13:73–83PubMedCrossRef DePew CL, Noll ML (1994) Inline closed-system suctioning: a research analysis. Dimens Crit Care Nurs 13:73–83PubMedCrossRef
18.
Zurück zum Zitat Blackwood B (1998) The practice and perception of intensive care staff using the closed suctioning system. J Adv Nurs 28:1020–1029PubMedCrossRef Blackwood B (1998) The practice and perception of intensive care staff using the closed suctioning system. J Adv Nurs 28:1020–1029PubMedCrossRef
19.
Zurück zum Zitat Glass C, Grap MJ, Sessler CN (1999) Endotracheal tube narrowing after closed-system suctioning: prevalence and risk factors. Am J Crit Care 8:93–100PubMed Glass C, Grap MJ, Sessler CN (1999) Endotracheal tube narrowing after closed-system suctioning: prevalence and risk factors. Am J Crit Care 8:93–100PubMed
20.
Zurück zum Zitat Lindgren S, Almgren B, Hogman M, Lethvall S, Houltz E, Lundin S, Stenqvist O (2004) Effectiveness and side effects of closed and open suctioning: an experimental evaluation. Intensive Care Med 30:1630–1637PubMedCrossRef Lindgren S, Almgren B, Hogman M, Lethvall S, Houltz E, Lundin S, Stenqvist O (2004) Effectiveness and side effects of closed and open suctioning: an experimental evaluation. Intensive Care Med 30:1630–1637PubMedCrossRef
21.
Zurück zum Zitat Witmer MT, Hess D, Simmons M (1991) An evaluation of the effectiveness of secretion removal with the Ballard closed-circuit suction catheter. Respir Care 36:844–848 Witmer MT, Hess D, Simmons M (1991) An evaluation of the effectiveness of secretion removal with the Ballard closed-circuit suction catheter. Respir Care 36:844–848
22.
Zurück zum Zitat Lasocki S, Lu Q, Sartorius A, Fouillat D, Remerand F, Rouby J (2006) Open and closed-circuit endotracheal suctioning in acute lung injury: efficiency and effects on gas exchange. Anesthesiology 104:39–47PubMedCrossRef Lasocki S, Lu Q, Sartorius A, Fouillat D, Remerand F, Rouby J (2006) Open and closed-circuit endotracheal suctioning in acute lung injury: efficiency and effects on gas exchange. Anesthesiology 104:39–47PubMedCrossRef
23.
Zurück zum Zitat Steuer JD, Stone KS, Nickel J, Steinfeld Y (2000) Methodologic issues associated with secretion weight as a dependent variable in research using closed-system suction catheters. Nurs Res 49:295–299PubMedCrossRef Steuer JD, Stone KS, Nickel J, Steinfeld Y (2000) Methodologic issues associated with secretion weight as a dependent variable in research using closed-system suction catheters. Nurs Res 49:295–299PubMedCrossRef
24.
Zurück zum Zitat Higgins J, Estetter B, Holland D, Smith B, Derdak S (2005) High-frequency oscillatory ventilation in adults: Respiratory therapy issues. Crit Care Med 33:S196–S203PubMedCrossRef Higgins J, Estetter B, Holland D, Smith B, Derdak S (2005) High-frequency oscillatory ventilation in adults: Respiratory therapy issues. Crit Care Med 33:S196–S203PubMedCrossRef
25.
Zurück zum Zitat Sweeney A-M, Lyle J, Ferguson ND (2005) Nursing and infection-control issues during high-frequency oscillatory ventilation. Crit Care Med 33:S204–S208PubMedCrossRef Sweeney A-M, Lyle J, Ferguson ND (2005) Nursing and infection-control issues during high-frequency oscillatory ventilation. Crit Care Med 33:S204–S208PubMedCrossRef
26.
Zurück zum Zitat List SJ, Findlay BP, Forstner GG, Forstner JF (1978) Enhancement of the viscosity of mucin by serum albumin. Biochemical Journal 175:565–571PubMed List SJ, Findlay BP, Forstner GG, Forstner JF (1978) Enhancement of the viscosity of mucin by serum albumin. Biochemical Journal 175:565–571PubMed
27.
Zurück zum Zitat Gordon MJ, Tingay DG, Johnston L, Copnell B (2006) Endotracheal suctioning practices of neonatal and paediatric intensive care nurses (abstract). Aust Crit Care 19:146 Gordon MJ, Tingay DG, Johnston L, Copnell B (2006) Endotracheal suctioning practices of neonatal and paediatric intensive care nurses (abstract). Aust Crit Care 19:146
28.
Zurück zum Zitat Cereda M, Villa F, Colombo E, Greco G, Nacoti M, Pesenti A (2001) Closed system endotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation. Intensive Care Med 27:648–654PubMedCrossRef Cereda M, Villa F, Colombo E, Greco G, Nacoti M, Pesenti A (2001) Closed system endotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation. Intensive Care Med 27:648–654PubMedCrossRef
29.
Zurück zum Zitat Choong K, Chatrkaw P, Frndova H, Cox PN (2003) Comparison of loss in lung volume with open versus in-line catheter endotracheal suctioning. Pediatr Crit Care Med 4:69–73PubMedCrossRef Choong K, Chatrkaw P, Frndova H, Cox PN (2003) Comparison of loss in lung volume with open versus in-line catheter endotracheal suctioning. Pediatr Crit Care Med 4:69–73PubMedCrossRef
30.
Zurück zum Zitat Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X, Richard JC, Mancebo J, Lemaire F, Brochard L (2003) Prevention of endotracheal suctioning-induced alveolar derecruitment in acute lung injury. Am J Respir Crit Care Med 167:1215–1224PubMedCrossRef Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X, Richard JC, Mancebo J, Lemaire F, Brochard L (2003) Prevention of endotracheal suctioning-induced alveolar derecruitment in acute lung injury. Am J Respir Crit Care Med 167:1215–1224PubMedCrossRef
31.
Zurück zum Zitat Tingay DG, Copnell B, Mills JF, Morley CJ, Dargaville PA (2007) Effects of open endotracheal suction on lung volume in infants receiving HFOV. Intensive Care Med 33:689–693PubMedCrossRef Tingay DG, Copnell B, Mills JF, Morley CJ, Dargaville PA (2007) Effects of open endotracheal suction on lung volume in infants receiving HFOV. Intensive Care Med 33:689–693PubMedCrossRef
32.
Zurück zum Zitat Williams R, Rankin N, Smith T, Galler D, Seakins P (1996) Relationship between the humidity and temperature of inspired gas and the function of the airway mucosa. Crit Care Med 24:1920–1929PubMedCrossRef Williams R, Rankin N, Smith T, Galler D, Seakins P (1996) Relationship between the humidity and temperature of inspired gas and the function of the airway mucosa. Crit Care Med 24:1920–1929PubMedCrossRef
33.
Zurück zum Zitat Blackwood B (1999) Normal saline instillation with endotracheal suctioning: primum non nocere (first do no harm). J Adv Nurs 29:928–934PubMedCrossRef Blackwood B (1999) Normal saline instillation with endotracheal suctioning: primum non nocere (first do no harm). J Adv Nurs 29:928–934PubMedCrossRef
34.
Zurück zum Zitat Harshbarger SA, Hoffman LA, Zullo TG, Pinsky MR (1992) Effects of a closed tracheal suction system on ventilatory and cardiovascular parameters. Am J Crit Care 1:57–61PubMed Harshbarger SA, Hoffman LA, Zullo TG, Pinsky MR (1992) Effects of a closed tracheal suction system on ventilatory and cardiovascular parameters. Am J Crit Care 1:57–61PubMed
35.
Zurück zum Zitat Lee CK, Ng KS, Tan SG, Ang R (2001) Effect of different endotracheal suctioning systems on cardiorespiratory parameters of ventilated patients. Ann Acad Med Singapore 30:239–244PubMed Lee CK, Ng KS, Tan SG, Ang R (2001) Effect of different endotracheal suctioning systems on cardiorespiratory parameters of ventilated patients. Ann Acad Med Singapore 30:239–244PubMed
36.
Zurück zum Zitat Fernandez MDM, Piacentini E, Blanch L, Fernandez R (2004) Changes in lung volume with three systems of endotracheal suctioning with and without pre-oxygenation in patients with mild-to-moderate lung failure. Intensive Care Med 30:2210–2215PubMedCrossRef Fernandez MDM, Piacentini E, Blanch L, Fernandez R (2004) Changes in lung volume with three systems of endotracheal suctioning with and without pre-oxygenation in patients with mild-to-moderate lung failure. Intensive Care Med 30:2210–2215PubMedCrossRef
37.
Zurück zum Zitat Paul-Allen J, Ostrow CL (2000) Survey of nursing practices with closed-system suctioning. Am J Crit Care 9:9–19PubMed Paul-Allen J, Ostrow CL (2000) Survey of nursing practices with closed-system suctioning. Am J Crit Care 9:9–19PubMed
38.
Zurück zum Zitat Demir F, Dramali A (2005) Requirement for 100% oxygen before and after closed suction. J Adv Nurs 51:245–251PubMedCrossRef Demir F, Dramali A (2005) Requirement for 100% oxygen before and after closed suction. J Adv Nurs 51:245–251PubMedCrossRef
Metadaten
Titel
A comparison of the effectiveness of open and closed endotracheal suction
verfasst von
Beverley Copnell
David G. Tingay
Nicholas J. Kiraly
Magdy Sourial
Michael J. Gordon
John F. Mills
Colin J. Morley
Peter A. Dargaville
Publikationsdatum
01.09.2007
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 9/2007
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0635-x

Weitere Artikel der Ausgabe 9/2007

Intensive Care Medicine 9/2007 Zur Ausgabe

Tipps für den Umgang mit Behandlungsfehlern

01.05.2024 DGIM 2024 Kongressbericht

Es ist nur eine Frage der Zeit, bis es zu einem Zwischenfall kommt und ein Behandlungsfehler passiert. Doch wenn Ärztinnen und Ärzte gut vorbereitet sind, schaffen es alle Beteiligten den Umständen entsprechend gut durch diese Krise. 

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.