Erschienen in: Intensive Care Medicine 2/2011

01.02.2011 | Physiological and Technological Notes

Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure

verfasst von: P.-E. Gancel, E. Roupie, L. Guittet, S. Laplume, N. Terzi

Erschienen in: Intensive Care Medicine | Ausgabe 2/2011

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Transcutaneous CO2 monitors are widely used in neonatal ICUs. Until recently, these devices performed poorly in adults. Recent technical modifications have produced transcutaneous CO2 monitors that have performed well in adults with chronic illnesses. We evaluated the accuracy of one of these devices, the TOSCA® 500, in adults admitted to an emergency department for acute respiratory failure.

Methods

We prospectively collected 29 pairs of simultaneous transcutaneous arterial CO2 (PtcCO2) and arterial CO2 (PaCO2) values in 21 consecutive adults with acute respiratory failure (acute heart failure, n = 6; COPD exacerbation, n = 8; acute pneumonia, n = 6; and pulmonary embolism, n = 1). Agreement between PaCO2 and PtcCO2 was evaluated using the Bland-Altman method.

Results

Mean arterial oxygen saturation was 90%, arterial oxygen tension ranged from 32 to 215 mmHg, and PaCO2 ranged from 23 to 84 mmHg. The mean difference between PaCO2 and PtcCO2 was 0.1 mmHg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from −6 to 6.2 mmHg. None of the patients experienced adverse effects from heating of the device clipped to the earlobe.

Conclusion

PtcCO2 showed good agreement with PaCO2 in adults with acute respiratory failure.
Literatur
1.
Zurück zum Zitat Crawford A (2004) An audit of the patient’s experience of arterial blood gas testing. Br J Nurs 13:529–532PubMed Crawford A (2004) An audit of the patient’s experience of arterial blood gas testing. Br J Nurs 13:529–532PubMed
2.
Zurück zum Zitat Zaloga GP (1990) Evaluation of bedside testing options for the critical care unit. Chest 97:185S–190SPubMed Zaloga GP (1990) Evaluation of bedside testing options for the critical care unit. Chest 97:185S–190SPubMed
3.
Zurück zum Zitat Thorson SH, Marini JJ, Pierson DJ, Hudson LD (1983) Variability of arterial blood gas values in stable patients in the icu. Chest 84:14–18CrossRefPubMed Thorson SH, Marini JJ, Pierson DJ, Hudson LD (1983) Variability of arterial blood gas values in stable patients in the icu. Chest 84:14–18CrossRefPubMed
4.
Zurück zum Zitat Eberhard P (2007) The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg 105:S48–S52CrossRefPubMed Eberhard P (2007) The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg 105:S48–S52CrossRefPubMed
5.
Zurück zum Zitat Rosner V, Hannhart B, Chabot F, Polu JM (1999) Validity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure. Eur Respir J 13:1044–1047CrossRefPubMed Rosner V, Hannhart B, Chabot F, Polu JM (1999) Validity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure. Eur Respir J 13:1044–1047CrossRefPubMed
6.
Zurück zum Zitat Sanders MH, Kern NB, Costantino JP, Stiller RA, Strollo PJ Jr, Studnicki KA, Coates JA, Richards TJ (1994) Accuracy of end-tidal and transcutaneous pco2 monitoring during sleep. Chest 106:472–483CrossRefPubMed Sanders MH, Kern NB, Costantino JP, Stiller RA, Strollo PJ Jr, Studnicki KA, Coates JA, Richards TJ (1994) Accuracy of end-tidal and transcutaneous pco2 monitoring during sleep. Chest 106:472–483CrossRefPubMed
7.
Zurück zum Zitat Cuvelier A, Grigoriu B, Molano LC, Muir JF (2005) Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation. Chest 127:1744–1748CrossRefPubMed Cuvelier A, Grigoriu B, Molano LC, Muir JF (2005) Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation. Chest 127:1744–1748CrossRefPubMed
8.
Zurück zum Zitat Maniscalco M, Zedda A, Faraone S, Carratu P, Sofia M (2008) Evaluation of a transcutaneous carbon dioxide monitor in severe obesity. Intensive Care Med 34:1340–1344CrossRefPubMed Maniscalco M, Zedda A, Faraone S, Carratu P, Sofia M (2008) Evaluation of a transcutaneous carbon dioxide monitor in severe obesity. Intensive Care Med 34:1340–1344CrossRefPubMed
9.
Zurück zum Zitat McVicar J, Eager R (2009) Validation study of a transcutaneous carbon dioxide monitor in patients in the emergency department. Emerg Med J 26:344–346CrossRefPubMed McVicar J, Eager R (2009) Validation study of a transcutaneous carbon dioxide monitor in patients in the emergency department. Emerg Med J 26:344–346CrossRefPubMed
10.
Zurück zum Zitat Parker SM, Gibson GJ (2007) Evaluation of a transcutaneous carbon dioxide monitor (“Tosca”) in adult patients in routine respiratory practice. Respir Med 101:261–264CrossRefPubMed Parker SM, Gibson GJ (2007) Evaluation of a transcutaneous carbon dioxide monitor (“Tosca”) in adult patients in routine respiratory practice. Respir Med 101:261–264CrossRefPubMed
11.
Zurück zum Zitat Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA (2005) Transcutaneous pco2 monitoring in critically ill adults: clinical evaluation of a new sensor. Crit Care Med 33:2203–2206CrossRefPubMed Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA (2005) Transcutaneous pco2 monitoring in critically ill adults: clinical evaluation of a new sensor. Crit Care Med 33:2203–2206CrossRefPubMed
12.
Zurück zum Zitat Kagawa S, Otani N, Kamide M, Gisiger PA, Eberhard P, Severinghaus JW (2004) Initial transcutaneous PCO2 overshoot with ear probe at 42 degrees c. J Clin Monit Comput 18:343–345CrossRefPubMed Kagawa S, Otani N, Kamide M, Gisiger PA, Eberhard P, Severinghaus JW (2004) Initial transcutaneous PCO2 overshoot with ear probe at 42 degrees c. J Clin Monit Comput 18:343–345CrossRefPubMed
13.
Metadaten
Titel
Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure
verfasst von
P.-E. Gancel
E. Roupie
L. Guittet
S. Laplume
N. Terzi
Publikationsdatum
01.02.2011
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 2/2011
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-010-2076-1

Weitere Artikel der Ausgabe 2/2011

Intensive Care Medicine 2/2011 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.