01.08.2012 | Physiological and Technical Notes

Bench testing of a new hyperbaric chamber ventilator at different atmospheric pressures

verfasst von: Jean-Claude Lefebvre, Aissam Lyazidi, Miguel Parceiro, Giuseppe F. Sferrazza Papa, Evangelia Akoumianaki, Deborah Pugin, Didier Tassaux, Laurent Brochard, Jean-Christophe M. Richard

Erschienen in: Intensive Care Medicine | Ausgabe 8/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Providing mechanical ventilation is challenging at supra-atmospheric pressure. The higher gas density increases resistance, reducing the flow delivered by the ventilator. A new hyperbaric ventilator (Siaretron IPER 1000) is said to compensate for these effects automatically. The aim of this bench test study was to validate the compensation, define its limits and provide details on the ventilator’s output at varied atmospheric pressures.

Methods

Experiments were conducted inside a multiplace hyperbaric chamber at 1, 2.2, 2.8 and 4 atmospheres absolute (ATA), with the ventilator connected to a test lung. Transducers were recalibrated at each ATA level. Various ventilator settings were tested in volume and pressure control modes. Measured tidal volumes were compared with theoretical predictions based on gas laws.

Results

Results confirmed the ventilator’s ability to provide compensation, but also identified its limits. The compensation range could be predicted and depended on the maximal flow attainable, decreasing linearly with increasing atmospheric pressure. With settings inside the range, tidal volumes approximated set values (mean error 10 ± 5 %). With settings outside the range, the volume was limited to the predicted maximal value calculated from maximal flow. A practical guide for clinicians is provided.

Conclusion

The IPER 1000 ventilator attempted to deliver stable tidal volume by adjusting the opening of the inspiratory valve in proportion to atmospheric pressure. Adequate compensation was observed, albeit only within a predictable range, which can be reliably predicted for each setting and ATA level combination. Setting a tidal volume outside this range can result in an unwanted decrease in minute ventilation.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
2.
Zurück zum Zitat Skinner M (1998) Ventilator function under hyperbaric conditions. Diving Hyperb Med J 28:62–71 Skinner M (1998) Ventilator function under hyperbaric conditions. Diving Hyperb Med J 28:62–71
3.
Zurück zum Zitat Workman WT (1999) Hyperbaric facility safety: a practical guide. Best Pub. Co., Flagstaff Workman WT (1999) Hyperbaric facility safety: a practical guide. Best Pub. Co., Flagstaff
4.
Zurück zum Zitat Stahl W, Radermacher P, Calzia E (2000) Functioning of ICU ventilators under hyperbaric conditions–comparison of volume- and pressure-controlled modes. Intensive Care Med 26:442–448PubMedCrossRef Stahl W, Radermacher P, Calzia E (2000) Functioning of ICU ventilators under hyperbaric conditions–comparison of volume- and pressure-controlled modes. Intensive Care Med 26:442–448PubMedCrossRef
5.
Zurück zum Zitat Stanga D, Beck G, Chimiak J (2003) Evaluation of respiratory support devices for use in the hyperbaric chamber. NEDU TR 03-18 Panama City, FL, Department of Defense, Navy Experimental Diving Unit Stanga D, Beck G, Chimiak J (2003) Evaluation of respiratory support devices for use in the hyperbaric chamber. NEDU TR 03-18 Panama City, FL, Department of Defense, Navy Experimental Diving Unit
6.
Zurück zum Zitat Blanch PB, Desautels DA, Gallagher TJ (1989) Mechanical ventilator function under hyperbaric conditions. Undersea Hyperb Med 16 (Supp):17 Blanch PB, Desautels DA, Gallagher TJ (1989) Mechanical ventilator function under hyperbaric conditions. Undersea Hyperb Med 16 (Supp):17
7.
Zurück zum Zitat Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC (2010) Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med 36:2074–2080PubMedCrossRef Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC (2010) Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med 36:2074–2080PubMedCrossRef
8.
Zurück zum Zitat Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376PubMedCrossRef Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376PubMedCrossRef
9.
Zurück zum Zitat West JB (1999) The original presentation of Boyle’s law. J Appl Physiol 87:1543–1545PubMed West JB (1999) The original presentation of Boyle’s law. J Appl Physiol 87:1543–1545PubMed
10.
Zurück zum Zitat Tassaux D, Jolliet P, Thouret JM, Roeseler J, Dorne R, Chevrolet JC (1999) Calibration of seven ICU ventilators for mechanical ventilation with helium-oxygen mixtures. Am J Respir Crit Care Med 160:22–32PubMed Tassaux D, Jolliet P, Thouret JM, Roeseler J, Dorne R, Chevrolet JC (1999) Calibration of seven ICU ventilators for mechanical ventilation with helium-oxygen mixtures. Am J Respir Crit Care Med 160:22–32PubMed
11.
Zurück zum Zitat Weaver LK, Greenway L, Elliot CG (1988) Performance of the Sechrist 500A hyperbaric ventilator in a monoplace hyperbaric chamber. J Hyperb Med 3:215–225 Weaver LK, Greenway L, Elliot CG (1988) Performance of the Sechrist 500A hyperbaric ventilator in a monoplace hyperbaric chamber. J Hyperb Med 3:215–225
12.
Zurück zum Zitat Churchill S, Weaver LK, Haberstock D (1999) Performance of the Omni-vent mechanical ventilator for use with the monoplace hyperbaric chamber. Undersea Hyperb Med 26(Suppl):70–71 Churchill S, Weaver LK, Haberstock D (1999) Performance of the Omni-vent mechanical ventilator for use with the monoplace hyperbaric chamber. Undersea Hyperb Med 26(Suppl):70–71
13.
Zurück zum Zitat Moon RE, Bergquist LV, Conklin B, Miller JN (1986) Monaghan 225 ventilator use under hyperbaric conditions. Chest 89:846–851PubMedCrossRef Moon RE, Bergquist LV, Conklin B, Miller JN (1986) Monaghan 225 ventilator use under hyperbaric conditions. Chest 89:846–851PubMedCrossRef
14.
Zurück zum Zitat Hennessey PJ (2007) Evaluation of the UNI-VENT Eagle Model 754 ventilator for use in the surface recompression chambers (SRC) of the submarine rescue diving and recompression system (SRDRS). NEDU TR 07-03 Panama City, FL, Department of Defense, Navy Experimental Diving Unit Hennessey PJ (2007) Evaluation of the UNI-VENT Eagle Model 754 ventilator for use in the surface recompression chambers (SRC) of the submarine rescue diving and recompression system (SRDRS). NEDU TR 07-03 Panama City, FL, Department of Defense, Navy Experimental Diving Unit
Metadaten
Titel
Bench testing of a new hyperbaric chamber ventilator at different atmospheric pressures
verfasst von
Jean-Claude Lefebvre
Aissam Lyazidi
Miguel Parceiro
Giuseppe F. Sferrazza Papa
Evangelia Akoumianaki
Deborah Pugin
Didier Tassaux
Laurent Brochard
Jean-Christophe M. Richard
Publikationsdatum
01.08.2012
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 8/2012
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-012-2590-4

Weitere Artikel der Ausgabe 8/2012

Intensive Care Medicine 8/2012 Zur Ausgabe

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.