Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 1/2018

25.01.2017 | Original Research

Feasibility of automated titration of vasopressor infusions using a novel closed-loop controller

verfasst von: Joseph Rinehart, Michael Ma, Michael-David Calderon, Maxime Cannesson

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2018

Einloggen, um Zugang zu erhalten

Abstract

Blood pressure management is a central concern in critical care patients. For a variety of reasons, titration of vasopressor infusions may be an ideal use-case for computer assistance. Using our previous experience gained in the bench-to-bedside development of a computer-assisted fluid management system, we have developed a novel controller for this purpose. The aim of this preliminary study was to assess the feasibility of using this controller in simulated patients to maintain a target blood pressure in both stable and variable blood-pressure scenarios. We tested the controller in two sets of simulation scenarios: one with stable underlying blood pressure and a second with variable underlying blood pressure. In addition, in the variable phase of the study, we tested infusion-line delays of 8–60 s. The primary outcome for both testing conditions (stable and variable) was % case time in target range. We determined a priori that acceptable performance on the first phase of the protocol would require greater than 95% case-time in-target given the simple nature of the protocol, and for the second phase of the study 80% or greater given the erratic nature of the blood pressure changes taking place. 250 distinct cases for each simulation condition, both managed and unmanaged, were run over 4 days. In the stable hemodynamic conditions, the unmanaged group had an MAP of 57.5 ± 4.6 mmHg and spent only 5.6% of case time in-target. The managed group had an MAP of 70.3 ± 2.6 and spent a total of 99.5% of case time in-target (p < 0.00001 for both comparisons between groups). In the variable hemodynamic conditions, the unmanaged group had an MAP of 53.1 ± 5.0 mmHg and spent 0% of case time in-target. The managed group had an MAP of 70.5 ± 3.2 mmHg (p < 0.00001 compared to unmanaged group) and spent 88.6% of case time in-target (p < 0.00001 compared to unmanaged group), with 6.4% of case time over and 5.1% of case time under target. Increasing infusion lag increased coefficient of variation by about 10% per 15 s of lag (p = 0.001). This study demonstrated that this novel controller for vasopressor administration is able to main a target mean arterial pressure in a simulated physiologic model in the face of random disturbances and infusion-line lag.
Literatur
1.
Zurück zum Zitat Rinehart J, et al. Review article: closed-loop systems in anesthesia: is there a potential for closed-loop fluid management and hemodynamic optimization? Anesth Analg. 2012;114(1):130–43.CrossRefPubMed Rinehart J, et al. Review article: closed-loop systems in anesthesia: is there a potential for closed-loop fluid management and hemodynamic optimization? Anesth Analg. 2012;114(1):130–43.CrossRefPubMed
3.
Zurück zum Zitat Battelino T, Omladic JS, Phillip M. Closed loop insulin delivery in diabetes. Best Pract Res Clin Endocrinol Metab. 2015;29(3):315–25.CrossRefPubMed Battelino T, Omladic JS, Phillip M. Closed loop insulin delivery in diabetes. Best Pract Res Clin Endocrinol Metab. 2015;29(3):315–25.CrossRefPubMed
4.
Zurück zum Zitat Rinehart J, Canales C. Closed-loop pharmacology in anesthesia and critical care: benefits and limitations. Int Anesthesiol Clin. 2015;53(2):91–101.CrossRefPubMed Rinehart J, Canales C. Closed-loop pharmacology in anesthesia and critical care: benefits and limitations. Int Anesthesiol Clin. 2015;53(2):91–101.CrossRefPubMed
5.
Zurück zum Zitat Dumont GA, Ansermino JM. Closed-loop control of anesthesia: a primer for anesthesiologists. Anesth Analg. 2013;117(5):1130–8.CrossRefPubMed Dumont GA, Ansermino JM. Closed-loop control of anesthesia: a primer for anesthesiologists. Anesth Analg. 2013;117(5):1130–8.CrossRefPubMed
7.
Zurück zum Zitat Bighamian R, et al. Is there opportunity for automated decision-support and closed-loop control in ICU patients receiving vasopressor infusion? Conf Proc IEEE Eng Med Biol Soc. 2014;2014:1949–52.PubMed Bighamian R, et al. Is there opportunity for automated decision-support and closed-loop control in ICU patients receiving vasopressor infusion? Conf Proc IEEE Eng Med Biol Soc. 2014;2014:1949–52.PubMed
8.
Zurück zum Zitat Asfar P, et al. High versus low blood-pressure target in patients with septic shock. N Engl J Med. 2014;370(17):1583–93.CrossRefPubMed Asfar P, et al. High versus low blood-pressure target in patients with septic shock. N Engl J Med. 2014;370(17):1583–93.CrossRefPubMed
9.
Zurück zum Zitat Takala J. Should we target blood pressure in sepsis? Crit Care Med. 2010;38(10 Suppl):S613–S9.CrossRefPubMed Takala J. Should we target blood pressure in sepsis? Crit Care Med. 2010;38(10 Suppl):S613–S9.CrossRefPubMed
11.
Zurück zum Zitat Upton RN, Ludbrook G. A physiologically based, recirculatory model of the kinetics and dynamics of propofol in man. Anesthesiology. 2005;103(2):344–52.CrossRefPubMed Upton RN, Ludbrook G. A physiologically based, recirculatory model of the kinetics and dynamics of propofol in man. Anesthesiology. 2005;103(2):344–52.CrossRefPubMed
12.
Zurück zum Zitat Rinehart J, et al. Evaluation of a novel closed-loop fluid-administration system based on dynamic predictors of fluid responsiveness: an in silico simulation study. Crit Care. 2011;15(6):R278.CrossRefPubMedPubMedCentral Rinehart J, et al. Evaluation of a novel closed-loop fluid-administration system based on dynamic predictors of fluid responsiveness: an in silico simulation study. Crit Care. 2011;15(6):R278.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Rinehart J, et al. Closed-loop fluid resuscitation: robustness against weight and cardiac contractility variations. Anesth Analg. 2013;117:1110–8CrossRefPubMed Rinehart J, et al. Closed-loop fluid resuscitation: robustness against weight and cardiac contractility variations. Anesth Analg. 2013;117:1110–8CrossRefPubMed
Metadaten
Titel
Feasibility of automated titration of vasopressor infusions using a novel closed-loop controller
verfasst von
Joseph Rinehart
Michael Ma
Michael-David Calderon
Maxime Cannesson
Publikationsdatum
25.01.2017
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2018
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-017-9981-6

Weitere Artikel der Ausgabe 1/2018

Journal of Clinical Monitoring and Computing 1/2018 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.