Validation of noninvasive continuous arterial pressure measurement by ClearSight System™ during induction of anesthesia for cardiovascular surgery

verfasst von: Tadashi Tanioku, Akari Yoshida, Yuichi Aratani, Keisuke Fujii, Tomoyuki Kawamata

Erschienen in: BMC Anesthesiology | Ausgabe 1/2020

Abstract

Background

Since blood pressure tends to be unstable during induction of anesthesia in patients undergoing cardiovascular surgery, an artery catheter is often inserted before induction to continuously monitor arterial pressure during induction of anesthesia. ClearSight System™ enables noninvasive continuous measurement of beat-to-beat arterial pressure via a single finger cuff without pain using photoplethysmographic technology. If ClearSight System™ can replace intra-arterial pressure measurement, blood pressure could be easily and noninvasively assessed. However, the validity of ClearSight System™ during induction of anesthesia in patients undergoing cardiovascular surgery has not been evaluated. The aim of this study was to compare blood pressure measured by ClearSight System™ with intra-arterial pressure during induction of anesthesia for cardiovascular surgery.

Methods

This study was registered retrospectively. Data during induction of anesthesia for elective cardiovascular surgery were obtained for patients in whom noninvasive arterial pressure was measured by ClearSight System™ (APcs) and invasive radial arterial pressure (APrad) was measured simultaneously. According to the widely used criteria formulated by international standards from the Association for the Advancement of Medical Instrumentation, the acceptable bias and precision for arterial pressure measurements were fixed at < 5 mmHg and 8 mmHg, respectively.

Results

Data for 18 patients were analyzed. For 3068 analyzed paired measurements, values of APcs vs APrad bias (precision) were 13.2 (17.5), − 9.1 (7.3) and − 3.9 (7.8) mmHg for systolic, diastolic, and mean arterial pressures, respectively.

Conclusions

Mean arterial pressure measured by ClearSight System™ could be considered as an alternative for mean radial arterial pressure during induction of anesthesia for elective cardiovascular surgery.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
APcs
Noninvasive arterial pressure measured by ClearSight System™
APrad
Invasive radial arterial pressure
LOA
Limits of agreement
SAP
Systolic arterial pressure
DAP
Diastolic arterial pressure
MAP
Mean arterial pressure
AAMI
Association for the Advancement of Medical Instrumentation

Background

Since blood pressure tends to be unstable during induction of anesthesia in patients undergoing cardiovascular surgery, an artery catheter is often inserted before induction to continuously monitor arterial pressure during induction of anesthesia. The success rate of the first attempt at arterial cannulation using palpation has been reported to be less than 50% and sometimes cannulation still fails despite the use of ultrasound [1]. Therefore, arterial cannulation in an awake condition may cause suffering for patients.
ClearSight System™ (previously named ccNexfin system™, Edwards Lifesciences Corp, Irvine CA, USA) enables noninvasive continuous measurement of beat-to-beat arterial pressure via a single finger cuff without pain using photoplethysmographic technology. If ClearSight System™ can replace intra-arterial pressure measurement, blood pressure could be continuously, easily, and noninvasively assessed. Previous studies have shown that this device is reliable in pregnant women [2], children [3], and patients undergoing upper abdominal surgery [4]. On the other hand, it has been reported that it is not reliable in critically ill patients [5] and patients undergoing neurosurgery in a sitting position [6]. Accordingly, the validity of ClearSight System™ may depend on the clinical situation including the type of surgery or the patient’s condition. However, the validity of ClearSight System™ during induction of anesthesia in patients undergoing cardiovascular surgery has not been evaluated. The aim of this study was to compare blood pressure measured by ClearSight System™ with intra-arterial pressure during induction of anesthesia for cardiovascular surgery.

Methods

Study design and setting

This retrospective observational study was approved by the medical ethics committee of Wakayama Medical University prior to its initiation (reference number 1919). The study was conducted at Wakayama Medical University Hospital.

Data collection

In this retrospective analysis, data were collected from all patients in whom noninvasive arterial pressure was measured by ClearSight System™ (APcs) and invasive radial arterial pressure (APrad) was measured simultaneously during induction of anesthesia for elective cardiovascular surgery between November 2016 and November 2017 at Wakayama Medical University Hospital. The use of ClearSight System™ depended on the anesthesiologists in charge. Paired values of systolic, diastolic, and mean arterial pressures obtained by both methods were recorded at the rate of 1 sample every 3 s in the institution’s Anesthesia Information Management System (PrimeGAIA™, Nihon Kohden Co, Tokyo, Japan). Data from 5 min before tracheal intubation to 5 min after tracheal intubation based on anesthetic records were analyzed.
In our hospital, the induction of cardiovascular anesthesia has been standardized. Two anesthesiologists are generally involved in one case: one anesthesiologist for managing the anesthesia, and other for recording. Before induction of anesthesia, an catheter is inserted into right radial artery in the most of patients to continuously measure blood pressure. Then anesthesia is induced by target-controlled infusion of propofol (1.5–3.0 μg/ml), remifentanil (0.1–0.3 μg/kg/min), and rocuronium (0.6–1.0 mg/kg). The doses of anesthetics depend on the decision of the anesthesiologist in charge. When blood pressure decreases during the induction, ephedrine (4 mg or 8 mg) or phenylephrine (0.1 mg or 0.2 mg) is intravenously administered according to the decision of the anesthesiologist in charge.

Statistical analysis

There is no established knowledge of how many patients should be included and how many measurements should be analyzed for each when performing a repeated measurement. In most of the studies using Bland-Altman analysis, the sample size was not examined. In this study, we collected over 3000 pairs of data based on similar previous studies in which radial arterial pressure was compared with blood pressure measured by ClearSight System™ [7, 8].
Data considered to be artifacts were excluded based on ClearSight auto-calibration, radial artery artifacts, and ClearSight artifacts. Auto-calibration is performed at least once every 70 heartbeats to keep the finger arteries open and of constant diameter. In addition, auto-calibration is performed when the measurement of blood pressure is temporarily interrupted for two or more beats. When auto-calibration is performed, systolic, diastolic and mean blood pressures become the same values, and the values increase step by step. Therefore, it is possible to discriminate such data as artifacts. Radial artery artifacts, which result from blood sampling and flushing, could be discriminated since systolic and diastolic pressures become the same values. ClearSight artifacts, which occur due to external pressure to the ClearSight cuff, can be recognized as extreme outliers.
Data are expressed as means (SD) or medians (interquartile range) as appropriate. For whole repeated paired measurements from all patients, correlations between APrad and APcs were determined by a linear regression. In addition, Bland-Altman analysis was used to study agreement between APrad and APcs. In this analysis, bias and precision were defined as the mean difference between APrad and APcs and as the SD of bias, respectively. In addition, limits of agreement (LOA) were calculated as bias ±2SD of bias. According to the widely used criteria formulated by international standards from the Association for the Advancement of Medical Instrumentation (AAMI), the acceptable bias and precision for arterial pressure measurements were fixed at < 5 mmHg and 8 mmHg, respectively [9].
For each patient, the SDs of averages of APrad and APcs (“within-subject variability”) were calculated to quantify the ranges of different pressures. The SDs of differences between APrad and APcs (“within-subject precision”) were also calculated to quantify tracking for systolic arterial pressure (SAP), diastolic arterial pressure (DAP), and mean arterial pressure (MAP). In addition, correlations between APrad and APcs were determined by linear regression.
A two-sided P-value of 0.05 was considered statistically significant. All analyses were performed using JMP statistical software (version 12.2; SAS Institute, Cary NC, USA).

Results

Data for 18 patients were obtained in this study. The characteristics of the patients are shown in Table 1. Given the retrospective nature of this study, all perioperative management was at the direction of the attending clinicians. In all patients, a 22-guage catheter was used for monitoring radial arterial pressure. Both APrad and APcs were measured on the right side in all patients, and noninvasive blood pressure measurement by a cuff was performed on the left arm. Although we obtained 3600 pairs of APcs and APrad, 532 pairs among them were excluded. Of the 532 measurements excluded, 297 measurements were excluded due to ClearSight auto-calibration. In addition, 115 measurements were excluded due to radial artery artifacts, and 120 measurements were excluded due to ClearSight artifacts. The percentage of exclusion data in our data (14.7%) was similar to the percentages in previous prospective studies [7, 8]. Thus, a total of 3068 valid pairs of simultaneous APcs and APrad measurements were analyzed. The median number of paired measurements per patient was 170 (170–200). The ranges of APrad measured during the observational period were 53–225 mmHg for SAP, 27–114 mmHg for DAP, and 41–144 mmHg for MAP. Continuous administration of phenylephrine was started from the beginning of anesthetic induction in 9 of the 18 patients.
Table 1
Baseline characteristics of the subjects
Age (yr)
70 (64–79)
Sex (M/F)
11/7
BMI (kg・m-2)
24 (21.9–24.9)
Type of surgery (%)
 Coronary artery bypass grafting
11
 Aortic valve replacement for aortic valve stenosis
7
 Aortic valve replacement for aortic regurgitation
2
 Thoracic ascending aortic graft replacement
1
Comorbidities (%)
 Hypertension
18
 Diabetes mellitus
9
 Renal insufficiency
10
 Dialysis
2
N = 18. Values are medians (interquartile range) or numbers
BMI Body mass index
Figure 1 shows individual scatter plots for SAP, DAP, and MAP. Correlation coefficients, within-subject variability, and within-subject precision are summarized in Table 2. Mean differences of pressure in paired data were 13.1 ± 15.5, − 8.5 ± 6.1, and − 3.4 ± 6.2 mmHg for SAP, DAP, and MAP, respectively. Figure 2 shows the correlations between APcs and APrad. APcs for SAP, DAP and MAP were significantly correlated with APrad. The correlation coefficients between APcs and APrad for SAP, DAP, and MAP were 0.85, 0.85 and 0.92, respectively. The results of Bland-Altman analysis between APcs and APrad are shown in Fig. 3. Bias and precision were 13.2 and 17.5 mmHg in SAP, − 9.1 and 7.3 mmHg in DAP, and − 3.9 mmHg 7.8 mmHg in MAP. Upper and lower LOAs were 47.4 and − 21.1 mmHg in SAP, 5.2 and − 23.4 mmHg in DAP, and 11.4 and − 19.2 mmHg in MAP. Accordingly, only MAP fulfilled the criteria of AAMI.
Table 2
Within-subject data averaged over the group
 
r, Median (25-75%)
Within-subject Variability
Within-subject Precision
SAP (mmHg)
0.95 (0.89–0.96)
17.3 (4.7)
7.0 (2.6)
DAP (mmHg)
0.91 (0.86–0.95)
8.6 (2.3)
4.1 (1.4)
MAP (mmHg)
0.95 (0.91–0.96)
12.1 (2.7)
4.6 (1.4)
Data are presented as medians (25th-75th percentiles) for correlations and as means (SD) for within-subject variability and within-subject precision in 18 subjects. r, coefficient of correlation
SAP Systolic arterial pressure, DAP Diastolic arterial pressure, MAP Mean arterial pressure

Discussion

In this study, Pearson’s correlation coefficients showed that SAP, DAP and MAP measured by ClearSight System™ were significantly correlated with radial arterial pressure. Bland-Altman analysis showed that ClearSight System™ had acceptable bias and precision in MAP but not in SAP and DAP for radial arterial pressure measurements. Our results suggest that changes in SAP, DAP and MAP measured by ClearSight System™ reflect those in radial arterial pressure and that MAP measured by ClearSight System™ is interchangeable with radial arterial pressure during induction of anesthesia for elective cardiovascular surgery.
Our study showed that MAP measured by ClearSight System™ statistically matched the AAMI criteria. On the other hand, neither SAP nor DAP matched the criteria. A previous study also showed that MAP, but not SAP and DAP, could be considered as an alternative for radial artery blood pressure during carotid endarterectomy, based on AAMI criteria [8]. In general, the arterial pressure waveform changes gradually from the brachial artery to the finger arteries [10]. Accordingly, SAP at a distal site to the heart is higher than that at a proximal site, while DAP at a distal site is lower than that at a proximal site. ClearSight System™ reconstructs brachial artery pressure from finger artery pressure for calculating blood pressure [10]. Therefore, there might be a significant difference between SAP/DAP measured by ClearSight System™ and radial artery pressure. On the other hand, as blood flows from the aorta to the radial artery, mean pressure decreases only slightly because there is little resistance to flow in the major conducting arteries [11]. In our study, the mean difference in MAP between APcs and APrad was − 3.4 ± 6.2 mmHg, which was small compared to the differences in SAP and DAP (13.1 ± 15.5 and − 8.5 ± 6.1 mmHg, respectively).
The aim of managing hemodynamics is to maintain adequate organ perfusion. MAP is widely used as an index for optimal blood pressure, and it reflects driving pressure at the organ level [12]. MAP is the value that has most often been used for assessing autoregulation of renal blood flow [13] and cerebral blood flow [14]. Measurement of MAP by ClearSight System™ would be useful for maintaining organ perfusion during induction of anesthesia for cardiovascular surgery. During induction of anesthesia for patients with coronary artery disease, maintain of DAP is also important. Our results showed that diastolic pressure measured by ClearSight System™ is not interchangeable with radial diastolic pressure but correlates well with it. Therefore, we can pay attention to coronary perfusion by assessing the change in diastolic pressure but not absolute values measured by ClearSight System™.
The limitation of this study would result from a retrospective nature. The quality of data in a prospective study are generally higher than that in a retrospective study. When APcs is compared to APrad, the advantage of a prospective study is that study conditions including patient’s bias, data collection period, and exclusion of data influenced by artifacts can be controlled. In this study, we decided to analyzed preserved data for the following reasons. First, when we reviewed our preserved data from 18 patients before data analysis, the characteristics of patients were similar to those in previous prospective studies. The exclusion criteria in previous studies included peripheral arterial disease, preoperative atrial fibrillation, and obesity (> BMI 30) [4, 7, 15, 16], and such patients were also not included in our study. Second, when data are collected from the anesthetic record, the time to intubate may not be accurate. In our hospital, two anesthesiologists are generally involved in one case: one anesthesiologist for managing the anesthesia, and other for recording. Therefore, we considered the time of events and data collection period (from 5 min before intubation to 5 min after intubation) would be accurate. Third, in a prospective study, data are excluded due to auto-calibration of APcs, unreliable radial artery wave, flushing arterial line, or APcs artifacts. Among them, auto-calibration of APcs, flushing an arterial line and APcs artifacts can be retrospectively discriminated. Accordingly, the percentage of exclusion data in our data (14.7%) was similar to the percentages in previous prospective studies [7, 8]. In addition, within-subject variability and within-subject precision in our data (Table 2) were similar to those in the previous prospective studies [8, 17], and we therefore considered that the quality of our data is adequate for analysis. For the above reasons, we considered that analysis of preserved data would be less inferior to analysis of prospective data when APcs is compared to APrad. However, a prospective study will be needed to obtain a more precise evaluation of ClearSight System™.

Conclusions

MAP measured by ClearSight System™ could be considered as an alternative for mean radial arterial pressure during induction of anesthesia for elective cardiovascular surgery. SAP and DAP may be useful for inferring changes in systolic and diastolic radial arterial pressures.

Acknowledgements

Not applicable.
This study was approved by the medical ethics committee of Wakayama Medical University prior to its initiation (reference number 1919).
Consent to participate was waved according to our local ethical committee policy.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Shiloh AL, Savel RH, Paulin LM, Eisen LA. Ultrasound-guided catheterization of the radial artery: a systematic review and meta-analysis of randomized controlled trials. Chest. 2011;139:524–9.CrossRef Shiloh AL, Savel RH, Paulin LM, Eisen LA. Ultrasound-guided catheterization of the radial artery: a systematic review and meta-analysis of randomized controlled trials. Chest. 2011;139:524–9.CrossRef
2.
Zurück zum Zitat Akkermans J, Diepeveen M, Ganzevoort W, van Montfans GA, Westerhof BE, Wolf H. Continuous non-invasive blood pressure monitoring, a validation study of Nexfin in a pregnant population. Hypertens Pregnancy. 2009;28:230–42.CrossRef Akkermans J, Diepeveen M, Ganzevoort W, van Montfans GA, Westerhof BE, Wolf H. Continuous non-invasive blood pressure monitoring, a validation study of Nexfin in a pregnant population. Hypertens Pregnancy. 2009;28:230–42.CrossRef
3.
Zurück zum Zitat Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C. Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth. 2012;109:609–15.CrossRef Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C. Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth. 2012;109:609–15.CrossRef
4.
Zurück zum Zitat de Wilde RB, de Wit F, Geerts BF, van Vliet AL, Aarts LP, Vuyk J, Jansen JR. Non-invasive continuous arterial pressure and pulse pressure variation measured with Nexfin® in patients following major upper abdominal surgery: a comparative study. Anaethesia. 2016;71:788–97.CrossRef de Wilde RB, de Wit F, Geerts BF, van Vliet AL, Aarts LP, Vuyk J, Jansen JR. Non-invasive continuous arterial pressure and pulse pressure variation measured with Nexfin® in patients following major upper abdominal surgery: a comparative study. Anaethesia. 2016;71:788–97.CrossRef
5.
Zurück zum Zitat Ameloot K, Van De Vijver K, Van Regenmortel N, De Laet I, Schoonheydt K, Dits H, Broch O, Bein B, Malbrain ML. Validation study of Nexfin® continuous non-invasive blood pressure monitoring in critically ill adult patients. Minerva Anestesiol. 2014;80:1294–301.PubMed Ameloot K, Van De Vijver K, Van Regenmortel N, De Laet I, Schoonheydt K, Dits H, Broch O, Bein B, Malbrain ML. Validation study of Nexfin® continuous non-invasive blood pressure monitoring in critically ill adult patients. Minerva Anestesiol. 2014;80:1294–301.PubMed
6.
Zurück zum Zitat Schramm P, Tzanova I, Gööck T, Hagen F, Schmidtmann I, Engelhard K, Pestel G. Noninvasive hemodynamic measurements during neurosurgical procedures in sitting position. J Neurosurg Anesthesiol. 2017;29:251–7.CrossRef Schramm P, Tzanova I, Gööck T, Hagen F, Schmidtmann I, Engelhard K, Pestel G. Noninvasive hemodynamic measurements during neurosurgical procedures in sitting position. J Neurosurg Anesthesiol. 2017;29:251–7.CrossRef
7.
Zurück zum Zitat Weiss E, Gayat E, Dumans-Nizard V, Le Guen M, Fischler M. Use of the Nexfin™ device to detect acute arterial pressure variations during anaesthesia induction. Brit J Anaesth. 2014;113:52–60.CrossRef Weiss E, Gayat E, Dumans-Nizard V, Le Guen M, Fischler M. Use of the Nexfin™ device to detect acute arterial pressure variations during anaesthesia induction. Brit J Anaesth. 2014;113:52–60.CrossRef
8.
Zurück zum Zitat Heusdens JF, Lof S, Pennekamp CW, Specken-Welleweerd JC, de Borst GJ, van Klei WA, van Wolfswinkel L, Immink RV. Validation of non-invasive arterial pressure monitoring during carotid endarterectomy. Br J Anaesth. 2016;117:316–23.CrossRef Heusdens JF, Lof S, Pennekamp CW, Specken-Welleweerd JC, de Borst GJ, van Klei WA, van Wolfswinkel L, Immink RV. Validation of non-invasive arterial pressure monitoring during carotid endarterectomy. Br J Anaesth. 2016;117:316–23.CrossRef
9.
Zurück zum Zitat Vos JJ, Poterman M, Mooyaart EAQ, Weening M, Struys MMRF, Scheeren TWL, Kalmar AF. Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia. Br J Anaesth. 2014;113:67–74.CrossRef Vos JJ, Poterman M, Mooyaart EAQ, Weening M, Struys MMRF, Scheeren TWL, Kalmar AF. Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia. Br J Anaesth. 2014;113:67–74.CrossRef
10.
Zurück zum Zitat Gizdulich P, Prentza A, Wesseling KH. Models of brachial to finger pulse wave distortion and pressure decrement. Cardiovasc Res. 1997;33:698–705.CrossRef Gizdulich P, Prentza A, Wesseling KH. Models of brachial to finger pulse wave distortion and pressure decrement. Cardiovasc Res. 1997;33:698–705.CrossRef
11.
Zurück zum Zitat Schroeder B, Barbeito A, Bar-Yosef S, Mark JB. Cardiovascular monitoring. In: Miller RD, editor. Miller’s anesthesia eighth edition. Philadelphia: WB Saunders; 2015. p. 1356. Schroeder B, Barbeito A, Bar-Yosef S, Mark JB. Cardiovascular monitoring. In: Miller RD, editor. Miller’s anesthesia eighth edition. Philadelphia: WB Saunders; 2015. p. 1356.
12.
Zurück zum Zitat Leone M, Asfar P, Radermacher P, Vincent JL, Martin C. Optimizing mean arterial pressure in septic shock: a critical reappraisal of the literature. Crit Care. 2015;101:1–7. Leone M, Asfar P, Radermacher P, Vincent JL, Martin C. Optimizing mean arterial pressure in septic shock: a critical reappraisal of the literature. Crit Care. 2015;101:1–7.
13.
Zurück zum Zitat Bersten AD, Holt AW. Vasoactive drugs and the importance of renal perfusion pressure. New Horiz. 1995;3:650–61.PubMed Bersten AD, Holt AW. Vasoactive drugs and the importance of renal perfusion pressure. New Horiz. 1995;3:650–61.PubMed
14.
Zurück zum Zitat Strandgaard S. Autoregulation of cerebral blood flow in hypertensive patients. The modifying influence of prolonged antihypertensive treatment on the tolerance to acute, drug-induced hypotension. Circulation. 1976;53:720–7.CrossRef Strandgaard S. Autoregulation of cerebral blood flow in hypertensive patients. The modifying influence of prolonged antihypertensive treatment on the tolerance to acute, drug-induced hypotension. Circulation. 1976;53:720–7.CrossRef
15.
Zurück zum Zitat Rogge DE, Nicklas JY, Schon G, Grothe O, Haas SA, Reuter DA, Saugel B. Continuous noninvasive arterial pressure monitoring in obese patients during bariatric surgery: an evaluation of the vascular unloading technique. Anesth Analg. 2019;128:477–83.CrossRef Rogge DE, Nicklas JY, Schon G, Grothe O, Haas SA, Reuter DA, Saugel B. Continuous noninvasive arterial pressure monitoring in obese patients during bariatric surgery: an evaluation of the vascular unloading technique. Anesth Analg. 2019;128:477–83.CrossRef
16.
Zurück zum Zitat Blazer F, Habicher M, Sander M, Sterr J, Scholz S, Feldheiser A, Muller M, Perka C, Treskatsch S. Comparison of the non-invasive Nexfin® monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients. J Int Med Res. 2016;44:832–43.CrossRef Blazer F, Habicher M, Sander M, Sterr J, Scholz S, Feldheiser A, Muller M, Perka C, Treskatsch S. Comparison of the non-invasive Nexfin® monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients. J Int Med Res. 2016;44:832–43.CrossRef
17.
Zurück zum Zitat Martina JR, Westerhof BE, van Goudoever J, de Beaumont EMFH, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BAJM, van Lieshout JJ. Noninvasive continuous arterial blood pressure monitoring with Nexfin®. Anesthesiology. 2012;116:1092–103.CrossRef Martina JR, Westerhof BE, van Goudoever J, de Beaumont EMFH, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BAJM, van Lieshout JJ. Noninvasive continuous arterial blood pressure monitoring with Nexfin®. Anesthesiology. 2012;116:1092–103.CrossRef
Metadaten
Titel
Validation of noninvasive continuous arterial pressure measurement by ClearSight System™ during induction of anesthesia for cardiovascular surgery
verfasst von
Tadashi Tanioku
Akari Yoshida
Yuichi Aratani
Keisuke Fujii
Tomoyuki Kawamata
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Anesthesiology / Ausgabe 1/2020
Elektronische ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-020-01091-x

Weitere Artikel der Ausgabe 1/2020

BMC Anesthesiology 1/2020 Zur Ausgabe

Mit dem Seitenschneider gegen das Reißverschluss-Malheur

03.06.2024 Urologische Notfallmedizin Nachrichten

Wer ihn je erlebt hat, wird ihn nicht vergessen: den Schmerz, den die beim Öffnen oder Schließen des Reißverschlusses am Hosenschlitz eingeklemmte Haut am Penis oder Skrotum verursacht. Eine neue Methode für rasche Abhilfe hat ein US-Team getestet.

Schlaganfall: frühzeitige Blutdrucksenkung im Krankenwagen ohne Nutzen

31.05.2024 Apoplex Nachrichten

Der optimale Ansatz für die Blutdruckkontrolle bei Patientinnen und Patienten mit akutem Schlaganfall ist noch nicht gefunden. Ob sich eine frühzeitige Therapie der Hypertonie noch während des Transports in die Klinik lohnt, hat jetzt eine Studie aus China untersucht.

Reanimation bei Kindern – besser vor Ort oder während Transport?

29.05.2024 Reanimation im Kindesalter Nachrichten

Zwar scheint es laut einer Studie aus den USA und Kanada bei der Reanimation von Kindern außerhalb einer Klinik keinen Unterschied für das Überleben zu machen, ob die Wiederbelebungsmaßnahmen während des Transports in die Klinik stattfinden oder vor Ort ausgeführt werden. Jedoch gibt es dabei einige Einschränkungen und eine wichtige Ausnahme.

Nicht Creutzfeldt Jakob, sondern Abführtee-Vergiftung

29.05.2024 Hyponatriämie Nachrichten

Eine ältere Frau trinkt regelmäßig Sennesblättertee gegen ihre Verstopfung. Der scheint plötzlich gut zu wirken. Auf Durchfall und Erbrechen folgt allerdings eine Hyponatriämie. Nach deren Korrektur kommt es plötzlich zu progredienten Kognitions- und Verhaltensstörungen.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.