Comparison of the TOFscan and the TOF-Watch SX during pediatric neuromuscular function recovery: a prospective observational study

verfasst von: Hyung-Been Yhim, Young-Eun Jang, Ji-Hyun Lee, Eun-Hee Kim, Jin-Tae Kim, Hee-Soo Kim

Erschienen in: Perioperative Medicine | Ausgabe 1/2021

Abstract

Background

TOFscan is a three-dimensional acceleromyography neuromuscular monitoring device that does not require initial calibration before muscle relaxant injection. This study aimed to compare TOFscan with TOF-Watch SX, the currently widely accepted uni-dimensional acceleromyography, for use among the pediatric population. We aimed to assess the agreement between TOFscan with TOF-Watch SX in the pediatric population’s neuromuscular recovery.

Methods

A total of 35 children aged 6–12 years were enrolled. Prior to any muscle relaxant injection, TOFscan and TOF-Watch SX were applied at each opposite arm and monitoring began concurrently throughout neuromuscular recovery. Calibration was performed for TOF-Watch SX, and train-of-four values were recorded every 15 s. Agreement between the two devices was evaluated with Modified Bland-Altman analysis.

Results

The bias between TOF-Watch SX and TOFscan were all within the 95% limits of agreement. The bias and standard deviation were smaller and the limit of agreement was narrower in the normalized group than in the non-normalized group [normalized bias −0.002 (95% CI, −0.013 to 0.010), standard deviation (SD) 0.111 vs non-normalized bias 0.010 (95% CI, −0.003 to 0.0236), SD 0.127].

Conclusions

TOFscan reliably demonstrated lack of bias and good concordance with TOF-Watch SX throughout the neuromuscular recovery, especially when normalized. Despite technical limitations, the two devices were unbiased along the path of spontaneous and pharmacological reversal in pediatric patients.

Trial registration

ClinicalTrials.​gov NCT03775603. Registered on 13 March 2018
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s13741-021-00215-2.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
NMBA
Neuromuscular blocking agent
IV
Intravenous cannulation
APB
Adductor pollicis brevis
OP
Opponens pollicis
FPB
Flexor pollicis brevis
CI
Confidence interval
WFSA
World Federation of Societies of Anaesthesiologists
SFAR
French Society for Anesthesia and Intensive Care

Introduction

When using neuromuscular blocking agents (NMBA), quantitative neuromuscular monitoring is mandatory to optimize intubation time, monitor intraoperative muscle relaxation, determine adequate pharmacologic reversal agents, and reduce postoperative residual paralysis (Murphy et al., 2013; Naguib et al., 2017; Brull & Kopman, 2017; Fortier et al., 2015; Naguib et al., 2018). National guidelines, such as World Federation of Societies of Anaesthesiologists (WFSA) International Standards for Safe Practice of Anesthesia, and French Society of Anesthesia and Intensive Care (SFAR), recommend incorporating objective neuromuscular monitoring into daily practice (Plaud et al., 2020; Gelb et al., 2018; Nemes & Renew, 2020). Although many anesthesiologists rely on subjective evaluation, dependence on such clinical parameters has limited success (Naguib et al., 2018).
Clinical assessment of adequate reversal is challenging in children given difficulties with communication and non-compliance with instructions. Previous evidence indicates that 10–28% of children experience postoperative residual block (i.e., train-of-four (TOF) ratio <0.9), with 6.5% displaying severe block (i.e., TOF ratio <0.7) (Ledowski et al., 2015; de Souza et al., 2011). Complications due to residual block can be detrimental given that children have smaller oxygen reserves and are more vulnerable to airway collapse (Fortier et al., 2015; von Ungern-Sternberg et al., 2006; Hardman & Wills, 2006).
The current standard device for determining the depth of muscle relaxation is the TOF-Watch SX (Organon, Swords Co., Dublin, Ireland), which is a one-dimensional acceleromyograph requiring initial calibration (Colegrave et al., 2016). Without calibration, TOF-Watch SX transmits erroneous information regarding the degree of residual neuromuscular blockade (Martin-Flores et al., 2012). Acceleromyography-based devices frequently overestimate TOF ratio more than mechanomyography or electromyography (Suzuki et al., 2006). Therefore, when using acceleromyography, “normalization” of taking baseline TOF ratio into account by dividing the expected TOF ratio with baseline TOF ratio at each interval throughout the neuromuscular recovery is recommended (Claudius et al., 2009).
Recently, TOFscan (Drager Technologies, Canada), a new device also using accelerometry, has emerged on the market. TOFscan does not require onerous calibration and differs from the TOF-Watch as it measures three-dimensional acceleration. A previous evaluation of the agreement between the two devices during neuromuscular recovery in adults indicated good agreement when TOF-Watch SX was calibrated and normalized (Murphy et al., 2018).
This study aimed to compare the performance of TOF-Watch SX and TOFscan in children. We hypothesized that TOFscan measures would be comparable to TOF-Watch SX’s among children aged ≤12 years.

Materials and methods

Patient recruitment

A prospective, observational clinical trial was conducted between December 2018 and August 2019 at a single tertiary medical center. The study was approved by the SNUH Institutional Review Board (1811-137-989) and was registered at ClinicalTrials.​gov (https://​clinicaltrials.​gov/​ct2/​show/​NCT03074968). Each participant and their parents were provided a verbal explanation of the study and given the opportunity to ask questions. Written informed consent was obtained from participants aged ≥7 years and one of their parents. Verbal consent was obtained from participants aged <7 years, in addition to written informed consent from one of their parents. All procedures were conducted in compliance with the principles of the Helsinki Declaration.
In total, 39 children aged ≤12 years were screened, of whom 35 were enrolled. All children were classified as American Society of Anesthesiologists physical status I–II and scheduled for elective surgery under general anesthesia. Exclusion criteria were as follows: body mass index ≥30 kg∙m−2; presence of neuromuscular disease, myopathy, susceptibility to malignant hyperthermia, renal insufficiency (i.e., estimated glomerular filtration rate ≤60 ml/min), or liver disease; surgery with expected duration <60 min; surgery involving the arms; need for rapid sequence intubation; surgery that required absolute immobility or prone position; and conditions requiring postoperative mechanical ventilation.

Anesthesia

All patients arrived in the operating room with a peripheral intravenous (IV) line. Standard monitors were applied with oxygen saturation measured on the IV limb and non-invasive blood pressure on the contralateral limb. Anesthesia was induced with administration of 5 mg·kg−1 of thiopental (aged <3 years) or 0.5 mg kg−1 of 1% lidocaine, followed by 2–2.5 mg kg−1 of propofol (aged ≥3 years). After loss of consciousness, calibration was performed followed by administration of 0.6 mg kg−1 of rocuronium. Patients were manually ventilated with sevoflurane in 100% oxygen at 6 L min−1 of fresh gas flow. After confirmation of full relaxation by neuromuscular monitoring, endotracheal intubation was performed. During surgery, anesthesia was maintained at 1–1.5 minimum alveolar concentration of sevoflurane. Remifentanil was administered at 0.1–0.2 μg kg−1 min−1. Additional boluses of rocuronium were administered as required. During anesthesia, blood pressure and heart rate were maintained according to the individual ward measurement. Ventilation was adjusted to a tidal volume of 7 ml−1 kg−1, and the respiratory rate was adjusted to maintain ETCO2 of 35–40 mmHg. Intraoperative hypotension was managed with additional fluid bolus, 5–10 mg kg−1 of calcium gluconate, 0.05–0.1 mg kg−1 of ephedrine, or continuous infusion of dopamine (5 mcg kg−1 min−1) as clinically indicated. Hypertension was managed by increasing the concentration of inhalation anesthesia or the infusion rate of remifentanil after determining the etiology. Temperature was monitored using either an esophageal or axillary temperature probe. Temperature was maintained between 35.6 and 37.5°C using an over-body forced-air warmer and under-body warming mattress.
During anesthetic recovery, TOF ratio measurements were obtained. After clinical evaluation suggesting adequate airway patency (Davis, 2017) and TOF ratio reaching 0.9, tracheal extubation was performed by an anesthesiologist with >1 year of experience in pediatric anesthesia.

Monitoring of neuromuscular blockade

After loss of consciousness, neuromuscular monitoring was simultaneously initiated using TOF-Watch SX in one forearm and TOFscan in the opposite, with the patient in the supine position. Upper forearms with supinated palm were passively extended and fixed to an arm board to ensure sole movement of adductor pollicis brevis (APB) (Fig. 1). After abrasion and cleansing with an alcohol swab, the skin surface was allowed to dry. Two surface electrodes were placed along the course of the ulnar nerve, with the negative electrode at a distal location near the styloid process of the radius and the positive electrode at 3 cm proximally. An adult or a pediatric hand sensor of TOFscan was used according to the size of the patient’s hand. TOF-Watch SX was calibrated using the CAL2 function (Martin-Flores et al., 2012). Time to acquire the baseline TOF-Watch SX calibration was measured. TOFscan was initiated with non-calibrated intensity of stimulation fixed at 50 mA for all patients (Colegrave et al., 2016). TOF stimuli from the two devices were repeated every 15 s until after surgery. Prior to extubation, neuromuscular blockade antagonism was made with atropine 15 mcg kg−1 and neostigmine 30 mcg kg−1 if reversal administration was indicated. TOF ratio ≥0.9 was considered complete neuromuscular function recovery for safe extubation.

Data analysis

Data were collected using the manufacturer-provided software. The time to reach TOF count 0 was measured. Then, the baseline TOF ratio in TOF-Watch SX’s deviation from 1.0 was adjusted for normalization (Murphy et al., 2018). Likewise, if the baseline TOF ratio with TOF-Watch SX after calibration was 0.9, all block levels (0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, and 1.0) were divided by 0.9, yielding the following TOF-Watch values representative of TOF ratio: 0.11, 0.22, 0.33, 0.44, 0.56, 0.67, 0.78, 0.89, 1.00, and 1.11. The TOFscan’s TOF ratio was recorded at the time when the representative values of TOF-Watch SX value were reached. However, non-normalization values do not take into account baseline TOF ratio. Non-normalization only detects matching absolute values of TOF ratio between the two devices at levels of TOF ratios 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, and 1.0. Additionally, intraoperative NMBA use data, including the initial and additional doses, was collected. If additional doses of NMBAs were administered, the TOF ratio was collected again, and the same process was repeated.
The primary endpoint was the measure of TOFscan value at TOF-Watch SX’s ratio of 0.7. The final analysis determined bias with limits of agreement (95% confidence interval, CI) for both normalized and non-normalized pairs throughout the TOF ratio of 0.1–1.0. Secondary outcomes included results of the expanded Bland-Altman method, referred to as modified true value varies analysis (Olofsen et al., 2014).

Sample size estimation

The sample size to assess the agreement between TOF-Watch SX and TOFscan was calculated according to a previously established methodology (Murphy et al., 2018). The results were bias of 0.021 (standard deviation [SD] = 0.009) for non-normalized and 0.015 (SD=0.008) for normalized values. With these results, we assumed the reliable mean of bias as 0.02 (SD=0.01) regardless of the presence or absence of normalization. We predefined the maximum allowed bias (δ) as 0.05, while setting the minimum power as 0.90. Overall, 35 pediatric patients were required; assuming a dropout rate of 10%, we planned to enroll 39 participants.

Statistical analysis

Bland-Altman analysis was used to evaluate the agreement over repeated measurements at each block level. Limits of agreement were calculated as the means of differences between two measurements ± 1.96×SD, resulting in 95% CI upper and lower limits. Furthermore, the bias was tested for normal distribution by Shapiro-Wilk test, along with probability quantile-quantile plot. To evaluate homogeneity of variances across the measurement range, correlation between the size of the bias and the mean values was measured through Spearman rank correlation and Kendall rank correlation coefficient.
Statistical analyses were performed using R programming version 3.6.1 and blandr (version 0.5.1) package and SPSS version 25.0 (IBM Corp., Armonk, NY). Continuous variables, such as patient characteristics and intraoperative measurements, were analyzed using Student’s t-test. Categorical variables were analyzed using Pearson’s chi-squared test. Values are expressed as mean ± SD with corresponding 95% CI, or median (interquartile rage). Normality of the distribution was determined using the Shapiro-Wilk test. A P-value <0.05 was considered statistically significant.

Results

In total, 39 children were screened, of whom 35 were enrolled. Demographic characteristics are shown in Table 1. Four patients were excluded due to missing data, and 16 required additional NMBA maintenance. Although the supramaximal current was secured, the TOF-Watch SX ratio of 12 children ended before reaching 1.0 due to short surgical time. Therefore, 12 data pairs of TOF ratio 1.0 in both normalized and non-normalized groups were missing. However, all 35 data pairs were secured until a TOF ratio of 0.9.
Table 1
Demographics of study population
Characteristics
N = 35
Age
9 [8–11]
Height (cm)
140.0 [131.0–151.5]
Weight (kg)
38.0 [28.0–46.5]
M/F
14/21
Surgery time (min)
125.0 [90.0–175.0]
Anesthesia time (min)
160.0 [117.5–217.5]
Values are numbers or median [IQR, 25–75%]
Baseline TOF ratio was 0.029 higher (P=0.026) for TOF-Watch SX than TOFscan (1.026±0.073 and 0.997±0.009, respectively). The time to reach TOF count 0 was shorter in TOF-Watch SX (100.5±34.6 s) than in TOFscan (112.1±34.6 s, P=0.018). Average calibration time for TOF-Watch SX was 36.88 s. At a TOF ratio of 0.7 in TOF-Watch SX, the TOF ratio of TOFscan was similar (0.712±0.116, P=0.545; Table 2).
Table 2
Baseline TOF ratio with the onset of NMBA between TOF-Watch SX and TOFscan
Data
TOF-Watch SX
TOFscan
Difference (95% CI)
P-value
Baseline TOF ratio
1.026±0.073
0.997±0.009
0.029 (0.004, 0.054)
0.026
 Calibration time (s)
38.68±17.08
   
Time to TOF count 0 (s)
100.5±34.6
112.1±34.6
11.5 (−7.7, 30.8)
0.018
TOFscan value at TOF-Watch 0.7
0.700±0.000
0.712±0.116
−0.012 (−0.052, 0.028)
0.545
Data are reported as mean ± SD and were compared using the paired t test. n = 35
In non-normalized TOF ratios, the bias was 0.010 (95% CI, −0.003 to 0.0236) with SD of 0.127. The 95% limits of agreement were −0.239 to 0.259 for non-normalized ratios. The CI for lower limit of agreement was −0.263 to −0.216, and CI for the upper limit of agreement was 0.235 to 0.282. Normalized TOF-Watch SX and TOFscan data showed bias of −0.002 (95% CI, −0.013 to 0.010) with SD of 0.111. The 95% limits of agreement were −0.219 to 0.216 for normalized ratios. The CI for lower limit of agreement was −0.239 to −0.119, and CI for the upper limit of agreement was 0.196 to 0.236 (Table 3). A discrepancy was detected in seven datasets because of additional TOF ratio gained at TOF ratio of 1.0 in the normalization process (non-normalized, 338, vs normalized, 345). Each measurement is displayed in the Bland-Altman plot along with each 95% CI shown in green (upper 95%) and red (lower 95%) (Fig. 2, Supplement 1).
Table 3
Bias and 95% limits of agreement in non-normalized and normalized TOF-Watch SX and TOFscan
TOF-Watch SX measurements
Bias ± standard error (TOF-Watch SX – TOFscan; 95% CI)
SD ± standard error of the differences
95% Limits of agreement
95% CI, lower limit of agreement
95% CI, upper limit of agreement
Non-normalizeda
0.010±0.007
(−0.003, 0.0236)
0.127±0.0120
−0.239 to 0.259
−0.263 to −0.216
0.235 to 0.282
Normalizedb
−0.002±0.006
(−0.013, 0.010)
0.111±0.010
−0.219 to 0.216
−0.239 to −0.199
0.196 to 0.236
a338 measurements in 35 individuals
b345 measurements in 35 individuals
The Spearman rank correlation coefficients (ρ) were −0.237 and −0.236 for non-normalized and normalized biases, respectively. Similarly, Kendall rank correlation coefficients (τ) were −0.164 and −0.160 for non-normalized and normalized groups, respectively. This suggests a weak relationship in the bias over the range of measurements, regardless of normalization. The intraclass correlation coefficients were 0.921 and 0.929 for non-normalized and normalized biases, respectively, indicating the high reliability of both devices, regardless of normalization (Table 4).
Table 4
Correlation coefficient and intraclass correlation in non-normalized and normalized TOF-Watch SX and TOFscan
TOF-Watch SX
Measurements
Spearman rank correlation a
coefficient (ρ)
Kendall rank correlation a
coefficient (τ)
Intraclass correlationc
(95% CI)
Non-normalized
−0.237b
−0.164b
0.921 (0.863 to 0.962)
Normalized
−0.236b
−0.160b
0.929 (0.914 to 0.943)
aThe correlation between the size of the bias and the mean values of corresponding TOF ratio
bCorrelation is significant at the 0.01 level (2-tailed)
cThe ratio of the between-subject variance and the total variance

Discussion

This study presents lack of bias and good concordance between TOF-Watch SX and TOFscan in children. The normalized group displayed less absolute value of bias with smaller SD and narrower bidirectional (upper and lower) limits of agreement than the non-normalized group. Normalized values coincided with those of TOFscan more accurately than the non-normalized values. The Spearman rank correlation coefficient (ρ) and Kendall rank correlation coefficient (τ) of normalized and non-normalized measurements indicated a weak correlation over the course of neuromuscular recovery.
Three-dimensional acceleromyography devices, such as TOFscan, Stimpod NMS 450 (Xavant, South Africa), and Mindray neuromuscular transmission transducer (Shenzhen, China), use three perpendicular piezoelectric probes to thoroughly measure freely moving target muscles. To date, only two studies, both in adults, have compared the performances of TOF-Watch SX and TOFscan. One reported no significant differences between the two but suggested that TOF-Watch SX is more sensitive during deep neuromuscular blockade and that better agreement was observed along the recovery (Colegrave et al., 2016). Another study detected minimal bias between these devices (Murphy et al., 2018). Our findings also indicate minimal bias; the absolute value of bias was also smaller. Also, Spearman rank and Kendall rank correlation analysis revealed a weak negative relationship. It indicates that size of bias did not significantly change but that a decreasing trend over bias was observed during neuromuscular recovery. This result supports previous observations of a similar but slight negative trend in adults (Murphy et al., 2018) and observations that better agreement was achieved at more complete levels of recovery (Colegrave et al., 2016).
Assessing three-dimensional acceleration in children is challenging due to difficulty in isolating APB movement. The APB originates from two heads (oblique head from the second and third metacarpal bones, and transverse head from the third metacarpal bone) and inserts at the base of the thumb’s proximal phalanx. Contraction of the APB brings the thumb’s tip to the center of the palm. However, adjacent muscles, such as the opponens pollicis (OP) and flexor pollicis brevis (FPB) muscles, also contribute to similar movements. Given the shallow skin-to-nerve and nerve-to-nerve distance in pediatric patients, applying same the “3-cm” distance between the positive and negative electrodes can cause extra stimulation to the median nerve, which innervates OP and FPB (Alanazy, 2017). The distance between the two electrodes determines the penetration depth (Fuchs-Buder, 2011). With the distance between the electrodes relatively far apart in the relatively short pediatric forearm, nearby muscles may also react. Strapping prevents only the four fingers, not the thumb from moving. OP and FPB cannot be restricted by mere strapping. Moreover, the deep part of the OP and FPB in 20% of the population is often innervated by the deep branch of the ulnar nerve (C8, T1) (Gupta & Michelsen-Jost, 2012). The erroneous extra-apposition of the thumb due to the mixed effect of unwanted muscle movement is likely to occur in a pediatric patient with smaller hands and more proximal nerve-to-nerve distance. This anatomical inevitability may cause unwanted acceleration and induce a larger correlation coefficient than that demonstrated in Murphy et al.’s previous report.
Normalization of the TOF-Watch values resulted in lower bias, narrower limits of agreement, and higher intraclass correlation between devices. TOF ratio >1 has been previously demonstrated in TOF-Watch SX (Suzuki et al., 2006; Claudius & Viby-Mogensen, 2008; Bowdle et al., 2019). Without normalization, if the initial TOF exceeds 1, subsequent TOF ratios will be overestimated. Alternatively, if initial TOF <1, subsequent ratios may be underestimated. Individual differences in distribution volume, muscle mass, NMBA clearance, and age-dependent maturation of neuromuscular junction may contribute to differences in initial TOF ratio. As normalization adjusts the inherent TOF ratio, the results support the use of normalized TOF ratios to provide a more accurate TOF ratio along the recovery in children (Suzuki et al., 2006; Claudius et al., 2009).
The completion of TOF count 0 in TOFscan was longer by approximately 12 s. According to previous studies, the time to TOF count 0 showed inconsistent results between TOF devices (Colegrave et al., 2016; Murphy et al., 2018). The reason for the discrepancy may be associated with the calibration process. The maximal value, the current that gives 100% response, is acquired with a calibration mode called CAL2, and 110% of maximal value is suggested as the supramaximal reference (Martin-Flores et al., 2012; Schreiber et al., 2011). Compared with TOFscan, which delivers uniform current intensity of 50 mA at 2 Hz, the calibration process in TOF-Watch provides 10 stimuli at 1 Hz until the optimal supramaximal current is detected. Such higher initial frequency may affect the gradual potentiation process—in other words, a staircase phenomenon (Zhou et al., 2013). The staircase phenomenon influences the onset time and duration of twitch depression by increasing T1 values (Martin-Flores et al., 2011). Nonetheless, some researchers argue that, despite the staircase phenomenon, T1 through T4 increase in the same proportion, therefore not affecting the TOF ratio (Suzuki et al., 2006). However, others reported the need for an extra-stabilization period after the staircase phenomenon to obtain a stable baseline (Martin-Flores et al., 2011). Therefore, calibration process may either effect T1 size or time to reach a stable T1 to T4 baseline. As such, further investigation is required to determine the reliability of the measurement devices at TOF count 0.
This study has some limitations. First, we did not compare acceleromyography with mechanomyography, despite acceleromyography itself not differing significantly from mechanomyography, especially when calibrated and normalized (Bowdle et al., 2019; Claudius et al., 2010). Second, no randomization was performed between the right and left arm, but previous research indicates that arm-to-arm variations did not display significant bias gap (Claudius et al., 2010; Hohenauer et al., 2017). Third, preload and device size were not individualized. However, acceleromyography is well reported to be precise with preload application ranging from 75–120 g (Claudius et al., 2009). Although we used a pediatric sensor, the TOFscan’s sensor might be too big for very small children. To correctly measure acceleration, the index finger should fit in the sensor’s hole after the curvature. Our youngest patient was a 16-month-old child whose finger length fit the sensor. However, to expand TOFscan’s validity to infants and neonates, a smaller sensor is required. Fourth, we did not identify the potential effect of calcium in neuromuscular monitoring. In motor neurons, presynaptic voltage-gated calcium channel activation produces neuromuscular junction’s synaptic vesicle to release acetylcholine, which eventually causes muscle contraction. Some studies report that increased ionized calcium levels decrease sensitivity to non-depolarizing NMBA and enhance neuromuscular recovery (Ju et al., 2017; Munir et al., 2003). When comparing TOF ratio, equalizing the use of calcium between groups would be helpful. Finally, our results reliably indicated concordance and lack of bias but no agreement between the devices. The limits of agreement’s absolute range exceeded 0.2, and this difference in TOF ratio indicates poor agreement. Careful interpretation is required since a 0.2 difference practically results in diverse decisions regarding extubation time and reversal dose. In conclusion, TOFscan demonstrated good concordance and is unbiased with TOF-Watch SX in children’s neuromuscular recovery, especially when normalized.

Acknowledgements

We would like to thank the Anesthesia and Pain Department in Seoul National University Children’s Hospital for their cooperation during the data collection.

Declarations

Obtained
Not applicable

Competing interests

The authors declare no conflict of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
Zurück zum Zitat Davis PJCF. Smith’s anesthesia for infants and children: Elsevie; 2017. Davis PJCF. Smith’s anesthesia for infants and children: Elsevie; 2017.
Zurück zum Zitat Fuchs-Buder T. Neuromuscular monitoring: Springer Science & Business Media; 2011. Fuchs-Buder T. Neuromuscular monitoring: Springer Science & Business Media; 2011.
Zurück zum Zitat Ledowski T, O’Dea B, Meyerkort L, Hegarty M, von Ungern-Sternberg BS. Postoperative residual neuromuscular paralysis at an Australian tertiary children’s hospital. Anesthesiol Res Pract. 2015;2015:1–4. Ledowski T, O’Dea B, Meyerkort L, Hegarty M, von Ungern-Sternberg BS. Postoperative residual neuromuscular paralysis at an Australian tertiary children’s hospital. Anesthesiol Res Pract. 2015;2015:1–4.
Metadaten
Titel
Comparison of the TOFscan and the TOF-Watch SX during pediatric neuromuscular function recovery: a prospective observational study
verfasst von
Hyung-Been Yhim
Young-Eun Jang
Ji-Hyun Lee
Eun-Hee Kim
Jin-Tae Kim
Hee-Soo Kim
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
Perioperative Medicine / Ausgabe 1/2021
Elektronische ISSN: 2047-0525
DOI
https://doi.org/10.1186/s13741-021-00215-2

Weitere Artikel der Ausgabe 1/2021

Perioperative Medicine 1/2021 Zur Ausgabe

Mit dem Seitenschneider gegen das Reißverschluss-Malheur

03.06.2024 Urologische Notfallmedizin Nachrichten

Wer ihn je erlebt hat, wird ihn nicht vergessen: den Schmerz, den die beim Öffnen oder Schließen des Reißverschlusses am Hosenschlitz eingeklemmte Haut am Penis oder Skrotum verursacht. Eine neue Methode für rasche Abhilfe hat ein US-Team getestet.

Schlaganfall: frühzeitige Blutdrucksenkung im Krankenwagen ohne Nutzen

31.05.2024 Apoplex Nachrichten

Der optimale Ansatz für die Blutdruckkontrolle bei Patientinnen und Patienten mit akutem Schlaganfall ist noch nicht gefunden. Ob sich eine frühzeitige Therapie der Hypertonie noch während des Transports in die Klinik lohnt, hat jetzt eine Studie aus China untersucht.

Reanimation bei Kindern – besser vor Ort oder während Transport?

29.05.2024 Reanimation im Kindesalter Nachrichten

Zwar scheint es laut einer Studie aus den USA und Kanada bei der Reanimation von Kindern außerhalb einer Klinik keinen Unterschied für das Überleben zu machen, ob die Wiederbelebungsmaßnahmen während des Transports in die Klinik stattfinden oder vor Ort ausgeführt werden. Jedoch gibt es dabei einige Einschränkungen und eine wichtige Ausnahme.

Nicht Creutzfeldt Jakob, sondern Abführtee-Vergiftung

29.05.2024 Hyponatriämie Nachrichten

Eine ältere Frau trinkt regelmäßig Sennesblättertee gegen ihre Verstopfung. Der scheint plötzlich gut zu wirken. Auf Durchfall und Erbrechen folgt allerdings eine Hyponatriämie. Nach deren Korrektur kommt es plötzlich zu progredienten Kognitions- und Verhaltensstörungen.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.