Native T1 mapping compared to ultrasound elastography for staging and monitoring liver fibrosis: an animal study of repeatability, reproducibility, and accuracy
verfasst von:
Jinning Li, Huanhuan Liu, Caiyuan Zhang, Shuyan Yang, Yanshu Wang, Weibo Chen, Xin Li, Dengbin Wang
Erschienen in:
European Radiology
|
Ausgabe 1/2020
Einloggen, um Zugang zu erhalten
Abstract
Objectives
To investigate the repeatability, reproducibility, and staging and monitoring of the performance of native T1 mapping for noninvasively assessing liver fibrosis in comparison with acoustic radiation force impulse (ARFI) elastography.
Methods
The repeatability and reproducibility were explored in 8 male Sprague-Dawley rats with intraclass correlation coefficient (ICC). Different degrees of fibrosis were induced in 52 rats by carbon-tetrachloride (CCl4) insult. Another 16 rats were used to build fibrosis progression and regression models. The native T1 values and shear wave velocity (SWV) were quantified by using native T1 mapping and ARFI elastography, respectively. The METAVIR system (F0–F4) was used for the staging of fibrosis. The area under the receiver operating characteristic curve (AUC) was determined to assess the performance of quantitative parameters for staging and monitoring fibrosis.
Results
Native T1 values shared similar good repeatability (ICC = 0.93) and reproducibility (ICC = 0.87) with SWV (ICC = 0.84–0.93). The AUC of native T1 values were 0.84, 0.84, and 0.75 for diagnosing significant fibrosis (≥ F2) and liver cirrhosis (F4) and detecting fibrosis progression, and those of SWV were 0.81, 0.86, and 0.7, respectively. No significant difference in performance was found between the two quantitative parameters (p ≥ 0.496). For detecting fibrosis regression, native T1 values had a better accuracy (AUC = 0.99) than SWV (AUC = 0.56; p = 0.002).
Conclusion
Native T1 mapping may be a reliable and accurate method for noninvasively assessing liver fibrosis. Compared with ARFI elastography, it provides similar good repeatability and reproducibility, a similar high accuracy for staging fibrosis, and a better accuracy for detecting fibrosis regression.
Key Points
• Native T1 mapping is a valuable tool for noninvasively assessing liver fibrosis and can be measured on virtually all clinical MRI machines without additional hardware or gadolinium chelate injection.
• Compared with acoustic radiation force impulse elastography, native T1 mapping yields similar good repeatability and reproducibility and a similar high accuracy for staging fibrosis.
• Native T1 mapping provides a significantly better performance for detecting fibrosis regression than acoustic radiation force impulse elastography.