Skip to main content
Erschienen in: Molecular Imaging and Biology 6/2018

Open Access 30.04.2018 | Research Article

Noise-Induced Variability of Immuno-PET with Zirconium-89-Labeled Antibodies: an Analysis Based on Count-Reduced Clinical Images

verfasst von: Yvonne W. S. Jauw, Dennis F. Heijtel, Josée M. Zijlstra, Otto S. Hoekstra, Henrica C. W. de Vet, Danielle J. Vugts, Henk M. Verheul, Ronald Boellaard, Sonja Zweegman, Guus A. M. S. van Dongen, C. Willemien Menke-van der Houven van Oordt, Adriaan A. Lammertsma, Marc C. Huisman

Erschienen in: Molecular Imaging and Biology | Ausgabe 6/2018

Abstract

Purpose

Positron emission tomography (PET) with Zirconium-89 (Zr-89)-labeled antibodies can be used for in vivo quantification of antibody uptake. Knowledge about measurement variability is required to ensure correct interpretation. However, no clinical studies have been reported on measurement variability of Zr-89 immuno-PET. As variability due to low signal-to-noise is part of the total measurement variability, the aim of this study was to assess noise-induced variability of Zr-89 -immuno-PET using count-reduced clinical images.

Procedures

Data were acquired from three previously reported clinical studies with [89Zr]antiCD20 (74 MBq, n = 7), [89Zr]antiEGFR (37 MBq, n = 7), and [89Zr]antiCD44 (37 MBq, n = 13), with imaging obtained 1 to 6 days post injection (D0–D6). Volumes of interest (VOIs) were manually delineated for liver, spleen, kidney, lung, brain, and tumor. For blood pool and bone marrow, fixed-size VOIs were used. Original PET list mode data were split and reconstructed, resulting in two count-reduced images at 50 % of the original injected dose (e.g., 37 MBq74inj).
Repeatability coefficients (RC) were obtained from Bland-Altman analysis on standardized uptake values (SUV) derived from VOIs applied to these images.

Results

The RC for the combined manually delineated organs for [89Zr] antiCD20 (37 MBq74inj) increased from D0 to D6 and was less than 6 % at all time points. Blood pool and bone marrow had higher RC, up to 43 % for 37 MBq74inj at D6. For tumor, the RC was up to 42 % for [89Zr]antiCD20 (37 MBq74inj). For [89Zr]antiCD20, (18 MBq74inj), [89Zr]antiEGFR (18 MBq37inj), and [89Zr]antiCD44 (18 MBq37inj), measurement variability was independent of the investigated antibody.

Conclusions

Based on this study, noise-induced variability results in a RC for Zr-89-immuno-PET (37 MBq) around 6 % for manually delineated organs combined, increasing up to 43 % at D6 for blood pool and bone marrow, assuming similar biodistribution of antibodies. The signal-to-noise ratio leads to tumor RC up to 42 %.
Begleitmaterial
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s11307-018-1200-4) contains supplementary material, which is available to authorized users.

Introduction

Antibody imaging is of interest to improve efficacy and limit toxicity of antibody treatment by providing a predictive imaging biomarker for antibody uptake. Zr-89-PET can be used for in vivo quantification of antibody biodistribution and tumor uptake [1, 2]. Knowledge about measurement variability is required for clinical application. Usually, a test-retest study is performed for novel tracers to assess repeatability. However, for Zr-89 -immuno-PET, repeatability is unknown. A classical test-retest study design with two tracer injections is challenging in case of Zr-89 -immuno-PET because of the long half-life of Zr-89 (78.4 h). This requires more than 10 days between two injections to have less than 10 % of the radioactivity due to the first injection remaining in the body. In addition, radiation exposure is significant (0.5 mSv/MBq) [3]. To date, most clinical PET studies using Zr-89-labeled monoclonal antibodies (mAbs) are performed with an injected dose of 37 MBq, resulting in an effective dose of 18.5 mSv. We hypothesize that the relatively low signal -to- noise ratio for Zr-89 -immuno-PET acquisition (due to the low injected dose and low positron abundance of Zr-89) results in a considerable source of measurement variability. The primary objective of this study was therefore to assess noise-induced variability of quantitative uptake measures derived from Zr-89 -immuno-PET for an injected dose of 37 MBq using count-reduced images.
For this purpose, previously acquired clinical datasets can be used to assess noise-induced variability at 50 % of the original injected dose. Raw PET data (list mode data) can be split in two equal parts (Fig. 1). The split list mode data can be reconstructed into two count-reduced images [4]. Each of the two count-reduced images is considered to be a count statistically independent estimate of an image that would have been obtained with 50 % of the original injected dose for the same scan time. For example, an original image acquired with an injected dose of 74 MBq results in two count-reduced images representing an injected dose of 37 MBq (denoted as 37MBq74inj). The count-reduced images are not independent with respect to other factors such as procedural variations or scanner drift.
In addition, we hypothesize that noise-induced variability is independent of the investigated mAb; therefore, the secondary objective was to investigate clinical datasets with three different Zr-89-labeled mAbs. Finally, the tertiary objective was to assess noise-induced variability of Zr-89 -immuno-PET for an injected dose of 18 MBq. This is the lowest injected dose used in clinical Zr-89 -immuno-PET studies for non-oncological indications, e.g., rheumatoid arthritis and multiple sclerosis [5, 6], as further limiting radiation exposure (to < 10 mSv) is necessary for these patient categories.
Knowledge of measurement variability is of interest as a small measurement error is required for the detection of even small changes over time (e.g., response evaluation, within patients). In the current study, we assessed noise-induced variability as source of measurement error (expressed as repeatability coefficient (RC) in %). For clinical relevance, assessment of reliability is important as this indicates the ability to divide patients in groups of interest despite measurement errors (e.g., response prediction, between patients). Therefore, an explorative reliability analysis was performed (expressed as intraclass correlation coefficients (ICC)) reflecting the contribution of this source of measurement variability to the observed differences (total variance) in biodistribution and tumor uptake in these datasets.
Potential clinical applications of Zr-89 -immuno-PET include use as a quantitative imaging biomarker to assess antibody uptake in normal tissue and tumor to guide individualized treatment and/or drug development [3].

Materials and Methods

Data Sources

Original list mode data were taken from three clinical Zr-89 -immuno-PET studies: [89Zr]antiCD20 mAb in non-Hodgkin lymphoma [7], [89Zr]anti-epidermal growth factor receptor (EGFR) mAb in colorectal cancer [8], and [89Zr]antiCD44 mAb in solid tumors [9]. The original injected dose was 74 MBq for [89Zr]antiCD20 mAb, 37 MBq for [89Zr]antiEGFR mAb, and 37 MBq for [89Zr]antiCD44 mAb. Scans were scheduled at the following time points: 1, 72, and 144 h p.i. (D0, D3, D6) for [89Zr]antiCD20 mAb and [89Zr]antiEGFR mAb, and 1, 24, and 96 h p.i. (D0, D1, D4) for [89Zr]antiCD44 mAb. Study procedures, including image acquisition protocols, have been reported previously [79]. All data were acquired on a Philips Gemini 64 or Ingenuity 128 PET/CT scanner (Philips Healthcare). The number of bed positions (for a scan trajectory of mid-thigh to vertex of the skull) was 10–12, with a 50 % bed overlap. Data were acquired for 5 min per bed. Data were normalized; corrected for decay, randoms, dead time, scatter, and attenuation; and reconstructed using 3D BLOB-OS-TF (3 iterations, 33 subsets). A 7 mm Gaussian post reconstruction filter was applied, in line with the recommendation for multicenter Zr-89 -immuno-PET studies [10].

Image Generation and Analysis

Using the original scans, VOIs were defined manually for liver, spleen, kidney, brain, and lung (brain and lung on the low dose CT, liver, spleen, kidney on the PET image). In addition, fixed-size VOIs of 8.6 and 2.9 mL were placed in the lumbar vertebrae to estimate bone marrow activity concentration (AC), and in the aortic arch to estimate blood pool AC, respectively. Tumor uptake was defined as focal uptake exceeding local background reported by the nuclear medicine physician. Tumors were manually delineated on the immuno-PET scan, using the low dose CT for anatomical reference, using in-house developed software (ACCURATE tool, developed by RB).
Original list mode data were split (all counts, including delayed): even counts were placed in one data set (e.g., H1), while the odd counts were placed in the second data set (e.g., H2), creating interleaved datasets that are count statistically independent of each other, while preserving identical scan conditions (e.g., patient movement) (Fig. 1). Next, these two split data sets were reconstructed (including scatter and attenuation correction).
To assess the noise-induced variability of Zr-89 -immuno-PET at an injected dose of 37 MBq74inj, the original [89Zr]antiCD20 dataset (74 MBq) was split in two equal parts (H1 and H2), which were reconstructed in count-reduced images of 37 MBq74inj. To assess whether noise-induced variability is independent of the investigated mAb, the original [89Zr]antiEGFR mAb dataset (37 MBq) and [89Zr]antiCD44 mAb dataset (37 MBq) were used to produce count-reduced images of 18 MBq37inj (H1 and H2). In addition, the [89Zr]antiCD20 mAb dataset (74 MBq) was split again (H1) resulting in count-reduced images at 25 % of the injected dose (18 MBq74inj) (Q1 and Q2). To keep the statistical analysis similar for all three datasets, we did not split and analyze the second dataset (H2) (effectively creating Q3, Q4) as the [89Zr]antiEGFR mAb and [89Zr]antiCD44 mAb datasets only had the availability of two splits (H1, H2).
All tissue and tumor VOIs were applied to the count-reduced images, resulting in AC1 from the first count-reduced image (H1) and AC2 from the second count-reduced image (H2). For the [89Zr]antiCD20 mAb dataset, all VOIs were also applied to the count-reduced images at 25 % of the injected dose (Q1 and Q2).
For normal tissue ACmean was derived, for tumors ACmax, ACpeak and ACmean were derived [11]. All ACs were converted into standardized uptake values (SUV), correcting for injected dose and body weight.

Noise-Induced Variability Analysis

Noise-induced variability was assessed as described by Bland and Altman [12, 13]. The measured uptake from each of the two count-reduced images was considered as repeat measures of the same quantity.
Per group of n VOIs, the mean percentage difference and standard deviation (SD) over the percent differences were calculated. The mean percentage difference is expected to be negligible, as the noise between images is assumed to be normally distributed around zero.
Repeatability coefficients (RC) in percent were calculated according to Eq. 1:
$$ RC\left(\%\right)=1.96\cdot {SD}_{\left(100\%\cdot \frac{\Delta}{MV}\right)}=1.96\cdot {CV}_w\left(\%\right), $$
(1)
where ∆ is (AC1 − AC2) per VOI, MV is the mean value 0.5 × (AC1 + AC2) per VOI, 100 % is multiplication by 100, SD the standard deviation over all percentage differences for n VOI in the group, and CVw the coefficient of variation within scan.
RC were expressed as a percentage instead of absolute value, as the difference between AC1 and AC2 scaled linearly with the mean value of AC1 and AC2. Furthermore, the mean percentage difference was calculated as well by taking the average ∆ over all VOIs in the group. The mean percentage difference and the RC combined define the limits of agreement (LoA). As a result, the LoA were directly related to the coefficient of variation. Use of a relative unit allows for comparison with other studies, irrespective of the measurement unit used (e.g., type of normalization used to calculate SUV).

Reliability Analysis

To estimate the contribution of noise-induced variability to the observed differences between patients or tumor lesions, the ICC was calculated in addition to the RC. The ICC was calculated as the proportion of the total variance that is due to the true variance. True variance reflects biological differences, for example, between patients or between tumor lesions, while the total variance comprises both true variance as well as the measurement variance (Eq. 2).
$$ ICC=\frac{\sigma_{voi}^2}{\left({\sigma}_{voi}^2+{\sigma}_{\Delta }^2\right)}, $$
(2)
where σvoi2 is the variance between the n VOI per group, σ2 is the variance over the differences between AC1 and AC2. For a reliable measure, a high ICC is expected, as an ICC of 1 reflects that all measured variance can be attributed to biological differences (the contribution of measurement variability is negligible). An ICC of 0 signifies that all measured variance can be attributed to measurement variability (no detection of biological differences beyond measurement variability). In general, an ICC of > 0.7 is considered acceptable [14]. The 95 % confidence interval of the ICC was obtained to estimate the precision of the ICC. ICCs and 95 % confidence intervals (two-way random model, single measure, absolute agreement) were calculated using SPSS software (SPSS).

Results

Raw list mode data was retrieved for [89Zr]antiCD20 mAb (7 patients, 20 scans), [89Zr]antiEGFR mAb (7 patients, 21 scans), and [89Zr]antiCD44 mAb (13 patients, 39 scans). List mode data was available for all scans at all time points, except for 1 patient in the [89Zr]antiCD20 mAb cohort who was originally only scanned at D0 and D6, and for which no D3 scan was available.
All original datasets were split and reconstructed (Fig. 1). However, not all splits could be reconstructed due to technical limitations, leading to the exclusion of 7 out of 80 scans. Technical limitations consisted of reconstruction failure (2 scans) and missing dicom information in the original data (5 scans). Acquisition time in h p.i. (average ± SD (range min-max)) for the included data was 1.3 ± 0.5 (0.9–2.4), 72.5 ± 3.6 (65.7–77.5), and 147.6 ± 8.0 (138.9–165.7) for [89Zr]antiCD20 mAb; 1.3 ± 0.5 (1.0–2.4), 68.4 ± 3.0 (64.6–72.8), and 143.2 ± 2.9 (138.1–145.9) for [89Zr]antiEGFR; and 1.5 ± 0.3 (1.1–2.1), 22.1 ± 1.5 (19.9–25.0), and 97.7 ± 5.3 (91.2–114.3) for [89Zr]antiCD44. The number of VOIs used for each analysis is denoted in Table 1, and delineated organ and tumor volumes are presented in Suppl. Table 1 (see Electronic Supplementary Material, ESM).
Table 1
Repeatability coefficients (%) of Zr-89-labeled mAbs
VOI type
[89Zr]antiCD20
[89Zr]antiEGFR
[89Zr]antiCD44
37 MBq74inj
18 MBq74inj
18 MBq37inj
18 MBq37inj
D0
D3
D6
D0
D3
D6
D0
D3
D6
D0
D1
D4
n = 7a
n = 6b
n = 6c
n = 7b
n = 6a
n = 6c
n = 6d
n = 6
n = 6
n = 12
n = 12
n = 12
Brain
3.5
6.0
5.2
2.4
6.5
7.6
3.0
7.6
13.8
3.7
2.9
7.3
Kidney
1.2
4.8
5.2
2.6
6.5
6.8
3.5
7.1
12.5
3.7
3.5
9.1
Lung
1.1
0.6
0.8
1.5
1.6
2.7
1.6
1.8
3.4
1.0
1.6
6.7
Spleen
1.9
4.7
8.2
3.4
8.8
17.2
5.2
8.4
13.9
4.4
3.7
7.8
Liver
1.5
1.6
1.4
1.6
0.8
3.9
1.4
1.6
1.6
1.2
2.4
1.7
Combined
2.3
4.7
5.5
2.5
6.3
11.1
4.2
6.4
10.7
3.4
3.0
7.1
Blood pool
8.3
17.1
42.7
10.6
29.9
38.5
13.4
28.8
47.4
10.4
18.3
32.3
Bone marrow
15.1
12.4
20.6
23.9
21.0
38.4
20.7
19.6
26.6
13.0
17.8
20.7
Tumor
D3
D6
D3
D6
D3
D6
D1
D4
n = 26
n = 32
 
n = 26
n = 32
 
n = 7
n = 7
 
n = 19
n = 19
SUVmax
41.6
41.5
39.1
45.5
34.9
54.1
29.5
33.5
SUVpeak
35.2
31.6
35.7
37.1
28.8
48.2
20.7
28.0
SUVmean
26.7
24.5
26.8
26.6
35.4
32.7
20.1
24.2
Combined = all VOIs of brain, kidney, lung, spleen, and liver, analyzed together as one group. Data marked in gray is presented as Bland-Altman plots in Fig. 2 (normal tissue) and 4 (tumor)
n number of VOIs per group
aNo brain VOI obtained in patient 2 (outside field of view) and 6 (tumor localization in the nasopharynx)
bNo brain VOI obtained in patient 6 (tumor localization in the nasopharynx)
cNo kidney, lung, spleen, liver, blood pool, and bone marrow in patient 2 (outside field of view). No brain VOI obtained in patient 6 (tumor localization in the nasopharynx). No brain VOI obtained in patient 1 (mismatch between low dose CT and PET image due to patient movement)
dNo brain VOI obtained in patient 8 (outside field of view)

Noise-Induced Variability Analysis

Normal Tissue Uptake
For [89Zr]antiCD20 mAb, examples of the Bland Altman plots for normal tissue are shown in Fig. 2. The corresponding RC for all datasets are presented in Table 1, corresponding SUV are presented in Suppl. Table 2 (see ESM). Liver and lung show the best RC of all normal tissue VOI, < 2 % at all time points for [89Zr]antiCD20 (37 MBq74inj). Bone marrow, a fixed-size VOI, shows a relative large RC, ranging from 12 to 21 % over time for [89Zr]antiCD20 mAb (37 MBq74inj). This pattern, showing smaller RC for the manually delineated organs, and larger RC for blood pool and bone marrow, was observed for [89Zr]antiCD20 (37 MBq74inj and 18 MBq74inj) (Table 1 and Fig. 2). The accompanying mean percentages differences were all near zero and are presented in Suppl. Table 3 (see ESM). The RC for [89Zr]antiCD20 mAb (37 MBq74inj) for all manually delineated organs combined (liver, spleen, kidney, lung, brain) increased from D0 to D6, but remained within 6 % for all measured time points (Fig. 3a).
For 18 MBq74inj, the RC for the manually delineated large organs remained within 12 % for all time points (Fig. 3a). For the image-derived blood pool (Fig. 3b), RC increased from 8 to 43 % from D0 to D6. We observed similar RC for [89Zr]antiCD20 (18 MBq74inj), [89Zr]antiEGFR (18 MBq37inj), and [89Zr]antiCD44 (18 MBq37inj) for the manually delineated organs (Fig. 3c) and for the image-derived blood pool (Fig. 3d). For the individual normal tissue VOI (brain, kidney, lung, spleen, liver, bone marrow), we found no apparent differences in RC for the various antibodies, except for kidney D6.
A decrease in RC was observed for increasing SUV and volume (Suppl. Fig. 1) (see ESM). For example, for the fixed-size blood pool VOI (2.9 mL) for [89Zr]antiCD20 mAb (37MBq74inj), a decrease in RC from 42, 17, to 8 % was observed, with a corresponding increase in SUV from 3.9, 5.8, to 11.7. For VOI with similar SUV (3.1–3.3) for [89Zr]antiCD44 mAb (18MBq37inj), a decrease in RC from 32, 18, to 4 % was observed, with a corresponding increase in volume from 2.9 mL (blood pool), 8.6 mL (bone marrow), to 309 mL (kidney).
Tumor Uptake
Bland Altman plots of tumor uptake of [89Zr]antiCD20 mAb (37MBq74inj) are shown in Fig. 4. Tumor RC are presented in Table 1, corresponding SUV in Suppl. Table 2, for mean percentages differences, see Suppl. Table 3. In contrast to the RC for organs, RC for tumor uptake did not increase consistently over time for [89Zr]anti-CD20 mAb. For tumor uptake of [89Zr]anti-CD20 mAb (37 MBq74inj), the best RC at D6 was obtained for SUVmean (26 %), followed by SUVpeak (34 %) and SUVmax (41 %) (Table 1). The same rank order (increasing RC for SUVmean, SUVpeak, and SUVmax) was observed for [89Zr]antiCD44 mAb at D4 (18 MBq37inj) and for [89Zr]antiEGFR mAb at D6 (18 MBq37inj). For [89Zr]anti-CD44 mAb, the RC for SUVpeak increased from 21 to 28 % RC (D1 to D4). These values were lower than for [89Zr]anti-CD20 mAb (Table 1). However, data for [89Zr]antiCD44 mAb were acquired at different time points after injection (D1 and D4) compared to [89Zr]antiCD20 mAb and [89Zr]antiEGFR mAb (D3 and D6). No differences were observed in RC for SUVpeak between all three mAbs at D3–D4, and between [89Zr]antiCD20 mAb and [89Zr]antiEGFR mAb at D6. The differences between SUVmean and SUVmax were significant for the three Zr-89 -mAbs combined (D3–D4), as well as on D6 for [89Zr]antiCD20 and [89Zr]antiEGFR combined.

Reliability Analysis

The ICCs for normal tissue and tumor uptake of Zr-89-mAbs are shown in Table 2 and Suppl. Table 4 (see ESM). For the manually delineated organs, ICCs > 0.9 were obtained at all time points for [89Zr]antiCD20 mAb (37 MBq74inj and 18 MBq74inj). For [89Zr]antiCD20 mAb (37 MBq74inj), blood pool and bone marrow ICCs were lower, with values as low as 0.74 and with wider 95 % confidence intervals than for the manually delineated organs.
Table 2
ICC for blood pool, bone marrow, and tumor uptake of Zr-89-labeled mAbs
ICC (lower-upper 95 % CI)
[89Zr]antiCD20 (37 MBq74inj)
D0
D3
D6
Blood pool
0.91 (0.56–0.98)
0.78 (−0.22–0.97)
0.74 (−0.27–0.97)
Bone marrow
0.81 (0.26–0.96)
0.98 (0.86–1.00)
0.97 (0.59–0.99)
Tumor (peak)
NA
0.92 (0.83–0.96)
0.98 (0.96–0.99)
 
[89Zr]antiCD20 (18 MBq74inj)
 
D0
D3
D6
Blood pool
0.92 (0.59–0.99)
0.25 (−0.60–0.84)
0.79 (−0.19–0.98)
Bone marrow
0.72 (0.11–0.94)
0.90 (0.50–0.99)
0.72 (−0.05–0.97)
Tumor (peak)
NA
0.89 (0.78–0.95)
0.97 (0.93–0.98)
 
[89Zr]antiEGFR (18 MBq37inj)
 
D0
D3
D6
Blood pool
0.98 (0.87–1.00)
0.96 (0.75–0.99)
0.89 (0.43–0.98)
Bone marrow
0.93 (0.59–0.99)
0.98 (0.88–1.00)
0.91 (0.55–0.99)
Tumor (peak)
NA
0.94 (0.47–0.99)
0.88 (0.47–0.99)
 
[89Zr]antiCD44 (18 MBq37inj)
 
D0
D1
D4
Blood pool
0.99 (0.95–1.00)
0.97 (0.89–0.99)
0.95 (0.83–0.99)
Bone marrow
0.96 (0.86–0.99)
0.96 (0.86–0.99)
0.87 (0.60–0.96)
Tumor (peak)
NA
0.97(0.91–0.99)
0.96 (0.90–0.99)
For tumor uptake of [89Zr]mAbs, ICCs of > 0.9 were obtained in all datasets, despite RC of 40 %. Similar tumor ICCs of 0.9 were obtained for [89Zr]antiCD20 mAb (18 MBq74inj), [89Zr]antiEGFR mAb (18 MBq37MBq), and [89Zr]antiCD44 mAb (18MBq37MBq). Increase in the range of tumor uptake between tumor lesions over time for [89Zr]antiCD20 (37 MBq74inj) (Fig. 5), resulted in a higher ICC at D6. This trend was not observed for [89Zr]antiEGFR (18 MBq37inj).

Discussion

In this study, noise-induced variability of quantitative uptake measures was assessed for Zr-89 -immuno-PET for an injected activity of 37 MBq based on count-reduced images. As expected, a variable increase in RC was observed from D0 to D6 for the manually delineated organs (Table 1). In general, the RC for the manually delineated large organs combined (liver, spleen, kidney, lung, and brain) was within 6 % at all time points (D0, D3, D6) for [89Zr]anti-CD20 mAb (37MBq74inj). Larger measurement variability was observed for the bone marrow and blood pool VOIs with RC up to 40 % at D6 for [89Zr]anti-CD20 (37 MBq74inj). These results are in line with an increase in RC for a lower total activity in the VOI.
For tumor uptake of [89Zr]antiCD20 mAb (37 MBq74inj), the lowest variability was obtained for SUVmean (27 %), followed by SUVpeak (35 %), while SUVmax (42 %) resulted in the highest measurement variability (Table 2). This is as expected, since for a given VOI the mean value takes all voxels into account, in contrast to the peak value (based on a limited number of voxels) and max value (based on a single voxel).
RC for the three Zr-89 -mAbs were similar. Therefore, the observed measurement variability is independent of the differences in biodistribution between these three Zr-89 -mAbs. These results suggest that similar noise-induced variability can be expected for other Zr-89 -mAbs, assuming harmonized image quality and a biodistribution within the same range as the Zr-89 -mAbs investigated in this study.
With the count-reduced images, noise-induced variability for an injected dose of 18 MBq was assessed for all three mAbs. RC for combined organs were up to 12 % at all time points (range D0–D6). Tumor RC varied between 20 and 54 %, depending on time point and VOI delineation method.
Despite relatively large RC for tumor, blood pool, and bone marrow, overall reliability for the three clinical Zr-89 -immuno-PET studies previously reported [79] was excellent (ICCs approximately 0.9). ICC values are given to show, for these Zr-89 -mAbs in their respective patient cohorts, that the measured differences in tumor uptake do exceed the variability induced by noise. ICC values, however, cannot be extrapolated to other Zr-89 -mAbs or even different patient cohorts imaged with the same Zr-89 -mAb.
As our study provides an assessment of measurement variability due to the signal-to-noise ratio, other sources of measurement error are not represented in this reliability assessment (Eq. 2). Factors affecting 2-deoxy-2-[18F]fluoro-d-glucose ([18F]FDG)-PET quantification have been described previously [15]. Repeatability assessed by a classical test-retest study contains intra-patient variability between the test and the retest scan as well. For a test-retest study with Zr-89 -immuno-PET, the following factors are expected to play a role (1) biological factors: e.g., uptake period (higher uptake levels at increasing time interval between injection and start of PET study), estimated at 1 % for ± 1 h [16, 17] and (2) physical factors: VOI definition, estimated at < 1 % for max and peak AC to 8 % for mean [18]. Based on the results of our study, we expect that noise-induced variability will be a major contribution to measurement variability in a Zr-89 -immuno-PET test-retest study (3 % for large organs, 20 % for tumor, values presented as SD instead of RC for comparison).
In previously reported test-retest studies of repeatability of [18F]FDG-PET, RC of < 10 % have been reported for tumor uptake on [18F]FDG-PET, resulting in thresholds of 10–15 % needed to detect therapy-induced changes in patients with non-small cell lung cancer [1921]. For Zr-89 -immuno-PET (37 MBq74inj), we obtained similar RC of less than 10 % for mean activity concentrations of manually delineated organs. However, for tumor uptake on Zr-89 -immuno-PET, RC up to 42 % for SUVmax reflect the much lower signal -to- noise ratio in Zr-89 -immuno-PET scans in comparison to [18F]FDG-PET scans. When immuno-PET with Zr-89-labeled mAbs is used to assess response to treatment (e.g., by alteration of antigen expression [22]), knowledge on measurement variability should be applied to set corresponding thresholds, following the example of the use of thresholds for response assessment with [18F]FDG-PET. For data already obtained, RC from our study are relevant to allow correct interpretation, as differences smaller than the measurement variability cannot be attributed to biological effects. Measurement variability (given as RC) is independent of the study population. In addition, RC of tumor and blood pool are of interest when selecting an appropriate uptake measure and/or VOI delineation method.
Tumor-to-blood ratios, which are commonly used, will result in even worse RC due to propagation of the individual RC as both numerator and denominator contain measurement variability. Future work may include an investigation to optimize delineation methods for image-derived blood pool (e.g., delineation of a larger region in the left ventricular cavity compared to a smaller fixed-size VOI in the aortic arch) to reduce noise-induced variability. We designed this study based on patient data to provide a clinically relevant assessment of measurement error. This study provides the first description of noise-induced variability for Zr-89 -immuno-PET. In literature, more sophisticated applications of the technique of count-reduced images have been described [23, 24]. By using bootstrapping, the difference between the two image estimates can be obtained more precisely, which may allow for an examination of the change of noise-induced variability with patient-specific factors such as tumor size, uptake, patient size, and location of the tumor in the patient. In addition, through application of this technique, the effect of lowering the injected dose on quantitative accuracy can be assessed.
Our results suggest that, as expected, measurement error may be dependent on volume and uptake (Suppl. Fig. 1). In future work, phantoms could be used to study dependencies of noise-induced variability on, e.g., VOI volume and activity concentration. Such phantom experiments could be combined with tumor characteristics as derived from clinical Zr-89 -immuno-PET studies (tumor volume, SUV, tumor -to- background ratio and localization) to obtain a detection limit for tumor identification. The trade-off between noise and contrast will determine the optimal time point for tumor imaging. In addition, tumor contrast on clinical images should be defined and characterized, allowing for a recommendation on the optimal time points for future clinical studies.
For potential application of immuno-PET with Zr-89-labeled antibodies as a quantitative imaging biomarker to predict which patients are likely to respond to antibody treatment, the ability to distinguish biological differences in antibody uptake between patients is required. We observed excellent ICCs (lesion-based analysis), suggesting that an injected dose of 18-37 MBq was sufficient for the datasets included. In general, justification of the injected dose should be based on the expected effect size in the population of interest.
Another clinical application of Zr-89 -immuno-PET is in vivo quantification of antibody biodistribution and tumor uptake to assess novel antibodies during early phases of drug development. For this purpose, noise-induced variability as source of measurement error as reported in our study (expressed as RC) can be applied to estimate the minimal measurement variability (e.g., for sample size calculations for novel studies).

Conclusion

Based on this study, noise-induced variability results in a RC for Zr-89 -immuno-PET (37 MBq) of less than 6 % for manually delineated organs combined, increasing up to 43 % at D6 for blood pool and bone marrow, assuming similar biodistribution of mAbs. The signal -to- noise ratio in Zr-89 -immuno-PET scans leads to tumor RC up to 42 %.

Acknowledgements

We thank Maqsood Yaqub (Department of Radiology & Nuclear Medicine, VU University Medical Center, Amsterdam, The Netherlands) for the PET analysis software and Birgit Witte (Department of Epidemiology and Biostatistics, VU University Medical Center, Amsterdam, the Netherlands) for statistical support. We thank F. Hoffman-La Roche Ltd., Basel, Switzerland for providing support of the [89Zr]-antiCD44 imaging study.

Compliance with Ethical Standards

Conflicts of Interest

DH is employed by Philips Healthcare. No potential conflicts of interest relevant to this article were reported.

Ethical Approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent was obtained from all individual participants included in this study.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Literatur
1.
Zurück zum Zitat Lamberts LE, Williams SP, Terwisscha van Scheltinga AG, et al (2015) Antibody positron emission tomography imaging in anticancer drug development. J Clin Oncol 33:1491–1504CrossRef Lamberts LE, Williams SP, Terwisscha van Scheltinga AG, et al (2015) Antibody positron emission tomography imaging in anticancer drug development. J Clin Oncol 33:1491–1504CrossRef
2.
Zurück zum Zitat Pandit-Taskar N, O’Donoghue JA, Beylergil V, et al (2014) 89Zr-huJ591 immuno-PET imaging in patients with advanced metastatic prostate cancer. Eur J Nucl Med Mol Imaging 41:2093–2105CrossRef Pandit-Taskar N, O’Donoghue JA, Beylergil V, et al (2014) 89Zr-huJ591 immuno-PET imaging in patients with advanced metastatic prostate cancer. Eur J Nucl Med Mol Imaging 41:2093–2105CrossRef
3.
Zurück zum Zitat Jauw YWS, Menke-van der Houven van Oordt CW, Hoekstra OS et al (2016) Immuno-positron emission tomography with zirconium-89-labelled monoclonal antibodies in oncology: what can we learn from initial clinical trials? Front Pharmacol 7:131CrossRef Jauw YWS, Menke-van der Houven van Oordt CW, Hoekstra OS et al (2016) Immuno-positron emission tomography with zirconium-89-labelled monoclonal antibodies in oncology: what can we learn from initial clinical trials? Front Pharmacol 7:131CrossRef
4.
Zurück zum Zitat Armstrong IS, James JM, Williams HA, et al (2015) The assessment of time-of-flight on image quality and quantification with reduced administered activity and scan times in 18F-FDG PET. Nucl Med Commun 36:728–737CrossRef Armstrong IS, James JM, Williams HA, et al (2015) The assessment of time-of-flight on image quality and quantification with reduced administered activity and scan times in 18F-FDG PET. Nucl Med Commun 36:728–737CrossRef
5.
Zurück zum Zitat Bruijnen S, Tsang-A-Sjoe M, Raterman H, et al (2016) B-cell imaging with zirconium-89 labelled rituximab PET-CT at baseline is associated with therapeutic response 24 weeks after initiation of rituximab treatment in rheumatoid arthritis patients. Arthritis Res Ther 18:266 Bruijnen S, Tsang-A-Sjoe M, Raterman H, et al (2016) B-cell imaging with zirconium-89 labelled rituximab PET-CT at baseline is associated with therapeutic response 24 weeks after initiation of rituximab treatment in rheumatoid arthritis patients. Arthritis Res Ther 18:266
7.
Zurück zum Zitat Jauw YWS, Zijlstra JM, de Jong D, et al (2017) Performance of 89Zr-labelled-rituximab-PET as an imaging biomarker to assess CD20 targeting: a pilot study in patients with relapsed/refractory diffuse large B cell lymphoma. PLoS One 12:e0169828CrossRef Jauw YWS, Zijlstra JM, de Jong D, et al (2017) Performance of 89Zr-labelled-rituximab-PET as an imaging biomarker to assess CD20 targeting: a pilot study in patients with relapsed/refractory diffuse large B cell lymphoma. PLoS One 12:e0169828CrossRef
8.
Zurück zum Zitat Menke-van der Houven van Oordt CW, Gootjes EC, Huisman MC, et al (2015) 89Zr-cetuximab PET imaging in patients with advanced colorectal cancer. Oncotarget 30:30384–30393 Menke-van der Houven van Oordt CW, Gootjes EC, Huisman MC, et al (2015) 89Zr-cetuximab PET imaging in patients with advanced colorectal cancer. Oncotarget 30:30384–30393
9.
Zurück zum Zitat Menke-van der Houven van Oordt CW, Gomez-Roca C, van Herpen C, et al (2016) First-in-human phase I clinical trial of RG7356, an anti-CD44 humanized antibody, in patients with advanced, CD44-expressing solid tumours. Oncotarget 43:80046–80058 Menke-van der Houven van Oordt CW, Gomez-Roca C, van Herpen C, et al (2016) First-in-human phase I clinical trial of RG7356, an anti-CD44 humanized antibody, in patients with advanced, CD44-expressing solid tumours. Oncotarget 43:80046–80058
10.
Zurück zum Zitat Makris NE, Boellaard R, Visser EP, et al (2014) Multicenter harmonization of 89Zr PET/CT performance. J Nucl Med 55:264–267CrossRef Makris NE, Boellaard R, Visser EP, et al (2014) Multicenter harmonization of 89Zr PET/CT performance. J Nucl Med 55:264–267CrossRef
11.
Zurück zum Zitat Boellaard R, Delgado-Bolton R, Oyen WJ, et al (2015) FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging 42:328–354CrossRef Boellaard R, Delgado-Bolton R, Oyen WJ, et al (2015) FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging 42:328–354CrossRef
12.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRef
13.
Zurück zum Zitat Giavarina D (2015) Understanding Bland Altman analysis. Biochem Med 25:141–151CrossRef Giavarina D (2015) Understanding Bland Altman analysis. Biochem Med 25:141–151CrossRef
14.
Zurück zum Zitat Nunnally JC (1994) Psychometric theory, 3rd edn. McGraw-Hill, New York Nunnally JC (1994) Psychometric theory, 3rd edn. McGraw-Hill, New York
15.
Zurück zum Zitat Boellaard R (2009) Standards for PET image acquisition and quantitative data analysis. J Nucl Med 50:Suppl 1:11S–20SCrossRef Boellaard R (2009) Standards for PET image acquisition and quantitative data analysis. J Nucl Med 50:Suppl 1:11S–20SCrossRef
16.
Zurück zum Zitat Shah DK, Betts AM (2012) Towards a platform PBPK model to characterize the plasma and tissue disposition of monoclonal antibodies in preclinical species and human. J Pharmacokinet Pharmacodyn 39:67–86CrossRef Shah DK, Betts AM (2012) Towards a platform PBPK model to characterize the plasma and tissue disposition of monoclonal antibodies in preclinical species and human. J Pharmacokinet Pharmacodyn 39:67–86CrossRef
17.
Zurück zum Zitat Klettering P, Maaß C, Reske S, et al (2015) Physiologically based pharmacokinetic modeling is essential in 90Y-labelled anti-CD66 radioimmunotherapy. PLoS One 5:e0127934CrossRef Klettering P, Maaß C, Reske S, et al (2015) Physiologically based pharmacokinetic modeling is essential in 90Y-labelled anti-CD66 radioimmunotherapy. PLoS One 5:e0127934CrossRef
18.
Zurück zum Zitat Jauw Y, Hoekstra O, Mulder E, et al (2016) Inter-observer agreement for tumour uptake quantification of 89Zr-labelled anti-CD20 antibodies with PET. J Nucl Med 57(Suppl 2):1412 Jauw Y, Hoekstra O, Mulder E, et al (2016) Inter-observer agreement for tumour uptake quantification of 89Zr-labelled anti-CD20 antibodies with PET. J Nucl Med 57(Suppl 2):1412
20.
Zurück zum Zitat Kramer GM, Frings V, Hoetjes N, et al (2016) Repeatability of quantitative whole-body 18F-FDG PET/CT uptake measures as function of uptake interval and lesion selection in non-small cell lung cancer patients. J Nucl Med 30:1343–1349CrossRef Kramer GM, Frings V, Hoetjes N, et al (2016) Repeatability of quantitative whole-body 18F-FDG PET/CT uptake measures as function of uptake interval and lesion selection in non-small cell lung cancer patients. J Nucl Med 30:1343–1349CrossRef
21.
Zurück zum Zitat Hoekstra CJ, Hoekstra OH, Stroobants SG, et al (2002) Methods to monitor response to chemotherapy in non-small cell lung cancer with 18F-FDG PET. J Nucl Med 43:1304–1309 Hoekstra CJ, Hoekstra OH, Stroobants SG, et al (2002) Methods to monitor response to chemotherapy in non-small cell lung cancer with 18F-FDG PET. J Nucl Med 43:1304–1309
22.
Zurück zum Zitat Gaykema SB, Schroder CP, Vitfell-Rasmussen J, et al (2014) 89Zr-trastuzumab and 89Zr-bevacizumab PET to evaluate the effect of the HSP90 inhibitor NVP-AUY922 in metastatic breast cancer patients. Clin Can Res 20:3945–3954CrossRef Gaykema SB, Schroder CP, Vitfell-Rasmussen J, et al (2014) 89Zr-trastuzumab and 89Zr-bevacizumab PET to evaluate the effect of the HSP90 inhibitor NVP-AUY922 in metastatic breast cancer patients. Clin Can Res 20:3945–3954CrossRef
23.
Zurück zum Zitat Lartizien C, Aubin J-B, Buvat I (2010) Comparison of bootstrap resampling methods for 3-D PET imaging. IEEE Trans Med Img 29:1442–1454CrossRef Lartizien C, Aubin J-B, Buvat I (2010) Comparison of bootstrap resampling methods for 3-D PET imaging. IEEE Trans Med Img 29:1442–1454CrossRef
24.
Zurück zum Zitat Markiewicz PJ, Reader AJ, Matthews JC (2015) Assessment of bootstrap resampling performance of PET data. Phys Med Biol 60:279–299CrossRef Markiewicz PJ, Reader AJ, Matthews JC (2015) Assessment of bootstrap resampling performance of PET data. Phys Med Biol 60:279–299CrossRef
Metadaten
Titel
Noise-Induced Variability of Immuno-PET with Zirconium-89-Labeled Antibodies: an Analysis Based on Count-Reduced Clinical Images
verfasst von
Yvonne W. S. Jauw
Dennis F. Heijtel
Josée M. Zijlstra
Otto S. Hoekstra
Henrica C. W. de Vet
Danielle J. Vugts
Henk M. Verheul
Ronald Boellaard
Sonja Zweegman
Guus A. M. S. van Dongen
C. Willemien Menke-van der Houven van Oordt
Adriaan A. Lammertsma
Marc C. Huisman
Publikationsdatum
30.04.2018
Verlag
Springer International Publishing
Erschienen in
Molecular Imaging and Biology / Ausgabe 6/2018
Print ISSN: 1536-1632
Elektronische ISSN: 1860-2002
DOI
https://doi.org/10.1007/s11307-018-1200-4

Weitere Artikel der Ausgabe 6/2018

Molecular Imaging and Biology 6/2018 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.