Optimization of Keyhole Imaging Parameters for Glutamate Chemical Exchange Saturation Transfer MRI at 7.0 T

verfasst von: Dong-Hoon Lee, Do-Wan Lee, Chul-Woong Woo, Jae-Im Kwon, Jeong Kon Kim, Kyung Won Kim, Dong-Cheol Woo

Erschienen in: Molecular Imaging and Biology | Ausgabe 4/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the effects of a reference image and keyhole factor (Kf) selections for high-frequency substitution on keyhole imaging technique for applications in glutamate chemical exchange saturation transfer (GluCEST) imaging.

Procedures

The CEST data were obtained using a 7.0 T MRI scanner. We used varied Kf ranges that constituted from 16.67 to 75 % of the full k-space. The reference image was respectively selected for − 3 and + 3 ppm images that associated with the GluCEST calculation and the unsaturated image. The zero-padding algorithm was applied for the missing k-space lines in the low-frequency data collected to compare the results obtained from using the keyhole imaging technique. All the techniques were evaluated using a healthy rat group and extended to the status epilepticus rat group to explore their applicability and usability.

Results

The calculated GluCEST signals and visually inspected results from the reconstructed GluCEST maps indicated that the combination of unsaturated image as a reference image, and over 50 % of Kf showed consistent signals and image quality compared with the fully sampled CEST data.

Conclusions

Combining the keyhole imaging technique with GluCEST imaging enables stable image reconstruction and quantitative evaluation, and this approach is potentially applicable in various CEST imaging applications.
Literatur
1.
Zurück zum Zitat Dula AN, Smith SA, Gore JC (2013) Application of chemical exchange saturation transfer (CEST) MRI for endogenous contrast at 7 tesla. J Neuroimaging 23:526–532CrossRef Dula AN, Smith SA, Gore JC (2013) Application of chemical exchange saturation transfer (CEST) MRI for endogenous contrast at 7 tesla. J Neuroimaging 23:526–532CrossRef
2.
Zurück zum Zitat Jones KM, Pollard AC, Pagel MD (2018) Clinical applications of chemical exchange saturation transfer (CEST) MRI. J Magn Reson Imaging 47:11–27CrossRef Jones KM, Pollard AC, Pagel MD (2018) Clinical applications of chemical exchange saturation transfer (CEST) MRI. J Magn Reson Imaging 47:11–27CrossRef
3.
Zurück zum Zitat van Zijl PCM, Yadav NN (2011) Chemical exchange saturation transfer (CEST): what is in a name and what isn’t? Magn Reson Med 65:927–948CrossRef van Zijl PCM, Yadav NN (2011) Chemical exchange saturation transfer (CEST): what is in a name and what isn’t? Magn Reson Med 65:927–948CrossRef
4.
Zurück zum Zitat Zhou J, Blakeley JO, Hua J et al (2008) Practical data acquisition method for human brain tumor amide proton transfer (APT) imaging. Magn Reson Med 60:842–849 Zhou J, Blakeley JO, Hua J et al (2008) Practical data acquisition method for human brain tumor amide proton transfer (APT) imaging. Magn Reson Med 60:842–849
5.
Zurück zum Zitat Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRef Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRef
6.
Zurück zum Zitat Hu X (1994) On the “keyhole” technique. J Magn Reson Imaging 4:231CrossRef Hu X (1994) On the “keyhole” technique. J Magn Reson Imaging 4:231CrossRef
7.
Zurück zum Zitat Hu X, Parrish T (1994) Reduction of field of view for dynamic imaging. Magn Reson Med 31:691–694CrossRef Hu X, Parrish T (1994) Reduction of field of view for dynamic imaging. Magn Reson Med 31:691–694CrossRef
8.
Zurück zum Zitat She H, Greer JS, Zhang S et al (2019) Accelerating chemical exchange saturation transfer MRI with parallel blind compressed sensing. Magn Reson Med 81:504–513 She H, Greer JS, Zhang S et al (2019) Accelerating chemical exchange saturation transfer MRI with parallel blind compressed sensing. Magn Reson Med 81:504–513
9.
Zurück zum Zitat Jaspan ON, Fleysher R, Lipton ML (2015) Compressed sensing MRI: a review of the clinical literature. Br J Radiol 88:20150487CrossRef Jaspan ON, Fleysher R, Lipton ML (2015) Compressed sensing MRI: a review of the clinical literature. Br J Radiol 88:20150487CrossRef
10.
Zurück zum Zitat Heo HY, Zhang Y, Lee DH, Jiang S, Zhao X, Zhou J (2017) Accelerating chemical exchange saturation transfer (CEST) MRI by combining compressed sensing and sensitivity encoding techniques. Magn Reson Med 77:779–786CrossRef Heo HY, Zhang Y, Lee DH, Jiang S, Zhao X, Zhou J (2017) Accelerating chemical exchange saturation transfer (CEST) MRI by combining compressed sensing and sensitivity encoding techniques. Magn Reson Med 77:779–786CrossRef
11.
Zurück zum Zitat Varma G, Lenkinski RE, Vinogradov E (2012) Keyhole chemical exchange saturation transfer. Magn Reson Med 68:1228–1233CrossRef Varma G, Lenkinski RE, Vinogradov E (2012) Keyhole chemical exchange saturation transfer. Magn Reson Med 68:1228–1233CrossRef
12.
Zurück zum Zitat Cai KJ, Haris M, Singh A et al (2012) Magnetic resonance imaging of glutamate. Nat Med 18:302–306 Cai KJ, Haris M, Singh A et al (2012) Magnetic resonance imaging of glutamate. Nat Med 18:302–306
13.
Zurück zum Zitat Singh A, Cai KJ, Haris M, Hariharan H, Reddy R (2013) On B1 inhomogeneity correction of in vivo human brain glutamate chemical exchange saturation transfer contrast at 7T. Magn Reson Med 69:818–824CrossRef Singh A, Cai KJ, Haris M, Hariharan H, Reddy R (2013) On B1 inhomogeneity correction of in vivo human brain glutamate chemical exchange saturation transfer contrast at 7T. Magn Reson Med 69:818–824CrossRef
14.
Zurück zum Zitat Du YP, Parker DL, Davis WL, Cao G (1994) Reduction of partial-volume artifacts with zero-filled interpolation in three-dimensional MR angiography. J Magn Reson Imaging 4:733–741CrossRef Du YP, Parker DL, Davis WL, Cao G (1994) Reduction of partial-volume artifacts with zero-filled interpolation in three-dimensional MR angiography. J Magn Reson Imaging 4:733–741CrossRef
15.
Zurück zum Zitat Lee DH, Lee DW, Kwon JI et al (2019) In vivo mapping and quantification of creatine using chemical exchange saturation transfer imaging in rat models of epileptic seizure. Mol Imaging Biol 21:232–239 Lee DH, Lee DW, Kwon JI et al (2019) In vivo mapping and quantification of creatine using chemical exchange saturation transfer imaging in rat models of epileptic seizure. Mol Imaging Biol 21:232–239
16.
Zurück zum Zitat Sperk G (1994) Kainic acid seizures in the rat. Prog Neurobiol 42:1–32CrossRef Sperk G (1994) Kainic acid seizures in the rat. Prog Neurobiol 42:1–32CrossRef
17.
Zurück zum Zitat Park J, Zhang Q, Jellus V, Simonetti O, Li D (2005) Artifact and noise suppression in GRAPPA imaging using improved k-space coil calibration and variable density sampling. Magn Reson Med 53:186–193CrossRef Park J, Zhang Q, Jellus V, Simonetti O, Li D (2005) Artifact and noise suppression in GRAPPA imaging using improved k-space coil calibration and variable density sampling. Magn Reson Med 53:186–193CrossRef
18.
Zurück zum Zitat Xiao Z, Hoge WS, Mulkern RV, Zhao L, Hu G, Kyriakos WE (2008) Comparison of parallel MRI reconstruction methods for accelerated 3D fast spin-echo imaging. Magn Reson Med 60:650–660CrossRef Xiao Z, Hoge WS, Mulkern RV, Zhao L, Hu G, Kyriakos WE (2008) Comparison of parallel MRI reconstruction methods for accelerated 3D fast spin-echo imaging. Magn Reson Med 60:650–660CrossRef
19.
Zurück zum Zitat Mogatadakala KV, Bankson JA, Narayana PA (2008) Three-element phased-array coil for imaging of rat spinal cord at 7T. Magn Reson Med 60:1498–1505CrossRef Mogatadakala KV, Bankson JA, Narayana PA (2008) Three-element phased-array coil for imaging of rat spinal cord at 7T. Magn Reson Med 60:1498–1505CrossRef
20.
Zurück zum Zitat Herrmann KH, Schmidt S, Kretz A et al (2012) Possibilities and limitations for high resolution small animal MRI on a clinical whole-body 3T scanner. Magn Reson Mater Phy 25:233–244 Herrmann KH, Schmidt S, Kretz A et al (2012) Possibilities and limitations for high resolution small animal MRI on a clinical whole-body 3T scanner. Magn Reson Mater Phy 25:233–244
21.
Zurück zum Zitat Lee DH, Lee DW, Kwon JI et al (2019) Retrospective brain motion correction in glutamate chemical exchange saturation transfer (GluCEST) MRI. Mol Imaging Biol 21:1064–1070 Lee DH, Lee DW, Kwon JI et al (2019) Retrospective brain motion correction in glutamate chemical exchange saturation transfer (GluCEST) MRI. Mol Imaging Biol 21:1064–1070
23.
Zurück zum Zitat Jones CK, Polders D, Hua J et al (2012) In vivo three-dimensional whole-brain pulsed steady-state chemical exchange saturation transfer at 7 T. Magn Reson Med 67:1579–1589 Jones CK, Polders D, Hua J et al (2012) In vivo three-dimensional whole-brain pulsed steady-state chemical exchange saturation transfer at 7 T. Magn Reson Med 67:1579–1589
24.
Zurück zum Zitat Zaiss M, Xu JZ, Goerke S et al (2014) Inverse Z-spectrum analysis for spillover-, MT-, and T-1-corrected steady-state pulsed CEST-MRI - application to pHweighted MRI of acute stroke. NMR Biomed 27:240–252 Zaiss M, Xu JZ, Goerke S et al (2014) Inverse Z-spectrum analysis for spillover-, MT-, and T-1-corrected steady-state pulsed CEST-MRI - application to pHweighted MRI of acute stroke. NMR Biomed 27:240–252
Metadaten
Titel
Optimization of Keyhole Imaging Parameters for Glutamate Chemical Exchange Saturation Transfer MRI at 7.0 T
verfasst von
Dong-Hoon Lee
Do-Wan Lee
Chul-Woong Woo
Jae-Im Kwon
Jeong Kon Kim
Kyung Won Kim
Dong-Cheol Woo
Publikationsdatum
19.12.2019
Verlag
Springer International Publishing
Erschienen in
Molecular Imaging and Biology / Ausgabe 4/2020
Print ISSN: 1536-1632
Elektronische ISSN: 1860-2002
DOI
https://doi.org/10.1007/s11307-019-01461-z

Weitere Artikel der Ausgabe 4/2020

Molecular Imaging and Biology 4/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.